ability of measures obtained using a custom-made ... was designed in an attempt to develop a simple, ... Military Performance Laboratory, Center for the Intrepid, Brooke Army Medical Center, Fort Sam Houston, TX. ..... of force application.
[

technical note

]

JASON WILKEN, MPT, PhD1 • SMITA RAO, PT, PhD2 • MIRIAM ESTIN3 CHARLES SALTZMAN, MD4 • H. JOHN YACK, PT, PhD5

A New Device for Assessing Ankle Dorsiflexion Motion: Reliability and Validity

C

linical and epidemiological studies have identified limitation in ankle dorsiflexion and increased ankle stiffness as key contributors to the evolution of foot and ankle pathology.8,16,26 A lack of robust measurement techniques to quantify ankle range of motion (ROM) and stiffness, however, has limited our ability to assess TTSTUDY DESIGN: Clinical measurement.

TTOBJECTIVE: To determine the validity and reli-

ability of measures obtained using a custom-made device for assessing ankle dorsiflexion motion and stiffness.

TTBACKGROUND: Limited dorsiflexion has

been implicated in the evolution of foot pain in a number of clinical populations. Assessment of ankle dorsiflexion range of motion (ROM) is, therefore, commonly performed as part of a foot and ankle examination. Conventional goniometric assessment methods have demonstrated limited intertester reliability, while alternative methods of measurements are generally more difficult to use. The Iowa ankle range of motion (IAROM) device was designed in an attempt to develop a simple, clinically relevant, and time- and cost-effective tool to measure ankle dorsiflexion range of motion and stiffness.

TTMETHODS: Validity and intertester reliability of

dorsiflexion range-of-motion measures using the IAROM device were assessed at 10, 15, 20, and 25 Nm of passively applied dorsiflexion torque, with both the knee extended and flexed approximately 20°. Stiffness (change in torque/change in dorsiflexion angle) values were determined using the angular change obtained between the 15- and

25-Nm torque levels. Convergent validity (n = 12) was assessed through comparison of ankle dorsiflexion angles measured simultaneously with the IAROM device and an optoelectronic motion analysis system. Intertester reliability (n = 17) was assessed by 2 testers who took measurements within the same day.

TTRESULTS: Validity testing demonstrated excel-

lent agreement (intraclass correlation coefficient [ICC] values ranging from 0.95 to 0.98). Reliability testing demonstrated good to excellent intertester agreement (ICC values ranging from 0.90 to 0.95). The ICCs for ankle joint dorsiflexion stiffness were .71 and .85 for the knee in an extended and flexed position, respectively.

TTCONCLUSION: The IAROM device provides valid and reliable measurement of ankle dorsiflexion ROM. The IAROM device also allows calculation of stiffness by measuring ROM at multiple torque levels, although the reliability of the measurement is not optimal. J Orthop Sports Phys Ther 2011;41(4):274-280. doi:10.2519/jospt.2011.3397

TTKEY WORDS: IAROM, plantar flexors, stiffness, talocrural joint

the functional consequences of relatively small changes in ankle ROM. Treatment strategies ranging from stretching exercises to surgical lengthening of the gastrocnemius-soleus complex are prescribed based on assessments of dorsiflexion ROM, despite the less-than-adequate reliability of these assessments.5,9,20,22,35 Recent research has identified the importance of being able to accurately measure relatively small changes in ankle ROM to accurately identify impairments.8,17,24,34 Currently used methods to obtain ankle ROM may be broadly classified into 3 categories: goniometry (based on tester’s perception of end feel),6,9 weight-bearing (based on the individual’s perception of end feel),4,21 and instrumented (torquereferenced) techniques.3,25,32 The primary objections to obtaining ROM measures using goniometry are that it is not adequately reliable, it is tester dependent, and recording angular displacement alone does not allow the calculation of stiffness.9,22,35 Better measures of ankle dorsiflexion must, therefore, be developed to allow simultaneous assessment of stiffness and to make appropriate decisions about treatment options and outcomes. To address the less-than-optimal reliability and validity of handheld goniometric measurements, several alternative

Director, Military Performance Laboratory, Center for the Intrepid, Brooke Army Medical Center, Fort Sam Houston, TX. 2Assistant Professor, Department of Physical Therapy, New York University, New York, NY. 3Research Assistant, Graduate Program in Physical Therapy and Rehabilitation Science, The University of Iowa, Iowa City, IA. 4Professor and Chairman, Department of Orthopaedics, University of Utah School of Medicine, Salt Lake City, UT. 5Associate Professor, Graduate Program in Physical Therapy and Rehabilitation Science, The University of Iowa, Iowa City, IA. This protocol for this study was approved by The University of Iowa Human Subjects Office, Institutional Review Board. We, the authors, affirm that we have no financial affiliation (including research funding) or involvement with any commercial organization in any matter included in this manuscript. This project was supported through departmental funding. Jason Wilken received partial support from the Foundation for Physical Therapy, Inc during the time the study was completed. Address correspondence to Dr Smita Rao, Department of Physical Therapy, New York University, 380 2nd Ave, 4th Floor, New York, NY. E-mail: [email protected] 1

274  |  april 2011  |  volume 41  |  number 4  |  journal of orthopaedic & sports physical therapy

methods of assessing ankle dorsiflexion ROM have been recently proposed. One functional approach for assessing ankle ROM has been to ask patients to perform a controlled lunge.4,21 While this approach has been shown to be reliable,4 weight-bearing methods of assessing dorsiflexion may have limited applicability in certain clinical situations, such as the early phases of ulcer healing or in postsurgery or conditions in which weight bearing is restricted. An alternative, relatively direct method of assessing ankle dorsiflexion ROM has been to photograph the foot and leg as a given force is applied to the metatarsal heads, then to measure the orientation of the foot relative to the leg.18 Beyond the practical difficulties of applying this technique in a clinic, failure to stabilize the foot and the localized application of force may not adequately replicate loading of the foot during functional tasks. Computerized approaches that isolate ankle joint motion and measure the applied external torque, making it possible to measure not only ankle ROM but ankle stiffness (changes in ankle dorsiflexion angle as a function of applied external torque), have also been proposed.3,7,8,10,32 However, the increased accuracy of these more sophisticated measurement approaches comes at a financial cost. These systems also tend to be difficult to transport and complicated to use, raising questions as to their clinical usefulness and cost effectiveness. Most importantly, few of these ROM devices have demonstrated reliability,4,18 and none, to our knowledge, has demonstrated validity. Further, while most studies report ankle ROM data from an average of 3 readings, none has quantified the potential benefit of single versus average measures of ROM. Given the shortcomings of these previous techniques, the Iowa ankle range of motion (IAROM) device was designed in an attempt to develop a simple, clinically relevant, and time- and cost-effective measurement technique. The purpose of this project was, first, to assess intratester reliability of single versus average

FIGURE. The Iowa ankle range of motion testing device positioned with the knee flexed. Key features of the device include the ability to alter knee flexion angle (A) and device axis of rotation (B), as well as determine angular displacement (C) at known torque levels (D).

measures (as to answer the question, “Do I need to take and average of 3 measures, or is 1 sufficient?”) and, second, to determine the intertester reliability and convergent validity of measures obtained with a new, portable, practical, and costeffective device for testing ankle dorsiflexion ROM and stiffness.

METHODS Participants

T

his study was approved by The University of Iowa Institutional Review Board for Experimentation on Human Subjects, and consent was obtained from all participants. For validity testing, 12 participants (6 male, 6 female; mean  SD age, 23  3 years; height, 1.7  0.1 m; body mass, 72  12 kg) were simultaneously evaluated using the IAROM device and a motion analysis system (Optotrak 3020; Northern Digital Inc, Waterloo, ON, Canada). To evaluate intertester reliability, dorsiflexion ROM was assessed in 17 participants (7 male, 10 female; mean  SD age, 52  15 years; height,1.7  0.1 m; body mass, 88  21 kg). Patients receiving orthopaedic care for unilateral foot or ankle injury or pathology at the University of Iowa Hospitals and Clinics were recruited to participate in the study. The limb contralateral to the one requiring orthopaedic care was tested.

Instrumentation The IAROM device (FIGURE 1) was de-

signed to be easy to assemble, inexpensive to produce, and easy to use. The device consisted of a 30-by-30-cm Plexiglas footplate, attached to a 30-by-40cm base plate. Two 10-cm-wide Velcro straps, passed through lateral slots in the base plate, secured the lower leg during testing. A 3-cm-high semirigid foam block was placed under the distal leg to support the tibia in a position perpendicular to the foot plate. Adjustable supports on the proximal end of the device allowed knee flexion angle to be increased to approximately 20° during testing.

Procedures Participants were positioned in the device with the long axis of their tibia perpendicular to the foot plate, with the foot vertical and the sole in contact with the transparent Plexiglas foot plate. The axis of rotation of the device was then adjusted in the anterior/posterior and superior/ inferior directions to approximate the ankle axis of rotation, as determined by palpation of the distal tips of the medial and lateral malleoli.12 To ensure that the participant was properly aligned in the device, to precondition the soft tissue, and to minimize potential measurement variability due to previous activity, 25 Nm of external dorsiflexion torque was applied approximately 20 times prior to testing. The rate of external torque application was kept extremely low to replicate the quasi-static testing procedures used clinically6 and to avoid eliciting a stretch reflex. The total time for each torque repetition took up to 5 seconds. Angular measurement was performed using a digital inclinometer (Checkpoint Inc, Torrance, CA), zeroed on the middle third of the tibial crest to provide a consistent anatomical reference point (perpendicular to the crest was used as 0° dorsiflexion) and then mounted on the foot plate. The moment of force (torque) about the ankle joint was controlled by applying 45, 67, 89, and 111 N of force (10, 15, 20, and 25 lb of force) perpendicular to the foot plate with a handheld force gauge (FDK 40; Wagner Instruments,

journal of orthopaedic & sports physical therapy  |  volume 41  |  number 4  |  april 2011  |  275

[

TABLE 1

technical note

]

Intrarater Reliability of Single Versus the Average of 3 Measures and 95% Confidence Intervals for Ankle Dorsiflexion Measured by the 2 Raters 10 Nm

15 Nm

20 Nm

25 Nm

Single measure Rater 1

0.964 (0.859, 0.990)

0.978 (0.883, 0.995)

0.988 (0.898, 0.997)

0.992 (0.957, 0.998)

Rater 2

0.946 (0.775, 0.985)

0.985 (0.882, 0.987)

0.963 (0.906, 0.988)

0.988 (0.946, 0.997)

Average of 3 measures Rater 1

0.988 (0.948, 0.997)

0.992 (0.920, 0.998)

0.996 (0.963, 0.999)

0.997 (0.985, 0.999)

Rater 2

0.981 (0.912, 0.995)

0.985 (0.933, 0.996)

0.987 (0.967, 0.996)

0.996 (0.981, 0.999)

Greenwich, CT) at a distance of 22.5 cm from the axis of rotation. Testing was performed by using one hand to apply the external torque, and placing the other hand on the participant’s lower leg to palpate for any muscle activation and ensure participant relaxation. The sequential application of external dorsiflexion torque from 10 to 25 Nm was considered a single trial. Following 3 trials, with the knee in an extended position, the device was inclined to approximately 20° of knee flexion. Flexing the knee reduced the ability of the bi-articular gastrocnemius to passively resist dorsiflexion of the ankle relative to the unaffected mono-articular soleus muscle. Inclination of the IAROM base alters the orientation of the tibia, which serves as the reference point for calculating ankle position. To account for this change in position, the inclination of the tibial crest was remeasured with the digital inclinometer.

Reliability Testing To evaluate intertester reliability, participants were tested by a physical therapist and a research assistant who was trained in the use of the measurement system but had no previous experience in ROM assessment. Training for the inexperienced tester consisted of a total of approximately 1 hour of data collection, under direction, and discussion of proper alignment and data recording using the device. Same-day reliability testing was performed such that tester order alternated with each consecutive participant to minimize any potential order effect. To assess intratester reliability of single

versus average measures, each tester performed 3 consecutive trials (each including all 4 force levels), first with the knee extended, then repeated the testing with the knee flexed. Testing was then performed by the second tester, after the device was completely removed and reset so that cues could not be taken from prior testing.

Validity Testing For validity testing, ankle dorsiflexion motion was simultaneously evaluated using the IAROM device and an optoelectronic motion analysis system (Optotrak 3020; Northern Digital, Inc, Waterloo, ON, Canada). Three markers on the tibial crest and 3 markers on the foot (fifth metatarsal, dorsum, and lateral heel) were used to generate a rigid-body representation for both segments, to determine the dorsiflexion angle of the foot relative to the tibia. Markers were placed over bony areas, where movement of underlying soft tissue could be minimized, to provide accurate measures of ankle ROM. Simultaneous measurements of ankle dorsiflexion ROM were acquired with the IAROM device and motion analysis system for 3 trials (all 4 force levels) at each of the 2 knee flexion angles. Using motion analysis as a reference, we assessed the convergent validity of our dorsiflexion ROM measures using the IAROM.

Controlling for Bias We controlled for selection bias by recruiting a random selection of patients who were seeking care from a foot and

ankle orthopaedic specialist. Both the digital inclinometer and handheld dynamometer in the IAROM device can be zeroed to a person’s specific reference prior to starting data collection, thus eliminating the introduction of calibration errors or instrument bias. Expectation bias was minimized by having the testers use the device in random order, and by blinding the second tester to the measures obtained by the first tester. Verification bias was minimized by testing the contralateral ankle of patients who were seeking care from an orthopaedic foot and ankle specialist, and by not restricting recruitment to patients with a specific pathology or diagnosis. In addition, bias was minimized through randomization, blinding, prospective design, and consecutive recruitment.

Dependent Variables Ankle dorsiflexion angle values were recorded for each force application level. Stiffness was computed as change in torque divided by change in ankle dorsiflexion angle over the 15-to-25-Nm interval. The 15-to-25-Nm torque interval was selected for calculating stiffness to avoid the more nonlinear “toe” region that exists at lower torque levels.19

Data Analysis Mean, SD, and standard error of the measurement (SEM) were used to summarize dependent variables. SEM was calculated as follows: SEM = SD × (1 – r)1/2, where r is the reliability coefficient. The 90% confidence bound of the minimal detectable change (MDC90) was used

276  |  april 2011  |  volume 41  |  number 4  |  journal of orthopaedic & sports physical therapy



TABLE 2

Intertester Reliability of Ankle Range of Motion (deg) and Stiffness   (Nm/deg) Measurements, With the Knee Extended and Flexed Mean ± SD

ICC2,1 (95% CI)

SEM

MDC90

Knee extended Ankle dorsiflexion at 10 Nm

1.0  8.8

0.924 (0.886, 0.961)

2.4

5.6

Ankle dorsiflexion at 15 Nm

10.5  7.5

0.943 (0.875, 0.967)

1.8

4.2

Ankle dorsiflexion at 20 Nm

16.9  6.8

0.901 (0.861, 0.922)

2.1

5.0

Ankle dorsiflexion at 25 Nm

21.2  6.7

0.943 (0.876, 0.984)

1.6

3.7

0.711 (0.518, 0.852)

0.08

0.19

Stiffness

1.04  0.15

Knee flexed Ankle dorsiflexion at 10 Nm

6.6  7.8

0.904 (0.874, 0.954)

2.4

5.6

Ankle dorsiflexion at 15 Nm

15.6  8.1

0.951 (0.891, 0.970)

1.8

4.2

Ankle dorsiflexion at 20 Nm

21.3  8.5

0.944 (0.901, 0.972)

2.0

4.7

Ankle dorsiflexion at 25 Nm

25.7  8.7

0.952 (0.853, 0.990)

1.9

4.4

0.851 (0.696, 0.921)

0.07

0.16

Stiffness

1.00  0.18

Abbreviations: CI, confidence interval; ICC, intraclass correlation coefficient; MDC, minimum detectable change; SEM, standard error of the measurement.

to examine clinically relevant change in peak dorsiflexion. The MDC90 represents the minimal change that is not due to measurement error.11 The MDC90 was computed as the product of SEM, the z score for the 90% level of confidence, and √2. Intratester reliability of single versus average measures was assessed using intraclass correlation coefficients (ICC2,1 and ICC2,3). To assess convergent validity, ICC2,3 was calculated for comparison between the values obtained with the IAROM device and those determined via motion analysis. To assess intertester reliability, ICC2,1 was calculated for each level of force application. The ability to determine ankle stiffness in a reliable manner was also assessed using ICC2,1. In addition to providing mean data for both ankle ROM and stiffness, rootmean-square (RMS) error was calculated to provide information regarding the absolute magnitude of difference that can be expected between methods of testing or between testers.

RESULTS Reliability

I

ntratester reliability of single versus the average of 3 measures is summarized in TABLE 1 and indicates excellent reliability (greater than 0.94) for

single measures.23 Therefore, single measures were used for subsequent analyses. The mean  SD RMS differences in peak dorsiflexion between testers at 10, 15, 20, and 25 Nm were 2.5°  0.5°, 2.3°  0.5°, 2.7°  0.6°, and 2.5°  0.5° with the knee extended and 2.9°  0.4°, 2.8°  0.4°, 3.4°  0.4°, and 3.7°  0.3° with the knee flexed, respectively. Mean  SD, ICC2,1, SEM, and MDC90 values are presented in TABLE 2, and demonstrate high levels of intertester agreement. Low SDs indicate that ankle joint stiffness was very homogenous within the group of participants tested. Though intertester differences were small for the calculation of ankle joint stiffness, the ICCs (95% confidence intervals) were 0.711 (0.518, 0.852) and 0.851 (0.696, 0.921) for the knee extended and knee flexed, respectively (TABLE 2).

Validity For convergent validity, we noted mean  SD RMS differences in peak dorsiflexion, at 10, 15, 20, and 25 Nm of 1.7°  0.4°, 1.3°  0.3°, 1.4°  0.4°, and 1.7°  0.4° with the knee extended and 2.1°  0.3°, 1.8°  0.4°, 1.5°  0.4°, and 1.5°  0.4° with the knee flexed, respectively, between testing methods. ICC2,3 and 95% confidence intervals for peak dorsiflexion measured with the knee extended and

flexed to approximately 20° are summarized in TABLE 3.

DISCUSSION

T

he IAROM device was developed in an attempt to provide a level of accuracy similar to that of more technologically advanced ankle ROM testing systems, while being more affordable and easier to use. We determined that the IAROM device provides both valid and reliable assessment of ankle dorsiflexion ROM. The IAROM’s design characteristics, which include the application of predetermined levels of external dorsiflexion torque and the ability to monitor foot positioning and the distribution of surface contact along the sole of the foot, likely contributed to the high validity and reliability measures seen in this study. To our knowledge, this is the only ankle passive dorsiflexion ROM device that has been evaluated for both reliability and validity. Preliminary testing with the IAROM device identified a few difficulties in testing certain populations that are likely common to other ROM testing methods. The ability to accurately measure ankle ROM was compromised in participants who were morbidly obese (body mass index greater than 50 kg/m2), as it was not

journal of orthopaedic & sports physical therapy  |  volume 41  |  number 4  |  april 2011  |  277

[

TABLE 3 Trial

technical note

]

Convergent Validity of Peak Dorsiflexion, as Determined Using Optoelectronic Tracking and the Iowa Ankle Range of Motion Device (n = 12)* 10 Nm

15 Nm

20 Nm

25 Nm

Knee extended

0.957 (0.882, 0.973)

0.967 (0.867, 0.997)

0.954 (0.847, 0.983)

0.966 (0.892, 0.979)

Knee flexed

0.971 (0.871, 0.981)

0.968 (0.894, 0.978)

0.967 (0.851, 0.992)

0.977 (0.858, 0.990)

*Values are ICC2,3 and 95% confidence intervals for testing performed with the knee extended and flexed to approximately 20°.

possible to adequately stabilize the leg in these participants. In addition, a small subset of participants initially had difficulty relaxing, as detected with palpation of the dorsiflexor tendons and increased trial-to-trial variability, and required some coaching to ensure the absence of muscle activation. Finally, a few elderly and frail participants, typically with a body mass of less than 55 kg, were unable to tolerate an applied torque of greater than 20 Nm. The validity testing demonstrated the importance of selecting landmarks when defining segments (APPENDIX). Using the crest of the tibia, as opposed to the long axis of the shank, as is typical in goniometric testing, resulted in an offset of approximately 5° between measures, which has been previously observed by Stebbins et al.31 After accounting for this offset, mean values obtained with the knee extended closely matched the goniometric values reported by Kaufman et al,14 as well as the device-measured values reported by Moseley et al.18 The IAROM device provided superior intertester reliability compared to previously reported values (ICCs ranging from 0.50 to 0.80) obtained from goniometric methods of ankle dorsiflexion ROM assessment.1,2,13-15,20,28 Intertester reliability noted with the IAROM device was comparable to that of other computerized systems3,32 and to the weight-bearing lunge methods.4,21 The poor reliability of ankle dorsiflexion assessment is problematic when deciding on a treatment or assessing treatment effects, as it provides the false perception of intervention effectiveness or obscures actual differences. The digital inclinometer has a resolution of 0.1°, and the handheld dynamometer

has a resolution of 0.45 kg (1 lb). Given the high accuracy of the instruments, we anticipated that between-trial variability would constitute a greater source of measurement error. Consistent with this expectation, we noted an overall intertester reliability of 0.96, SEM of 2°, and MDC90 of approximately 5°. SEM values obtained during reliability testing of the IAROM device were less than a fourth of those determined in the previously published goniometric study showing the best intertester reliability,20 which underscores the greater precision with the IAROM. One reason for the better intertester reliability values reported in this study may be the IAROM protocol’s standardized application of external dorsiflexion torque. Such control is likely key to the collection of reliable and clinically relevant measures, as this study data and others18 have shown that a 5-Nm change in torque can result in a 5° to 10° change in measured ankle dorsiflexion ROM. This magnitude of difference can represent as much as 25% of the total motion available in dorsiflexion. The application of predetermined torque levels, as compared to determining the torque required to reach the predetermined dorsiflexion angle, is also more practical for the assessment of a wide variety of patient populations in which ROM may be limited. The ability to determine ankle dorsiflexion ROM through a range of predetermined force levels also has the advantage of documenting ankle stiffness.19,32 Ankle stiffness can be calculated from the dorsiflexion angle change as a function of applied torque. While the stiffness values obtained using the IAROM device are in agreement with those reported in the literature,27,29,30,33

reliability coefficients were not as high and had wider 95% confidence intervals (TABLE 2). For stiffness, we noted low SD and SEM, as well as low ICC values. This counterintuitive combination of low SD and low ICCs may be due to the intrinsic homogeneity of the stiffness measure in our sample.23 Consistent with this contention, the intersubject variability in the stiffness measures was less than 10% in our sample. Perhaps recruiting participants from a population with greater variability in stiffness, such as individuals with diabetes or other clinical populations, would result in improved ICC values. A limitation of this study was that the reliability data were not obtained from ankles with a documented pathology. In our experience, the device has served well in determining ankle ROM limitations and changes in joint stiffness in patients with diabetes.25 Further study in participant groups with differing impairments is needed. As in all other ROM approaches, the measures obtained with the IAROM device may be affected by patients who have foot or ankle pain and/or muscle splinting. However, because one hand is free when obtaining the measures, it is possible to monitor the soft tissue around the ankle to detect soft tissue changes. The acrylic footplate also allows the investigator to monitor changes in foot contact that might be associated with secondary compensations. We, therefore, believe that the reliability demonstrated in this study substantiates the potential value of this approach for assessing ankle ROM; however, as with any other assessment, tester attentiveness is required to obtain valid results.

278  |  april 2011  |  volume 41  |  number 4  |  journal of orthopaedic & sports physical therapy

CONCLUSION

A

nkle dorsiflexion ROM can be reliably and validly measured in clinical or research settings using the IAROM device. The device’s simple and relatively low-tech design allows individuals who have minimal previous knowledge and skill in range-of-motion assessment to use the IAROM device to determine ankle dorsiflexion ROM, with little training, and to do so with a higher level of reliability and confidence than is possible with conventional goniometry. t

11.

12. 13.

14.

15.

16.

REFERENCES 1. A  merican Academy of Orthopaedic Surgeons. Joint Motion: Method of Measuring and Recording. Chicago, IL: AAOS; 1965. 2. American Medical Association. Guides to the Evaluation of Permanent Impairment. 3rd ed. Chicago, IL: American Medical Association; 1990. 3. Assal M, Shofer JB, Rohr E, Price R, Czerniecki J, Sangeorzan BJ. Assessment of an electronic goniometer designed to measure equinus contracture. J Rehabil Res Dev. 2003;40:235-239. 4. Bennell KL, Talbot RC, Wajswelner H, Techovanich W, Kelly DH, Hall AJ. Intra-rater and inter-rater reliability of a weight-bearing lunge measure of ankle dorsiflexion. Aust J Physiother. 1998;44:175-180. 5. Diamond JE, Mueller MJ, Delitto A. Effect of total contact cast immobilization on subtalar and talocrural joint motion in patients with diabetes mellitus. Phys Ther. 1993;73:310-315. 6. Diamond JE, Mueller MJ, Delitto A, Sinacore DR. Reliability of a diabetic foot evaluation. Phys Ther. 1989;69:797-802. 7. Digiovanni CW, Holt S, Czerniecki JM, Ledoux WR, Sangeorzan BJ. Can the presence of equinus contracture be established by physical exam alone? J Rehabil Res Dev. 2001;38:335-340. 8. DiGiovanni CW, Kuo R, Tejwani N, et al. Isolated gastrocnemius tightness. J Bone Joint Surg Am. 2002;84-A:962-970. 9. Elveru RA, Rothstein JM, Lamb RL. Goniometric reliability in a clinical setting. Subtalar and ankle joint measurements. Phys Ther. 1988;68:672-677. 10. Greisberg J, Drake J, Crisco J, DiGiovanni C. The reliability of a new device designed to as-

17.

18.

19.

20.

21.

22.

23.

24.

25.

sess gastrocnemius contracture. Foot Ankle Int. 2002;23:655-660. Haley SM, Fragala-Pinkham MA. Interpreting change scores of tests and measures used in physical therapy. Phys Ther. 2006;86:735-743. Hicks JH. The mechanics of the foot. I. The joints. J Anat. 1953;87:345-357. Hoppenfeld S. Physical Examination of the Spine and Extremities. New York, NY: AppletonCentury-Crofts; 1976. Kaufman KR, Brodine SK, Shaffer RA, Johnson CW, Cullison TR. The effect of foot structure and range of motion on musculoskeletal overuse injuries. Am J Sports Med. 1999;27:585-593. Kendall FP, McCreary EK. Muscles: Testing and Function. 3rd ed. Baltimore, MD: Williams & Wilkins; 1983. Lavery LA, Armstrong DG, Boulton AJ. Ankle equinus deformity and its relationship to high plantar pressure in a large population with diabetes mellitus. J Am Podiatr Med Assoc. 2002;92:479-482. Martin RL, McPoil TG. Reliability of ankle goniometric measurements: a literature review. J Am Podiatr Med Assoc. 2005;95:564-572. Moseley AM, Adams R. Measurement of passive ankle dorsiflexion: procedure and reliability. Aust J Physiother. 1991;37:175-181. Moseley AM, Crosbie J, Adams R. Normative data for passive ankle plantarflexion--dorsiflexion flexibility. Clin Biomech (Bristol, Avon). 2001;16:514-521. Mueller MJ, Diamond JE, Delitto A, Sinacore DR. Insensitivity, limited joint mobility, and plantar ulcers in patients with diabetes mellitus. Phys Ther. 1989;69:453-459; discussion 459-462. Munteanu SE, Strawhorn AB, Landorf KB, Bird AR, Murley GS. A weightbearing technique for the measurement of ankle joint dorsiflexion with the knee extended is reliable. J Sci Med Sport. 2009;12:54-59. http://dx.doi.org/10.1016/j. jsams.2007.06.009 Pandya S, Florence JM, King WM, Robison JD, Oxman M, Province MA. Reliability of goniometric measurements in patients with Duchenne muscular dystrophy. Phys Ther. 1985;65:1339-1342. Portney LG, Watkins MP. Foundations of Clinical Research: Applications to Practice. 2nd ed. Upper Saddle River, NJ: Prentice Hall; 2000. Radford JA, Burns J, Buchbinder R, Landorf KB, Cook C. Does stretching increase ankle dorsiflexion range of motion? A systematic review. Br J Sports Med. 2006;40:870-875; discussion 875. http://dx.doi.org/10.1136/ bjsm.2006.029348 Rao SR, Saltzman CL, Wilken J, Yak HJ. Increased passive ankle stiffness and re-

26.

27.

28.

29.

30.

31.

32.

33.

34.

35.

duced dorsiflexion range of motion in individuals with diabetes mellitus. Foot Ankle Int. 2006;27:617-622. Riddle DL, Pulisic M, Pidcoe P, Johnson RE. Risk factors for plantar fasciitis: a matched case-control study. J Bone Joint Surg Am. 2003;85-A:872-877. Riemann BL, DeMont RG, Ryu K, Lephart SM. The effects of sex, joint angle, and the gastrocnemius muscle on passive ankle joint complex stiffness. J Athl Train. 2001;36:369-375. Rome K, Cowieson F. A reliability study of the universal goniometer, fluid goniometer, and electrogoniometer for the measurement of ankle dorsiflexion. Foot Ankle Int. 1996;17:28-32. Salsich GB, Brown M, Mueller MJ. Relationships between plantar flexor muscle stiffness, strength, and range of motion in subjects with diabetes-peripheral neuropathy compared to age-matched controls. J Orthop Sports Phys Ther. 2000;30:473-483. Salsich GB, Mueller MJ, Sahrmann SA. Passive ankle stiffness in subjects with diabetes and peripheral neuropathy versus an agematched comparison group. Phys Ther. 2000;80:352-362. Stebbins J, Harrington M, Giacomozzi C, Thompson N, Theologis T. Measurement of Foot Kinematics and Plantar Pressure in Children Using the Oxford Foot Model. Bethesday, MD: Proceedings of Pediactric and Adult Foot and Ankle: New Horizons in Clinical Treatment and Innovative Technology; 2003. Trevino SG, Buford WL, Jr., Nakamura T, John Wright A, Patterson RM. Use of a torque-rangeof-motion device for objective differentiation of diabetic from normal feet in adults. Foot Ankle Int. 2004;25:561-567. Vandervoort AA, Chesworth BM, Cunningham DA, Rechnitzer PA, Paterson DH, Koval JJ. An outcome measure to quantify passive stiffness of the ankle. Can J Public Health. 1992;83 Suppl 2:S19-23. Wrobel JS, Armstrong DG. Reliability and validity of current physical examination techniques of the foot and ankle. J Am Podiatr Med Assoc. 2008;98:197-206. Youdas JW, Bogard CL, Suman VJ. Reliability of goniometric measurements and visual estimates of ankle joint active range of motion obtained in a clinical setting. Arch Phys Med Rehabil. 1993;74:1113-1118.

@

MORE INFORMATION WWW.JOSPT.ORG

journal of orthopaedic & sports physical therapy  |  volume 41  |  number 4  |  april 2011  |  279

[

technical note

]

APPENDIX

The standard method for measuring ankle dorsiflexion using a goniometer uses the fibula as a reference for the stationary arm and the plantar aspect of the heel or fifth metatarsal for the moving arm (solid black lines). The Iowa ankle range of motion device uses the tibia crest as a reference (dashed black line), similar to the reference used by Stebbins et al, 31 and the plantar aspect of the foot for the moving arm. The angular difference between using the lateral aspect of the fibula versus the tibial crest as a reference is approximately 5°. The offset does not affect our measurements but helps in interpreting our results relative to those reported in the literature.

This image depicts the Iowa ankle range of motion device setup and the position of the tibial crest, with the knee extended (solid line) and knee flexed (dashed line). Position of the tibia relative to the device does not change. However, change in inclination of the IAROM base flexes the knee by about 20° and also changes inclination of the tibial crest relative to the foot plate.

280  |  april 2011  |  volume 41  |  number 4  |  journal of orthopaedic & sports physical therapy