An Intelligent Decision Support System for the Treatment of Patients Receiving Ventricular Assist Device Support E. C. Karvounis1,2; M. G. Tsipouras1,2; A. T. Tzallas1,2; N. S. Katertsidis1,2; K. Stefanou1,2; Y. Goletsis1,3; M. Frigerio4; A. Verde4; R. Caruso5; B. Meyns6; J. Terrovitis7; M. G. Trivella5; D. I. Fotiadis1,2 1Biomedical
Research Institute-FORTH, Ioannina, Greece; of Medical Technology and Intelligent Information Systems, Dept. of Material Science and Engineering, University of Ioannina, Ioannina, Greece; 3Department of Economics, University of Ioannina, Ioannina, Greece; 4Cardiology 2 – Heart Failure and Transplantation Division, „A. De Gasperis“ Cardiothoracic and Vascular Department, Niguarda Ca’ Granda Hospital, Milan, Italy; 5CNR Institute of Clinical Physiology, Pisa, Italy; 6Department of Cardiac Surgery, University Hospital Leuven, Leuven, Belgium; 73rd Cardiology Department, University of Athens, School of Medicine, Athens, Greece 2Unit
Keywords Heart failure, ventricular assist devices, decision support system
Summary Background: Heart failure (HF) is affecting millions of people every year and it is characterized by impaired ventricular performance, exercise intolerance and shortened life expectancy. Despite significant advancements in drug therapy, mortality of the disease remains excessively high, as heart transplant
Corespondence to: Dimitrios I. Fotiadis Unit of Medical Technology and Intelligent Information Systems Dept. of Material Science and Engineering University of Ioannina PO Box 1186 451 10 Ioannina Greece E-mail:
[email protected]
1. Introduction Heart failure (HF) can be defined as an abnormality of cardiac structure or function leading to failure of the heart to deliver oxygen at a rate commensurate with the requirements of the metabolizing tissues, © Schattauer 2014
remains the gold standard treatment for endstage HF when no contraindications subsist. Traditionally, implanted Ventricular Assist Devices (VADs) have been employed in order to provide circulatory support to patients who cannot survive the waiting time to transplantation, reducing the workload imposed on the heart. In many cases that process could recover its contractility performance. Objectives: The SensorART platform focuses on the management and remote treatment of patients suffering from HF. It provides an inter-
Methods Inf Med 2014; 53: ■–■ doi: 10.3414/ME13-01-0047 received: April 29, 2013 accepted: January 1, 2014 prepublished: ■■■
despite normal filling pressures (or only at the expense of increased filling pressures) [1]. Common causes of heart failure include myocardial infarction and other forms of ischemic heart disease, hypertension, valvular heart disease and cardiomyopathy. It is a common, costly, disabling
operable, extendable and VAD-independent solution, which incorporates various hardware and software components in a holistic approach, in order to improve the quality of the patients’ treatment and the workflow of the specialists. This paper focuses on the description and analysis of Specialist’s Decision Support System (SDSS), an innovative component of the SensorART platform. Methods: The SDSS is a Web-based tool that assists specialists on designing the therapy plan for their patients before and after VAD implantation, analyzing patients’ data, extracting new knowledge, and making informative decisions. Results: SDSS offers support to medical and VAD experts through the different phases of VAD therapy, incorporating several tools covering all related fields; Statistics, Association Rules, Monitoring, Treatment, Weaning, Speed and Suction Detection. Conclusions: SDSS and its modules have been tested in a number of patients and the results are encouraging.
and potentially deadly condition. In developed countries, around 2% of adults suffer from heart failure, but in those over the age of 65, this rate reaches up to 6 –10% [2]. For these reasons, together with the difficulty of having a sufficient number of heart donors, it is recognized that the deMethods Inf Med 2/2014
2
E. C. Karvounis et al.: An Intelligent Decision Support System for the Treatment of Patients Receiving Ventricular ■
vice-based therapeutic approaches assume an increasing and important role in the treatment of HF patients, not only as bridge to transplant, but also as destination therapy. Additionally, in recent years, there is important evidence that patients assisted by Ventricular Assist Devices (VADs) can potentially recover the natural heart functions. VAD is a mechanical pump used to support heart function and blood flow in people who have weakened hearts. In this framework, the SensorART platform focuses on the management and remote treatment of patients suffering from heart failure by offering an interoperable, extendable and VAD-independent (i.e. working with different VAD types/brands) solution, which incorporates hardware and software components in a holistic approach, in order to improve the quality of the patients’ treatment and the workflow of the specialists [3]. In addition, SensorART enables specialists to better understand the patientdevice interactions, and get an insight into new knowledge. The only decision support component of the SensorART platform is the Specialist’s Decision Support System (SDSS) [4]. The SensorART platform in general and SDSS in particular comes to fill-in the gap of data analysis and related decision support in a rather new medical field where although an abundance of data is created, there is no systematic way to store, access and analyse it. Even more, in such patient data, hidden knowledge usually resides; still there are neither available tools to mine such data, nor related databases. Even statistical analysis and hypotheses testing are not supported by specialised tools and researchers have to create manually input files from medical records. Also, there are no systems capable of simultaneous monitoring and analysis of patient measurements. Such a tool could serve as a prognostic/alerting tool in everyday patient management and could significantly improve patient treatment by early identifying patient risk. Patient risk is currently estimated by generalised scoring systems which do not consider individual patient profiles. The VAD performance, customizable throughout pump speed selection, is mainly set empirically following a trialerror procedure without the support of inMethods Inf Med 2/2014
telligent process that uses the patient profile to elaborate the suggested speed. In that framework, as all medical DSSs that use electronic clinical data and provide patient-specific information to clinicians with the aim of reducing medical errors [5], the proposed SDSS offers support towards three main directions: i) Analysis of patient data, hypothesis testing and extraction of hidden associations among patient variables. ii) Support on treatment decisions by providing estimation of adverse events risk and identification of patients that could be considered as candidates for weaning from VAD, and iii) Support on optimal setting of VAD speed based on patient profiles and estimated cardiac output obtained through a specially built VADcirculatory system simulator [3]. Suction detection is also employed towards this direction. In order to create and integrate our SDSS into the clinical practise, the proposed SDSS should act pro-active and produce reminders, warnings, ask questions, analyse trends based on the data of the electronic record and on up-to-date medical knowledge while trained by the outcome of their decisions [6]. The proposed SDSS provides a holistic approach by supporting medical decisions and actions in the whole patient life-cycle. Before VAD implantation, using the SDSS the specialist is able to design the therapy plan by assessing patient status and the risk of adverse events associated to the implantation. This plan could even include the decision whether to implant a VAD or not. After the implantation, the specialist can use the SDSS to get a suggestion on the optimal VAD speed, can monitor patient and VAD parameters and assess the probability for an adverse event the following day. Eventually, the specialist can assess improvements in patient status that could denote adequate cardiac recovery, which in turn could lead to weaning from the VAD. Specialist’s research is also supported though hypotheses testing on a specially designed repository built with heterogeneous data coming multiple sources (laboratories, implantable and wearable sensors). In this paper we present the functionalities of the SDSS and its sub-components and we demonstrate their application on
real patient data. The authors have presented an initial SDSS design in [3]; the current paper describes in detail the final implementation, including evaluation results on real patient datasets. The paper is organized as follows: In Section 2 an extended state of the art of related DSSs and methodologies in the field of VADs is presented. In Section 3 tools and components of the SDSS are presented in detail, while, in Section 4, the obtained results on a real patient dataset are provided. Finally, extensions and concluding remarks on the application of SDSS tools are discussed in Section 5.
2. Related Work Current clinical practice related to treatment support of left VAD (LVAD) patients is characterized by none to limited application of information technology and intelligent decision support systems. In the literature only research works that support individual aspects of the problem have been referred. Analysis of clinical studies often necessitates complex processes and multiple graphical representation of the results. An exploratory data analysis includes a variety of techniques, in order to discover hidden relationships, such as patterns or clusters in a dataset, identification of most important parameters, etc. Most professional software packages (IBM SPSS, Mathematica, SAS, etc) are either only commercially available or hard to use, especially if one aims to generate or customize a huge number of similar graphical outputs. In addition, most statistics programs force users to import data from a database before they can use them, while in many cases there is no direct connection to the database. Association rule mining has been applied in order to extract interesting relations between variables in medical databases. A Priori [7] is a well-known method using support-based pruning to systematically control the exponential growth of candidate itemsets. Doddi et al [8] have investigated the application of association rules in medical data. Hu [9] utilized the association rules technique (using an Im© Schattauer 2014
E. C. Karvounis et al.: An Intelligent Decision Support System for the Treatment of Patients Receiving Ventricular ■
proved A Priori algorithm) to conduct data mining on 285 cases of breast disease patients to create rules between tumor recurrence and other attributes such as age and tumor size. Sriraam et al [10] used a Malaysian hospital dataset in order to decide on kidney dialysis treatment using association rules. Immamura et al [11] utilized an association rules technique on 477 patients in order to investigate the three most useful clinical findings for chronic diseases. Lee et al. [12] have proposed an association rule mining method on 1,247 young Korean adults to extract patterns related to acute myocardial infarction. Still, most commercial DSSs focus mainly on extracting statistical measurements from a given patient database and not on extracting new knowledge. For VAD treated patients, several risk scores and assessment tools have been presented in the literature [13 –30]. Some of them are mentioned here. The Heart Failure Survival Score (HFSS) [13] has been proposed for patient selection for LVAD support based on the estimation for expected survival during the next 1–3 years. It has been developed using statistical likelihood analysis. In the same context, the Seattle Heart Failure Model (SHFM) [14] and the Randomized Evaluation of Mechanical Assistance in Treatment of Chronic Heart Failure trial (REMATCH) [15] have been also employed. The Interagency Registry for Mechanically Assisted Circulatory Support registry (INTERMACS) [16] has been used for patient classification in risk groups, interval analysis [17], and timing of implant assessment [18]. Also, patient classification related to the risk of developing other diseases when undergoing LVAD implantation has been addressed with the Model for End-Stage Liver Disease (MELD) [19]. In addition, the RVF risk score (RVFRS) pre-operative tool [27] and the pre-operative RV stroke work index (RVSWI) [28] have also been presented in the literature. Still, most of the aforementioned studies are based on statistical analysis techniques only, while [25] and [30] employ data mining techniques. Current clinical practice of LVAD patients seems to be lacking advanced visualization tools that can be used for monitoring day-to-day patient data. Such tools can © Schattauer 2014
prove to be significant assistants to patient monitoring, ideally linked to prognostic systems that can predict adverse events. The value of day-to-day data recorded in a patient diary is demonstrated in several studies presented in the literature, for both heart failure patients or for patients suffering from other chronic diseases and conditions. Kirchner et al [32] proposed a heart failure patient monitoring system with implantable defibrillators. White et al. [33] performed diary data analysis for Heart Failure symptom recognition. Hayes et al [34] included walking program diary data to study the effects of exercise training on the quality of life in LVAD patients. As, in some cases, LVAD devices can lead to heart recovery – at least this being the vision for the future- a hard decision to be taken is patient weaning. A few models have been presented in the literature. Santelices et al. [35] proposed a model derived from retrospective medical experience, through a series of structured interviews and questionnaires of 11 members of the multidisciplinary Artificial Heart Program at the University of Pittsburgh medical center. Dandel et al. [36] analyzed data on cardiac morphology and function collected before VAD implantation, echocardiographic parameters recorded during “off-pump” trials, duration of HF before implantation, and stability of recovery before and early after VAD removal. To assess the predictability of post-weaning outcome without heart transplantation or other VAD implantation and to identify risk factors for post-weaning HF recurrence, the authors used the data collected before and during VAD support. Birks et al. [37] proposed a set of minimum criteria with the LVAD at 6000 rpm for 15 minutes for explantation. In the case of parameter improvement, the combination therapy was continued until the maximum improvement had been achieved in each patient. All above weaning models propose a set of simple crisp rules in order to identify weaning candidate patients and the final outcome of the related models is a yes/no suggestion without any quantification of the strength of this suggestion. In the case of Pump Speed Selection, a few sensor-less methods for determining the pump speed were developed by using
pump variables such as current, voltage and speed [38 – 40]. These methods were based on the observation that pressure across, and flow through, an LVAD can be inferred or estimated from the electrical current and frequency of the pump’s motor. Several other researchers have adopted a similar approach without the use of implantable sensors [41– 43]. Control schemes to keep the average pressure across the pump (or between the aorta and the left ventricle) constant have also been developed [44 – 46]. These can provide a pump speed in the safe zone or optimal zone for a certain systemic vascular resistance. Other control approaches, such as using the heart rate as input have been reported [47]. Chen presented an improved method that incorporates the heart rate and the systemic vascular resistance (SVR), and responds to the physiological changes of the body instantaneously, based on the baroreflex and the built in cardiovascular regulation system [48]. McConahy shows the constraints on cardiac output, left atrial pressure, and arterial pressure [49]. However, the proposed methodologies for pump speed selection present significant drawbacks like: they are reliable only in a relatively narrow range of pump variables, some methods do not take into account the change in SVR, etc. Concerning detection of suction events in the available related signals (i.e. pump flow or pump current signals), several approaches have been proposed to evaluate it [50–53]. These approaches are based on the empirical observation of certain variables. Thus, some suction indices are based on time-domain features [54 –57], frequency-domain features [55 –57], and time-frequency-domain features [55 –57]. Among them, there exist methods which extract features from the pump flow signal being one of very few signals that can be easily measured and use powerful pattern recognition algorithms to classify the signal into different pump states [57– 61]. However, in the majority of the above described approaches, a large number of features are employed. Also, suction detection algorithms are used only in LVAD controllers [62].
Methods Inf Med 2/2014
3
4
E. C. Karvounis et al.: An Intelligent Decision Support System for the Treatment of Patients Receiving Ventricular ■
3. Material and Methods The SensorART platform continuously monitors and evaluates patient-device interactions and optimizes the heart unloading and support. In this direction, the proposed SDSS enhances the SensorART platform with advanced tools, thus enabling the realization of an “intelligent” VAD implementation. By facilitating such an approach, the SDSS aims at providing a better understanding of the patient condition and progress in order for the specialists to personalize and optimize the corresponding treatment support, determine the capacity of the natural heart to develop major or minor delivery according to the assisting time, identify risk factors and recovery mechanisms in order to recognize potential outcomes etc. ▶Figure 1 presents the role of the SDSS within the overall SensorART platform. The SDSS resides of the server side, accesses the SensorART repository where patient data are continuously stored, and communicates with the VAD Simulation
Platform (in order to acquire customized simulation sessions). The SensorART repository itself is EUROMACS (www. euromacs.org) compliant so as to ensure interoperability; a special wrapper application ensures HL7 compatibility, as well. Data are generated on the client side either directly by the VAD itself, by implantable sensors and by wearable sensors (Continua alliance certified). An Autoregulation Unit used for VAD control and data collection, manages the data which, in turn, are sent through a portable device (e.g. a tablet) (running the Patients’ Monitoring Application) to the data repository. The SDSS and the Specialist’s Monitoring Application access this repository. They form a resourceful environment that provides valuable outcomes (patient condition /parameters, data analyses, risk factors, suggestions, estimations, etc.) to the specialists (i.e. clinicians, researchers, etc.). The SDSS offers support to medical and VAD experts through seven tools for Association Rules extraction, Statistics analysis, Treatment decision Support, Monitoring, Weaning
decision support, VAD Speed Selection and Suction Detection respectively. The SDSS tools communicate with the following system parts: i) the SDSS User Interface that includes a set of pages/forms and controls allowing advanced interactivity with the users of the SDSS, ii) the VAD Simulation Platform providing a hybrid ventricular and circulatory model allowing specialists to simulate the behaviour of a patient’s circulatory system using different VAD types and functional parameters [63], iii) an R environment for statistical computing, and most important iv) the SensorART repository that provides storage and retrieval functionalities for all the data used in the SensorART platform and was developed to support the management of data that are used by the corresponding tools. The SersorART repository was designed and implemented with the following data entities: users, user groups, alerts and notifications, sensor measurements, VAD measurements, laboratory measurements, patient profiles etc. Various screenshots of the platform are presented in ▶ Figure 2
Figure 1 A general overview of the SensorART Platform
Methods Inf Med 2/2014
© Schattauer 2014
E. C. Karvounis et al.: An Intelligent Decision Support System for the Treatment of Patients Receiving Ventricular ■
while all seven SDSS tools are described in detail in the following paragraphs.
3.1 Association Rules Tool Using the Association Rules Tool, the specialist can extract association rules, linking different patient variables based on patient data over a specific observation period. In particular, the Specialist can be assisted in analysis and research, by using data mining techniques in order to discover associations among different variables and discover new knowledge. Associations are in the form of Rules: IF (variable 1 > value 1) AND (variable 2 value 3) AND …(variable n > value n). By using this tool, various hypotheses on relationships among variables can be now examined (if confirmed) by existing data. The specialist can select the variables that will be included in the IF and the THEN parts, the patients that will be included and the time period (start and end date) of the analysis. Furthermore, he can filter the extracted rules by setting support and confidence thresholds. Rules together with their confidence and support values are presented to the specialist. Moreover, in order to acquire as much meaningful results as possible, a double filtering approach is followed by (i) ignoring features that have a high percentage of missing values in the database and (ii) ignoring records for which a high number of feature values is missing. The specialist can define the relevant thresholds. The A Priori algorithm [7] was chosen as the rule mining algorithm, because of its robustness and ability to solve large problems using small computational power and time. The A Priori algorithm attempts to find subsets which are common to at least a minimum number of the itemsets, given an initial set of itemsets. The algorithm can find associations between different sets of data and the sets of rules (the algorithm’s outcome) implies how often items are contained in sets of data. It should be mentioned that, although several different approaches have been presented in the literature, A Priori remains the benchmark algorithm of the association rule mining field, while most of the other algorithms improve © Schattauer 2014
the computational time for mining the association rules from the dataset, still mainly producing the same set of rules [7]. The A Priori algorithm in general uses breadth-first search and a Hash tree structure to keep track of the candidate item sets. Then, it generates these candidate item sets of length N from the initial itemsets of length N – 1. Afterwards, the candidates which have an infrequent sub pattern are pruned. Finally, the ultimate candidate set contains all frequent-length itemsets. The problem of association rule mining is defined as: Let I = i1, i2 , ... , id be the set of all items in a dataset, and T = t1, t2 , ... , tN be the set of all transactions. Each transaction has a unique identifier and contains a set of items, called an itemset. An itemset with k items is called a k-itemset. An important property of an itemset is its support count, which refers to the number of transactions which contain this particular itemset. An association rule is a conditional implication among itemsets, of the form X Þ Y where X and Y are disjoint itemsets, i.e. X Ç Y = Æ. The strength of the rule can be measured in terms of its support and confidence. The rule has support s, if s % of transactions include all the items in both X and Y, and confidence c if c % of transactions containing also contain Y. Support is an important measure, because a rule that has very low support may occur simply by chance. Confidence, on the other hand, measures the reliability of the inference made by a rule. Hence, the data mining task for association rules can be broken into two steps. The first step consists of finding all frequent itemsets, i.e. itemsets that occur in the database with a certain user specified frequency, called minimum support. The second step consists of forming the rules among the frequent itemsets. Initially, the data and the assigned support and confidence values are read. Then, the data are preprocessed by a parsing tool, and the preprocessed data are used as input to the A Priori algorithm and exports the results to a file. The parsing tool operates as follows: taking into account the percentage of the non-missing values that the user has filled in, the algorithm counts the number of missing values per variable and if the sum of the missing values exceeds the de-
sired user percentage, then the whole variable is not taken into account when the A Priori rule mining algorithm runs, producing results only for the variables that fulfill the user defined criteria.
3.2 Statistics Tool The multitude of data generated in the medical field requires the adoption of intelligent analysis techniques that allows specialists to summarize and present their knowledge, get insight into the data, test hypotheses, draw conclusions and directly interact with all the available information. Thus, the Statistics Tool provides specialists with instruments for: i) Analyzing and interpreting large patient data directly from the SensorART repository through powerful statistical techniques and Kaplan-Meier analysis, ii) Examining the results of previous therapeutic regimens and obtaining quantitative explanations of the observations through relevant statistical tests and iii) Generating efficient reports with intelligent data visualization. Using the tool the medical expert has multiple options, including the selection of population of interest, to work only with data from specific visits, to choose the variables of interest (e.g. heart related, sensor related, laboratory related) and the method of interest from a complete set of basic and advanced features covering both clinical and research needs of clinical partners (basic statisticsmean, variance, std, quantile, length, paired T-test, unpaired T-test, F-test, χ2 tests and Kaplan-Meier analysis). The Statistics Tool is based on the R environment (http://www.r-project.org) for its back-end functionalities except for the Kaplan-Maier component. The latter allows specialists to perform survival analysis and generate a Kaplan-Meier plot. In clinical studies, the researchers are usually interested in the time until the patients in a study present a specific event or endpoint. In the context of SensorART, these were defined as death, cerebral bleeding, gastrointestinal bleeding, ischemic stroke, TIA (transient ischemic attack) and thromboembolic events. The Kaplan-Meier tool is based on a standard algorithm for determing, directly from the SensorART repositories, the fraction of patients living Methods Inf Med 2/2014
5
6
E. C. Karvounis et al.: An Intelligent Decision Support System for the Treatment of Patients Receiving Ventricular ■
Figure 2
Screenshots from (a) Association Rules Tool, (b) Statistics Tool, (c) Treatment Support Tool and (d) Weaning Tool
Methods Inf Med 2/2014
© Schattauer 2014
E. C. Karvounis et al.: An Intelligent Decision Support System for the Treatment of Patients Receiving Ventricular ■
up to the occurrence of a specific event. The starting time is always considered the time of VAD implantation and thus the start of treatment.
3.3 Treatment Support Tool The Treatment Support Tool supports the specialists on the most suitable treatment plan according to the condition/phase of the patients (stabilized clinical state, normal phase in home conditions, worsening phase and/or reacutisation phase). Its functionality includes the calculation of several acknowledged risk scores along with the prediction of alternative treatments outcome with respect to adverse events appearance. More specifically, the Treatment Support Tool provides two functionalities, calculation of known risk scores and treatment prediction based on risk for adverse event appearance. In the first case, several known scores of survival are incorporated into the tool to allow rapid decisions and to foresee possible complications after VAD implant. All risk scores are calculated for similar and partially overlapped – but not identical – objectives including prediction of: i) Expected survival/the risk of death on medical therapy, ii) Expected survival/the risk of death after LVAD implantation and iii) Probability of specific complications (e.g. right ventricular failure) after LVAD implantation. Unfortunately, some risk factors for death without operation are also associated with worse postoperative survival and/or higher probability of complications, making difficult to define the risk/benefit profile for individual patients. In this context, four risk scores are included in the Treatment Support Tool: • the Heart Failure Survival Score (HFSS), which provides an estimation of expected 1-year survival [13], • the Seattle Heart Failure Model (SHFM), which provides an estimation of expected 1-, 2- and 3-year patient survival [14], • the Model for End-Stage Liver Disease (MELD), which estimates risk fro multisystem malfunctions and post-operative complications [19],
© Schattauer 2014
•
the RVF risk score (RVFRS), which estimates risk for right ventricular failure [27].
Concerning risk prediction based on adverse events appearance, this modality assesses the risk of adverse events in the case of LVAD implantation. It has been developed by applying machine learning techniques in an annotated dataset. In model building, several widely known classification methodologies have been tested: Naïve Bayes classifier (NB), k-nearest neighbor (kNN), Decision trees (DT), Random forests (RF), Multilayer perceptron (MLP) neural networks and Support Vector Machines (SVMs), using commonly used parameters, proposed in the literature: • NB classifier was implemented by modeling numeric values using normal distributions. • kNN was implemented with k = 3 and based on the Euclidean distance. • DT were implemented using the C4.5 algorithm, using the pessimistic error rate based method (sub-tree replacement) for pruning, with confidence factor 0.25 and 2 as minimum instances in a leaf. • MLP architecture was implemented with 1 hidden layer, 0.3 learning rate and 500 maximum number of epochs. • RFs were based on 10 decision trees. • SVM was implemented with polynomial kernel. Based on the obtained results (presented in Section 4) and the fact that the medical experts preferred a mechanism that can offer interpretability of the classification outcome, the DT have been finally selected.
3.4 Monitoring Tool The Monitoring Tool offers monitoring of day-to-day LVAD and patient parameters and their association with the appearance of specific adverse events. The monitoring parameters used are: Pump flow, Pump speed, Pump index, Pump power, Temperature, Systolic blood pressure, Diastolic blood pressure, Pulses, Weight, INR and Anticoagulant treatment. Those parameters can come directly from the patient or,
in the future implementation of the SensorART platform, can be automatically obtained through a set of implantable and other sensors. In addition, several adverse events are also recorded including bleeding, arrhythmia, heart failure, thromboembolism (major), thromboembolism (minor) and pump thrombosis. The patterns that were used to develop the prognostic model were extracted from the dataset. Again, NB classifier, kNN, DT, RF and MLP classifiers were tested, while DT have been finally selected based on the obtained results (Section 4) and the fact that the medical experts preferred the idea of a transparent decision mechanism against a black-box approach (such as e.g. in the case of MLPs).
3.5 Weaning Tool The Weaning Tool can be used to identify the most appropriate candidates for weaning from the VAD support. As already mentioned in Section 2, all proposed approaches reported in the literature for supporting optimal ventricular assist device weaning make use only of strict rules. In our proposed tool, a Fuzzy Knowledge Subsystem is incorporated in order mainly to be more flexible on the decision boundaries and closer to the human logic compared to classical binary (crisp) logic. More specifically, the Weaning tool enables specialists to create and modify expert rules for weaning, in the form of comprehensive and personalized IF-THEN rules. The tool combines expert knowledge with fuzzy analysis, in order to support the specialists on the weaning decision, i.e. the selection of patients with adequate cardiac recovery that may be removed from the VAD therapy. The tool is based on the crisp engine and the fuzzy knowledge subsystem [64], in order to create initial knowledgebased models and then transform them into fuzzy models. All patient data, such as patient information (e.g. demographics, medical history, medication), VAD parameters, sensor measurements, laboratory measurements and clinical evaluations, are used as inputs into the fuzzy models, in order to provide the status of each patient (candidate for weaning or not). Methods Inf Med 2/2014
7
8
E. C. Karvounis et al.: An Intelligent Decision Support System for the Treatment of Patients Receiving Ventricular ■
(3) Symbol “˅” denotes the binary OR operator. Based on the above, a crisp model can be defined as: (4)
Figure 3
Concerning the Fuzzy Knowledge Subsystem, being the most innovative part of the Weaning tool, the main idea is to produce a fuzzy model for a specific domain of application based on initial knowledge for this area provided by domain experts. This knowledge is formatted in a set of initial crisp rules, i.e. a collection of “IF …THEN …” rules. More specifically, each crisp classification rule ric (x, θic) is expressed as: ric (x, θic): dic (x, θic) → yi , where ric is the i-th crisp rule, x is a feature vector comprised from a number of features aj , θic is a vector of thresholds, dic is the crisp rule’s precondition, containing a conjunction of feature tests, and yi is the predicted class. Precondition is defined as:
Figure 4 tool [63]
(1)
where tion, with
is the crisp membership funcand
opj = {=, ≠, , ≤, ≥} and defined as can be seen in ▶Figure 3. The symbol “˄” denotes the binary AND operator. All crisp rules from the medical rule set that have as consequent the same class are combined in a single class rule, defined as:
where M c is the crisp model, Θc is the threshold vector containing all thresholds used in the model, Φc is a decision function and N is the number of classes in the problem, and thus the number of class rules Rc. The crisp model Mc can be transformed into an equivalent fuzzy model M f as follows: The crisp membership function g c is replaced by a fuzzy one g f. The binary AND and OR operators are replaced by Tnorm and Snorm functions, respectively. The decision function Φc is replaced by a defuzzyfication function Φ f. Based on these changes, the fuzzy model is defined as: (5) where, θ f is the parameter vector containing all parameters used in the fuzzy model and Rk f is the fuzzy class rules, defined as:
(a) Speed selection flowchart, (b) Internal connectivity of the VAD-Heart Simulation platform with the suction detection and speed selection
Methods Inf Med 2/2014
© Schattauer 2014
E. C. Karvounis et al.: An Intelligent Decision Support System for the Treatment of Patients Receiving Ventricular ■ Table 1
(6) where ri is the fuzzy rule defined as: f
[7] Mf
The obtained fuzzy model is optimized with respect to Θ f parameters. For this purpose an objective function must be defined using a training dataset (such as the square error function), and then minimized with respect to a Θ f using a local or global optimization technique. Local optimization techniques start from an initial point and result to the respective local minimum while global optimization techniques attempt to find all local minima (which are potentially global) of the objective function inside a bounded set. The described methodology for fuzzy expert systems creation is based on the fuzzyfication of an initial (crisp) medical rule set. Thus, the final fuzzy expert system is the initial rule set with more “flexible” boundaries and optimized based on a training set. The methodology can be applied to any given rule set. In the framework of the Weaning Tool, the crisp and fuzzy rule engines have been developed for two relevant models in the literature i.e. the Dandel et al. [36] set of rules and Birks et al. [37] flow chart. Although, the original versions of the models include only crisp rules, fuzzy models have also been developed since the definition and parameter set of the fuzzy model allows it to be extremely more flexible and thus being able to cope with the complexity of the respective medical decision.
3.6 Speed Selection Tool Setting the pump speed of the VAD is an important parameter in order to achieve the optimal cardiac output, and ensure the patient’s quality of life. In a critical care setting, the desired operating point of the VAD may be determined by a specialist or technical personnel and adjusted to provide more or less cardiac output depending on the status of the patient. However, when © Schattauer 2014
No
Features in the dataset for the treatment support tool
Feature
mean ± std
No
Feature
mean ± std
1
Age
54.16 ± 10.65
13
Reversible PH
1 (median)
2
INTERMACS profile
2 (median)
14
Heart Rate
84.88 ± 11.09
15
PA
98.96 ± 13
3
Platelets
267.04 ± 119.5
16
International Normalized Ratio
1.46 ± 0.69
4
Hemoglobin
12.38 ± 1.5
17
Bilirubine
1.28 ± 1.17
5
Hematocrit
37.76 ± 4.84
18
Creatinine
1.18 ± 0.44
6
White blood cells
9.66 ± 4.86
19
Urea
267.04 ± 119.5
7
Right atrial Pressure
7.58 ± 5.33
20
Na+
138.42 ± 7.98
8
Pulmonary Artery Pressure – max
51.76 ± 18.18
21
Model for End-Stage Liver Disease (MELD)
10.72 ± 5.62
9
Pulmonary Artery Pressure – min
23.14 ± 8.66
22
MELD UNOS
12.91 ± 4.18
10
Pulmonary Artery Pressure – mean
35.5 ± 10.75
23
MELD U+Age
62.85 ± 12.56
11
Pulmonary Capillary Wedge
25.71 ± 9.32
24
Aspartate Anotransferase
49.9 ± 96.45
12
Cardiac Index
1.78 ± 0.47
25
Inotropes
0.63 ± 0.87
the patient leaves the critical care setting, a clinician is no longer readily available and the device must provide adequate cardiac output to sustain the patient’s level of activity without clinical supervision. In addition, as the patient recovers and regains his strength, a change in the VAD speed might be required. Cardiac output can be increased by increasing the pump speed. However, two important constraints should be taken into account regarding the pump speed. First, if the speed is too low, blood may regurgitate from the aorta to the left ventricle through the pump (backflow). Second, if the pump speed is too high, the pump will attempt to draw more blood from the ventricle than is available, resulting in the suction phenomenon. A solution to maximize cardiac output while operating the pump at a safe speed is to operate the pump at a speed just below suction. The drawback to this solution is that a pump speed that does not cause suction may still have adverse effects on other physiological parameters. In this framework, the proposed Speed Selection Tool has been developed enabling the specialists to analyze simulation sessions from the VAD Simulation plat-
form, investigate the potential effect on important hemodynamic variables (such as cardiac output and arterial pressure) and determine a pump speed that provides adequate cardiac output for the patient to maintain his current level of activity. In this framework, we designed a flowchart (▶ Figure 4a) for the speed selection process currently followed after the operation. The Left Atrial Pressure (LAP) check is substitute for echo examination, which is used to check if the aortic valve is opening correctly. As shown above in the flowchart, the automated identification of VAD pump speed depends on the automated detection of suction events in the pump flow signal. In this framework, a tool has been developed (Suction Detection Tool), connected with the Simulation Platform, in order to provide to the specialists a powerful assistant in their attempt to analyze data from simulation sessions, identify different pump states and possible issues regarding to the suction phenomenon. The process is shown graphically in ▶ Figure 4b. The Suction Detection Tool processes simulation sessions from the Simulation Platform, allowing specialists to simulate the beMethods Inf Med 2/2014
9
10
E. C. Karvounis et al.: An Intelligent Decision Support System for the Treatment of Patients Receiving Ventricular ■ Table 2 Treatment Support Tool Classification Results for Sensitivity (Sens), Positive Predictive Value (PPV) and Accuracy (Acc), for Naïve Bayes (NB), k-nearest neighbor (kNN), Decision trees (DT), Random forests (RF), Multilayer perceptron (MLP) neural networks and Support Vector Machines (SVMs) classifiers (a)
Initial dataset (D1)
(b)
Initial dataset with replaced missing values (D2)
Sens (%)
PPV (%)
Acc (%)
Sens (%)
PPV (%)
Acc (%)
NB
37.66
36.89
67.35
NB
38.61
38.89
69.39
kNN
32.9
43.15
57.14
kNN
38.79
41.89
65.31
DT
35.58
35.14
67.35
DT
44.68
46.19
71.43
RF
30.48
23.19
65.31
RF
31.43
23.4
67.35
MLP
32.9
31.5
57.14
MLP
33.85
32.41
59.18
SVM
44.85
46.46
69.39
SVM
45.8
48.43
73.47
(c)
Resampled dataset (D3)
(d)
Resampled dataset with replaced missing values (D4)
Sens (%)
PPV (%)
Acc (%)
Sens (%)
PPV (%)
Acc (%)
NB
85.85
86.19
85.71
NB
83.63
84.37
83.67
kNN
69.59
69.24
67.35
kNN
78.36
77.08
77.55
DT
79.65
RF
84.91
79.41
79.59
DT
93.8
93.78
93.88
87.34
85.71
RF
81.87
81,89
81.63
MLP
80.12
80.12
79.59
MLP
81.87
82.19
81.63
SVM
87.6
88.35
87.76
SVM
93.8
93.78
93.88
haviour of a patient’s circulatory system, using different VAD types and functional parameters in order to be supported to the speed decision.
3.7 Suction Detection Tool Initially, an interconnected upload mechanism allows the submission of simulation sessions from the VAD Simulation Platform. The dataset requested is composed of cardiac and circulatory variables, and other parameters important to obtain the specific object’s (patient or animal) patho-physiological condition. The operator, adjusting the circulatory model parameters is able to obtain the desired hemodynamic status. Some of the data transmitted from SDSS to VHSP are: patient’s weight, cardiac output, mean central venous pressure, duration of simulation etc, selected according to clinical partner’s suggestions. Next, the simulated pump flow signal is returned back to SDSS, together with the following traces: Left Atrial Pressure, Left Ventricular pressure, Aortic Pressure, VAD Flow, heart rate, Total Cardiac Output and pump speed in RPM. Methods Inf Med 2/2014
A new suction detector algorithm has been developed based on the detection of the sudden decreases in signal’s baseline. The methodology is based on online estimation of a Gaussian Mixture Model (GMM) with two mixtures corresponding to non-suction & suction classes. More specifically, the proposed methodology is consisted of three steps: i) signal windowing, ii) GMM based classification and iii) GMM parameter adaptation. In our GMM model the only time varying parameter that was selected is the mean of the non-suction mixture [64]. Reducing degrees of freedom in the model ensures higher reliability.
4. Datasets and Results In order to evaluate each proposed tool, related datasets were used. In the case of Treatment Support Tool and especially of the risk of adverse events prediction, sample data from 49 patients have been used. The variables recorded are mentioned in ▶ Table 1. The re-hospitalizations for all patients for the first year and follow-up data have
been used to determine the occurrence of adverse events: among 49 patients treated with VADs; 35 had no relevant adverse events; 3 had bleeding episodes and 11 died. The dataset have been obtained from the Heart Failure and Transplantation Division, „A. De Gasperis“ Cardiothoracic & vascular Department, Niguarda Ca’ Granda Hospital, Milan, Italy. Although being very common, infections of the entry site of the driveline were not considered as relevant adverse events, and thus were excluded from the study. The dataset included several missing values thus replacement of missing values has been applied using the 3-Nearest Neighbors technique. As the number of the prototypes per category is unbalanced, resampling from the normal distribution of the minority class procedure has been employed, i.e. a normal distribution was calculated for each feature, using all patterns belonging to the same class (the minority class that needs to be resampled). Then, N random values were generated from each calculated distribution, with N being the number of additional samples for the class, and thus N additional patterns belonging to this class were generated. In addition, a feature selec© Schattauer 2014
E. C. Karvounis et al.: An Intelligent Decision Support System for the Treatment of Patients Receiving Ventricular ■ Table 3 Monitoring Tool Classification Results for Sensitivity (Sens), Positive Predictive Value (PPV) and Accuracy (Acc), for Naïve Bayes (NB), k-nearest neighbor (kNN), Decision trees (DT), Random forests (RF), Multilayer perceptron (MLP) neural networks and Support Vector Machines (SVMs) classifiers (a)
Initial dataset (D1)
(b)
Resampled dataset (D2)
Sens (%)
PPV (%)
Acc (%)
Sens (%)
PPV (%)
Acc (%)
NB
99.9
91.7
99.9
NB
100
100
100
kNN
98.9
41.4
96.8
kNN
99.1
98.4
98.6
DT
99.7
59.1
99.1
DT
99.5
99
99.2
RF
99.8
80.9
99.6
RF
99.9
99.8
99.8
MLP
99.7
52.7
99.1
MLP
99.5
99.2
99.3
tion, using exhaustive search and Chisquare statistics, was applied in order to select the most informative features for this problem, resulting to the subset of features: Hemoglobin , Hematocrit, Right Atrial Pressure, cardiac Index, International Normalized Ratio and Aspartate Anotransferase. As already mentioned in Section 2, NB, 3NN, DT using the C4.5 algorithm, MLP neural networks and RF classification methodologies have been tested. Evaluation was performed for i) the initial dataset (D1), ii) the initial dataset with replaced missing values (D2), iii) the resampled dataset (D3) and iv) resampled dataset with replaced missing values (D4). The leaveone-patient-out cross-validation technique was employed. The respective confusion matrices (cmii) were obtained, while the metrics of classification accuracy (Acc), sensitivity for each class (Sensi) and average positive predictive value (PPV) for each class (PPVi) for each class, are calculated:
(8)
Based on the above, the average sensitivity for all classes (Sens) and average positive predictive value (PPV) for all classes are calculated. All results are presented in ▶Table 2. © Schattauer 2014
In order to develop the prognostic procedure that assesses the risk of an adverse event appearance in the next day (Monitoring Tool), based on the recorded data for the last three days, data from six patients were included. The dataset have been obtained from the Department of Cardiac Surgery, University Hospital Leuven, Leuven, Belgium. The patterns that were used to develop the prognostic model were extracted from this dataset. In order to extract a pattern, four consecutive days should exist in the recording and in this case the data from the three days are used as features and the event from the fourth day as the annotation. Thus, the dataset is formulated as: D = {X, T }where X being a 33xN matrix with each row being one data pattern of the dataset with 33 features (i.e. the 11 variables for three consecutive days), and T being the annotation of the respective pattern. Using this procedure, a dataset with 1026 patterns has been created, including patterns covering three classes, being normal (1022), ventricular tachycardia (3) and heart failure (1). Initially, since the number of the patterns per category is unbalanced, the same resampling procedure (resampling from the normal distribution of the minority class) was applied to create a balanced dataset. The classification methodologies that were employed include NB classifier, kNN, C4.5, RF and MLP neural networks, using the parameters already mentioned previously. Evaluation was performed using: i) the 10-fold stratified cross validation method and ii) the initial dataset (before the resampling) and the respective confusion matrices were obtained, while metrics such as classification accuracy and sensitivity/positive predictive value per
class are calculated. The obtained results are shown in ▶ Table 3. In the framework of the Weaning Tool and the optimization of the fuzzy model, data from 11 patients are used, collected from the 3rd Cardiology Department, School of Medicine of the University of Athens in Greece from 2004 to 2009. Almost 70 parameters are recorded. Four (4) of the patients were successfully weaned from VAD. Recordings were made in three main patient conditions, i) on pump, ii) off pump for 5 mins, and iii) off pump for 15 min. The set of rules presented by Birks et al. [37] were used as input to the above described methodology and the results are presented using the above described dataset. The on-pump data and the off pump for 5 mins data were used for optimization. These included 9 patterns for weaning and 23 patterns for non-weaning. The training dataset is formulated as: Dmax = {X, T} where X being an 6x32 matrix with each row being one data pattern of the dataset (i.e. Xi = x, with i = 1, ... , 32 ), and T being an 2x32 matrix with each row being the annotation of the respective pattern (i.e. T i = t with t = 1, ... , 32). A target vector for a specific data pattern (t) is defined as: (9) The optimization function is based on the sum of square errors function: (10) A local optimization technique has been used. As mentioned above, by setting: θi,b f = θi c, θi,a f = 1 it is: M f(x, Θ f ) → Methods Inf Med 2/2014
11
12
E. C. Karvounis et al.: An Intelligent Decision Support System for the Treatment of Patients Receiving Ventricular ■
Table 4
Weaning Tool Confusion Matrices for the Initial (Θfmax) and the Optimized (Θf*) Fuzzy Models
Initial fuzzy model Θfinitial
Optimized fuzzy model Θf*
annotation wean
fuzzy model
no wean
wean
4
0
no wean
6
22
M c(x, Θc). Thus, the initial starting point Θ finitial is defined by setting θi,a f = 2 and θi,b f = θi c, i.e.: Θ finitial = {2,60,2,50,2,45,2,12,2,2.8,2,16} In order to restrain the local optimization from drifting far from the Θ finitial (and thus maintaining the properties of the initial medical rule set) upper and lower bound for each one of the optimization parameters (Θ f ) are set so as to limit the optimization procedure inside this search area. These bounds are set as: θi,a f Î [0,5] and θi,b f Î [0.9θi c, 1.1θi c]. The optimization procedure results to a local minimum: Θ f * = {0.5, 59, 1, 51, 0.17, 45, 0.5, 13.5, 0.5, 2.8, 2.5, 16}. For Θ finitial and Θ f * the obtained confusion matrices are presented in ▶ Table 4 while square error, sensitivity, specificity and classification accuracy results in ▶ Table 5.
Table 5 Weaning Tool Classification Results for Sensitivity (Sens), Positive Predictive Value (PPV) and Accuracy (Acc), for the Initial (Θfmax) and the Optimized (Θf*) Fuzzy Models Θfinitial
Θf*
Square error
10.87
4.22
Sensitivity (%)
40
100
Specificity (%)
100
100
Classification accuracy (%) 81.25
100
wean fuzzy model
wean no wean
Finally, concerning the Suction Detection Tool, two different datasets, provided by the Institute of Biocybernetics and Biomedical Engineering of Polish Academy of Sciences (IBBE-PAS) and annotated be medical experts, are used in order to test our methodology: i. 10 pump flow signals with suction events approximately 46 minutes in total duration are collected from the VAD Simulation Platform which enables the specialists to simulate the behaviour of a patient’s circulatory system with connected a real assist device (e.g. nonpulsatile blood pump) [64]. ii. 26 pump flow signals approximately 20 hours are produced from a numerical simulator, simulating different medical cases with predefined pathologies. A large number of medical realistic cases with patient pathologies were defined and determined by medical doctors [64]. The confusion matrices, sensitivities, specificities and the total accuracy are given in ▶Table 6 and ▶Table ■ for the two datasets, respectively.
5. Discussion In recent years, the use of VADs for the treatment of end stage heart failure has
Table 6 Results from 10 signals with approximately 46 minutes (dataset I) and from 26 signals with approximately 20 hours (dataset II), where GMM online estimation was applied [63]. DATASET I
DATASET II
GMM No Suction
GMM Suction
GMM No Suction
GMM Suction
No-Suction
1904
71
63547
1568
Suction
117
510
3480
4555
Sensitivity
0.96
0.81
0.98
0.57
Specificity
0.94
0.88
0.95
0.74
Accuracy
Methods Inf Med 2/2014
0.93
annotation
0.93
no wean
10
0
0
22
steadily increased. This widespread application has increased research in the field, however, based on the literature review, decision support systems concerning VAD patients have not been proposed over the past years; only research works that support individually aspects of the problem have been presented. In this context, the SDSS is innovative, supporting medical and VAD experts through the different phases of VAD therapy. This paper focuses on the presentation of the SDSS, the main decision support component of the SensorART system. SDSS assists specialists on designing the best treatment plan for their patients before and after VAD implantation, analysing patients’ data, testing hypotheses, extracting new knowledge and making informative decisions. The proposed SDSS Tools have been integrated in a unified environment, covering all the specialist’s requirements and providing a unique selection of features. More specifically, all Tools are web-based, accessible through an intuitive graphical interface allowing secured access to available data. In addition, different types of data and detailed reports are provided to the medical experts from the majority of the Tools. More important, the proposed system doesn’t address just a single issue but assists the specialists on remote monitoring and VAD control and receive information on patient status. It enables the design of the most proper treatment plan for every patient before and after VAD implantation and enables the analysis of patients’ data for the extraction of new knowledge and making informative decisions. Finally, the SDSS enables the potential to translate valuable expert knowledge into standardized, personalized and optimized VAD therapy. Summarizing, the proposed system presents several advantages: © Schattauer 2014
E. C. Karvounis et al.: An Intelligent Decision Support System for the Treatment of Patients Receiving Ventricular ■
•
•
•
It includes tools covering all aspects of VAD patients’ management, thus providing an overall VAD patient management system and decision support mechanism and not just a problemspecific technique. It follows a holistic approach, including all models presented in the literature along with user-defined models, thus providing a “mixture-of-experts” approach. It provides an automated and easy-touse platform that can be operated using either data from the database or by data-input forms.
Besides the advantages derived from the overall system, each individual tool of the SDSS is either innovative or presents advantages and novelties compared to other approaches proposed in the literature. • The knowledge discovery tools (Association Rules Tool and Statistics Tool) allow the medical expert to extract new knowledge and data associations or analyze the data using statistical techniques. Such an analysis has neither presented as research in the literature (data mining on VAD data) nor publicly offered as an automated tool. • In the same manner, currently there is no other research in the literature that monitors LVAD patients using day-today data and employs these data for automated prognosis of adverse events. Thus, the Monitoring Tool constitutes one of the innovations of the SDSS, providing valuable assistance to patient management. • The Treatment Support Tool includes all medical scores presented in the literature along with alternative support with respect to adverse events appearance, based on patient pre-operation and post-operation data. • Regarding VAD Weaning, the SDSS provides a holistic approach, including all related models presented in the literature allowing the addition of newer models that may be derived. In addition, all available models are automatically transformed into fuzzy ones, thus overcoming fundamental limitations of the crisp logic. The Weaning Tool provides a “mixture-of-experts” approach, © Schattauer 2014
Table 7 Classification Accuracy Results for Treatment Support, Monitoring, Weaning and Suction Detection Tools of the SDSS SDSS component
Dataset
Origin
Technique
Results (Classification Accuracy)
Treatment Support Tool
49 patients
Niguarda Ca’ Granda Hospital
DT
84%
Monitoring Tool
6 patients / 1026 patterns
University Hospital Leuven
DT
99%
Weaning Tool
11 patients / 32 patterns
University of Athens Optimized Fuzzy Model
100%
Suction Detection Tool
36 simulated signals / approximately 23 hours
IBBE-PAS
93%
•
allowing the medical user to assess if a patient is candidate for weaning using all available knowledge weaning models and thus leading to the most informative decision. The Speed Optimization and Suction Detection Tools are connected with a VAD-Heart simulator providing the ability to test several different hypotheses. More specifically, they provide to the medical users a powerful assistant in their attempt to effectively plan the treatment strategy for a patient. In this framework, medical users are able to load different simulation sessions from the VAD Simulation Platform: i) Analyzing data in terms of the suction problem, ii) investigating the potential effect on important hemodynamic variables and iii) determining a pump speed that maintains a safe physiological operating point while providing adequate cardiac output for the patient.
In order to evaluate our risk prediction tools (namely the Treatment Support Tool and the Monitoring Tool) a wide selection of classifiers following different approaches (probabilistic, neural, non-parametric, decision tree, ensemble) were tested. As each of the classification approaches can capture distinctive data characteristics, the wide testing of algorithms ensures that selected models are the ones best adapting to the problem. Furthermore, well-known cross-validation techniques were employed (leave-one-patient-out cross-validation and 10-fold-stratified cross-validation),
GMM
thus ensuring the quality of the obtained results. Especially in the case of the leaveone-patient-out cross-validation technique the algorithms each time are validated with new, unseen data from a new patient, evaluating in this way the generalization ability of the classifier. In addition, existing treatment scores included in the system (namely the HFSS, SHFM, MELD, RVFRS) have been already validated in related studies [13, 14, 19, 27] respectively). The validation of the proposed platform as a system is carried out separately for each component. In order to verify that the proposed modules meet user specifications and fulfil user intended purpose, we recorded the conclusion derived after a trial process. The medical users used personalized annotated data to test the reliability of the exported results. A specific questionnaire based on the System Usability Scale (SUS) methodology was used [65]. The validation process verified that all modules actually meet user needs, the majority of the functional specifications were correct in the first place while verification ensured that the product has been built according to the initial requirements and design specifications. An average SUS score of 72.5% was identified from medical experts. During the tests, fine tuning and suggestions were mentioned by the users, contributing to the robustness of the modules. Validation of the obtained results (in terms of accuracy, etc.) has been carried out with initial patient sets provided by Niguarda Ca’ Granda Hospital, University Hospital Leuven and University of Athens, as deMethods Inf Med 2/2014
13
14
E. C. Karvounis et al.: An Intelligent Decision Support System for the Treatment of Patients Receiving Ventricular ■
scribed in Section 4. Data-use agreements have been signed with respective bodies. Still, additional patient data/trials are needed to verify the accuracy of the tools. Currently the SDSS is populated with data from participating hospitals/centers. As the number of new VAD implanted patients per year is not very high and patients have to be monitored for a long period so as to include potential adverse events in the dataset, large scale trials with the collaboration of additional institutions are needed before the SDSS is put into everyday clinical practice. The powerful and complete nature of the SDSS implies that specialized personnel of cardiovascular centers are expected to primarily utilize this system: i) Cardiac and vascular surgeons who are responsible for the implantation and weaning of the VAD devices. ii) Cardiologists and general clinicians who are responsible for the on-going care and management of the patients and require advanced expertise in order to make specific adjustments to the treatment plan, e.g. regulate drug therapy, titrate medical therapy to the functioning of the VAD device etc., (iii) Biologists and similar researchers interested in investigating and understanding the patient/VAD hemodynamic relationships during assistance. Furthermore, the proposed system will be potentially used to transfer the success and deductions of the most experienced centers to those less established. In general, VAD therapy entails processing complex, uncertain and incomplete data which are dynamically evolving. As a consequence, the medical centers with a greater patient volume are at an advantage compared with those that treat only a few VAD patients per year. The introduction of the proposed system will enable the potential to translate valuable expert knowledge into standardized, personalized and optimized VAD therapy. It should be noted that a major limitation of the proposed system is that some SDSS components have been trained with rather limited datasets. Future work and main focus will be the additional development/validation of the tools based on richer datasets from different medical centers, in order to reinforce the credibility Methods Inf Med 2/2014
and the efficiency of the proposed SDSS application.
6. Conclusions In recent years, the use of VADs for the treatment of end stage heart failure has steadily increased. This widespread application has increased research in the field, but a limited number of DSSs have been proposed over the past years. In this context, the proposed SDSS creates a hallmark in the field, supporting medical and VAD experts through the different phases of VAD therapy. Seven complete tools have been developed and integrated in a unified environment, covering all the specialist’s requirements and providing a unique selection of features. Using the SDSS management of VAD patients can be changed. Treatment decisions are supported by models and tools that can estimate patient risk (for death or adverse events). VAD speed selection currently done mainly with a trial-error process is optimized using simulations, therefore minimizing patient risk. Monitoring patient measurements and VAD parameters altogether is a major enhancement in current patient management and is further exploited for the early identification of potential adverse events. Research is also supported by tools that mine the specialized repository and extract knowledge to complement the knowledge already embedded to the system. As a result, the SDSS fulfils all the functional requirements proposed by the medical experts while assists specialists on designing the best treatment plan for their patients before and after VAD implantation, analysing patients’ data, hypothesis testing, extracting new knowledge and making informative decisions. Acknowledgements
This work is part funded by the European Commission (SensorART project: “A remote controlled Sensorized ARTificial heart enabling patients empowerment and new therapy approaches”, FP7-248763).
References 1. Dickstein K, Cohen-Solal A, Filippatos G, McMurray JJ, Ponikowski P, Poole-Wilson PA, et al. ESC guidelines for the diagnosis and treatment of acute and chronic heart failure 2008: the Task Force for the diagnosis and treatment of acute and chronic heart failure 2008 of the European Society of Cardiology. Developed in collaboration with the Heart Failure Association of the ESC (HFA) and endorsed by the European Society of Intensive Care Medicine (ESICM). Eur J Heart Failure 2008; 10: 933–989. 2. McMurray JJ, Pfeffer MA. Heart failure. Lancet 2005; 365 (9474): 1877–1889. 3. Karvounis EC, Katertsidis NS, Exarchos TP, Fotiadis DI. An intelligent Decision Support System for the treatment of patients receiving Ventricular Assist Device (VAD) support. Proceedings of the 33rd Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC ’11); 2011 Aug 30 – Sep 3; Boston, USA; 2011. pp 8295– 8298. 4. Karvounis EC, Stefanou K, Exarchos TP, Tzallas AT, Tsipouras MG, Fotiadis DI. A Treatment Decision Support System for patients receiving Ventricular Assist Device (VAD) therapy. Proceedings of the IEEE-EMBS International Conference on Biomedical and Health Informatics (BHI); 2012 Jan. 5–7; Hong Kong, Shenzhen; 2012. pp 695 – 698. 5. Kawazoe Y, Ohe K. An ontology-based mediator of clinical information for decision support systems: a prototype of a clinical alert system for prescription. Methods Inf Med 2008; 47 (6): 549 –559. 6. Pfeiffer KP. Future Development of Medical Informatics from the Viewpoint of Health Telematics. Methods Inf Med 2009; 48 (1): 55– 61. 7. Agrawal R, Srikant R. Fast algorithms for mining association rules in large databases. Proceedings of the 20th International Conference on Very Large Data Bases; 1994 Sept. 12–15; Santiago, Chile; 1994. pp 487– 499. 8. Doddi S, Marathe A, Ravi SS, Torney DC. Discovery of association rules in medical data. Med Inform Internet Med 2001; 26: 25 –33. 9. Hu R. Medical Data Mining Based on Association Rules. Computer and Information Science 2010; 3 (4): 104. 10. Sriraam N, Natasha V, Kaur H. Data mining approaches for kidney dialysis treatment. J Mech Med Biol 2006; 6: 109 –121. 11. Immamura T, Matsumoto S, Kanagawa Y, Tajima B, Matsuya S, Furue M, et al. A technique for identifying three diagnostic findings using association analysis. Med Biol Eng Comput 2007; 45: 51– 59. 12. Lee DG, Ryu KS, Bashir M, Bae J-W, Ryu KH. Discovering Medical Knowledge using Association Rule Mining in Young Adults with Acute Myocardial Infarction. J Med Syst 2013; 37: 9896. 13. Aaronson KD, Schwartz JS, Chen TM, Wong KL, Goin JE, Mancini DM. Development and prospective validation of a clinical index to predict survival in ambulatory patients referred for cardiac transplant evaluation. Circulation 1997: 95: 2660 –2667. 14. Levy WC, Mozaffarian D, Linker DT, Sutradhar SC, Anker SD, Cropp AB, et al. The Seattle Heart
© Schattauer 2014
E. C. Karvounis et al.: An Intelligent Decision Support System for the Treatment of Patients Receiving Ventricular ■
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
Failure Model: prediction of survival in heart failure. Circulation 2006; 113: 1424 –1433. Levy WC, Mozaffarian D, Linker DT, Farrar DJ, Miller LW. Can the Seattle heart failure model be used to risk-stratify heart failure patients for potential left ventricular assist device therapy? J Heart Lung Transplant 2009; 28: 231–236. Kirklin JK, Naftel DC, Kormos RL, Stevenson LW, Pagani FD, Miller MA, et al. Second INTERMACS annual report: More than 1,000 primary left ventricular assist device implants. J Heart Lung Transplant 2010; 29: 1–10. Holman WL, Pae WE, Teutenberg JJ, Acker MA, Naftel DC, Sun BC, et al. INTERMACS: interval analysis of registry data. J Am Coll Surg 2009; 208: 755 –761. Stevenson LW, Pagani FD, Young JB, Jessup M, Miller L, et al. INTERMACS profiles of advanced heart failure: the current picture. J Heart Lung Transplant 2009; 28: 535 –541. Matthews JC, Koelling TM, Pagani FD, Aaronson KD. Model for End-Stage Liver Disease Score Predicts Left Ventricular Assist Device Operative Transfusion Requirements, Morbidity, and Mortality. Circulation 2010; 121: 214 –220. Farrar DJ. Preoperative predictors of survival in patients with Thoratec ventricular assist devices as a bridge to heart transplantation. Thoratec Ventricular Assist Device Principal Investigators. J Heart Lung Transplant 1994; 13 (1 Pt 1): 93 –100. Oz MC, Goldstein DJ, Pepino P, Weinberg AD, Thompson SM, Catanese KA, et al. Screening scale predicts patients successfully receiving long-term implantable left ventricular assist devices. Circulation 1994; 92: 69 –73. Rao V, Oz MC, Flannery MA, Catanese KA, Argenziano M, Naka Y. Revised screening scale to predict survival after insertion of a left ventricular assist device. J Thorac Cardiovasc Surg. 2003; 125 (4): 855–862. Lin CY, Tsai FC, Tian YC, Jenq CC, Chen YC, Fang JT, Yang CW. Evaluation of outcome scoring systems for patients on extracorporeal membrane oxygenation. Ann Thorac Surg 2007; 84 (4): 1256 –1262. Lietz K, Long JW, Kfoury AG, Slaughter MS, Silver MA, Milano CA, Rogers JG, Naka Y, Mancini D, Miller LW. Outcomes of left ventricular assist device implantation as destination therapy in the post-REMATCH era: implications for patient selection. Circulation 2007; 116 (5): 497–505. Wang Y, Rosé CP, Ferreira A, McNamara DM, Kormos RL, Antaki JF. A Classification Approach for Risk Prognosis of Patients on Mechanical Ventricular Assistance. Proceedings of Machine Learning and Applications (ICMLA); 2010 Dec. 12–14; Boca Raton, Florida, USA: 2010. pp 293 –298. Fitzpatrick JR, Frederick JR, Hsu VM, Kozin ED, O’Hara ML, Howell E, et al. Risk score derived from pre-operative data analysis predicts the need for biventricular mechanical circulatory support. J Heart Lung Transplant 2008; 27 (12): 1286 –1292. Matthews JC, Koelling TM, Pagani FD, Aaronson KD. The right ventricular failure risk score a preoperative tool for assessing the risk of right ventricular failure in left ventricular assist device candidates. J Am Coll Cardiol 2008; 51(22): 2163–2172.
© Schattauer 2014
28. Ochiai Y, McCarthy PM, Smedira NG, Banbury MK, Navia JL, Feng J, et al. Predictors of severe right ventricular failure after implantable left ventricular assist device insertion: analysis of 245 patients. Circulation 2012; 106 (12 Suppl 1): I198 –202. 29. Kormos RL, Teuteberg JJ, Pagani FD, Russell SD, John R, Miller LW, et al. Right ventricular failure in patients with the HeartMate II continuous flow left ventricular assist device: incidence, risk factors, and impact on outcomes. J Thorac Cardiovasc Surg 2010; 139 (5): 1316 –1324. 30. Wang Y, Simon M, Bonde P, Harris BU, Teuteberg JJ, Kormos RL, et al. Prognosis of Right Ventricular Failure in Patients with Left Ventricular Assist Device based on Decision Tree with SMOTE. IEEE Trans Info Tech Biomed 2012; 16 (3): 383 –390. 31. Slaughter MS, Pagani FD, Rogers JG, Miller LW, Sun B, Russell SD, et al. Clinical management of continuous-flow left ventricular assist devices in advanced heart failure. J Heart Lung Transplant 2010; 29: S1-S39. 32. Kirchner J, Arnold M, Beckendorf C, Paule S, Czygan G. Heart failure monitoring with implantable defibrillators. Biomedical Engineering/Biomedizinische Technik 2012; 57 (suppl. 1): 1079 –1082. 33. White M, Howie-Esquivel J, Caldwell M. Improving Heart Failure Symptom Recognition: A Diary Analysis. Journal of Cardiovascular Nursing 2010; 25 (1): 7–12. 34. Hayes K, Leet AS, Bradley SJ, Holland AE. Effects of exercise training on exercise capacity and quality of life in patients with a left ventricular assist device: A preliminary randomized controlled trial. J Heart Lung Transplant 2012; 31 (7): 729 –734. 35. Santelices LC, Wang Y, Severyn D, Druzdzel MJ, Kormos RL, Antaki JF. Development of a Hybrid Decision Support Model for Optimal Ventricular Assist Device Weaning. Ann Thorac Surg 2010; 90 (3): 713–720. 36. Dandel M, Weng Y, Siniawski H, Stepanenko A, Krabatsch T, Potapov E, at al. Heart failure reversal by ventricular unloading in patients with chronic cardiomyopathy: criteria for weaning from ventricular assist devices. Eur Heart J 2011; 32 (9): 1148 –1160. 37. Birks EJ, George RS, Hedger M, Bahrami T, Wilton P, Bowles CT, Webb C, Bougard R, Amrani M, Yacoub MH, Dreyfus G, Khaghani A. Reversal of severe heart failure with a continuous-flow left ventricular assist device and pharmacological therapy: a prospective study. Circulation. 2011; 123 (4): 381– 390. 38. Boston JR, Antaki JF, Simaan MA. Hierarchical Control of Heart-Assist Devices. IEEE Robotics & Automation Magazine. 2003; 10: 54 – 64. 39. Antaki JF, Choi S, Boston JR, Butler KC, Thomas DC, Amin DV. Speed control system for implanted blood pumps. US Patent 6,066,086: Nimbus Inc, 2000. 40. Konishi H, Antaki JF, Amin DV, Boston JR, Kerrigan JP, et al. Controller for an axial flow blood pump. Artificial Organs 1996; 20 (6): 618 – 620. 41. Waters T, Allaire P, Tao G, Adams M, Bearnson G, et al. Motor feedback physiological control for a continuous flow ventricular assist device. Artificial Organs 1999; 23 (6): 480 – 486.
42. Giridharan GA, Skliar M. Physiological control of blood pumps without implantable sensors. Proceedings of the American Control Conference; 2003 June 4–6; Denver, Colorado USA: 2003. pp 471– 476. 43. Oshikawa M, Araki K, Nakamura K, Anai H, and Onitsuka T. Detection of total assist and sucking points based on the pulsatility of a continuous flow artificial heart: in vivo evaluation. ASAIO J 1998; 44 (5): 704 –707. 44. Waters T, Allaire P, Tao G, Adams M, Bearnson G, et al. Motor feedback physiological control for a continuous flow ventricular assist device. Artificial Organs 1999; 23 (6): 480 – 486. 45. Giridharan GA, Skliar M. Nonlinear controller for ventricular assist devices. Artificial Organs 2002; 26 (11): 980 –984. 46. Giridharan GA, Skliar M, Olsen DB, Pantalos GM. Modeling and control of a brushless DC axial flow ventricular assist device. Asaio J 2002; 48 (3): 272–289. 47. Parnis SM, Conger JL, Fuqua JM Jr., Jarvik RK, Inman RW, et al. Progress in the development of a transcutaneously powered axial flow blood pump ventricular assist system. Asaio J 1997; 43 (5): 576 –580. 48. Chen SH. Baroreflex-Based Physiological Control of a Left Ventricular Assist Device. Thesis. The School of Engineering, University of Pittsburgh, 2006. 49. McConahy DC. Application of multiobjective optimization to determining an optimal left ventricular assist device (LVAD) pump speed. Thesis. The School of Engineering, University of Pittsburgh, 2007. 50. Konishi H, Antaki JF, Amin DV, Boston JR, Kerrigan JP, et al. Controller for an axial flow blood pump. Artificial Organs 1996; 20 (6): 618 – 620. 51. Yuhki A, Nogawa E, Miura M, Shimazaki Y, Takatani S. Detection of Suction and Regurgitation of the Implantable Centrifugal Pump Based on the Motor Current Waveform Analysis and Its Application to Optimization of Pump Flow. Artificial Organs. 1999; 23 (6): 532–537. 52. Boston JR, Antaki JF, Simaan MA. Hierarchical Control of Heart-Assist Devices. IEEE Robotics & Automation Magazine 2003; 10: 54 – 64. 53. Liu D, Boston JR, Lin H, Antaki JF, Simaan MA, Wu J. Monitoring the development of suction in an LVAD. Proceedings of the BMES-EMBS 1st joint Conference; 1999; 2003. pp 400 – 401. 54. Vollkron M, Schima H, Huber L, Benkowski R, Morello G, Wieselthaler G. Advanced Suction Detection System for an Axial Blood Pumps. Artificial Organs 2006; 30 (9): 665– 670. 55. Ferreira A. A rule-based controller based on suction detection for rotary blood pumps. Thesis. The School of Engineering, University of Pittsburgh, 2007. 56. Ferreira A, Chen S, Simaan MA, Boston JR, Antaki JF. A discriminant analysis-based suction detection system for rotary blood pumps. Proceedings of the Engineering in Medicine and Biology Society; 2006 Aug. 30–Sept. 3; New York City, USA: 2006. pp 5382–5385. 57. Wang Y, Faragallah G, Divo E, Simaan MA. Detection of Ventricular Suction in an Implantable Rotary Blood Pumps Using Support Vector Machines. Proceedings of the IEEE Engineering in
Methods Inf Med 2/2014
15
16
E. C. Karvounis et al.: An Intelligent Decision Support System for the Treatment of Patients Receiving Ventricular ■ Medicine and Biology Conference; 2011 Aug. 30-Sept. 3; Boston, Massachusetts, USA: 2011. pp 3318 –3321. 58. Oshikawa Ì, Araki Ê, Endo G, Anai H, Sato M. Sensor less controlling method for a continuous flow left ventricular assist device. Artif Organs 2000; 24 (8): 600 – 605. 59. Karantonis DM, Lovell NH, Ayre PJ, Mason DG, Cloherty SL. Identification and Classification of Physiologically Significant Pumping States in an Implantable Rotary Blood Pump. Artificial Organs 2006; 30 (9): 671– 679. 60. Karantonis DM, Cloherty SL, Lovell NH, Mason DG, Salamonsen RF, et al. Noninvasive detection of suction in animplantable rotary blood pump
Methods Inf Med 2/2014
using neural networks. Int. J. Computation Intelligence and Appl 2008; 7 (3): 237– 247. 61. Morello, Gino F. Blood pump system and method of operation. US Patent: 0215050, 2004. 62. AlOmari AHH, Savkin AV, Stevens M, Mason DG, Timms DL, et al. Developments in control systems for rotary left ventricular assist devices for heart failure patients: a review. Physiol Meas 2013; 34 (1): R1–27. 63. Kozarski M, Ferrari G, Zieliñski K, Górczyñska K, Pa3ko KJ, et al. A hybrid (Hydro-Numerical) Cardiovascular Model Application to Investigate Continuous-Flow Pump Assistance Effect. Biocybernetics and Biomedical Engineering 2012; 32 (4): 77– 91.
64. Tzallas AT, Rigas G, Karvounis EC, Tsipouras MG, Goletsis Y, et al. A Gaussian Mixture Model to Detect Suction Events in Rotary Blood Pumps. Proceedings of the 12h Annual International Conference of the IEEE Bioinformatics & Bioengineering (BIBE); 2012 Nov. 11–13; Larnaka, Cyprus; 2012. pp 127–131. 65. Brooke J. SUS: a “quick and dirty” usability scale. In: PJordan PW, Thomas B, Weerdmeester BA, McClelland AL. Usability Evaluation in Industry. London: Taylor and Francis; 2006.
© Schattauer 2014