Mobile C-arms using standard real-time fluoroscopy: General fluoro techniques ...... BV Libra. BV Endura. BV Pulsera. Pulsera R2? Philips. No, full lead shutters.
FOR INTERNAL USE ONLY. Data shown are from published specifications which are subject to change. This document updated 5/16/06.

Comments

Ziehm Quantum

Vista Plus

Vision

Vision R

real-time fluoro

real-time fluoro

pulse based

pulse based

2.2 kW

2.2 kW

2.2 kW

7.5 kW

Generator Mobile C-arms using standard real-time fluoroscopy: General fluoro techniques are typically 1-5 mA @ 60-100 kV, or .06kW - .5kW. Occasionally higher power is needed in dense antomy imaging (obese patients) in which case a High Level Fluoro (HLF) technique would be used. In HLF the federal limit on dose rate is 20R/min. Therefore, HLF max. power would typically not exceed 1 - 1.2 kW (example: 80 kV @ 12.5 mA). In summary, most surgical imaging is accomplished at power outputs of less than 2 kW. The 15kW rating of the OEC 9800 would only be used during coronary angiography - where Max. output power very high mA, short pulse widths are uniquely required to stop motion. (One could argue that even 15kW may not be sufficient for coronary angiography when compound imaging angles are required). Pulsed Fluoroscopy systems like the Ziehm Vision function differently from standard fluoroscopy. The Ziehm vision may run at a higher fluoro mA (to improve anatomic visualization) while decreasing overall patient dose using pulsed fluoroscopy. For example, when using a 30 ms. pulse width, actual dose or mA to the patient is reduced by a factor of roughly 1/3. kV

40 - 110 KV

40 - 110 KV

40 - 110 KV

40 - 120 KV

Higher mA output provides better visualization of dense anatomy (reduces noise in the image), given appropriate penetration by kV output . However, max. mA values by themselves can be misleading. Max. wattage values allow calculation of max. mA at a given kV. Ziehm's Vision R has 120 kV, maximum output, allowing maximum penetration for dense anatomy such as deep oblique views, and lateral views of the lumbar spine.

Continuous Fluoro Mode mA, incl. HLF

0.2-6 mA

0.2-6 mA

Vision is always in pulse mode

Vision is always in pulse mode

DR mode also known as Snapshot or One-shot mode. A single, brief exposure is made at higher than normal mA to capture a higher quality single image.

Digital Radiography (DR) or Snapshot Mode mA

up to 8 mA "snapshot" mode

up to 8 mA "snapshot" mode

Up to 20 mA

up to 75 mA

The primary purpose of pulsed fluoro is for dose reduction. The secondary purpose is for reduction of blurring while imaging moving anatomy. On some systems like the Ziehm Vision, pulse mode may also be used with higher mA values to improve image quality.

Pulsed Fluoro mA

up to 6mA

0.2-6 mA

0.1 - 20 mA

up to 75 mA

The key here is that the best spec's would be those that gave a wide range of pulse rate selections - particularly at 8 ppsand higher because this is the rate at which the image starts to appear continuous to the naked eye. 15 and 30 pps rates are used to acquire cine for vascular, cardiac and bolus chase or "runoff" applications.

Pulsed fluoro rates

1-3 pps

1 pps.

1-30 pps.

1-30 pps.

Pulsed Fluoro mA + HQ output

N/A

N/A

Up to 20 mA when using HQ mode

Up to 75 mA when using HQ mode

Digital Cine Pulse Widths

N/A

N/A

4-30 ms, varied by anatomic program

4-30 ms, varied by anatomic program

Used for dynamic vascular studies. Combines High Level Fluoro & short pulses to stop motion. High Quality mode can be used with digital cine to produce high quality bolus chase or "runoff" studies. Shorter pulse widths stop motion blurring during dynamic vascular imaging procedures. HQ Mode in the Ziehm vision always uses a short 10 mS pulse width. Pulse widths vary by anatomic program for optimal image quality.

Page 1 of 26

FOR INTERNAL USE ONLY. Data shown are from published specifications which are subject to change. This document updated 5/16/06.

Quantum

Vista Plus

Vision

Vision R

real-time fluoro

real-time fluoro

pulse based

pulse based

Image Intensifier

Thomson

Thomson

Thomson

Thomson

Fields of View

Dual-mode 9"/6"

Dual-mode 9"/6"

Tri-mode 9"/6"/4.5" or 12"/9"6"

Tri-mode 9"/6"/4.5" or 12"/9"/6"

Comments The Ziehm system uses a quicker "rise and fall" time when producing pulses. Therefore, ziehm reduces unnecessary radiation output to the patient while lowering heat build-up. A longer "rise and fall" time will emit "soft" radiation that contributes dose to the patient while not providing usable mA to produce an image. Smallest field of view (4.5") is not clinically necessary and not worth the tradeoff of higher dose. However, Ziehm uses technical aspects of the camera & lens assembly in order to more efficiently provide an image at comparatevely lower dose in the 4.5" field of view

Ziehm

Page 2 of 26

FOR INTERNAL USE ONLY. Data shown are from published specifications which are subject to change. This document updated 5/16/06.

Ziehm Quantum

Vista Plus

Vision

Vision R

real-time fluoro

real-time fluoro

pulse based

pulse based

Yes

Yes

Yes

Yes

Focal spot fluoro (mm)

0.5/1.5 or 0.6/1.5

0.5/1.5 or 0.6/1.5

0.6

0.3/0.6mm

The combination of power (kW), anode angle and focal spot size dictate the use of stationary or rotating anodes. Mobile C-arm fluoroscopic applications are sufficiently handled by a stationary anode and 0.6mm focal spot. The use of a 0.3 mm focal spot mandates anode rotation.

Anode (Stat.or Rot.)

Stationary

Stationary

Stationary

Rotating

In mobile C-arm applications, anode heat capacity and dissipation are not the only limiting factors in heat management (see Tube Housing heat specifications below).

Anode Heat Capacity

76,000

76,000

45,000 HU

300,000 HU

Heat capacity of the tube housing

1,140,000 HU

1,140,000 HU option: active cooling 5 Million HU

5 Million HU

5 Million HU

Anode Cooling Rate, HU/min.

19,800

Comments Mu Metal Shielding of image intensifier prevents metallic fixation devices from appearing curved due to the magnetic distortion inherent in image intensifiers. Mu metal shielding compensates for lower magnetic fields inherent in exam rooms. Mu metal shielding is optional. Ziehm's ADCS - "active S-distortion correction package" uses real-time magnetic sensing with corrrection coils for eliminating S-distortion. This is standard on the Vision R 12" I.I.

Mu Metal Shielding

X-ray Tube Generally, smaller focal spot size provides better image resolution. However, the image digital matrix dictates the limits of resolution. A 1K2 matrix with a 0.3 mm focal spot will produce approximately the same resolution as a 0.6 mm focal spot at object-to-image distances of 40 cm or less. Therefore, in C-arm 2

imaging with a 1K matrix, the performance of a 0.6mm focal spot is equal to a 0.3mm.

Higher heat capacity in the housing generally translates to the ability to do many high dose procedures back-to-back.

The Vision R is continually extracting heat through its thermal management system, including generator housing and anode. This results in significantly extended fluoro procedure time possibilities in comparison to competitors. In addition, since Ziehm Vision systems are always using pulse mode rather than real-time fluoroscopy, less heat is emitted from the tube from the beginning.

19,800 - with optional With standard active active cooling, 5,000,000 cooling, 5,000,000 equiv. equiv.

85,000

Collimation

Semi-transparent (translucent) leading edge (tapered) shutters permit soft tissue visualization around long bone or spine and aids in placement of needle and screw alignment.

Iris collimation

yes

yes

yes

yes

Semi-transparent shutter

Translucent collimator blades.

Translucent collimator blades.

Translucent collimator blades.

Translucent collimator blades.

Page 3 of 26

FOR INTERNAL USE ONLY. Data shown are from published specifications which are subject to change. This document updated 5/16/06.

Comments Ziehm systems provide "virtual" collimation which allows the user to adjust collimation and see where collimator shutters are positioned using digital lines displayed on a held image on the monitor. This provides the ability to adjust collimation without requiring fluoroscopy, thus providing dose reduction to the patient.

Ziehm Quantum

Vista Plus

Vision

Vision R

real-time fluoro

real-time fluoro

pulse based

pulse based

Preview collimator indicators

No

Yes, both iris and parallel Yes, both iris and parallel Yes, both iris and parallel shutter position and shutter position and shutter position and rotation indicators on rotation indicators on rotation indicators on fluoro display fluoro display fluoro display

TV-System The Ziehm vision's camera is "spectrally responsive" to the I.I.'s light output. Light output from an Image Intensifier is inherently in the green portion of the color spectrum. Ziehm's CCD camera closely matches the spectral output of the I.I., which translates to a more sensitive camera, requiring less dose. In addition, the high quality lens provides more light to the camera.

TV Camera

CCD

CCD

Spectrally Responsive CCD

Spectrally Responsive CCD

The Siemens Arcadis Varic and Ziehm Vision are the only true 1K2 systems.

Pixel Count (resolution)

768 x 494

768 x 576

1024 x 1024

1024 x 1024

Page 4 of 26

FOR INTERNAL USE ONLY. Data shown are from published specifications which are subject to change. This document updated 5/16/06.

Ziehm Quantum

Vista Plus

Vision

Vision R

real-time fluoro

real-time fluoro

pulse based

pulse based

Monitor size

18.1"

18.1"

18.1"

18.1"

Lighting conditions vary in surgery. Bright overhead lights may be used, or lights may be dimmed, with only the surgical field well lit. Brighter monitors ensure improved image visualization in varied lighting conditions. Siemens' Syngo interface requires the use of a color monitor to display color-coded icons, at the expense of image brightness. However, B&W monitors are brighter than color. Ziehm separates its user interface from monitor displays.

Monitor luminescence

600 cd/m2

600 cd/m2

600 cd/m2

600 cd/m2

Digital image rotation contributes to dose and time savings during surgical procedures.

Digital image rotation

Digital Image Rotation and Reversal without Radiation

Digital Image Rotation and Reversal without Radiation

Digital Image Rotation and Reversal without Radiation

Digital Image Rotation and Reversal without Radiation

Free-space

30"

30"

30"

30"

Depth of immersion

27"

27"

27"

27"

Wheel distance (width base unit)

31.5"

31.5"

31.5"

31.5"

Weight of C-arm

560-580 lbs with options

550 lb, 572 with active cooling

752 lb.

836 lb (9" I.I), 876 lb (12: I.I.)

C-arm Size

67" H x 63" L x 31.5" W or 84.25" H x 71.65" L x 31.5" W

84.25" H x 71.65" L x 31.5" W

67" H x 63" L x 31.5" W or 84.25" H x 71.65" L x 31.5" W

70.5" H x 74" L x 31.5" W or 87.4" H x 82.7" L x 31.5" W

Weight of the monitor cart

NA

389

352

352

Comments

Larger monitors allow easier viewing at typical working distances.

C-arm The higher the free space, the more likely the user will be able to position the C around larger patients and tables. Also, more working space will be available for the surgeon. Better positioning clearance around bigger patients and tables A narrower wheel base allows the C-arm to fit easier through narrow doors Ligher weight generally means easier maneuverability.

Measurements monitor cart

NA

66.5" H x 22.8" W x 27.4" 69.7" H x 28.3" W x 26.4" 69.7" H x 28.3" W x 26.4" D D (MC6) D (MC6)

Movements

Increased orbital rotation provides more positioning flexibility. Allows the user to see more complex angles without needing to roll or reposition the patient.

Brakes

manual

manual

manual

manual

Orbital Rotation

90°/+45°

90°/+45°

90°/+45° (9" I.I.) 90°/+25° (12" I.I.)

90°/+45° (9" I.I.) 90°/+25° (12" I.I.)

Page 5 of 26

FOR INTERNAL USE ONLY. Data shown are from published specifications which are subject to change. This document updated 5/16/06.

Ziehm Quantum

Vista Plus

Vision

Vision R

real-time fluoro

real-time fluoro

pulse based

pulse based

Base image storage capacity

10,000

1,000

5,000

5,000

Optional memory

N/A

Optional up to 5,000

up to 55,000

up to 55,000

Digital Cine Acquisition Rates (in f/s)

1,2,4,8 & 15 f/s

1-8 frames/s, 8-12.5 frames/s

1-10 frames/s, 1-15 frames/s, 1-30 frames/s

1-10 frames/s, 1-15 frames/s, 1-30 frames/s

Cine Loop Playback

Auto playback, stop and start at C-arm touchpad

Cine loop start, stop and replay rate controls at workstation

Playback via touch screen at monitor cart or chassis. Auto playback

Playback via touch screen at monitor cart or chassis. Auto playback

Comments Memory

8-15 f/s is generally considered sufficient for vascular procedures. In some cases 30 f/s may be considered favorable for neurovascular or cardiac applications. Slower acquisition speeds may be acceptable for general surgery applications such as Laparoscopic Cholangiography or ERCP.

Other Functions a) children, extremities b) Bone, Heart, Abdomen, Extremities, Chest, spine, hip c) heart or lung Endo, Motion, Metal, Soft, Bone, Heart, Abdomen, Head/Spine, Abdominal, (preconfiguration done by Large Patient, Automatic Endo, Motion, Metal, Soft, Metal,1/2 Dose, and LPD technical staff) and LPD Motion Detection, Large Patient (Large Patient Diameter) (Large Patient Diameter) Automatic Metal key. Correction

Provide custom pre-set adjustments specific to the preferences of the surgeon or specific to anatomy.

Organ programs

ODDC (Object Detected Dose Control): - Detects the location of anatomy in the field of view from a 16 X 16 measurement grid to adjust dose and image presentation - Detects Motion in an Orthopedic image to adjust frame averaging filters real time - Detects Motion in an Orthopedic image and adjusts pulse rates accordingly to save patient dose - Automatically adjusts X-ray technique to compensate when metal objects are introduced into the imaging field.

ODDC

N/A

N/A

Yes

Selectable dose-rate*

1/2 dose button

1/2 dose button

1/2 dose button

Page 6 of 26

Yes

1/2 dose button

FOR INTERNAL USE ONLY. Data shown are from published specifications which are subject to change. This document updated 5/16/06.

Ziehm Quantum

Vista Plus

Vision

Vision R

real-time fluoro

real-time fluoro

pulse based

pulse based

Edge enhancement

yes (4 Steps - for Realtime and Post Processing)

yes (5 Steps - for Realtime and Post Processing)

yes (5 Steps - for Realtime and Post Processing)

yes (5 Steps - for Realtime and Post Processing)

Allow immediate "toggle-through" of pre-set brightness & contrast settings.

LUT (Look-up Tables)

Manual Window/ Level

toggle through 4 steps level/window

toggle through 4 steps level/window

toggle through 4 steps level/window

Labels may be added to an image for identification of anatomic structures before printing.

Image annotations

Standard

Standard

Standard

Standard

Once an image is magnified digitally, Roam allows the user to continuously vary the area of the image to be magnified.

Roam

Yes

Yes

Yes

Yes

Digital zoom

2x & 4x

yes (7 levels)

yes (7 levels)

yes (7 levels)

Vascular Pkg required

Vascular Pkg required

Vascular Pkg. required

Vascular Pkg. required

Yes No

Yes Yes

Yes Yes

Yes Yes

Remask

No

Yes

Yes

Yes

Pixel-shift

No

Yes

Yes

Yes

Peak opacification (MSA)

MSA can display at termination of exposure (service setting)

Yes

Yes

Yes

Min. Opacification (for CO2)

No

No

Optional

Optional

Landmarking

No

Yes

Yes

Yes

Stenosis evaluation

No

No

Yes, with Vascular Module

Yes, with Vascular Module

Distance Measurement

No

1 or 2 lengths

1 or 2 lengths

1 or 2 lengths

Angle Measurement

No

1 or 2 angles

1 or 2 angles

1 or 2 angles

Comments Improves visualization of bony edges and fracture lines digitally

Allows magnification of an image without needing an extra exposure at a smaller field of view.

Features for Subtraction: Subtraction (DSA) Roadmapping (RSA) Remask allows the user to use a mask already used to obtain a DSA run for a Roadmap, saving time and dose to the patient Allows re-alignment of the mask and the native image in situations where the patient may have moved during DSA acquisition. Without pixel shift, smaller vessels may be obsured by a mis-aligned mask. Siemens has unique automatic pixel-shift through a region of interest.

For some specialized vascular procedures CO2 is injected rather than fluid contrast into the vessels. Min. Opacification allows for visualization of CO2, which appears white in the vessels rather than black, like liquid contrast agents. In some situations a vascular surgeon may want to fade some of the subtracted mask away to allow the visualization of bony landmarks in an image. On-screen measurement of vessel stenosis compared to an open portion of a vessel displayed as % stenosis.

Often used for surgical planning in orthopedic procedures such as osteotomy. This feature allows the user to "draw an angle" on a fluoro image digitally to obtain a numerical angle measurement.

Page 7 of 26

FOR INTERNAL USE ONLY. Data shown are from published specifications which are subject to change. This document updated 5/16/06.

Ziehm Quantum

Vista Plus

Vision

Vision R

real-time fluoro

real-time fluoro

pulse based

pulse based

System Control

"Desk View" Control at Carm

Keyboard (Monitor cart), Keyboard on Carm Chassis

Mosaic display (image collage)

Patient oriented search function with 4 image thumbnail display

Search function via 16 image mosaic display (mosaic archiving with vascular configuration)

CD or DVD Burner

optional external USB DVD-RW

yes CD (1000 in TIFF /DICOM format)

Other manufacturer specific media

USB Hub/Port

USB Hub/Port

DICOM functions available

Storage, Print, Media Class, Worklist,

Comments Other Features The Vision CenterTouch Control System allows the technologist access to all fluoro and generator functions at either side of the C-arm or at the workstation. The Quantum's "desk view" offers similar interface with the absence of a workstation.

"Vision Center": Same UI "Vision Center": Same UI at Chassis as well as at Chassis as well as Monitor trolley (Extra Monitor trolley (Extra Touch Screen) Touch Screen)

Search function via 16 image mosaic display

Search function via 16 image mosaic display

Transfer / Connectivity

An optional “Query Retrieve” of other modality images allows users to retrieve a CT or MR image, for example, and display it as a reference on the right workstation monitor during surgery. Media class allows the users to archive images to external media in DICOM , tiff or jpeg formats.

DVD-RW 4.7 GB DVD-RW 4.7 GB capy.DICOM, TIFF, JPG, capy.DICOM, TIFF, JPG, Multimedia Multimedia USB Hub/Port

USB Hub/Port

Storage, Storage Commit, Storage, Storage Commit, Storage, Storage Commit, Print, Media Class, Print, Media Class, Print, Media Class, Worklist, Query-Retrieve, Worklist, Query-Retrieve, Worklist, Query-Retrieve, Verification Class Verification Class Verification Class

Accessory Options Video output allows Paper: sony 960/970 connection to stand-alone paper and film: Sony UP printer 980/990

Printer Options

paper and film: Sony UP 980/990

paper and film: Sony UP 980/990

Handheld Remote control for workstation functions

No

No

No

No

Laser light localizer tube side (alignment or targeting tool)*

Option

Option

Option

Option

Laser light localizer II side (aimer)*

Option

Option

Option

Option

Page 8 of 26

FOR INTERNAL USE ONLY. Data shown are from published specifications which are subject to change. This document updated 5/16/06.

Quantum

Vista Plus

Vision

Vision R

real-time fluoro

real-time fluoro

pulse based

pulse based

Remote Service Diagnostics

No

No

No

No

3D Image Reconstruction

No

No

No

No

Comments Siemens' ARCADIS C-arms have remote service diagnostics via network connection. Allows service to identify repair parts requirements prior to on-site diagnosis which enhances service ability to correct malfunction on the first call. However, in many cases, hospital IT specialists will not allow access to remote diagnostics through the hospital's eternet port due to security concerns.

Ziehm

Page 9 of 26

Siemens Vision FD

Vario 3D

pulse based

real-time fluoro

2.2 kW

GE OEC

Compact L

ARCADIS Varic

ARCADIS Orbic

ARCADIS Avantic

Fluorostar

FlexiView 8800

Series 9800 Plus

9900 Elite

2.2 kW

1.4 kW

2.3 kW

2.3 kW

20 kW

2.2 kW

2.2 kW

15 kW

15 kW

40 - 110 KV

40 - 110 KV

40 - 110 kV

40 - 110 kV

40 - 110 kV

40-125

36-110kV

40 - 110kV

40 - 120 kV

40 - 120 kV

Vision is always in pulse mode

0.2-6 mA

0.2 - 8.9 mA (max. 550 W)

0.2 - 15.2 mA (max. 0.2 - 15.2 mA (max. 0.2 - 15.2 mA (max. 1000 W) 1000 W) 1000 W)

0.2-8.0 mA (may be lower for cont. fluoro)

0.1 - 6 mA (max. 660 W)

Up to 20 mA

up to 8 mA "snapshot" mode

0.2 -12.2 mA (max. 850 W)

0.2 - 23 mA (max. 0.2 - 23 mA (max. 1000 W) 1000 W)

up to 250 mA

0.2-8.0 mA

up to 10 mA (max. W not specified)

up to 75 mA

up to 75 mA

0.1 - 20 mA

0.2-6 mA

0.2 - 8.9 mA (max. 550 W)

up to 23 mA (max. up to 23 mA (max. 2300 W) 2300 W)

up to 67 mA

0.2-8.0 mA

up to 16 mA (max. W not specified)

0.2 - 10, 1 - 40 mA HLF pulsed fluoro

0.2 - 10, 1 - 40 mA HLF pulsed fluoro

1-30 pps.

1 pps.

0.5, 1, 2 pps

0.5 to 8 pps

up to 15 pps.

not specified

1, 2, 4, 8 pps

1,2,4,8 pps

1,2,4,8 pps

Up to 20 mA when using HQ mode

N/A

No

0.5 - 8 pps std., 15 pps optional; up to 23 mA

No

1 - 8 pps; up to 16 mA

15 pps (30 pps on Cardiac) up to 150 mA

15 pps (30 pps on Cardiac) up to 150 mA

4-30 ms, varied by anatomic program

N/A

N/A

min. 7 ms

N/A

25 - 50 ms

10 ms

10 ms

0.5 to 8 pps

0.5 - 8 pps std., 15 Up to 30 pps., up to pps optional; up to 250 mA., 23 mA

min. 7 ms

Page 10 of 26

min. 7 ms

0.2 - 10mA, 1.0 - 20 0.2 - 10mA, 1.0 - 20 mA HLF mA HLF

Siemens Vision FD

Vario 3D

pulse based

real-time fluoro

Varian

19.8 X 19.8 cm square (7.8 X 7.8 inches)

GE OEC

Compact L

ARCADIS Varic

ARCADIS Orbic

Thomson

Siemens

Siemens

Siemens

Dual-mode 9"/6"

9"/6" dual mode

9"/6" dual mode

9"/6" dual mode

Page 11 of 26

ARCADIS Avantic

Fluorostar

FlexiView 8800

Series 9800 Plus

9900 Elite

Thales (Thomson) Thales (Thomson) Thales (Thomson) Thales (Thomson)

Quad-mode 13"/8"/6"/5"

Tri-mode 9"/6"/4.5"

Tri-mode 9"/6"/4.5" Tri-mode 9"/6"/4.5" Tri-mode 9"/6"/4.5" or 12"/9"/6" (12" not or 12"/9"/6" (12" not avail. on super C) avail. on super C)

Siemens Vision FD

Vario 3D

pulse based

real-time fluoro

Yes

Yes

GE OEC

Compact L

ARCADIS Varic

ARCADIS Orbic

ARCADIS Avantic

Fluorostar

FlexiView 8800

Series 9800 Plus

9900 Elite

Yes

Yes

Yes

Yes

Optional

Optional

Optional

Optional

Siemens

Siemens

Siemens

VARIAN

VARIAN

0.6

0.5/1.5 or 0.6/1.5

0.6

0.6

0.6

dual 0.3 and 0.5

0.6

0.6

dual 0.3 and 0.6

dual 0.3 and 0.6

Stationary

Stationary

Stationary

Stationary

Stationary

Rotating

Stationary

Stationary

Rotating

Rotating

45,000 HU

76,000

50,000 HU

50,000 HU

50,000 HU

270,000 HU

55,000

74,200 HU

300,000 HU

300,000 HU

5 Million HU

1,140,000 HU option: active cooling 5 Million HU

1,200,000 HU

1,200,000 HU

1,200,000 HU

2,565,000 HU

not specified

980,000 HU

1,600,000 HU

1,600,000 HU

19,800 - with optional With standard active active cooling, 5,000,000 cooling, 5,000,000 equiv. equiv.

85000 HU/Min

yes

yes

yes

yes

yes

yes

yes

yes

yes

Translucent collimator blades.

Translucent collimator blades.

Yes

yes

Yes

yes, semitransparent slot collimators

yes, semitransparent blades

yes, semitransparent blades

No, solid tungsten collimator shutters

Page 12 of 26

No, solid tungsten collimator shutters

Siemens Vision FD

Vario 3D

pulse based

real-time fluoro

Compact L

ARCADIS Varic

GE OEC ARCADIS Orbic

ARCADIS Avantic

Fluorostar

FlexiView 8800

Series 9800 Plus

9900 Elite

Yes, both iris and Yes, both iris and Yes, both iris and Yes, both iris and Yes, both iris and Yes, both iris and Yes, both iris and Yes, both iris and Yes, both iris and parallel Yes, both iris and parallel parallel shutter parallel shutter parallel shutter parallel shutter parallel shutter parallel shutter parallel shutter parallel shutter shutter position and shutter position and position and rotation position and rotation position and rotation position and rotation position and rotation position and rotation position and rotation position and rotation rotation indicators on rotation indicators on indicators on fluoro indicators on fluoro indicators on fluoro indicators on fluoro indicators on fluoro indicators on fluoro indicators on fluoro indicators on fluoro fluoro display fluoro display display display display display display display display display

Flat Detector

CCD

CCD

CCD

CCD

CCD

CCD

CCD

CCD

CCD

1024 x 1024

768 x 576

512 x 512

1024 x 1024

1024 x 1024

1024 x 1024

512 X 512

980 X 980

980 X 980

1280 X 1024

Page 13 of 26

Siemens Vision FD

Vario 3D

pulse based

real-time fluoro

18.1"

18.1"

GE OEC

Compact L

ARCADIS Varic

ARCADIS Orbic

ARCADIS Avantic

Fluorostar

FlexiView 8800

Series 9800 Plus

9900 Elite

17" TFT

18" Color TFT or Grayscale TFT

18" Color TFT

18" Color TFT

18.1" rectangular TFT

16" square CRT

16" square CRT

18" LCD Flat Panel mounted on an articulating arm

up to 600 cd/m2 (TFT greyscale); 250 cd/m2 (TFT color)

up to 600 cd/m2 (TFT greyscale); 250 cd/m2 (TFT color)

up to 600 cd/m2 (TFT greyscale); 250 cd/m2 (TFT color)

not specified

not specified

not specified

700 cd/m2 maximum brightness

Yes. Method of rotation not specified

Rotated image appears after movement of onscreen indicator

600 cd/m2

600 cd/m2

250 cd/m2

Digital Image Rotation and Reversal without Radiation

Digital Image Rotation and Reversal without Radiation

Rotated image appears after movement of onscreen indicator

33"

30"

31"

31"

31"

31"

30"

31"

27"

27"

29"

29"

29"

29"

26"

26"

31.5"

31.5"

31.5"

31.9"

31.5

33.5

31"

33"

33"

33"

752 lb.

630.5 lb

566 lbs.

520 lbs.

765 lbs.

770 lbs.

657 lbs.

593 lbs.

721 lbs. (9" I.I.)

721 lbs. (9" I.I.)

Rotation occurs in Rotation occurs in Rotation occurs in real-time as image real-time as image real-time as image is rotated via cursor is rotated via cursor is rotated via cursor

67" H x 63" L x 31.5" W or 84.25" H x 71.65" L x 31.5" W

84.25" H x 71.65" L x 31.5" W

68.9"H x 67.5" L x 31.5" W

68.1" H X 67.5" L x 31.9" W

70.8" H x 84.6" L x 31.5" W

72" H x 78" L x 33.5" W

64.5" H x 67" L x 31" W

352

389

325 lbs.

395 lbs.

418 lbs.

418 lbs.

N/A

69.7" H x 28.3" W x 26.4" 66.5" H x 22.8" W x 27.4" 65.4" H x 33.1" W x 68.3" H x 32.7" W x 68.3" H x 32.7" W x 68.3" H x 32.7" W x D (MC6) D 33.1" W 28.3" D 28.3" D 28.3" D

N/A

manual

manual for standard fluoro, motorized for 3D

manual

manual

electromagnetic

electromagnetic

manual

135 deg. 90°/+45°

90°/+45°

130° (+90°, -40°)

130° (+90°, -40°)

190° (+95°, -95°)

132° (+90°, -42°)

90°/+30°

Page 14 of 26

Rotated image appears after ? movement of onscreen indicator 3 C-arm Designs: a) 4 C-arm Designs: a) 9" I.I., b) 12" I.I., c) 9" I.I., b) 12" I.I., c) "Super C" (only with "Super C" (only with 31"

31"

26" (9" I.I.), 28" (12" 26" (9" I.I.), 28" (12" I.I.), 33" (Super C) I.I.), 33" (Super C)

69.8" H x 75.9" L x 69.8" H x 75.9" L x 33" W (9"I.I.); 72.5" 33" W (9"I.I.); 72.5" 69.5" H x 69.8" L x H x 78.8" L x 33" W H x 78.8" L x 33" W 33" W (12" I.I.); 70.5" H x (12" I.I.); 70.5" H x 78.3" L x 33"W 78.3" L x 33"W (Super C) (Super C) not specified

not specified

503 lbs

64" H x 27" W x 27" 64" H x 27" W x 27" 69.5" H x 35.5" W x D D 29.1" D Motorized C-arm Motorized C-arm Option ($60K) Option ($60K) provides motorized provides motorized manual manual manual

135°

Standard and Standard and Medium C: Medium C: 115°(90/25°), Super 115°(90/25°), Super C = 148° (93°/55°) C = 148° (93°/55°)

Siemens Vision FD

Vario 3D

pulse based

real-time fluoro

5,000

1,000

up to 55,000

Optional up to 4,850 (3D recon. Memory separate from 4,850 memory)

1-10 frames/s, 1-15 frames/s, 1-30 frames/s

Playback via touch screen at monitor cart or chassis. Auto playback

GE OEC

Compact L

ARCADIS Varic

ARCADIS Orbic

ARCADIS Avantic

200 std., 2,000 and 5,000 optional

10,000

10,000

40,000

1-8 frames/s, 8-12.5 frames/s

0.5,1,2,3,4,5,6

1,2,3,5,8 (15 optional)

1,2,3,5,8 (15 optional)

Cine loop start, stop and replay rate controls (at workstation)

Playback via manual button at chassis or monitor cart

Playback through syngo interface at workstation

Playback through syngo interface at workstation

Bone, Heart, Abdomen, Endo, Motion, Metal, Soft, a) children, extremities b) 5 Per User Mode (4) Large Patient, Automatic spine, hip c) heart or lung + 1x Roadmapping Motion Detection, (preconfiguration done by = 21 Automatic Metal technical staff) Correction

Yes

N/A

1/2 dose button

1/2 dose button

200 Exam Sets

200 Exam Sets

Fluorostar

FlexiView 8800

9900 Elite

63 (General 63 (General Surgical) 400 (all Surgical) 400 (all others) others) 400 image storage 400 image storage 400 image storage USB storage device on ESP & Vascular on ESP & Vascular on ESP & Vascular port, optional CD pkgs. Dynamic packages. Dynamic packages. Dynamic storage device, image storage = image storage = image storage = optional DICOM 3.0 2,400 images with 8 960 images with 8 960 images with 8 f/s disk and 9,000 f/s disk, 9,000 f/s disk, 9,000 60,000

63 (GSP = Basic Version)

1,2,3,5,8 (15 & 30 optional)

N/A

1,2,4,8 BasicVasc; 1,2,4,8,15 Vascular

Autoplayback

N/A

Autoplayback

200 Exam Sets

Series 9800 Plus

limited

None

1,2,4,8 BasicVasc; 1,2,4,8 BasicVasc; 1,2,4,8,15 Vascular; 1,2,4,8,15,30 1,2,4,8,15,30 Vascular MTS & NeuroVasc & Cardiac Cardiac

Autoplayback

Autoplayback

None

Dynamic Range Management (DRM) for General, Cardiac, Ortho, Vascular, Bolus Chase & Spine

The Auto Trak feature uses a moving sampling window to "find" patient anatomy within the imaging field and adjust technique accordingly. Yes

Yes

Yes

Page 15 of 26

Yes

"low dose" mode

"low dose" key

"low dose" key

"low dose" key

Siemens Vision FD

Vario 3D

pulse based

real-time fluoro

yes (5 Steps - for Realtime and Post Processing)

GE OEC

Compact L

ARCADIS Varic

ARCADIS Orbic

yes (5 Steps - for Realtime and Post Processing)

Yes, Realtime & Post-Processing

Yes; Realtime & Post-Processing

toggle through 4 steps level/window

toggle through 4 steps level/window

4 Steps + Auto Window

4 Steps + Auto Window

4 Steps + Auto Window

Standard

Standard

Yes

Yes

Yes

Yes

Yes

yes (7 levels)

yes (7 levels)

Vascular Pkg. required

Vascular Pkg. required

Yes Yes

ARCADIS Avantic

Fluorostar

FlexiView 8800

Series 9800 Plus

9900 Elite

Yes

only with ESP Package

Yes

Yes

5 Steps + Auto Window

No

No

No

No

Yes

Yes

Not Specified

Yes. (not on GSP pkg.)

Yes

Yes

Yes

Yes

Yes

Yes

Yes. Not on GSP pkg.)

Yes

Yes

Yes, 2 x

Yes, continously variable

Yes, continously variable

Yes, continously variable

Yes, up to 1600%

Yes Yes

Vascular Pkg. required Yes Yes

Vascular Pkg. required Yes Yes

Vascular Pkg. required Yes Yes

Vascular Pkg. required Yes Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Optional

No

Yes

Yes,Realtime & Post- Yes,Realtime & PostProcessing Processing

Yes, 2X and 4X (not Yes, 2X and 4X (not Yes, 2X and 4X (not on GSP pkg) on GSP pkg) on GSP pkg)

No No

Vascular Pkg. required Yes Yes

Vascular Pkg. required Yes Yes

Vascular Pkg. required Yes Yes

Yes

No

Yes

Yes

Yes

Yes

Yes

No

Yes

Yes

Yes

Yes

Yes

Yes

No

Yes

Yes

Yes

Yes

Yes

Yes

Yes

No

No

No

No

Yes

Yes

Yes

Yes

Yes

No

Yes

Yes

Yes

Yes, with Vascular Module

No

No

No

No

No

No

Yes

Yes

Yes

1 or 2 lengths

1 or 2 lengths

No

Yes

Yes

Yes

No

Yes

Yes

Yes

1 or 2 angles

1 or 2 angles

No

Yes

Yes

Yes

No

no

no

no

Page 16 of 26

Siemens Vision FD

Vario 3D

pulse based

real-time fluoro

"Vision Center": Same UI at Chassis as well as Monitor trolley (Extra Touch Screen)

GE OEC

Compact L

ARCADIS Varic

ARCADIS Orbic

ARCADIS Avantic

Keyboard (Monitor cart), Keyboard on Carm Chassis, Touch panel monitor at C-arm for 3D control

Keyboard (Monitor trolley) & Chassis

Syngo Interface, Keyboard and mouse (Monitor trolley) & Chassis

Syngo Interface, Keyboard and mouse (Monitor trolley) & Chassis

Syngo Interface, Keyboard and mouse (Monitor trolley) & Chassis

Search function via 16 image mosaic display

Search function via 16 image mosaic display (mosaic archiving with vascular configuration)

Yes 4 x 4

DVD-RW 4.7 GB capy.DICOM, TIFF, JPG, Multimedia

yes CD (1000 in TIFF /DICOM format)

No

yes, in DICOM & BMP formats

yes, in DICOM & BMP formats

USB Hub/Port

USB Hub/Port

MOD Option: 900 dicom format images

No

Storage, Storage Commit, Storage, Storage Commit, Print, Media Class, Print, Media Class, Send, Print, Worklist Worklist, Query-Retrieve, Worklist, Query-Retrieve, Verification Class Verification Class

Fluorostar

FlexiView 8800

Series 9800 Plus

9900 Elite

2 Touchscreens

Keyboard (Monitor trolley), Touchscreen (Monitor), Chassis (left & right)

Keyboard (Monitor trolley), Touchscreen (Monitor), Chassis (left & right)

Keyboard (Monitor Trolley) with touchpad, Touchscreen Monitor, Chassis (left & right)

Not Specified

9x

16x

16x

yes, in DICOM & BMP formats

Optional

No

Option, external drive

Onboard CD/DVD

No

No

USB storage device port

PC Memory Card

PC Memory Card

Send, Print, Worklist, MPPS, Storage Commit, Multi-Modality Query-Retrieve

Send, Print, Worklist, MPPS, Storage Commit, Multi-Modality Query-Retrieve

Send, Print, Worklist, MPPS, Storage Commit, Multi-Modality Query-Retrieve

DICOM 3.0 optional. Functions Not Print, Send, Worklist Print,Send, Worklist Specified.

Store, Print, Worklist Query

Sony 980 / 960, Sony 980 / 960, Sony thermal printer Codonics EP, Sony Codonics EP, Sony UP-D71XR UP-D71XR

Integrated film/paper printer, thermal printers

Yes, selectable 2x2, Yes, selectable 2x2, Yes, selectable 2x2, 3x3, 4x4 3x3, 4x4 3x3, 4x5

paper and film: Sony UP 980/990

Paper: sony 960/970 paper and film: Sony UP 980/990

Sony 980 / 960

Sony UP-D72

Sony UP-D72

Sony UP-D72

No

No

No

No

No

No

N/A

Option

Option

Option

Option

Option

Option

Option

Option

Option

Optional

No

No

No

Option

Option

Option

Option

Option

Option

Optional

Option

Option

Option

Page 17 of 26

Siemens Vision FD

Vario 3D

pulse based

real-time fluoro

No

No

GE OEC

Compact L

ARCADIS Varic

ARCADIS Orbic

ARCADIS Avantic

Fluorostar

FlexiView 8800

Series 9800 Plus

9900 Elite

No

No

Yes

Yes

Yes

No

No

No

"remote assessment"

YES, with 3D package

no

No

Yes. Requires 3D package. Performs 3D reconstructions from motorized

No

No

no

Page 18 of 26

Yes. Requires 3D Yes. Requires 3D navigation system. navigation system. Performs 3D Performs 3D reconstructions from reconstructions from

Philips BV Libra

BV Endura

BV Pulsera

3.15 kW

3.15 kW

7.5 kW

40 - 105 kV

40 - 105 kV

40 - 120 KV

0.1 - 7.2 mA

0.1 - 7.2 mA

0.1 - 8.3 mA. Up to 10mA HLF, up to 20 mA for High Definition Fluoro Mode

N/A

N/A

up to 75 mA "sharpshot" mode

Not specified

Not specified

up to 60 mA

1 pps

1, 2 pps

6.25 & 12.5 pps (pulsed fluoro, quarter & half dose mode)

No

No

3 - 30 pps with Cardiac Pulse Package

N/A

N/A

8 - 11 ms

Pulsera R2?

Page 19 of 26

Philips BV Libra

BV Endura

BV Pulsera

Philips

Philips

Philips

Tri-mode 9"/7"/5"

Tri-mode 9" (9/7/5") or 12" (12/9/7")

Tri-mode 9" (9/7/5") or 12" (12/9/7")

Not specified

Not specified

Not specified

Clinton

Clinton

Varian

0.6

0.6

0.3 & 0.6

Stationary

Stationary

Rotating Monobloc

50,000 HU

50,000 HU

300,000 HU

1,200,000 HU

1,200 000 HU

1,900,000 HU

30,600 HU/Min

30,600 HU/Min

70,000 HU/Min

yes

yes

yes

Pulsera R2?

Page 20 of 26

Philips BV Libra

No, full lead shutters

BV Endura

BV Pulsera

Pulsera R2?

No, 2 independently No, 2 independently moveable full lead moveable lead shutters shutters with wedge tip

Yes, both iris and Yes, both iris and Yes, both iris and parallel parallel shutter parallel shutter shutter position and position and rotation position and rotation rotation indicators on indicators on fluoro indicators on fluoro fluoro display display display

CCD

CCD

CCD

768 x 494

768 x 494

1024 x 1024

18" LCD

18" LCD

18" LCD

350 cd/m2

350 cd/m2

350 cd/m2, 500 cd/m2 optional

30.3"

30.3"

30.3"

25"

24"

24"

34.6"

34.6"

34.6"

605 lbs.

660 lbs.

672 lb.(9"), 683 lb.(12")

Page 21 of 26

Philips BV Libra

BV Endura

BV Pulsera

72" H x 72.4"L x 34.6" (9" I.I.)

72" H x 72.4"L x 34.6" (9" I.I.), 74.4" H x 72.4" L x 34.6" W (12" I.I.)

9": 68.5" H x 76" L x 31.9" W

Not Specified

Not Specified

Not Specified

74" H x 35.8" W x 27.6" D

74" H x 35.8" W x 27.6" D

74" H x 35.8" W x 27.6" D

manual

manual

manual

115° (+90 / -25°)

135° (+90 / -45°)

115° (+90 / -25°) / extended rotation option: 135°

Pulsera R2?

Page 22 of 26

Philips BV Libra

BV Endura

BV Pulsera

16

16

10,000 max

Pulsera R2?

Extension: 500 Extension: 50, 500 (5F/s) or 5,000 500 (5 fps) 5,000 (8fps) (3 F/s) or 1000 (5 (8F/s) or 10,000 (30 10,000 (30 fps) F/s) F/s)

1,2,3,5,8,15

1,2,3,5,8,15

3,5,8,15,30

Playback via manual button at monitor cart

Playback via manual button at monitor cart

Playback via manual button at monitor cart

10 freely programmable organprograms

yes: Anatom.Progr.Fluor o (APF)

yes: Anatom.Progr.Fluoro (APF)

Philips' Body Smart" feature uses a grid of fewer sampling windows to "look" for the location of actual anatomy within the imaging field. In cases where an orthopedic physician may not have centered an extremity correctly, this feature assists in providing the user with an appropriate exposure.

No

No

No

Yes

Yes (Realtime & Postprocessing)

Yes (Realtime & Postprocessing)

No

No

Not specified

Yes

Yes

Yes

option: extended option: extended processing package processing package

Yes (option: extended processing package) Page 23 of 26

Philips BV Libra

BV Endura

option: extended option: extended processing package processing package

BV Pulsera

Pulsera R2?

Yes (option: extended processing package)

Vascular Pkg. required Yes Yes

Vascular Pkg. required Yes Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Not Specified

Not specified

Not specified

Yes

Yes

Yes

Yes

Yes

Yes

Vascular Pkg. required Yes Yes

Page 24 of 26

Philips BV Libra

Keyboard with Trackball (Monitor trolley), Chassis

BV Endura

BV Pulsera

Pulsera R2?

Keyboard with Keyboard with Trackball Trackball (Monitor (Monitor trolley), Chassis trolley), Chassis

16 x (only if memory 16 x (only if memory extension is extension is chosen) chosen)

16 x (only if memory extension is chosen)

yes - option

yes - option

yes - option

Print, Send, Worklist. Optional interface includes Storage Commit, MPPS

Print, Send, Worklist. Optional interface includes Storage Commit, MPPS

Multimodality Query Retrieve via View Forum onboard PC. Print, Send, Worklist. Optional interface includes Storage Commit, MPPS

Sony 980 / 960

Sony 980 / 960

Sony 980 / 960

No

Option

Option

Option

Option

Option

Option (for 9"I.I. only)

Option (for 9"I.I. only)

Option (for 9"I.I. only)

Page 25 of 26

Philips BV Libra

BV Endura

BV Pulsera

No

No

No

no

no

3-DRX option for 3D volume acquisition and reconstruction

Pulsera R2?

Page 26 of 26