Colonic perforation from left ventricular assist device - Semantic Scholar

12 downloads 0 Views 221KB Size Report
Meigs J, Moy C, Nichol G, O'Donnell C, Roger V, Sorlie P, Steinberger J,. Thom T, Wilson M, Hong Y; American Heart Association Statistics. Committee and ...
ARTICLE IN PRESS doi:10.1510/icvts.2010.237966 Editorial

www.icvts.org

Case report - Cardiac general

Department of Anesthesiology, Perioperative and Pain Medicine, Brigham and Women’s Hospital, 75 Francis Street, Boston, MA 02115, USA b Division of Cardiac Surgery, Brigham and Women’s Hospital, 75 Francis Street, Boston, MA 02115, USA c Division of General Surgery, Brigham and Women’s Hospital, 75 Francis Street, Boston, MA 02115, USA

a

Abstract

Historical Pages Brief Case Report Communication

䊚 2010 Published by European Association for Cardio-Thoracic Surgery

Nomenclature

*Corresponding author. Department of Anesthesiology, Perioperative and Pain Medicine, Brigham and Women’s Hospital, 75 Francis Street, Boston, MA 02115, USA. Tel.: q1 917 749-8332; fax: q1 617 582-6131. E-mail address: [email protected] (S.C. Herman).

Best Evidence Topic

A 50-year-old man with idiopathic non-ischemic cardiomyopathy underwent implantation of an intracorporeal

State-of-the-art

2. Case presentation

Follow-up Paper

Over 5.3 million Americans are affected by heart failure each year, but only approximately 2200 heart transplantations are performed for patients with end-stage heart failure due to the lack of suitable organs w1x. Left ventricular assist devices (LVAD) are often used to bridge endstage congestive heart failure patients to cardiac transplantation or as a destination therapy. During this time, the LVAD functions as the pump of the failing left ventricle. The body of the pump is frequently implanted in the left upper quadrant of the abdomen using an intraperitoneal or preperitoneal technique. Though extremely reliable, LVADs are associated with complications. The main complications precluding transplantation after LVAD placement are related to sepsis, postoperative bleeding, and thromboembolism w2, 3x. Although infectious complications following LVAD implantation are common, LVAD erosion leading to perforation of the gastrointestinal tract and sepsis is only rarely reported in the medical literature w2– 5x. This is the first report describing a LVAD-associated colonic perforation despite the use of a Gore-Tex䊛 wrap, implemented as a result of prior reports of perforation.

Thoratec HeartMate XVE LVAD, while awaiting cardiac transplantation. The LVAD was implanted in the peritoneal cavity. A Gore-Tex䊛 Dual Mesh Plus sheet was sutured along the left gutter, the inferior aspect of the diaphragm, and along the posterior right rectus fascia in order to effectively compartmentalize the device from intra-abdominal organs. This pocket housed the entire LVAD except the intraperitoneal section of the drive-line. The large and thick omentum was not dissected off the transverse colon. The tip of the omentum was wrapped around the intraperitoneal portion of the drive-line and sutured up to the peritoneum around the line. During the ensuing days and weeks, numerous reoperations were necessary for bleeding in the chest and abdomen, and for implantation and explantation of a centrifugal right (RVAD) and its cannulae. On the 25th day after LVAD implant, an AXIOM silicone drain was placed into a hematoma behind the Gore-Tex䊛 pocket. It was removed within a week upon resolution of the hematoma. The patient’s recovery was also complicated by pulmonary embolism and recurrent infections, including persistent Candida fungemia and methicillin-sensitive Staphylococcus aureus requiring chronic suppressive antibiotic treatment. The patient was admitted to our institution on multiple occasions in the seven months following LVAD implantation with hypotension, high-grade fevers, and positive blood cultures. On all admissions, no obvious vegetations were identified by transesophageal echocardiography (TEE), and imaging of the LVAD and sternal wound was equally unrevealing. On postoperative day 234 from initial LVAD implantation, the patient demonstrated a similar clinical picture of

Negative Results

1. Introduction

Proposal for Bailout Procedure

Keywords: Circulatory assist devices; Colonic perforation; Sepsis

ESCVS Article

A patient with idiopathic non-ischemic cardiomyopathy had a left ventricular assist device (LVAD) implanted, while awaiting cardiac transplantation. The patient had been stable following a complicated and prolonged postoperative course, but was admitted to the hospital for suspected low-grade LVAD-related infection. Work-up for sepsis was suspicious for perforated viscus. An exploratory laparotomy revealed a perforated transverse colon with gross spillage of succus. Although infectious complications following LVAD implantation are common, sepsis as a result of gastrointestinal perforation secondary to LVAD erosion is not. This first report of viscus perforation despite the use of a Gore-Tex䊛 wrap highlights a rare complication of LVAD therapy. 䊚 2010 Published by European Association for Cardio-Thoracic Surgery. All rights reserved.

Institutional Report

Received 16 March 2010; received in revised form 26 May 2010; accepted 1 June 2010

Protocol

Sara C. Hermana,*, Jochen D. Muehlschlegela , Gregory S. Couperb, Edward Kellyc

Work in Progress Report

Colonic perforation from left ventricular assist device: a rare complication

New Ideas

Interactive CardioVascular and Thoracic Surgery 11 (2010) 369–370

ARTICLE IN PRESS 370

S.C. Herman et al. / Interactive CardioVascular and Thoracic Surgery 11 (2010) 369–370

refractory hypotension, fluctuating fevers as high as 1038F, and positive blood cultures. TEE was again negative for vegetations, but computed tomography (CT)-scan of the chest showed gas around the LVAD pocket with an air-fluid level suspicious for underlying infection or fistula. On postoperative day 236, the patient was transferred to the operating theater for emergent exploratory laparotomy. Exploration revealed that the colon had migrated behind an intact Gore-Tex䊛 wrapped LVAD. After the Gore-Tex䊛 curtain was dissected away from the abdominal structures, a 3-cm colonic perforation, clearly caused by the weight of the device on the colon, was identified (Fig. 1). A partial colectomy was performed and an endoscopy was normal aside from an impression of the LVAD on the stomach. Given the patient’s severe sepsis, the abdomen was closed temporarily and the patient returned to the cardiothoracic intensive care unit on multiple vasopressors. The LVAD pocket was eventually reconstructed with Strattice bioprosthetic mesh after multiple abdominal debridements and antibiotic irrigation. An ileostomy was created and feeding jejunostomy was inserted during subsequent surgeries. However, the patient’s condition continued to worsen and he stopped responding to treatment. At this point, the patient and family determined to proceed with comfort measures only. The patient expired at home with hospice care approximately nine and a half months after the initial LVAD implantation.

be at particular risk due to the mild rotation of the pump housing associated with electric motor function. Previous reports of gastrointestinal perforation prompted the use of a Gore-Tex䊛 wrap during device implantation to protect our patient’s adjacent viscera from damage w3–5x. Unfortunately, even this extra precaution did not prevent colonic perforation. In our estimation, no other preventative surgical measures could have been taken with this type of LVAD implant. As LVAD therapy continues to improve patient outcomes w6–8x, more implanted devices will be utilized as long-term treatments for end-stage congestive heart failure w8x. Patients undergoing bridge-to-transplant therapy with VAD devices are usually supported with pulsatile, volume-displacement devices like the HeartMate XVE. The development of second-generation axial flow pumps has overcome many of the limitations of the first-generation pulsatile flow devices. Recent reports of improved survival with newer continuous flow devices have demonstrated less mortality compared to older, larger pulsatile pumps w9x. However, despite significant reduction in size and weight of new devices, serious complications from implantation continue to account for significant patient morbidity in second-generation devices w8–10x. These experiences highlight the importance of ongoing research efforts to strengthen patient selection criteria and advance device technology in order to fashion both a safe and effective prosthetic cardiac support system.

3. Comment

References

Despite ongoing improvements in LVAD technology, implantation of ventricular-assist devices continues to be associated with adverse events. Infectious complications following an LVAD implantation are common, but we describe a case of viscus perforation secondary to LVAD erosion leading to septic shock and eventually death. The literature acknowledges the potential for serious injury to adjacent organs secondary to the physical design, weight, and anatomical location of the implanted device w2–5x. Patients supported with the HeartMate XVE are thought to

Fig. 1. Intact Gore-Tex䊛 wrap dissected away from left ventricular assist device (LVAD) and abdominal structures revealing 3 cm colonic perforation.

w1x Rosamond W, Flegal K, Furie K, Go A, Greenlund K, Haase N, Hailpern SM, Ho M, Howard V, Kissela B, Kittner S, Lloyd-Jones D, McDermott M, Meigs J, Moy C, Nichol G, O’Donnell C, Roger V, Sorlie P, Steinberger J, Thom T, Wilson M, Hong Y; American Heart Association Statistics Committee and Stroke Statistics Subcommittee. Heart disease and stroke statistics – 2008 update: a report from the American Heart Association Statistics Committee and Stroke Statistics Subcommittee. Circulation 2008;117:e25–e146. w2x Piccione W. Left ventricular assist device implantation: short and longterm surgical complications. J Heart Lung Transplant 2000;19:S89–S94. w3x Phillips WS, Burton NA, Macmanus Q, Lefrak EA. Surgical complications in bridging to transplantation: the Thermo Cardiosystems LVAD. Ann Thorac Surg 1992;53:482–486. w4x Yannopoulos D. Subacute gastric perforation caused by a left ventricular assist device. World J Gastroenterol 2007;13:3253–3254. w5x Hou JK, Hampel H, Lukens FJ. Gastric ulceration and perforation as a complication of a left ventricular assist device. Gastrointest Endosc 2005;61:629–631. w6x Long JW, Kfoury AG, Slaughter MS, Silver M, Milano C, Rogers J, Delgado R, Frazier OH. Long-term destination therapy with the HeartMate XVE left ventricular assist device: improved outcomes since the REMATCH study. Congest Heart Fail 2005;11:133–138. w7x Long JW, Healy AH, Rasmusson BY, Cowley CG, Nelson KE, Kfoury AG, Clayson SE, Reid BB, Moore SA, Blank DU, Renlund DG. Improving outcomes with long-term ‘‘destination’’ therapy using left ventricular assist devices. J Thorac CardioVasc Surg 2008;135:1353–1360; discussion 1360–1361. w8x Badiwala MV, Rao V. Left ventricular device as destination therapy: are we there yet? Curr Opin Cardiol 2009;24:184–189. w9x John R, Kamdar F, Liao K, Colvin-Adams M, Boyle A, Joyce L. Improved survival and decreasing incidence of adverse events with the HeartMate II left ventricular assist device as bridge-to-transplant therapy. Ann Thorac Surg 2008;86:1227–1235. w10x Feller ED, Sorensen EN, Haddad M, Pierson RN, Johnson FL, Brown JM, Griffith BP. Clinical outcomes are similar in pulsatile and nonpulsatile left ventricular assist device recipients. Ann Thorac Surg 2007;83:1082– 1088.