Aerosol Science and Technology, 38:382–390, 2004 c American Association for Aerosol Research Copyright  ISSN: 0278-6826 print / 1521-7388 online DOI: 10.1080/02786820490442626

Determination of Total Aerosol Nitrogen by Thermal Evolution Qianfeng Li and Jian Zhen Yu Department of Chemistry, Hong Kong University of Science & Technology, Clear Water Bay, Kowloon, Hong Kong

An aerosol nitrogen analyzer has been developed for determination of total nitrogen (TN) content in aerosols collected on a filter substrate. It uses rapid thermoevolution of nitrogenous species at 800◦ C in a 2.5% O2 /97.5% He carrier gas. Evolved nitrogen is oxidized to nitrogen oxides on a manganese dioxide catalyst, converted to nitrogen monoxide on a molybdenum catalyst, and quantified with a chemiluminescence detector. The analyzer is built upon components of two commercially available instruments, a thermal aerosol carbon analyzer and a chemiluminescent NOx analyzer. This system is able to provide fast (3 min per sample) and highly sensitive (a detection limit of 26 ng N) TN measurements for aerosol samples of a small size (∼1 × 1 cm2 ) without any pretreatment. When the aerosol nitrogen analyzer was properly calibrated by ammonium sulfate, excellent agreement was obtained between TN measurements of twelve aerosol samples by the aerosol nitrogen analyzer and those obtained from an elemental analyzer. This instrument provides a useful and convenient tool for characterization of the organic nitrogen (ON) pool in atmospheric aerosols. The ON concentrations in the twelve aerosol samples, calculated from the difference between TN and the sum of the inorganic nitrogen species determined by ion chromatographic analysis, were found comparable in magnitude to particulate nitrate-N. The contribution of ON to TN ranged from 1.7% to 54% for the limited set of test samples. While twelve samples are insufficient to draw general conclusions about the relative distribution of ambient aerosol nitrogen species, they have demonstrated the applicability of the new method to ambient aerosol characterization and exposed a potential limitation in deriving at ON concentrations by taking a sometime small difference between two large numbers (i.e., TN and inorganic nitrogen). Differentiation of organic and inorganic nitrogen using thermal methods is also explored and found not to be feasible.

Received 12 August 2003; accepted 3 February 2004. This work was supported by the Research Grants Council of Hong Kong, China (DAG 00/01.SC14 and N HKUST612/01). The authors thank Eric C. H. Wan and Hong Yang for collecting the aerosol samples, Jinhui Xu for providing the IC analysis, and the Marine Coastal Lab at HKUST for allowing us to use its elemental analyzer. Address correspondence to Jian Zhen Yu, Department of Chemistry, Hong Kong University of Science & Technology, Clear Water Bay, Kowloon, Hong Kong. E-mail: [email protected]

382

INTRODUCTION Atmospheric particles can be an important source of fixed nitrogen in the environment because of their typically high nitrogen content (Finlayson-Pitts and Pitts 2000; Seinfeld and Pandis 1998). A large number of N-containing species have been identified to exist in ambient aerosols. To date, most studies have focused on inorganic nitrogen (IN) species. The IN species – – mainly are ammonium (NH+ 4 ), nitrate (NO3 ), and nitrite (NO2 ) (Seinfeld and Pandis 1998). They are most commonly determined by ion chromatography (IC) or flow injection analysis (Luke and Valigura 1997). These measurement techniques are established, and they have been well tested by researchers in this field. Organic nitrogen (ON) appears to be a ubiquitous and significant component in the atmospheric nitrogen budget (Cornell et al. 1995), but it is much less studied, partly due to a lack of simple analytical methods. ON is usually determined as the difference between total nitrogen (TN) and IN. As such, TN measurement techniques provide indirect solutions for ON determination. Several methods for the determination of TN in aqueous solutions are currently in widespread use. They include persulfate oxidation, UV photo-oxidation, and high-temperature oxidation or catalytic oxidation (Bronk et al. 2000). These methods fall into two categories, wet methods and combustion-based methods. In the first category, ON species are converted into IN ions either by the use of chemical oxidants such as persulfate or by UV photolysis with or without addition of oxidants (Bronk et al. 2000; Solorzano and Sharp 1980). The wet methods have the drawback of being laborious and often suffer high and variable nitrogen blank levels caused by the oxidization agents (Cornell and Jickells 1999). In the second category, the methods involve oxidation of nitrogenous species by high-temperature oxidation or catalytic oxidation, followed by detection of the N gas products (Suzuki et al. 1985; Sugimura and Suzuki 1988; Hansell 1993; Ammann et al. 2000; AlvarezSalgado and Miller 1998). These combustion-based methods analyze the entire sample content without sample preparation. They have the advantages of being simple and rapid and require

DETERMINATION OF TOTAL AEROSOL NITROGEN

a smaller amount of the sample, although the instrumentation is less readily available. Historically, the methods in both categories have been developed for the analysis of aqueous samples, particularly seawater matrices. Their applications to atmospheric samples are limited to rainwater and water extracts of aerosol samples (Cornell et al. 1995; Cornel and Jickells 1999; Zhang and Anastasio 2001; Zhang et al. 2002). Direct analysis of solid matrices, such as aerosol-laden filters, is not accommodated by methods in either category. On the other hand, combustion-based methods are routinely used in the direct analysis of the carbon content in aerosols (e.g., Chow et al. 1993; Birch and Cary 1996). These methods involve evolution of carbonaceous materials through thermal decomposition, volatilization, and oxidation under programmed heating followed by catalytic oxidation to CO2 for detection. A limited number of studies have explored the combustion-based approach for the determination of nitrogen in aerosols. Dod and his colleagues have demonstrated the feasibility of combining thermal evolution techniques and chemiluminescence detection for the measurement of aerosol nitrogen (Dod and Novakov 1982; Dod et al. 1984; Daisey and Gundel 1991), but no detailed method characterization and validation is available in their reports. More recently, Stolzenburg and his coworkers (Stolzenburg and Hering 2000; Stolzenburg et al. 2003) reported an automated method for one group of aerosol nitrogenous compounds, i.e., nitrates. Their method uses flash vaporization and selectively converts nitrate compounds in the aerosols to nitrogen oxides for subsequent chemiluminescent detection. We describe here a simple aerosol TN analysis system, called the aerosol nitrogen analyzer (ANA)

383

hereafter, which employs flash heating of aerosol-laden filters in a 2.5% O2 /97.5% He carrier gas with subsequent catalytic conversion to nitrogen monoxide for chemiluminescent detection. The system is built upon components of two commercially available instruments, a combustion-based aerosol carbon analyzer (Sunset Laboratory, Forest Grove, OR, USA) and a chemiluminescent NOx analyzer (Model 200A, API, San Diego, CA, USA). Presented below are details about the instrument and characterization of the instrument’s performance. Also shown are example data of aerosol TN and ON measurements made with the instrument. EXPERIMENTAL SECTION Apparatus The aerosol nitrogen analyzer is designed for determination of microgram levels of aerosol nitrogen. A block diagram of the instrument is shown in Figure 1. The major components of this apparatus are the following: 1. The quartz tube (0.5 cm o.d., 12 cm in length) is divided into two zones, the front oven zone and the back oven zone, each 4 cm in length and wrapped in separate 20-gauge nichrome wires. The front oven accepts a rectangular filter (1–1.5 cm2 ) cut from an aerosol-laden quartz filter. The back oven houses the manganese dioxide (MnO2 ) catalyst. A filter cut is first placed in the front oven, after which the system is closed and flushed with a 2.5% O2 /97.5% He mixture at 100 mL/min at room temperature. During analysis, the temperature of the front oven is programmed to allow thermal evolution of the aerosol materials into the

Figure 1. Block diagram of the aerosol nitrogen analyzer.

384

2.

3.

4.

5.

Q. LI AND J. Z. YU

carrier gas stream. The back oven temperature is fixed at a preset, programmable value, which is 870◦ C for all ensuing analyses. The MnO2 catalyst in the back oven oxidizes the nitrogen-containing gases that have evolved from the front oven to nitrogen oxides (e.g., NO, NO2 ). After the heating zones, the O2 /He mixture gas mix with a separate air stream supplied at ∼400 mL/min in the mixing box. The molybdenum converter is a stainless steel cartridge containing molybdenum chips heated to 315◦ C. The converter reduces various forms of higher nitrogen oxides (e.g., NO2 , N2 O5 ) to nitric oxide (NO) when the sample stream is routed through the heated molybdenum. The ozone generator subsystem consists of a 47 mm particulate filter, a permeation drier, and a generator module. Ozone is needed for the chemiluminscent detection of NO as described below. Excessive ozone is destroyed in a catalytic ozone scrubber. The detector module consists of a reaction cell, a photo multiplier tube (PMT), and a preamplifier assembly. In the reaction cell, ozone mixes with the NO gas generated after the carrier gas passes through the Mo converter to produce electronically excited NO2 molecules. The excited NO2 molecules emit chemiluminescent light when they drop to a lower energy level. The chemiluminescent light intensity is directly proportional to the NO concentration. Parts of a commercial chemiluminescence NOx analyzer are modified to provide components 2–4. The data acquisition module (Microlink 751, Biodata Limited, Manchester, UK) collects the analog voltage signal from the preamplifier assembly at a frequency of 0.2 Hz and sends the data to the computer for display. The microprocessor in the NOx analyzer has to be bypassed to achieve this high frequency.

Elemental Analyzer A PE2400 series II CHNS/O analyzer (Perkin Elemer, Shelton, CT, USA) was used in its CHN mode to provide independent N measurements for evaluating the accuracy of the measurements by the ANA. This elemental analyzer uses a combustion method to convert the measured elements (C, H, and N) to simple gases, i.e., CO2 , H2 O, and N2 . These gases are then quantified as a function of thermal conductivity. A known standard, acetanilide (C8 H9 NO), is first analyzed to calibrate the analyzer. The resulting calibration factor is then used to determine the samples. A filter cut sized 1 × 1 cm is placed on a tin disk, folded, and rolled into a cylinder. The cylinder is then dropped into the analyzer for elemental analysis. The elemental analyzer has a detection limit of 0.68 µgN. Reagents and Aerosol Materials The standard nitrogenous compounds for testing include ammonium chloride, potassium nitrate, sodium nitrite, ammonium nitrate, ammonium sulfate, ammonium hydrogen sulfate, methy-

lamine, urea, and a mixture of 20 individual amino acids. Ammonium hydrogen sulfate was made by mixing ammonium sulfate and sulfuric acid in stoichiometric proportions. The amino acid mixture was purchased from Waters (Milford, MA) and all the other standards were from Acros (Geel, Belgium). All chemicals were reagent grade or better and used without further purification. Aerosol samples were collected on quartz fiber filters (Pall Gelman, Ann Arbor, MI, USA) from three locations, Jeju Island, S. Korea, Hong Kong, and Nanjing, China. All quartz filters were prebaked in a muffle furnace at 550◦ C overnight before use. The Jeju samples were collected onto 20 cm quartz filters at a flow rate of 0.50 m3 /min using a high-volume particle trap impactor/denuder sampler with an impactor inlet to remove particles of larger than 5 µm (Mader et al. 2001). The Hong Kong and the Nanjing samples were collected onto 20 × 25 cm quartz filters at a flow rate of 1.13 m3 /min using a PM2.5 high-volume sampler (Andersen Instruments, Smyrna, GA, USA). All of the samples were collected under ambient temperature and relative humidity. No attempts were made to remove gaseous nitric acid upstream the filters. Determination of Organic and Inorganic Nitrogen − Concentrations of the IN species in aerosols, NH+ 4 , NO3 , − and NO2 , were determined by injecting water extracts of the aerosol samples into an ion chromatograph (DX500, Dionex, Sunnyvale, CA, USA). Each water extract was obtained by extracting 2–4 pieces of filter strips sized in 1 × 1.45 cm with two portions of water (2 ml each) in a sonication bath, each extraction for 15 min. The two extractions were then passed through a same syringe filter and combined in a 5.0 volumetric flask. An additional 1 ml of water was then passed through the same syringe filter and combined with the previous two extractions. The final volume of the combined water extracts was made to be 5.0 ml. An IonPac CG12 guard column (4 × 50 mm), an IonPac CS12 analytical column (4 × 250 mm), and an ED40 electrochemical detector were used in the cation analysis with isocratic elution at a flow rate of 1 mL/min. The eluent was 20 mM of methanesulfonic acid. In the anion analysis, an Ionpac AG11 guard column (4 × 50 mm), an IonPac AS11 analytical column (4 × 250 mm), and a CD20 conductivity detector were used. The eluent was NaOH with a programmed gradient from 0.2 to 5 mM, and the flow rate was 2 mL/min. The sampling loop was 50 µL in both cation and anion analyses. ON concentrations were calculated as the difference between TN and the sum of the three IN species. RESULTS AND DISCUSSION Calibration, Precision, and Detection Limit Prior to the commencement of an analysis, the front oven temperature is maintained at ca 50◦ C due to its proximity to the back oven, which is maintained at 870◦ C at all times. Once the filter sample is placed in the oven and the system is closed, the front oven temperature is set to 800◦ C and the analysis starts.

DETERMINATION OF TOTAL AEROSOL NITROGEN

385

Figure 2. N thermoevolution peaks of a few standard nitrogen species and an aerosol sample using a fast thermoevolution temperature program. Analysis conditions: Front oven, 800◦ C; MnO2 catalyst oven, 870◦ C; Mo converter, 315◦ C; Carrier gas flow 500 mL/min. Figure 2 shows the evolved peaks for the various tested standard compounds and an aerosol sample. The standard nitrogenous compounds vary in the N oxidation states and chemical structures. It is noted that they evolve at slightly different times and appear in different peak shapes. Ammonium chloride, methylamine, and urea appear as single symmetric peaks; ammonium sulfate appears as two overlapping peaks; the other tested standards show either leading or tailing peaks. Regardless of the peak shapes, the peak area is proportional to the amount of N present for each individual standard, suggesting that the conversion of N species to NO and the subsequent detection of NO is quantitative. As illustrated in Figure 2, the evolution of N is completed in less than 3 min. This demonstrates that the analyzer is capable of rapid analysis of aerosol N through thermal evolution. Ammonium sulfate is chosen to calibrate the aerosol nitrogen analyzer on the grounds that it is a known major nitrogen species in atmospheric aerosols (Seinfeld and Pandis 1998). The detector calibration factor, called the K-factor hereafter, is defined as the ratio of the peak area of the chemiluminescence signal and the spiked (NH4 )2 SO4 –N amount in µgN. Figure 3 shows an

example response curve of TN in the range of 0.5–50 µgN, introduced into the ANA by syringe injection of known amounts of an (NH4 )2 SO4 aqueous solution onto prebaked quartz filter strips. The detector response is linear up to 15 µgN. The K-factor is determined from the linear regression of the measurements below 15 µgN. The detection limit of the analyzer, calculated from 3SDb /K-factor, is estimated to be 0.026 µgN, where SDb is the standard deviation of replicate measurements of blanks. The relative standard deviation (RSD) for 4–5 replicate analyses of the standard samples is typically less than 5%, even at N levels as low as 0.1 µg N. Validation For assessment of the compound dependence of the instrument response, a number of standard N compounds of known amounts were analyzed to determine their recoveries. They are listed in Table 1. They were selected for analysis because of their established atmospheric presence. The characterization of the molecular composition of ON is still far from being comprehensive and complete. Available data indicate that urea,

386

Q. LI AND J. Z. YU

Figure 3. Response curve of the aerosol nitrogen analyzer to the amount of N in ammonium sulfate. The linear curve is generated using calibration standards of 0.5–15 µgN. aliphatic amines, and amino acids are present in atmospheric particulate matter (Zhang and Anastasio 2001; Zhang et al. 2002; Gorzelska and Galloway 1990; Gorzelska et al. 1997; Cornell et al. 1998). The above-mentioned ON species, together with the few IN compounds, cover the full range of oxidation states of nitrogen (from −3 to +5) and represent a variety of chemical structures. The recoveries of these compounds were determined to range from 80–129% (Table 1). A unity recovery for a given compound indicates that the instrument response for this compound is the same as the instrument response for the calibration compound (NH4 )2 SO4 . Six out of nine compounds had their recoveries deviate less than 6% from unity, demonstrating the

robustness of the oxidation catalyst and the Mo converter. The remaining three compounds, NH4 Cl, KNO3 , and KNO2 , exhibited deviations of –20%, 30%, and –15%, respectively. The recovery, and thereby the instrument response, is apparently dependent on both the N oxidation state and its chemical surroundings in the individual compound (e.g., NH4 Cl versus (NH4 )2 SO4 ). It is conceivable that both factors could affect the resulting composition of the nitrogen gas products and thereby the oxidation efficiency of the MnO2 catalyst and subsequently the efficiency of the Mo converter. The nitrogen pool in real aerosol particulate matter is poorly understood. In addition, the chemical composition of aerosols varies in space and time. Without knowledge of the detailed molecular composition, it is impractical to test every N species possibly present in the aerosols. In addition, it is unknown if the presence of numerous other chemicals alters thermal evolution behaviors and products of a given N compound. Consequently, it is difficult to assess the impact of the few N species that have moderate deviations from the reference compound (NH4 )2 SO4 on the measurement accuracy of the TN content. The alternative approach is to assess the extent of the agreement between measurements made by the ANA and a different technique. Elemental analyzers, which have been routinely used to determine the elemental composition of C, H, and N in chemical compounds for decades (Skoog et al. 1998), provide such an alternative technique for TN measurement. In this connection, we have compared the TN measurements of twelve ambient aerosol samples by the aerosol nitrogen analyzer and those by the Perkin Elemer elemental analyzer. The aerosol samples collected from different locations in the East Asian region were chosen to capture aerosols of different chemical composition. The results in µgN per cm2 of filter area are listed in Table 2 and depicted in Figure 4. The linear regression line in Figure 4 has a

Table 1 Recoveries of a few nitrogenous standards as measured by the aerosol nitrogen analyzer Nitrogenous standard Ammonium chloride Potassium nitrate Sodium nitrite Ammonium nitrate Ammonium sulfate Ammonium hydrogen sulfate Methylamine Urea Amino acid mixtureb

Spiked-N (µg N)

Recovery (n = 3)a (%)

4.69 5.07 5.00 2.74 5.66 5.04 2.95 6.61 4.54

79.9 ± 0.8 129.6 ± 0.6 85.5 ± 1.0 104.1 ± 1.4 99.2 ± 5.6 101.7 ± 1.2 105.1 ± 3.6 94.9 ± 0.4 106.4 ± 1.4

a A unity recovery indicates that a given compound has the same instrument response as the calibration compound (NH4 )2 SO4 . b A mixture of 20 individual amino acids.

Figure 4. Comparison of total nitrogen measurements by the aerosol nitrogen analyzer and the elemental analyzer.

387

1 2 3 4 5 6 7 8 9 10 11 12

Sample no.

Oct. 11, 2002 Oct. 12, 2002 Oct. 13, 2002 Oct. 14, 2002 Oct. 15, 2002 Oct. 16, 2002 Apr. 12, 2002 Apr. 16, 2002 Apr. 28, 2002 Sep. 22, 2001 Sep. 24, 2001 Sep. 28, 2001

Date Hong Kong (PM 2.5) Kong Hong (PM 2.5) Kong Hong (PM 2.5) Kong Hong (PM 2.5) Kong Hong (PM 2.5) Hong Kong (PM 2.5) Jeju, S. Korea (PM 5) Jeju, S. Korea (PM 5) Jeju, S. Korea (PM 5) Nanjing, China (PM 2.5) Nanjing, China (PM 2.5) Nanjing, China (PM 2.5)

Site

Sample information Elemental analyzer 6.21 ± 0.21 6.14 ± 0.24 6.14 ± 0.12 5.04 ± 0.12 8.21 ± 0.60 6.62 ± 0.21 4.35 ± 0.21 2.41 ± 0.12 1.73 ± 0.32 2.90 ± 0.21 11.79 ± 0.42 5.93 ± 0.24

Aerosol N analyzer 6.65 ± 0.34 6.42 ± 0.07 6.49 ± 0.19 4.92 ± 0.03 8.02 ± 0.11 6.72 ± 0.25 4.33 ± 0.02 1.91 ± 0.01 1.71 ± 0.12 2.83 ± 0.07 11.53 ± 0.06 5.81 ± 0.10

TN (µgN/cm2 )

1.75 1.69 1.71 1.29 2.11 1.75 2.09 0.78 0.76 1.49 5.83 3.05

TN (µgN/m3 ) 1.53 1.41 1.44 1.13 2.07 1.70 0.50 0.49 0.31 0.26 4.42 1.81

NH+ 4 -N (µgN/m3 )

Table 2 TN, ON, and IN concentrations in twelve ambient aerosol samples

0.03 0.04 0.04 0.02 0.01