Journal of Clinical Virology 106 (2018) 34–40
Contents lists available at ScienceDirect
Journal of Clinical Virology journal homepage: www.elsevier.com/locate/jcv
Diagnostic accuracy of digital RNA quantification versus real-time PCR for the detection of respiratory syncytial virus in nasopharyngeal aspirates from children with acute respiratory infection
T
⁎
Maiara L. Bouzasa, , Juliana R. Oliveiraa, Artur Queirozb, Kiyoshi F. Fukutania,c, Aldina Barrala,b,d, Annabel Rectore, Elke Wollantse, Els Keyaertse,f, Winke Van der Guchte, Marc Van Ranste,f, Kurt Beuselinckf, Camila I. de Oliveiraa,b, Johan Van Weyenberghe,1, Cristiana M. Nascimento-Carvalhoa,g,1, The Acute Respiratory Infection and Wheeze Study Group Phase I and II a
Postgraduate Program in Health Sciences, Federal University of Bahia School of Medicine, Salvador, Brazil Centro de Pesquisas Gonçalo Moniz (CPqGM), Fundação Oswaldo Cruz (FIOCRUZ), Salvador, Brazil c Department of Biochemistry, Immunology, and Cell Biology, Ribeirão Preto Medical School, University of São Paulo, Ribeirão Preto, Brazil d Department of Pathology, Federal University of Bahia School of Medicine, Salvador, Brazil e Department of Microbiology and Immunology, Laboratory for Clinical and Epidemiological Virology, Rega Institute for Medical Research, KU Leuven, Leuven, Belgium f Department of Laboratory Medicine, University Hospitals Leuven, Leuven, Belgium g Department of Pediatrics, Federal University of Bahia School of Medicine, Salvador, Brazil b
A R T I C LE I N FO
A B S T R A C T
Keywords: RSV Diagnostics Respiratory viruses Virus detection Viral respiratory infection Transcriptomics
Background: Virus-specific molecular assays such as real-time polymerase chain reaction (RT-PCR) are regularly used as the gold standard to diagnose viral respiratory tract infections, but simultaneous detection of multiple different pathogens is often challenging. A multiplex digital method of RNA quantification, nCounter (NanoString Technologies), can overcome this disadvantage and identify, in a single reaction, the presence of different respiratory viruses. Objectives: To evaluate the accuracy of nCounter to identify and quantify RSV-A and RSV-B in nasopharyngeal aspirates (NPA) of children (6–23-months-old) with acute respiratory infection. Study design: NPA was collected at enrolment in a prospective cross-sectional study conducted in Salvador, Brazil. A quantitative RT-PCR with a subgroup-specific primer and probeset for RSV-A and RSV-B was performed in parallel with a customized nCounter probeset containing viral targets in NPA. Results: Of 559 NPA tested, RSV was detected by RT-PCR in 139 (24.9%), by nCounter in 122 (21.8%) and by any method in 158 (28.3%) cases. Compared to the gold standard of qRT-PCR, sensitivity of nCounter was 74.3% (95%CI:63.3%–82.9% RSV-A) and 77.6% (95%CI:66.3%–85.9% RSV-B); specificity was 98.4% (95%CI:96.8%–99.2% RSV-A) and 97.8% (95%CI:96.0%–98.8% RSV-B); positive predictive value was 87.3% (95%CI:76.9%–93.4% RSV-A) and 82.5% (95%CI:71.4%–90.0% RSV-B) and negative predictive value was 96.1% (95%CI:94.1%–97.5% RSV-A), and 96.9% (95%CI:95.1%–98.2% RSV-B). Accuracy was 95.2% (95%CI:93.1%–96.7%) for RSV-A and 95.3% (95%CI:93.3%–96.9%) for RSV-B, while both methods significantly correlated for RSV-A (r = 0.44, p = 8 × 10−5) and RSV-B (r = 0.73, p = 3 × 10-12) quantification. Conclusions: nCounter is highly accurate in detecting RSV-A/B in NPA. Robustness and high-throughput multiplexing indicate its use in large-scale epidemiological studies.
⁎
Corresponding author at: Postgraduate Program in Health Sciences, Federal University of Bahia School of Medicine, Largo do Terreiro de Jesus, s/n, Centro Histórico, CEP 40025-010, Salvador, Brazil. E-mail address:
[email protected] (M.L. Bouzas). 1 JVW and CMN-C are both senior authors. https://doi.org/10.1016/j.jcv.2018.07.003 Received 13 March 2018; Received in revised form 12 June 2018; Accepted 13 July 2018 1386-6532/ © 2018 Elsevier B.V. All rights reserved.
Journal of Clinical Virology 106 (2018) 34–40
M.L. Bouzas et al.
1. Background
nostril until reaching the nasopharynx; negative pressure was applied and approximately 2 mL of nasal secretions were collected and deposited in a sterile tube with 1 mL of Nuclisens Lysis Buffer (Biomerieux, Boxtel, The Netherlands) and kept frozen at −70 °C until further processing.
Respiratory syncytial virus (RSV) is one of the most common etiological agents of acute respiratory infections (ARI) among children such as bronchiolitis and pneumonia [1,2]. RSV can be divided into two subgroups (A and B) that commonly produce annual epidemics characterized by the circulation of several genotypic strains [3]. The seasonality of RSV-A and RSV-B can be markedly different [4]. Virusspecific molecular assays such as real-time polymerase chain reaction (RT-PCR) are now considered the gold standard in the diagnosis of viral respiratory tract infections. They are rapid, relatively inexpensive and offer increased sensitivity and specificity over prior techniques such as virus culture and direct immunofluorescence, but simultaneous (multiplex) detection of different pathogens remains challenging. Upon comparison of four different commercial assays, Salez et al. [5] found considerable variation in sensitivity (20–100%), positive predictive value (50–100%) and Youden Index (0.20–1.00) between assays for fourteen respiratory viruses. Moreover, RNA content measure in all RTPCR protocols requires enzymatic amplification, which can be considered a major limitation, given the highly variable RNA quantity and quality in NPA [6]. A digital method of mRNA expression quantification, nCounter (NanoString Technologies), can overcome these disadvantages. It identifies by specific hybridization of the genetic material using two probes of 50 nucleotides each (both adjacent on the RNA target), one capture probe and one reporter probe anchored to a barcode [7,8]. Unique barcoding and gene-specific nucleic acid target capture allows up to 800 different targets to be simultaneously detected in a single experiment [9]. In addition, nCounter uses standardized internal control RNAs (8 negative and 6 positive) in each sample, allowing for efficient RNA quality control, as well as precise quantification of both host and viral transcripts up to (sub-)femtomolar level [7–10]. Our group and others have previously shown that nCounter detection is highly sensitive and robust for samples with degraded RNA, such as nasopharyngeal aspirates (NPA) and tissue biopsies [8,9,11,12]. RSV-A and RSV-B had been previously detected by nCounter in NPA in a small sample of patients (n = 65) [8]. However, a formal validation of the detection of RSV-A and RSV-B by this new method is lacking.
3.3. RNA extraction and nCounter digital quantification Total RNA was extracted using RNEasy (Qiagen´s, Hilden, Germany) following manufacturer´s instructions, and was subsequently hybridized against probes targeting RSV A–B, as part of a customized codeset containing other viral and bacterial pathogens and human target genes [8], designed by NanoString Technologies. For nCounter analysis, each sample was mixed with eight positive control RNAs in fixed amounts (0.2–2000 fg), as well as six negative control RNAs. Laboratory procedures and analysis were performed at the Laboratory for Clinical and Epidemiological Virology, Rega Institute for Medical Research, (KU Leuven, Belgium), where researchers were blinded to clinical data and to the results of the other method. 3.4. RSV-A and RSV-B quantification by real-time PCR The presence and viral load of RSV-A and RSV-B in the extracted RNA was measured by singleplex real-time PCR using subgroup-specific primers and probes (adapted from Zlateva et al.) [13]. The primers and probe for RSV-A are located in the F-gene of RSV-A, whereas the primers and probe for RSV-B are located in the N-gene of RSV-B (Table 1). Real-time qPCR reactions were performed using EuroScript reverse transcriptase and Reaction buffer of the OneStep RT qPCR MasterMix kit (Eurogentec,) with a concentration of 250 nM of each primer for RSV-A and 375 nM of each primer for RSV-B, and 125 nM of probe. To a reaction volume of 20 μL, 5 μL of RNA (extracted sample RNA or diluted cRNA for standard curve) or water (negative control) was added. Reactions were run on an ABI7500 Fast PCR instrument. Quantification of the RSV-A or RSV-B viral load was done by inference on a standard curve, generated using a dilution series containing 103-108 copies/per reaction of RSV-A or B cRNA standards. For generation of cRNA, the same genomic region that is used for the real-time PCR was first amplified by reverse trancriptase PCR, after which the PCR product was transcribed to cRNA using T7 MEGAshortscript. Quantification of cRNA was done by Nanodrop measurement.
2. Objective We aimed to evaluate the accuracy of nCounter (NanoString Technologies) to identify and quantify RSV-A and RSV-B in NPA of children with ARI (aged 6–23 months), using real-time PCR as the reference method.
3.5. Bioinformatic analysis Raw data were processed using nSolver 2.0 software (NanoString Technologies) sequentially correcting three factors: technical variation between samples and experiments (positive control RNAs), background correction (negative control RNAs) and RNA content by adjusting the counts (geometric mean) for the 3 human housekeeping genes (G6PD, GAPDH, GUSB), followed by normalization using logarithmic transformation (base 2). The nCounter probe sequences were compared with globally circulating RSV-A and RSV-B strains from the NR (non-redundant) database using BLAST software [14]. For visualization, 50 sequences were selected, and multiple alignments were performed with Muscle software [15]. Real-time PCR primers were aligned using Primer Blast software [16].
3. Study design 3.1. Patients selection This cross-sectional study evaluated community-dwelling children with ARI attending the Pediatric Emergency Room of the Federal University of Bahia Hospital, in Salvador, Northeastern Brazil, between September 2009 and October 2013. Inclusion criteria were children aged from 6 to 23 months with report of fever, sneeze, running nose, nasal blockage, or cough for up to seven days. Children transferred from other hospitals or reporting a previous episode of wheeze were excluded. Clinical and demographic data as well as NPA were collected at enrolment.
Table 1 Primers and probes used for real-time PCR quantification of RSV.
3.2. Samples collection
RSV-A
NPA samples were collected using the following protocol: the distance between the entrance of the nostril and the ear lobe was measured as an estimate of the distance from the entrance of the nostril to the nasopharynx; an aseptic plastic sputum catheter was inserted into the
RSV-B
35
FW primer REV primer MGB probe FW primer REV primer MGB probe
AFFW AFRV AFTP BNFW BNRV BNTP
5’ CTGTGATAGARTTCCAACAAAAGAACA 3’ 5’ AGTTACACCTGCATTAACACTAAATTCC 3’ FAM 5’ CAGACTACTAGAGATTACC 3’ MGB 5’ GGCTCCAGAATATAGGCATGATTC 3’ 5’ TGGTTATTASAAGRGCAGCTATACAYAGT 3’ FAM 5’ TATCATCCCACAGTCTG 3’ MGB
Journal of Clinical Virology 106 (2018) 34–40
M.L. Bouzas et al.
(data not shown). Since a small but considerable number of samples (27 for RSV-A, 26 for RSV-B) were detected by a single method only, a portion of the RSV epidemic might remain undetected depending on the method used. Therefore, we performed a comparative gene alignment for the sequences targeted by primers and probes in RT-PCR vs. nCounter using a total of 100 RSV-A and RSV-B strains representing the global epidemic. As shown in Fig. 3, a nearly perfect alignment with nCounter probes (> 98%) was obtained for RSV-B, whereas a slightly lower (> 95%) alignment was observed for RSV-A, which might partially explain the higher correlation coefficient for RSV-B (r = 0.73) vs. RSV-A (r = 0.44) when comparing both methods. In agreement with Houspie et al. 2013 [3] and Zlateva et al. 2007 [17], we found that the degenerated RT-PCR primers for both RSV-A and RSV-B performed equally well in gene alignment with globally circulating strains (Supplementary Fig. S1 and Fig. S2). To further explore if discrepancies between nCounter and RT-PCR might be due to technical variation or genetic differences within RSV-A and RSV-B strains, we were able to perform a confirmatory experiment in a subset of 35 (mostly discordant) samples, for which sufficient RNA was available for replication. Of those, ten samples were retested with the original RT-PCR protocol, showing concordant results in 8/10 samples (80%) compared with the first analysis by RT-PCR (one false positive for each RSV-A and RSV-B, data not shown). In contrast, replication of ten samples by nCounter resulted in 100% concordance for RSV-A and RSV-B. Another 15 samples were retested with an independent third assay, a real-time PCR targeting a different gene region (polymerase gene for both RSV-A and RSV-B [18]. Of those, 9/15 samples were concordant for RT-PCR (six false positive for RSV-A in the first analysis) and 10/15 for nCounter (three false positive for RSV-B and one false negative for each RSV-A and RSV-B, data not shown). However, all false positives were characterized by a low viral load, in either RT-PCR (659 ± 996 copies/ml) or nCounter (1.5 ± 2.0 counts), such that 13/15 samples (86.7%) were concordant in all three assays at viral loads > 1000 copies/ml (both RT-PCR protocols) or > 10 counts (nCounter). Only 2/15 samples (13.3%) with a moderate/high viral load were repeatedly discrepant (one in both RT-PCR assays: 0 vs. 3712 copies/ml, and one between both RT-PCRs and nCounter: 0 copies/ml vs. 35,546 counts), hinting at possible genetic differences between the samples. Considering the results of the third independent assay, sensitivity of the nCounter assay increased most for RSV-A, from 74.3% to 83.1% (95%CI:71.3%–89.5%) and to a lesser extent for RSV-B, from 77.6% to 78.8% (95%CI:67.5%–86.9%), accompanied by a rise in diagnostic accuracy to 96.4% and 95.5%, respectively. Taken together, this confirmatory experiment suggests technical variation might explain a minor portion of discrepant results for RT-PCR but not nCounter, whereas a larger part of discrepancies might be due to low viral load, as well as possible genetic variation between viral strains.
3.6. Statistical analysis Categorical variables were presented as absolute number (percentage). 95% Confidence Interval (CI) was included for each of the reported parameters: sensitivity, specificity, positive predictive value, negative predictive value and accuracy. Sensitivity was calculated as the number of cases with virus detected by nCounter among the number of cases with virus detected by RT-PCR. Specificity was calculated as the number of cases without the virus detected by nCounter among the number of cases without the same virus detected by RT-PCR. Positive predictive value was calculated as the number of cases with virus detected by both nCounter and RT-PCR in the numerator and the number of cases with the same virus detected by nCounter in the denominator. Negative predictive value was calculated as the number of cases without virus by both nCounter and RT-PCR in the numerator and the number without this virus as defined by nCounter in the denominator. Accuracy was calculated as the sum of cases with or without the virus as defined by both nCounter and RT-PCR in the numerator and the total number of studied cases in the denominator. All cases had RT-PCR performed and only one case did not have nCounter performed and was therefore excluded. Correlation between both assays was calculated using non-parametric (Spearman) correlation (GraphPad Prism 6.0), as viral loads did not follow a normal distribution. 4. Results Among 1154 evaluated children, 504 (43.7%) reported at least one previous episode of wheezing, 16 (1.41%) did not have NPA collected successfully, 1 (0.09%) did not have sufficient NPA for nCounter performance, 11 (1.0%) came from other hospitals and 63 (5.4%) did not consent (Fig. 1). Thus, this study group comprised 559 cases. The mean age was 11.4 ± 4.5 months and 286 (51.2%) were female. The most frequent complaints were cough (86.2%), running nose (86.9%), fever (81.6%), and sneeze (77.3%). On physical examination, the most frequent findings were fever (40.3%), ronchi (36.9%), and tachypnea (23.6%). Overall, RSV was detected by RT-PCR in 139 (24.9%) cases, RSV-A in 74 (13.2%) cases, and RSV-B in 67 (12.0%) cases (Fig. 1). Two (1.4%) were co-infected. Digital quantification by nCounter, the new diagnostic method, detected RSV in 122 (21.8%) samples, RSV-A in 63 (11.3%) and RSV-B in 63 (11.3%) (Fig. 1). Co-infections were detected in 4 (3.3%) cases. Interestingly, detection of RSV-A and RSV-B by any method occurred in 158 (28.3%) cases, indicating both methods might be complementary in detecting the complete RSV epidemic. The validation of nCounter as a qualitative measure (presence vs. absence), considering RT-PCR as the reference standard is shown in Table 2A for RSV-A and in Table 2B for RSV-B. The sensitivity of nCounter was 74.3% (95%CI:63.3%–82.9%) for RSV-A and 77.6% (95%CI:66.3%–85.9%) for RSV-B; the specificity was 98.4% (95%CI:96.8%–99.2%) for RSV-A and 97.8% (95%CI:96.0%–98.8%) for RSV-B; the positive predictive value was 87.3% (95%CI:76.9%–93.4%) for RSV-A and 82.5% (95%CI:71.4%–90.0%) for RSV-B and the negative predictive value was 96.1% (95%CI:94.1%–97.5%) for RSV-A and 96.9% (95%CI:95.1%–98.2%) for RSV-B. Overall, accuracy was 95.2% (95%CI:93.1%–96.7%) for RSV-A and 95.3% (95%CI:93.3%–96.9%) for RSV-B. Moreover, quantification of both RSV-A and RSV-B viral RNA was significantly correlated between nCounter and RT-PCR, as shown in Fig. 2A and B. Again, using RT-PCR as a reference, a significant correlation (Spearman r = 0.44, p = 8 × 10−5) was found in RSV-Apositive samples (n = 74), between quantitative detection by nCounter (measured as normalized counts) and RT-PCR (measured as copies/ml) (Fig. 2A). Similarly, in RSV-B RT-PCR-positive samples (n = 67), a significant correlation (Spearman r = 0.73, p = 3 × 10-12) was found between RSV-B quantitative detection by nCounter and RT-PCR (Fig. 2B). A similarly significant result was obtained when selecting only nCounter-positive samples or samples positive by any method
5. Discussion Using real-time PCR as a reference, nCounter was shown to detect RSV-A and RSV-B simultaneously with high accuracy, both qualitatively and quantitatively. To the best of our knowledge, this is the first study to assess the diagnostic accuracy of nCounter for the detection of RSV-A and B in NPA collected from a large sample of children with ARI. Considering the high specificity of nCounter, its absence of cross-hybridization to other viruses [8], as well the use of internal positive and negative control RNAs for each individual sample, we believe the additional 19 samples (8 RSV-A and 11 RSV-B, detected only by nCounter, Tables 2A and 2B) represent genuine RSV infections, which would have gone undetected if only RT-PCR was used for diagnosis. Replication experiments indeed indicate a larger technical variation for RT-PCR, either using the same or an independent protocol, in contrast to nCounter. This could be explained by a greater possibility of 36
Journal of Clinical Virology 106 (2018) 34–40
M.L. Bouzas et al.
Fig. 1. Flow-chart of recruitment of children with Acute Respiratory Infection and Venn diagramo f RSV-A and RSV-B investigation by RT-PCR and by nCounter.
contamination between samples, as well as the lack of internal positive and negative controls to correct for RNA quality e.g. the presence of enzymatic inhibitors in the case of RT-PCR. Of note, real-time PCR primers and probes were designed [3] and updated (this study) to fully capture the Belgian RSV epidemic, whereas nCounter probes were designed for and tested in Brazilian RSV samples [8]. In addition, primers and probe are located in the F-gene for RSV-A and in the N-gene for RSV-B, whereas the nCounter probes are located in the hypervariable region of the G protein for both RSV-A and RSV-B. It is reassuring that a third independent assay targeting the polymerase gene resolved most discrepancies between RT-PCR and nCounter, increasing both sensitivity and accuracy for the latter. However, it will be necessary to regularly check whether probes still match with circulating strains, in order to update the custom nCounter probe set when necessary. Our alignment of nCounter probes shows 95% and 98% identity, respectively, for RSV-A and RSV-B (Fig. 3), with a striking similarity between Brazilian and Belgian samples, as well as other strains circulating in Latin America, Europe, North America (USA) and Asia (South Korea, Saudi Arabia). Regarding the global spread of RSV-A and RSV-B epidemics, we believe that our study demonstrates that digital
Table 2A Diagnostic accuracy of nCounter to detect RSV-A. RT-PCR
nCounter
Yes No Total
Total
Yes
No
55 (74.3%) 19 (25.7%) 74 (13.2%)
8 (1.6%) 477 (98.4%) 485 (86.8%)
63 (11.3%) 496 (88.7%) 559 (100%)
Table 2B Diagnostic accuracy of nCounter to detect RSV-B. RT-PCR
nCounter
Yes No Total
Total
Yes
No
52 (77.6%) 15 (22.4%) 67 (12.0%)
11 (2.2%) 481 (97.8%) 492 (88.0%)
63 (11.3%) 496 (88.7%) 559 (100%)
37
Journal of Clinical Virology 106 (2018) 34–40
M.L. Bouzas et al.
and cost-effective option with larger sample sizes, e.g. cohort studies, or when less hands-on time is preferred, e.g. core laboratories, due to its high-throughput capacity. Some limitations of our study must be highlighted. First of all, this was a single-center study. Therefore, studies conducted at other sites are necessary. Secondly, the 95% CI precision for sensitivity and positive predictive value was wide, which implies the evaluation of these parameters requires a sample larger than the one presented here. Finally, the exclusion of children with a previous episode of wheezing may have selected a group of previously healthy children different from children with ARI typically seen in emergency rooms. This exclusion was due to the cross-sectional study design, within a prospective cohort study on the development of recurrent wheezing. Nevertheless, our study has several strengths. First, patients were recruited under daily surveillance in a specialized Pediatric Emergency Department, during the entire study period, when all patients who might fulfill the inclusion criteria were evaluated. Second, the 95% CI precision was narrow for specificity, negative predictive value, and accuracy. Third, the data collection period lasted 4 years and 1 month, which argues against a possible seasonality bias. In addition, nCounter technology has been shown to allow accurate mRNA expression quantification using low amounts of total RNA and is able to detect as little as 0.5 fM of mRNA transcripts. This advantage could reduce the amplification bias offered by the RT-PCR method [7]. In conclusion, digital RNA quantification of RSV-A and RSV-B by nCounter is highly accurate (> 95%), using real-time PCR as a reference. Its robustness, high-throughput multiplex capacity and detection of cases undetected by real-time PCR indicate its suitability for large-scale epidemiological studies.
Funding This work was supported by Fundação de Amparo à pesquisa do Estado da Bahia (FAPESB; grant PNX0019/2009). MLB was supported by fellowships from FAPESB. KFF was supported by fellowships from Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPQ), FAPESB and Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP no. 2017/03491-6). JRO was supported by fellowships from FAPESB and Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES). AB, CIO, CMN-C are senior investigators from CNPQ. JVW was supported by (Research Foundation - Flanders FWO;grant G0D6817N).
Fig. 2. Correlation between RSV-A and RSV-B levels quantified by nCounter and Real-Time PCR. (A) In RT-PCR-positive samples (n = 74), a significant correlation (Spearman r = 0.44, p = 8 × 10−5) was found between RSV-A quantitative detection by nCounter (measured as normalized counts) and RT-PCR (measured as copies/ml). (B) In RT-PCR-positive samples (n = 67), a significant correlation (Spearman r = 0.73, p = 3 × 10-12) was found between RSV-B quantitative detection by nCounter (normalized counts) and RT-PCR (copies/ml).
Ethical approval The study was approved by the Ethics Committee from the Federal University of Bahia (no 067/2009). Parents or legal guardians signed written informed consent before enrollment of each study participant.
quantification of respiratory viruses by nCounter will be useful in large epidemiological studies worldwide. Considering the clinical and research application of nCounter vs. RT-PCR assays, both turnaround time and cost should be considered, as well as the total number of samples to be processed. For a batch of 12–24 samples, total turnaround time is 24–36 h (30 min hands-on time) vs. 4 h (1.5 h hands-on time), for nCounter (full respiratory panel) vs. in-house RT-PCR (for RSV-A and RSV-B only), respectively. Cost per sample corresponds to 55 euro for nCounter and 50–70 euro for in-house and confirmatory RT-PCR protocols, which compare favorably to most routinely used commercial assays (65–140 USD/sample, although list prices obviously vary worldwide). RT-PCR protocols are usually the method of choice when sample sizes are small and fast diagnosis is preferred, e.g. emergency rooms. On the other hand, we propose nCounter technology as a valid
Competing interests Authors declare they have no conflicts of interest.
Acknowledgements The acute respiratory infection and wheeze study group phase I and II Elaine N. Pacheco,1 Eliana Silva,2 Eveline Xavier,1 Karen S. Miranda,2 Laíse R. Neri,2 Patrícia F. Silva,2 Ramom S. Amoedo,1 Rômulo B. Menezes,2 Sylvia Patrícia T. Ledo2 and Thamirys Marinho.1 Affiliations:1 Federal University of Bahia School of Medicine, Salvador, CEP 40020-210, Brazil. 2 Bahiana School of Medicine, Bahiana Foundation for Science Development, Salvador, CEP 40290-000, Bahia, Brazil. 38
Journal of Clinical Virology 106 (2018) 34–40
M.L. Bouzas et al.
Fig. 3. Alignment of nCounter probes (100 bp) with global strains of RSV-A (A) and RSV-B (B).
39
Journal of Clinical Virology 106 (2018) 34–40
M.L. Bouzas et al.
Appendix A. Supplementary data Supplementary material related to this article can be found, in the online version, at doi:https://doi.org/10.1016/j.jcv.2018.07.003.
[9]
References [10] [1] G.S. Durigon, D.B.L. Oliveira, M.C.C. Felicio, C. Finelli, M.F.B. Pereira, J.G. Storni, R.N. Caldeira, R.C. Brezin, E.L. Durigon, E.N. Berezin, Poor outcome of acute respiratory infection in young children with underlying health condition in Brazil, Int. J. Infect. Dis. 34 (2015) 3–7, https://doi.org/10.1016/j.ijid.2015.03.003. [2] C.M. Nascimento-Carvalho, M.R.A. Cardoso, A. Barral, C.A. Araujo-Neto, J.R. Oliveira, L.S. Sobral, A. Saukkoriipi, M. Paldanius, R. Vainionpaa, M. Leinonen, O. Ruuskanen, Seasonal patterns of viral and bacterial infections among children hospitalized with community-acquired pneumonia in a tropical region, Scand. J. Infect. Dis. 42 (2010) 839–844, https://doi.org/10.3109/00365548.2010.498020. [3] L. Houspie, P. Lemey, E. Reijmen, V. Vergote, A. Vankeerberghen, F. Vaeyens, H.D. Beenhouwer, M.V. Ranst, Circulation of HRSV in Belgium: multiple genotype circulation to prolonged circulation of predominant genotypes, PLoS One 8 (4) (2013) e60416, , https://doi.org/10.1371/journal.pone.0060416. [4] M.L. Bouzas, J.R. Oliveira, K.F. Fukutani, I.C. Borges, A. Barral, W. Van Der Gucht, E. Wollants, M. Van Ranst, C.I. Oliveira, J. Van Weyenbergh, C.M. NascimentoCarvalho, Respiratory syncytial virus a and b display different temporal patterns in a 4-year prospective cross-sectional study among children with acute respiratory infection in a tropical city, Medicine 95 (41) (2016) e5142, , https://doi.org/10. 1097/MD.0000000000005142. [5] N. Salez, A. Vabret, M. Leruez-Ville, L. Andreoletti, F. Carrat, F. Renois, X. de Lamballerie, Evaluation of four commercial multiplex molecular tests for the diagnosis of acute respiratory infections, PLoS One 10 (6) (2015) e0130378, , https:// doi.org/10.1371/journal.pone.0130378. [6] F. Thorburn, S. Bennett, S. Modha, D. Murdoch, R. Gunson, P.R. Murcia, The use of next generation sequencing in the diagnosis and typing of respiratory infections, J. Clin. Virol. 69 (2015) 96–100, https://doi.org/10.1016/j.jcv.2015.06.082. [7] G.K. Geiss, R.E. Bumgarner, B. Birditt, T. Dahl, N. Dowidar, D.L. Dunaway, H.P. Fell, S. Ferree, R.D. George, T. Grogan, J.J. James, M. Maysuria, J.D. Mitton, P. Oliveri, J.L. Osborn, T. Peng, A.L. Ratcliffe, P.J. Webster, E.H. Davidson, L. Hood, K. Dimitrov, Direct multiplexed measurement of gene expression with color-coded probe pairs, Nat. Biotechnol. 26 (March (3)) (2008) 317–325, https://doi.org/10. 1038/nbt1385. [8] K.F. Fukutani, C.M. Nascimento-Carvalho, W. Van der Gucht, E. Wollants, R. Khouri, T. Dierckx, M.Van Ranst, L. Houspie, M.L. Bouzas, J.R. Oliveira,
[11]
[12]
[13]
[14]
[15]
[16]
[17]
[18]
40
A. Barral, J. Van Weyenbergh, C.I. Oliveira, Pathogen transcriptional profile in nasopharyngeal aspirates of children with acute respiratory tract infection, J. Clin. Virol. 69 (2015) 190–196, https://doi.org/10.1016/j.jcv.2015.06.005. E. Speranza, L.A. Altamura, K. Kulcsar, S.L. Bixler, C.A. Rossi, R.J. Schoepp, E. Nagle, W. Aguilar, C.E. Douglas, K.L. Delp, T.D. Minogue, G. Palacios, A.J. Goff, J.H. Connor, Comparison of transcriptomic platforms for analysis of whole blood from ebola-infected cynomolgus macaques, Sci. Rep. 7 (November (1)) (2017) 14756, https://doi.org/10.1038/s41598-017-15145-7. B. Moens, C. Pannecouque, G. López, M. Talledo, E. Gotuzzo, R. Khouri, A. Bittencourt, L. Farré, B. Galvão-Castro, A.M. Vandamme, J. Van Weyenbergh, Simultaneous RNA quantification of human and retroviral genomes reveals intact interferon signaling in HTLV-1-infected CD4+ T cell lines, Virol. J. 23 (August (9)) (2012) 171, https://doi.org/10.1186/1743-422X-9-171. R. Khouri, G.S. Santos, G. Soares, J.M. Costa, A. Barral, M. Barral-Netto, J. Van Weyenbergh, SOD1 plasma level as a biomarker for therapeutic failure in cutaneous leishmaniasis, J. Infect. Dis. 210 (July (2)) (2014) 306–310, https://doi.org/10. 1093/infdis/jiu087. D.C. Andrade, I.C. Borges, M.L. Bouzas, J.R. Oliveira, K.F. Fukutani, A.T. Queiroz, C.I. de Oliveira, A. Barral, J. Van Weyenbergh, C. Nascimento-Carvalho, 10-valent pneumococcal conjugate vaccine (PCV10) decreases metabolic activity but not nasopharyngeal carriage of streptococcus pneumoniae and haemophilus influenzae, Vaccine 35 (July 33) (2017) 4105–4111, https://doi.org/10.1016/j.vaccine.2017. 06.048. K.T. Zlateva, P. Lemey, E. Moës, A.M. Vandamme, M. Van Ranst, Genetic variability and molecular evolution of the human respiratory syncytial virus subgroup B attachment G protein, J. Virol. 79 (July (14)) (2005) 9157–9167. S.F. Altschul, W. Gish, W. Miller, E.W. Myers, D.J. Lipman, Basic local alignment search tool, J. Mol. Biol. 215 (1990) 403–410, https://doi.org/10.1016/S00222836(05)80360-2. R.C. Edgar, MUSCLE: multiple sequence alignment with high accuracy and high throughput, Nucleic Acids Res. 32 (5) (2004) 1792–1797, https://doi.org/10.1186/ 1471-2105-5-113. J. Ye, G. Coulouris, I. Zaretskaya, I. Cutcutache, S. Rozen, T.L. Madden, PrimerBLAST: a tool to design target-specific primers for polymerase chain reaction, BMC Bioinf. 13 (June (134)) (2012), https://doi.org/10.1186/1471-2105-13-134. K.T. Zlateva, L. Vijgen, N. Dekeersmaeker, C. Naranjo, M. Van Ranst, Subgroup prevalence and genotype circulation patterns of human respiratory syncytial virus in Belgium during ten successive epidemic seasons, J. Clin. Microbiol. (September) (2007) 3022–3030, https://doi.org/10.1128/JCM.00339-07. J. Kuypers, N. Wright, R. Morrow, Evaluation of quantitative and type-specific realtime RT-PCR assays for detection of respiratory syncytial virus in respiratory specimens from children, J. Clin. Virol. 31 (October (2)) (2004) 123–129.