Feasibility of Using Intravascular Ultrasonography for Assessment of Giant Cavernous Aneurysm after Endovascular Treatment: a Technical Report Shahram Majidi, MD, Mikayel Grigoryan, MD, Wondwossen G Tekle, MD, Masaki Watanabe, MD, PhD, and Adnan I Qureshi, MD Zeenat Qureshi Stroke Research Center, University of Minnesota
Abstract Introduction—Intravascular ultrasonography (IVUS) has been shown as a valuable adjunct imaging tool during endovascular procedures but its value in detection of any recurrence during follow up after endovascular coil embolization of large and giant intracranial aneurysms is not reported. Methods—A 41 years old man who had been treated using stent assisted coil embolization for cavernous segment aneurysm of the left internal carotid artery underwent 60 month angiographic follow up. Concurrently, IVUS catheter was advanced under fluoroscopic guidance inside the cavernous portion of the left internal carotid artery. Then IVUS images were used to visualize the stent, coil loops, and aneurysm neck. Results—The angiographic images were limited because of superimposition of the aneurysm on the parent vessel in all projections. IVUS images demonstrated that the stent was patent along its whole length and there was no sign of stent deformity or in-stent thrombosis. Loops of the coil were visualized as hyperechoic signals inside the aneurysm and there was no sign of herniated loops of coil inside the stent. Conclusion—In this case report, we observed that adjunct use of IVUS can provide valuable information not ascertained by angiography during follow up assessment of coil embolized aneurysm.
Journal of Vascular and Interventional Neurology
Keywords Intravascular Ultrasonography; Aneurysm; Coil Embolization; Stent
Background Giant intracranial aneurysms with fundus diameters of 25mm or more comprise approximately 5% of all intracranial aneurysms.1, 2 While surgical clipping of these aneurysms is very challenging with high morbidity rate,3, 4 endovascular embolization has emerged as an acceptable first line treatment option with relatively low morbidity.5, 6 However, aneurysm re-growth, coil compaction, and/or new growth of aneurysm occurs frequently after endovascular embolization of large aneurysms. Therefore, long-term angiographic follow up is mandatory after the initial embolization procedure.7 However, angiographic images are often complicated by superimposition of the aneurysm on the parent vessel obscuring visualization of the aneurysm parent vessel junction. In this study, we report on the feasibility of using intravascular ultrasonography (IVUS) during fol-
low-up monitoring for carotid cavernous aneurysms treated with coil embolization and stent placement.
Case Description A 41 year old man with a history of coil embolization of the left cavernous segment large aneurysm 5 years ago and covered stent (Jo-Stent) placement across the neck of the aneurysm 4 years ago underwent a 60 month follow up angiogram. The patient consented for the angiogram and IVUS procedure. Under fluoroscopic guidance, 6F catheter was advanced into the left internal carotid artery and biplane and rotational angiography was performed over the cranium. Coil mass artifact was visualized in the cavernous segment. The stent was poorly visualized and both the intraluminal segment and the junction of the aneurysm parent
Published June, 2012. All Rights Reserved by JVIN. Unauthorized reproduction of this article is prohibited Corresponding author: Shahram Majidi, MD, Zeenat Qureshi Stroke Research Center, Department of Neurology, University of Minnesota, 420 Delaware Street S.E., MMC 295, Minneapolis, MN 55455, Phone: 612-626-1866, Email:
[email protected]
7
Journal of Vascular and Interventional Neurology
Figure-1. IVUS catheter advanced into the cavernous segment of left internal carotid artery and advanced until the scanner was distal to the aneurysm neck (A). Then it was slowly pulled back to visualize the whole length of stent and the aneurysm neck (B).
vessel were obscured by superimposition of the coil mass.
Journal of Vascular and Interventional Neurology
Under fluoroscopic guidance, IVUS catheter (Eagle Eye Gold, 20 MHZ Digital, 64 Bement, s5 Imaging System, Volcano Corp.) was introduced through the 6F catheter using 0.014 guide wire (Synchro2TM Standard, Boston Scientific) and advanced into the cavernous part of the left carotid artery (Figure-1A and B). After the scanner of IVUS catheter was placed distal to the aneurysm neck, catheter was slowing pulled back to visualize the stent, aneurysm neck and apposition of coil loops inside the aneurysm and any possible herniated loops of the coil into the parent artery. IVUS images demonstrated that the stent was patent and there was no sign of herniated loops of coil inside the parent vessel (Figure-2A and B). The procedure was completed without any complications and the patient was discharged home the same day. The 20 mm scanning diameter of the IVUS catheter was limited for visualization of the entirely of the coils inside the aneurysm.
Discussion Digital subtraction angiography is the gold standard imaging tool during endovascular interventional procedures. As a 2D silhouette of the vessel lumen, angiographic images do have some limitations such as fore-
shortening of vessels depending on the view and vessel overlap.8, 9 After coil placement in large aneurysms, it is not infrequent to see superimposition of the aneurysm with coils on the parent vessel obscuring visualization of the aneurysm parent vessel junction. Therefore, it is appropriate to find a reliable adjunct imaging tool which can be used during and/or after aneurysm embolization besides angiography. IVUS is an adjunct imaging tool that has been extensively used in interventional cardiology during the past two decades. It provides useful information during coronary artery stent placement and leads to reduced rates of stent thrombosis and need to repeat revascularization procedure.10, 11 In fact IVUS can provide valuable and reproducible information from the vessel wall, lumen, and devices during endovascular procedures which may improve the procedural outcome.12, 13 Considering the tortuous anatomy of intracranial vasculature, there are limited reports of using IVUS during neuro-interventional procedures. Wehman et al.14 have reported IVUS guidance during angioplasty and stent placement of two patients with occlusive dissection of the intracranial internal carotid artery and atherosclerotic restenosis of the basilar artery. They concluded that IVUS provided valuable information regarding lesion characteristics, and helped with appropriate stent selection and stent placement. In another study, Yoon et al.12 performed stent angioplasty of iatrogenic extracranial vertebral artery dissection under IVUS guidance and concluded
Journal of Vascular and Interventional Neurology
Majidi et al.
8
Journal of Vascular and Interventional Neurology
Figure-2. Coil loops are visualized as hyperechoic signal inside the aneurysm (arrow heads in A). In image B, stent is seen patent without any deformity and sign of herniated coil loops.
that IVUS is a useful adjunct imaging tool in confirmation of true lumen and ascertaining appropriate stent apposition.
Journal of Vascular and Interventional Neurology
Coil embolization has been proven as a reliable and safe alternative for surgically difficult cavernous carotid aneurysms.15, 16 Endovascular treatment of giant aneurysms carries less morbidity and provides favorable longterm resuluts.6 Because of the risk of aneurysm regrowth and reopening, coil compaction and assessment of stent patency in stent-assisted aneurysm embolization, it is routine to perform serial angiographic follow ups at regular intervals after an initial coil embolization procedure.4 These follow up imaging Studies help neurointerventionalists identify aneurysm recurrence and the need for repeat coil placement or other additional procedures.6, 7 In this case report, we showed that IVUS has the a potential to be used as an adjunct imaging tool during angiographic follow up of coil embolized aneurysms. By providing intraluminal direct visualization of the vessel wall and aneurysm neck from inside the vessel, IVUS provided us with valuable information regarding patency of the stent inside the vessel, in-stent thrombosis, and coil herniation inside the parent artery. The new generations of IVUS are expected to be more flexible with smaller profiles promising that intracranial use is feasible and relevant. Adjunct use of IVUS during
neuro-endovascular procedures has the potential to provide additional information not ascertained by angiography and to also reduce the need for multiple intra-procedural angiographic images, and radiation and contrast exposure.
References 1. Barrow DL, Alleyne C. Natural history of giant intracranial aneurysms and indications for intervention. Clin Neurosurg 1995;42:214– 44. 2. Choi IS, David C. Giant intracranial aneurysms: development, clinical presentation and treatment. Eur J Radiol 2003;46:178–94. 3. Lawton MT, Spetzler RF. Surgical strategies for giant intracranial aneurysms. Neurosurg Clin N Am 1998;9:725–42. 4. Sluzewski M, Menovsky T, van Rooij WJ, Wijnalda D. Coiling of very large or giant cerebral aneurysms: long-term clinical and serial angiographic results. AJNR Am J Neuroradiol 2003;24:257–62. 5. Gruber A, Killer M, Bavinzski G, Richling B. Clinical and angiographic results of endosaccular coiling treatment of giant and very large intracranial aneurysms: a 7-year, single-center experience. Neurosurgery 1999;45:793–803.discussion -4 6. van Rooij WJ, Sluzewski M. Endovascular treatment of large and giant aneurysms. AJNR Am J Neuroradiol 2009;30:12–8. 7. Thornton J, Debrun GM, Aletich VA, Bashir Q, Charbel FT, Ausman J. Follow-up angiography of intracranial aneurysms treated with endovascular placement of Guglielmi detachable coils. Neurosurgery 2002;50:239–49.discussion 49-50 8. Hong MK, Mintz GS, Popma JJ, et al. Limitations of angiography for analyzing coronary atherosclerosis progression or regression. Ann Intern Med 1994;121:348–54. 9. Topol EJ, Nissen SE. Our preoccupation with coronary luminology.
Journal of Vascular and Interventional Neurology
9
Journal of Vascular and Interventional Neurology
The dissociation between clinical and angiographic findings in ischemic heart disease. Circulation 1995;92:2333–42. 10. Fitzgerald PJ, Oshima A, Hayase M, et al. Final results of the Can Routine Ultrasound Influence Stent Expansion (CRUISE) study. Circulation 2000;102:523–30. 11. Oemrawsingh PV, Mintz GS, Schalij MJ, Zwinderman AH, Jukema JW, van der Wall EE. Intravascular ultrasound guidance improves angiographic and clinical outcome of stent implantation for long coronary artery stenoses: final results of a randomized comparison with angiographic guidance (TULIP Study). Circulation 2003;107:62–7. 12. Yoon WK, Kim YW, Kim SD, Park IS, Baik MW, Kim SR. Intravascular ultrasonography-guided stent angioplasty of an extracranial vertebral artery dissection. J Neurosurg 2008;109:1113–8. 13. Majidi S, Khatri R, Watanabe M, Siddiq F, Chaudhry SA, Qureshi
AI. Aneurysm Embolization Using Detachable Coils Under Intravascular Ultrasonography (IVUS) Guidance: an in-Vitro Feasibility Study. Neurosurgery 2012: Jun;70(6):1557–64. 14. Wehman JC, Holmes DR Jr, Hanel RA, Levy EI, Hopkins LN. Intravascular ultrasound for intracranial angioplasty and stent placement: technical case report. Neurosurgery 2006;59:ONSE481–3.discussion ONSE3 15. Bavinzski G, Killer M, Ferraz-Leite H, Gruber A, Gross CE, Richling B. Endovascular therapy of idiopathic cavernous aneurysms over 11 years. AJNR Am J Neuroradiol 1998;19:559–65. 16. Stiebel-Kalish H, Kalish Y, Bar-On RH, et al. Presentation, natural history, and management of carotid cavernous aneurysms. Neurosurgery 2005;57:850–7.discussion -7
Journal of Vascular and Interventional Neurology Journal of Vascular and Interventional Neurology