ARTHUR L. MALKANI, MICHAEL J. VOOR, KIRK A FEE,. C. SHANNON BATES. From the University of Louisville, USA. We have tested the axial and torsional ...
FEMORAL COMPONENT REVISION USING IMPACTED MORSELLISED CANCELLOUS GRAFT A BIOMECHANICAL STUDY OF IMPLANT STABILITY ARTHUR L. MALKANI, C. SHANNON BATES
MICHAEL J. VOOR,
KIRK A FEE,
From the University of Louisville, USA e have tested the axial and torsional stability of femoral components after revision arthroplasty in a cadaver model, using impacted morsellised cancellous graft and cement. Each one of six matched pairs of fresh frozen human femora had either a primary or a revision prosthesis cemented in place. For the ‘revision’ experiments, all cancellous bone was removed from the proximal femur which was then over-reamed to create a smooth-walled cortical shell. An MTS servohydraulic test frame was used to apply axial and torsional loads to each specimen through the prosthetic femoral heads with the femur submerged in isotonic saline solution at 37°C. The mean subsidence was 0.27 ± 0.17 mm for the primary and 0.52 ± 0.30 mm for the revision groups. The difference was statistically significant (p < 0.025), but the mean subsidence was 0.015). Based on our results we suggest that revision of the femoral component using morsellised cancellous graft followed by cementing with a collarless prosthesis with a polished tapered stem restores the integrity of the proximal femur and provides immediate stability of the implant.
W
J Bone Joint Surg [Br] 1996;78-B:973-8. Received 12 February 1996; Accepted after revision 19 June 1996
A. L. Malkani, MD, Assistant Professor, Chief, Adult Reconstruction M. J. Voor, PhD, Assistant Professor, Director of Research K. A. Fee, MD, Administrative Chief Resident C. S. Bates, BA, Medical Student Department of Orthopaedic Surgery, University of Louisville School of Medicine, HSC/ACB 3rd Floor Bridge, Louisville, Kentucky 40292, USA. Correspondence should be sent to Dr A. L. Malkani. ©1996 British Editorial Society of Bone and Joint Surgery 0301-620X/96/61288 $2.00 VOL. 78-B, NO. 6, NOVEMBER 1996
Revision of the femoral component of a hip arthroplasty is a challenging problem. Immediate stability of the implant at the time of revision may be difficult to achieve due to underlying osteolysis. An enlarged sclerotic proximal femur may be deficient in both cancellous and cortical bone (Hungerford and Jones 1988; Poss et al 1988; D’Antonio et al 1989). The choice of implant depends on several factors including the extent of loss of proximal femoral bone, the quality of the remaining host bone, the activity level of the patient, and the experience of the surgeon (Callaghan 1992). The implants have included uncemented, extensively porous-coated stems (Engh et al 1988; Paprosky, Bradford and Younger 1994), proximally-coated long-stem implants (Gustilo and Pasternak 1988; Hedley, Gruen and Ruoff 1988; Hungerford and Jones 1988; Mallory 1988; Malkani et al 1996) and cemented prostheses (Pellicci, Wilson and Sledge 1982; Callaghan et al 1985; Kavanagh, Ilstrup and Fitzgerald 1985; Rubash and Harris 1988; Marti et al 1990). The results of cemented revision have been variable, mainly due to the lack of interdigitation between the cement and the remaining sclerotic proximal femur (Dohmae et al 1988). Proximally porous-coated stems have shown high rates of failure due to the incidence of intraoperative fractures and difficulty in obtaining proximal ingrowth from the poor host bone (Malkani et al 1996). Extensively coated, cementless long stems with stable distal fixation show good intermediate results at the cost of proximal stress shielding (Engh et al 1988; Paprosky et al 1994). For massive proximal bone loss, methods of management have included the use of cortical strut grafts (Emerson et al 1992; Head et al 1993), allograft prosthetic composites (Gross et al 1993; Chandler et al 1994), and tumour-type prostheses (Malkani, Sim and Chao 1993). Cortical strut grafts have been useful in restoring structural bone loss around the proximal femur but their main function is to augment the existing bone stock and they cannot be used to provide implant stability. The use of impacted morsellised cancellous allograft in the proximal femur followed by insertion of a cemented, polished, tapered femoral component has been advocated for proximal femoral revision (Gie et al 1993; Ling, Timperley and Linder 1993). The ‘Ling technique’ is believed 973
974
A. L. MALKANI,
M. J. VOOR,
to reconstitute proximal femoral bone gradually through substitution and incorporation (Ling et al 1993). This creates interfaces between the allograft and host bone, the cement and the allograft, and the implant and the cement. This raises concern about the ability to achieve immediate stability of the implant and the planning of postoperative rehabilitation. The immediate implant stability which is achieved has not been tested biomechanically. Our aim was to test the axial and torsional stability of the femoral component after insertion of a cemented polished tapered stem into a bed of impacted morsellised cancellous graft in the cadaver femur.
MATERIALS AND METHODS Specimen preparation. The proximal two-thirds of six matched pairs of fresh frozen human cadaver femora, from patients of mean age 73 years (62 to 88), were sealed in plastic and stored at a temperature of –20°C. Each specimen was radiographed to identify bony abnormality. All were thawed in a refrigerator at 4°C for 24 hours and then at room temperature until tested. A primary and a revision arthroplasty was performed on each pair. On the primary side, a collarless polished tapered femoral component (CPT; Zimmer Inc, Warsaw, Indiana) was inserted using modern cementing techniques after standard
Fig. 1a
K. A. FEE,
C. S. BATES
reaming and broaching. The size of the implant was determined by using preoperative radiographs and templates. In the opposite paired femur all cancellous bone was removed followed by over-reaming of the femoral canal to create a smooth cortical envelope. The heads from each pair of femora were used as cancellous grafts for restoration of the bone loss. A bone mill (Tracer Designs, Santa Paula, California) was used to grind the heads into chips and the resulting morsellised graft was impacted into the femur using CPT instrumentation as described by the manufacturer (Zimmer Inc, Warsaw, Indiana). In the final phase of impaction, smooth tamps were used to prepare the site for the introduction of the cement (Fig. 1). An identical prosthesis was then inserted using the same cementing technique. The amount of morsellised material obtained from the two heads was sufficient to create an adequate graft to fill the over-reamed canal and to accommodate an implant of appropriate size and a cement mantle 1 mm thick. The graft varied in thickness from approximately 3 mm at the corners of the implant to over 10 mm along its face. Radiological markers were inserted on the shaft of both femora at the level of the distal end of the implant to indicate the position of the implant before and after mechanical testing. Mechanical testing. Each specimen was mounted on a stand on an MTS servohydraulic test frame (Model 858
Fig. 1b
Fig. 1c
The revision specimens were prepared from normal femora by over-reaming the proximal portion. Cancellous bone chips were ground and impacted (a) into the proximal canal to form a graft mantle (b) into which a prosthesis (c) was cemented. THE JOURNAL OF BONE AND JOINT SURGERY
FEMORAL COMPONENT REVISION USING IMPACTED MORSELLISED CANCELLOUS GRAFT
Fig. 2a
975
Fig. 2b
Figure 2a – Cyclic axial loading was applied through the prosthetic femoral head over 5000 cycles between 70% and 200% of average body-weight. A water chamber was used to keep the femur, graft, cement and prosthesis at 37°C to simulate clinical conditions. Figure 2b – Torsional loading was applied about the axis of the prosthesis and femoral shaft.
Bionix; MTS Systems Inc, Minneapolis, Minnesota). The cadaver femora were loaded through a 28 mm femoral head by direct contact with a concave indentation in a metal disc attached to the actuator of the MTS machine. The mounting stand was pinned at its base, allowing a rotational degree of freedom in the plane of the implant (Fig. 2). We performed axial cyclic loading and measurement of subsidence because the principal modes of failure in this technique are thought to be subsidence of the prosthesis and the inability of the graft to withstand repeated loading. An axial load cell (Model 62.10A-05; MTS Systems Inc, Minneapolis, Minnesota) was used to record the compressive forces applied to the femur and also for feedback control of the cyclic testing. In each test the implant was slowly loaded through the prosthetic head to 440 N, simulating 70% of the body-weight. The position was recorded and an anteroposterior radiograph obtained. Cyclic testing was then performed by increasing the compressive force to 1320 N (approximately 200% of the body-weight during the stance phase of gait) and returning to 440 N (the swing phase of gait) at a frequency of 1 Hz over 5000 cycles to simulate the first few days to weeks of limited load-bearing. The axial displacement sensor for the linear actuator of the MTS machine was used to measure the position of the femoral head. The maximum and minimum positions were recorded for each cycle. The total axial displacement was defined as the difference in the position of the actuator between the first and last cycles under 440 N load. The axial subsidence was defined as the difference in position between the fifth and last cycles under 440 N load. The first five cycles were excluded to avoid the rapid initial settling of the specimen which occurs on initial loading. After cyclic testing another radiograph was obtained to check for visual evidence of subsidence as compared with that measured mechanically. VOL. 78-B, NO. 6, NOVEMBER 1996
Table I. Results for cyclic loading in vitro of revision and primary hip prostheses in a cadaver model Total axial displacement after 5000 cycles (mm)
Axial subsidence from 5 to 5000 cycles (mm)
Specimen
Primary
Revision
Primary
Revision
1
-0.44
-1.35
0.36
0.48
2
-0.36
-1.15
0.25
0.36
3
-1.19
-2.33
0.51
1.04
4
-0.66
-0.66
0.16
0.32
5
-0.49
-0.79
0.09
0.40
Mean ±
SD
-0.63 ± 0.33* -1.25 ± 0.66*
0.27 ± 0.17† 0.52 ± 0.30†
*p < 0.05 † p < 0.025
To test the ultimate axial stability the final 500 cycles in each test were used to compute a final rate of subsidence taken as the slope of a ‘best-fit’ line through the last 500 data points. This slope indicates the incremental distance moved with each cycle at the end of the loading period. The femora were then loaded to failure in torsion using another rig (Fig. 2). We performed torsional tests as a definitive measure of the static stability. An axial load of 440 N was applied and the prosthesis then rotated at a constant angular rate of 2.0°/s. The angle and the applied torque were recorded simultaneously. The torsional strength (N-m) was determined as the maximum torque (N-m) applied to the specimen before failure. The torsional stiffness (N-m/°) was calculated as the slope of the linear portion of the curve generated by plotting applied torque (N-m) against the angle of deformation (°). Data analysis. The total axial displacement, the axial subsidence, and the rate of final subsidence were analysed to determine the extent of subsidence of the implant and to ascertain if there was any statistically significant difference
976
A. L. MALKANI,
M. J. VOOR,
K. A. FEE,
C. S. BATES
Fig. 3 A typical axial cyclic loading test indicates decreasing subsidence with increasing cycles in both primary and revision procedures. The curves represent the position of the prosthesis after each cycle during the swing (70% body-weight) portion of the simulated gait cycle.
between the primary and revision sides. The sinusoidal axial load varied between a maximum of 1320 N, representing the stance phase, and a minimum of 440 N, representing the swing phase of gait. Only measurements in the swing phase were analysed in order to minimise the influence of flexion of the femur and implant which is more pronounced in the stance phase. The axial displacement, the subsidence and the rate of subsidence on the primary side were compared with those on the revision side and tested for statistical significance using paired t-tests, as were the torsional strength and stiffness.
RESULTS Axial subsidence. An error in collection of the data required elimination of one set of results of axial testing of the six paired femora. The results of axial loading on the remainder are shown in Figure 3 and Tables I and II. Both primary and revision specimens showed rapid subsidence in the first few cycles followed by a steadily decreasing rate with further loading. In every case, nearly all of the total displacement was seen during the first 1000 cycles of loading with relatively little occurring later. The total axial displacement by 5000 cycles was 0.63 ± 0.33 mm for the primary and 1.25 ± 0.66 mm for the revision prostheses. The axial subsidence from 5 to 5000 cycles was 0.27 ± 0.17 mm for the primary and 0.52 ± 0.30 mm for the revision prostheses. The results of the paired t-tests showed a statistically significant difference in both the axial displacement and subsidence between the primary and revision groups (p < 0.025). The mean final rates of subsidence were 0.0108 ± 0.0082 m/cycle and 0.0143 ± 0.0097 m/ cycle for the primary and revision prostheses respectively; these were not significantly different (p > 0.05). Radiological analysis was inadequate since measurement was not able to distinguish the small displacements observed. Torsional testing. The results of torsional testing are presented in Figure 4 and Table III. A testing error eliminated one pair of specimens. In every case failure was seen as a longitudinal or spiral crack in the femoral cortex. Failure or gross rotational displacement within the graft was not seen
Table II. Final subsidence rate of revision and primary hip prostheses after 5000 cycles in a cadaver model Subsidence rate over last 500 cycles (µm/cycle) Specimen
Primary
Revision
1
0.0172
0.0229
2
0.0127
0.0108
3
0.0192
0.0251
4
0.0003
0.0014
0.0044
0.0115
0.0108 ± 0.0082
0.0143 ± 0.0097
5 Mean ±
SD
Table III. Torsional testing results for in vitro loading of revision and primary hip prostheses in a cadaver model Maximum torque (N-m)
Torsional stiffness (N-m/°)
Specimen
Primary
Revision
Primary
1
22.7
15.7
1.8
3.2
2
75.6
68.2
6.7
9.8
3
35.4
35.2
4.0
9.1
4
14.4
19.3
1.8
2.5
5
66.2
35.8
11.0
5.7
42.9 ± 26.9
34.8 ± 20.7
Mean ±
SD
5.1 ± 3.9
Revision
6.1 ± 3.3
in the revision specimens. The torque at failure was 42.86 ± 26.9 N-m for the primary and 34.82 ± 20.7 N-m for the revision specimens. The torsional stiffness was 5.06 ± 3.9 N-m/° for the primary and 6.06 ± 3.3 N-m/° for the revision specimens. The strength and stiffness of the primary and revision specimens were not significantly different (p > 0.15).
DISCUSSION Concern about the stability of the implant after revision of the femoral prosthesis using impacted morsellised cancellous allograft has led to uncertainty about the immediate postoperative management. We have tested the axial and torsional stability of the femoral component after revision in a cadaver model, and although we could not reproduce the conditions pertaining in vivo we were able to compare THE JOURNAL OF BONE AND JOINT SURGERY
FEMORAL COMPONENT REVISION USING IMPACTED MORSELLISED CANCELLOUS GRAFT
977
Fig. 4 Typical torsional tests for revision and primary specimens indicate similar strength and stiffness.
the immediate stability of the prosthesis after the Ling technique with that of a standard primary cemented component. The total axial subsidence was greater in the revision than in the primary group; most of the axial displacement in both occurred in the first few cycles. This probably represents settling and wedging of the polished implant into the cement mantle. In the revision group, rapid settling would also be expected in the morsellised bone which tended to expand slightly when left uncompressed for the period before the specimens were placed in the testing machine. Under clinical conditions, the hip is reduced intraoperatively and loaded immediately and continually by muscle contraction, so that there is no opportunity for graft expansion to take place. We therefore felt that it was more relevant to the clinical state to observe the subsidence beginning with the fifth cycle. The mean subsidence from 5 to 5000 cycles was 0.27 mm in the primary compared with 0.52 mm in the revision prostheses; neither would be considered clinically unstable. In both groups the subsidence steadily decreased with increasing loading cycles with no statistically significant difference in the final rate of subsidence (Table II), indicating increasing axial stability of the femoral component. Analysis of the torsional data showed no statistically significant difference in strength or stiffness between the revision and primary groups. There was a large variation in the torsional findings (strength range 14.4 to 75.6 N-m; stiffness range 1.8 to 11.0 N-m/°), but this was primarily due to the variability in the quality of the bone in individual specimens. Torsional strength and stiffness were more dependent on the pair of femora being tested than on whether the procedure was of the primary or revision type
VOL. 78-B, NO. 6, NOVEMBER 1996
(Table III). Statistically, the torsional stability was no different in the revision than in the primary group. Both failed by longitudinal or spiral cracking of the cortex, indicating that the implant/cement/graft/cortex interfaces were stronger in torsion than the bone itself. While increased axial stability from driving a wedge-shaped prosthesis into the graft material may be apparent, the excellent torsional characteristics strongly support the biomechanical rationale for this procedure. Rotational moments have been reported about the hip during stair-climbing which are similar to the torques observed at failure in our experiment (Bergmann, Graichen and Rohlmann 1995). Our experimental method may have created a tendency towards lower torsional strength than would be expected in a whole healthy femur but failure occurred at the cortex and not in the prosthesis/ cement/graft complex. The clinical implications of high torsional loading of the impacted allograft/host bone interface can only truly be addressed by further clinical investigation. It is interesting to speculate where the slightly greater subsidence occurred in the revision group. The size of the prosthesis was the same in both. The wedge-shaped polished implant attempts to subside in the cement mantle causing an outward or radial compressive effect into the layers. The subsidence within the cement was probably the same in the two groups and the difference lay in the interposition of the impacted morsellised graft between the cement mantle and the solid host bone in the revision model. Compression will cause radial displacement of the graft which results in subsidence of the implant. Shear forces along the interfaces may allow slippage of the cement along the graft or of the graft along the smooth
978
A. L. MALKANI,
M. J. VOOR,
cortical wall. Because of the wedge effect and interdigitation between the cement and the graft it is less likely that slippage will occur at the cement-graft interface. We therefore believe that the subsidence is due to a combination of slight slippage of the graft mantle against the smooth cortical wall of the host femur and compression as a result of the wedging of the implant and cement into it. The success of the technique is due to the increased friction developed by the high compressive forces; this results in progressively improved stability and reduced slippage. We feel that revision of the femoral component using impacted morsellised cancellous allograft can provide immediate stability of the implant. Firm impaction of the graft material is essential. Potential fractures due to high stresses caused by graft impaction may be managed by supplementing the host cortical bone with allograft cortical strut grafts. We were concerned about the immediate torsional stability under high torsional loading such as during stair-climbing and have allowed our patients to progress gradually to full weight-bearing by three months. Our experimental results are encouraging but must be confirmed by long-term clinical investigation. This research was supported by Zimmer Inc and the Fischer-Owen Orthopaedic Fund. Although none of the authors have received or will receive benefits for personal or professional use from a commercial party related directly or indirectly to the subject of this article, benefits have been or will be received but are directed solely to a research fund, foundation, educational institution, or other non-profit institution with which one or more of the authors is associated. REFERENCES Bergmann G, Graichen F, Rohlmann A. Is staircase walking a risk for the fixation of hip implants? J Biomech 1995;28:535-53. Callaghan JJ. Total hip arthroplasty: clinical perspective. Clin Orthop 1992;276:33-40. Callaghan JJ, Salvati EA, Pellicci PM, Wilson PD Jr, Ranawat CS. Results of revision for mechanical failure after cemented total hip replacement, 1979 to 1982: a two to five-year follow-up. J Bone Joint Surg [Am] 1985;67-A:1074-85. Chandler H, Clark J, Murphy S, et al. Reconstruction of major segmental loss of the proximal femur in revision total hip arthroplasty. Clin Orthop 1994;298:67-74. D’Antonio J, Capello M, Borden L, et al. Classification and management of acetabular abnormalities in total hip arthroplasty. Clin Orthop 1989;243:126-37.
K. A. FEE,
C. S. BATES
Dohmae Y, Bechtold JE, Sherman RE, Puno RM, Gustilo RB. Reduction in cement-bone interface shear strength between primary and revision arthroplasty. Clin Orthop 1988;236:214-20. Emerson RH Jr, Malinin TI, Cuellar AD, Head WC, Peters PC. Cortical strut allografts in the reconstruction of the femur in revision total hip arthroplasty: a basic science and clinical study. Clin Orthop 1992;285:35-44. Engh CA, Glassman AH, Griffin WL, Mayer JG. Results of cementless revision for failed cemented total hip arthroplasty. Clin Orthop 1988;235:91-110. Gie GA, Linder L, Ling RSM, et al. Impacted cancellous allografts and cement for revision total hip arthroplasty. J Bone Joint Surg [Br] 1993;75-B;14-21. Gross AE, Allan DG, Lavoie GJ, Oakeshott RD. Revision arthroplasty of the proximal femur using allograft bone. Orthop Clin North Am 1993;24:705-15. Gustilo RB, Pasternak HS. Revision total hip arthroplasty with titanium ingrowth prosthesis and bone grafting for failed cemented femoral component loosening. Clin Orthop 1988;235:111-9. Head WC, Wagner RA, Emerson RH Jr, Malinin TI. Restoration of femoral bone stock in revision total hip arthroplasty. Orthop Clin North Am 1993;24:697-703. Hedley AK, Gruen TA, Ruoff DP. Revision of failed total hip arthroplasties with uncemented porous-coated anatomic components. Clin Orthop 1988;235:75-90. Hungerford DS, Jones LC. The rationale of cementless revision of cemented arthroplasty failures. Clin Orthop 1988;25:12-24. Kavanagh BF, Ilstrup DM, Fitzgerald RH Jr. Revision total hip arthroplasty. J Bone Joint Surg [Am] 1985;67-A:517-26. Ling RSM, Timperley AJ, Linder L. Histology of cancellous impaction grafting in the femur: a case report. J Bone Joint Surg [Br] 1993;75-B:693-6. Malkani AL, Sim FH, Chao EY. Custom-made segmental femoral replacement prosthesis in revision total hip arthroplasty. Orthop Clin North Am 1993;24:727-33. Malkani AL, Lewallen DG, Cabanela ME, Wallrichs SL. Femoral component revision using an uncemented, proximally coated, longstem prosthesis. J Arthoplasty 1996;11:411-8. Mallory TH. Preparation of the proximal femur in cementless total hip revision. Clin Orthop 1988;235:47-60. Marti RK, Schuller ¨ HM, Besselaar PP, Vanfrank Haasnoot EL. Results of revision of hip arthroplasty with cement: a five to fourteenyear follow-up study. J Bone Joint Surg [Am] 1990;72-A:346-54. Paprosky WG, Bradford MS, Younger TI. Femoral reconstruction with massive allograft and cementless prosthesis. Chir Organi Mov 1994;79:313-8. Pellicci PM, Wilson PD Jr, Sledge CB. Revision total hip arthroplasty. Clin Orthop 1982;170:34-41. Poss R, Walker P, Spector M, et al. Strategies for improving fixation of femoral components in total hip arthroplasty. Clin Orthop 1988; 235:181-94. Rubash HE, Harris WH. Revision of nonseptic, loose, cemented femoral components using modern cementing techniques. J Arthroplasty 1988;3:241-8.