Flow velocity estimation using joint Spectral and Time domain Optical Coherence Tomography Maciej Szkulmowski, Anna Szkulmowska, Tomasz Bajraszewski, Andrzej Kowalczyk, and Maciej Wojtkowski* Institute of Physics, Nicolaus Copernicus University, ul. Grudziadzka 5, PL87-100 Torun, Poland *Corresponding author:
[email protected]
Abstract: We propose a modified method of acquisition and analysis of Spectral Optical Coherence Tomography (SOCT) data to provide information about flow velocities. The idea behind this method is to acquire a set of SOCT spectral fringes dependent on time followed by a numerical analysis using two independent Fourier transformations performed in time and optical frequency domains. Therefore, we propose calling this method as joint Spectral and Time domain Optical Coherence Tomography (joint STdOCT). The flow velocities obtained by joint STdOCT are compared with the ones obtained by known, phase-resolved SOCT. We observe that STdOCT estimation is more robust for measurements with low signal to noise ratio (SNR) as well as in conditions of close-to-limit velocity measurements. We also demonstrate that velocity measurement performed with STdOCT method is more sensitive than the one obtained by the phaseresolved SOCT. The method is applied to biomedical imaging, in particular to in vivo measurements of retinal blood circulation. The applicability of STdOCT different measurement modes for in vivo examinations, including 1, 5 and 40 μs of CCD exposure time, is discussed. ©2008 Optical Society of America OCIS codes: (170.4500) Optical coherence tomography; (170.3880) Medical and biological imaging; (170.4470) Ophthalmology; (280.2490) Flow diagnostics
References and links 1.
2. 3. 4.
5.
6.
7.
8.
D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Puliafito, and J. G. Fujimoto, "Optical coherence tomography," Science 254, 1178-1181 (1991). A. F. Fercher, C. K. Hitzenberger, G. Kamp, and S. Y. Elzaiat, "Measurement of Intraocular Distances by Backscattering Spectral Interferometry," Opt. Commun. 117, 43-48 (1995). M. Wojtkowski, R. Leitgeb, A. Kowalczyk, T. Bajraszewski, and A. F. Fercher, "In vivo human retinal imaging by Fourier domain optical coherence tomography," J. Biomed. Opt. 7, 457-463 (2002). N. Nassif, B. Cense, B. H. Park, S. H. Yun, T. C. Chen, B. E. Bouma, G. J. Tearney, and J. F. de Boer, "In vivo human retinal imaging by ultrahigh-speed spectral domain optical coherence tomography," Opt. Lett. 29, 480-482 (2004). M. Wojtkowski, V. Srinivasan, J. G. Fujimoto, T. Ko, J. S. Schuman, A. Kowalczyk, and J. S. Duker, "Three-dimensional retinal imaging with high-speed ultrahigh-resolution optical coherence tomography," Ophthalmology 112, 1734-1746 (2005). V. J. Srinivasan, M. Wojtkowski, A. J. Witkin, J. S. Duker, T. H. Ko, M. Carvalho, J. S. Schuman, A. Kowalczyk, and J. G. Fujimoto, "High-definifion and 3-dimensional imaging of macular pathologies with high-speed ultrahigh-resolution optical coherence tomography," Ophthalmology 113, 2054-2065 (2006). R. Leitgeb, M. Wojtkowski, A. Kowalczyk, C. K. Hitzenberger, M. Sticker, and A. F. Fercher, "Spectral measurement of absorption by spectroscopic frequency-domain optical coherence tomography," Opt. Lett. 25, 820-822 (2000). R. A. Leitgeb, L. Schmetterer, W. Drexler, A. F. Fercher, R. J. Zawadzki, and T. Bajraszewski, "Real-time assessment of retinal blood flow with ultrafast acquisition by color Doppler Fourier domain optical coherence tomography," Opt. Express 11, 3116-3121 (2003).
#92569 - $15.00 USD
(C) 2008 OSA
Received 7 Feb 2008; revised 7 Apr 2008; accepted 8 Apr 2008; published 14 Apr 2008
28 April 2008 / Vol. 16, No. 9 / OPTICS EXPRESS 6008
9.
10. 11. 12.
13. 14.
15.
16. 17. 18. 19.
20. 21.
22. 23. 24.
25. 26. 27.
Y. Yasuno, S. Makita, Y. Sutoh, M. Itoh, and T. Yatagai, "Birefringence imaging of human skin by polarization-sensitive spectral interferometric optical coherence tomography," Opt. Lett. 27, 1803-1805 (2002). M. A. Choma, A. K. Ellerbee, S. Yazdanfar, and J. A. Izatt, "Doppler flow imaging of cytoplasmic streaming using spectral domain phase microscopy," J. Biomed. Opt. 11 (2006). F. Rothenberg, A. M. Davis, and J. A. Izatt, "Non-invasive investigations of early embryonic cardiac blood flow with optical coherence tomography," FASEB J. 20, A451-A451 (2006). B. R. White, M. C. Pierce, N. Nassif, B. Cense, B. H. Park, G. J. Tearney, B. E. Bouma, T. C. Chen, and J. F. de Boer, "In vivo dynamic human retinal blood flow imaging using ultra-high-speed spectral domain optical Doppler tomography," Opt. Express 11, 3490-3497 (2003). S. Makita, Y. Hong, M. Yamanari, T. Yatagai, and Y. Yasuno, "Optical coherence angiography," Opt. Express 14, 7821-7840 (2006). Y. Zhao, Z. Chen, C. Saxer, S. Xiang, J. F. de Boer, and J. S. Nelson, "Phase-resolved optical coherence tomography and optical Doppler tomography for imaging blood flow in human skin with fast scanning speed and high velocity sensitivity," Opt. Lett. 25, 114-116 (2000). B. H. Park, M. C. Pierce, B. Cense, S. H. Yun, M. Mujat, G. Tearney, B. Bouma, and J. de Boer, "Real-time fiber-based multi-functional spectral-domain optical coherence tomography at 1.3 µm," Opt. Express 13, 3931-3944 (2005). S. H. Yun, G. J. Tearney, J. F. de Boer, and B. E. Bouma, "Motion artifacts in optical coherence tomography with frequency-domain ranging," Opt. Express 12, 2977-2998 (2004). A. H. Bachmann, M. L. Villiger, C. Blatter, T. Lasser, and R. A. Leitgeb, "Resonant Doppler flow imaging and optical vivisection of retinal blood vessels," Opt. Express 15, 408-422 (2007). R. K. Wang, S. L. Jacques, Z. Ma, S. Hurst, S. R. Hanson, and A. Gruber, "Three dimensional optical angiography," Opt. Express 15, 4083-4097 (2007). Y. Yasuno, S. Makita, T. Endo, G. Aoki, M. Itoh, and T. Yatagai, "Simultaneous BM-mode scanning method for real-time full-range Fourier domain optical coherence tomography," Appl. Opt. 45, 18611865 (2006). Z. Chen, T. E. Milner, D. Dave, and J. S. Nelson, "Optical Doppler tomographic imaging of fluid flow velocity in highly scattering media," Opt. Lett. 22, 64-66 (1997). J. A. Izatt, M. D. Kulkami, S. Yazdanfar, J. K. Barton, and A. J. Welch, "In vivo bidirectional color Doppler flow imaging of picoliter blood volumes using optical coherence tomography," Opt. Lett. 22, 1439-1441 (1997). S. Yazdanfar, C. H. Yang, M. V. Sarunic, and J. A. Izatt, "Frequency estimation precision in Doppler optical coherence tomography using the Cramer-Rao lower bound," Opt. Express 13, 410-416 (2005). B. Vakoc, S. Yun, J. de Boer, G. Tearney, and B. Bouma, "Phase-resolved optical frequency domain imaging," Opt. Express 13, 5483-5493 (2005). M. Wojtkowski, V. J. Srinivasan, T. H. Ko, J. G. Fujimoto, A. Kowalczyk, and J. S. Duker, "Ultrahighresolution, high-speed, Fourier domain optical coherence tomography and methods for dispersion compensation," Opt. Express 12, 2404-2422 (2004). M. Wojtkowski, A. Kowalczyk, R. Leitgeb, and A. F. Fercher, "Full range complex spectral optical coherence tomography technique in eye imaging," Opt. Lett. 27, 1415-1417 (2002). R. Leitgeb, C. K. Hitzenberger, and A. F. Fercher, "Performance of Fourier domain vs. time domain optical coherence tomography," Opt. Express 11, 889-894 (2003). J. W. Goodman, Statistical Optics (John Wiley & Sons, Inc., New York, 2000).
1. Introduction Optical Coherence Tomography (OCT) provides cross-sectional images of internal structure of objects with micrometer resolution, and can be either performed in time [1] or frequency domain [2]. The latter is performed by analysis of interferometric modulations of light intensity versus optical wavelengths (spectral fringe signal). In Spectral OCT known also as Spectral domain OCT (SOCT/SdOCT) spectral intereferometric fringes are registered by a spectrometer. This modality is particularly useful for ophthalmic examinations since it offers high speed of more than 20 000 A-scans per second and detection sensitivity of more than 95dB [3-6]. The spectral interferometric fringe signals are collected for each lateral position of the scanning beam and numerically processed to obtain two-dimensional cross-sectional images representing the amount of back-reflected light versus depth and lateral positions of the elements of internal structure of an object. In addition to morphological imaging, SOCT can provide visualization of physiological parameters [7-11]. At present the retinal blood flow attracts attention as a potentially important physical parameter in the functional OCT studies. The measurements of #92569 - $15.00 USD
(C) 2008 OSA
Received 7 Feb 2008; revised 7 Apr 2008; accepted 8 Apr 2008; published 14 Apr 2008
28 April 2008 / Vol. 16, No. 9 / OPTICS EXPRESS 6009
bidirectional flow by Spectral OCT have been demonstrated by many groups [8, 12, 13]. In all of these publications, authors used the phase-resolved technique based on linear relationship between the phase difference of consecutive spectral fringe signals and the velocity of the moving sample. This technique is analogous to phase-resolved approach applied in the OCT methods with the time domain detection [14]. Two main factors deteriorate and even preclude velocity recovery in the phase-resolved OCT techniques: its vulnerability to low signal-to-noise ratio (SNR) [15] and motion artifacts causing severe phase instabilities [12, 16]. Both problems frequently occur in OCT measurements of biomedical objects in vivo. Recently, a novel spectral method has been proposed in order to minimize the influence of phase instabilities so called resonant Doppler imaging [17], which determines flow velocities on the intensity basis without the need of extracting the signal phase. Moreover, this method overcomes a phenomenon of blurring of interference fringes caused by any sample movement during CCD camera integration time. Another phase independent method was proposed by Wang, et al., [18]. Since this technique separates the moving and static components within a sample, only blood perfusion imaging is possible without any flow velocity estimations. This optical angiography relies on introducing a constant Doppler frequency to modulate the spatial OCT spectral interferograms what was initially introduced to Spectral OCT by Yasuno, et al., in 2006 [19]. In this paper we present an alternative method of measuring and processing OCT signals providing information on the spatial distribution of flow velocities. We propose joint time and frequency domain detection of interferometric OCT signals. The information about the velocity is obtained directly – similarly to first velocity estimation techniques in TdOCT [20, 21] – from the time dependent beating frequency due to the Doppler shift between the reference and the sample light beams. Since the velocity estimation is not based on explicit phase information extracted from interferometric fringes, the proposed variant of Doppler SOCT is significantly less sensitive to undesired phase instabilities present in low SNR conditions. Since it does not require any phase wrapping and averaging procedure, it is accurate for flows close to the upper limit of measurable velocities. This approach does not require any modifications in hardware of a standard SOCT instrument. High sensitivity of this method facilitates flow velocity estimation within the time frame required by the regular OCT imaging. In many biomedical applications, especially in ophthalmology, there are severe limitations in optical power, which can be delivered to the sample. In such cases a preservation of high sensitivity requires fixed value of exposure time of the CCD camera collecting the spectral fringe data. In this case the multi-shot measurements required in STdOCT can be balanced by reduction of CCD exposure time. And the same it is possible to keep sensitivity and the optical power delivered to the object at the same levels like in regular SOCT imaging. In such case the spectral fringe signals can be first processed and then superimposed giving the structural reconstruction, while time dependent Fourier transformation will yield information about flow velocities. Comparing to phase-resolved techniques our method can operate in conditions of much lower SNR still preserving high accuracy in the whole velocity range, what is crucial in any quantitative measurements of biomedical samples in vivo in OCT functional studies. 2. Theory In Spectral OCT Fourier transformation (FT) of the spectral fringe signal measured by a single exposure of a CCD camera creates one line of a structural cross-sectional image (Ascan). In order to asses a velocity of a moving interface most of known methods require at least two spectral fringes acquired in the same lateral position of a sampling light beam [8] or almost the same lateral position [15]. The acquired set of spectral interferometric fringes can be described as a function of wavenumber k and time t according to the following equation:
#92569 - $15.00 USD
(C) 2008 OSA
Received 7 Feb 2008; revised 7 Apr 2008; accepted 8 Apr 2008; published 14 Apr 2008
28 April 2008 / Vol. 16, No. 9 / OPTICS EXPRESS 6010
⎛
I (k , t ) = I 0 (k )⎜ ⎝
∑ R + R + 2∑ l
r
l
⎞
Rl Rr cos(2 zl (t ) ⋅ k )⎟ ,
(1)
⎠
l
where I (k , t ) is the spectral fringe signal, I 0 (k ) is spectral density of the light source, Rl and Rr denote the reflectivity of the sample and reference mirror, respectively; zl (t ) denotes the optical path difference between the reference mirror and the l -th interface in the sample, which is time dependent due to the movement of the reference mirror and/or due to the displacement of the l -th interface in the sample. The displacement of the interfaces within the sample is usually caused either by a movement of the entire sample itself or by a motion of the specific interface z l within the sample. If we assume that both, reference mirror velocity and velocities in the sample are constant during the acquisition of the spectral fringes, Eq. (1), then the time-dependent position of the l -th interface zl (t ) can be expressed as: zl (t ) = zl + δz (t ) = zl + vl ⋅ t (2) In this relation, z l is the depth position of the l -th interface at the beginning of data acquisition and vl is the difference between the velocity of the reference mirror and an axial component of velocity (parallel to the direction of the probing beam propagation) of the l -th interface. If the l -th interface moves with velocity Vl at an angle α to the probing beam and the velocity of the reference mirror is equal to vr this can be expressed as: vl = Vl cos α − vr
(3)
Here velocities directed towards a beam-splitter are regarded positive. One can rewrite Eq. (1) making use of Eqs. (2) and (3).
((
) )⎞⎟
I (k , t ) = I 0 (k )⎜ ∑ Rl + Rr + 2∑ Rl Rr cos 2 zl + δz (t ) ⋅ k l ⎝l ⎛
⎛
I (k , t ) = I 0 (k )⎜ ⎝
∑ R + R + 2∑ l
r
l
l
(4)
⎠
⎞
Rl Rr cos(2 zl ⋅ k + ωl ⋅ t )⎟
(5)
⎠
Although the above equations represent the same interference pattern, they emphasize its different properties. Phase of the oscillatory component visible in Eq. (4) is a function of wavenumber and its modulation frequency depends on static position zl of l -th interface and small additional change of δz , that occurs if the l -th interface is moving. Equation (5) highlights the time-dependence of the interferometric fringes and shows that signal is modulated in time with frequency ωl . This beat frequency is caused by a Doppler effect, that arises for each l -th interface along the time axis. This frequency depends on the velocity vl and is different for each wavenumber k :
ωl = 2vl k
(6)
The phase-resolved methods of velocity estimation enables extracting and using the phase of the signal Eq. (4) while the joint Spectral and Time domain OCT uses the Fourier transformation to analyze the time-dependent frequency of the signal Eq. (5).
#92569 - $15.00 USD
(C) 2008 OSA
Received 7 Feb 2008; revised 7 Apr 2008; accepted 8 Apr 2008; published 14 Apr 2008
28 April 2008 / Vol. 16, No. 9 / OPTICS EXPRESS 6011
2.1 Velocity measurement using phase-resolved SOCT The idea behind the phase-resolved OCT is to determine the phase difference between points at the same depth in consecutive A-scans. Knowing the change of position δz of l -th interface, that arises during the time Δt between two consecutive measurements, the velocity
vl of the l -th interface can be calculated. Since δz is much smaller than zl , the difference between two consecutive measurements appears as a phase change ΔΦ of interferometric fringes: ΔΦ λ ΔΦ . (7) vl = = 2kΔt 4Δt π Here ΔΦ is the phase difference between successively recorded depth profiles at the same location of the probing beam. The time between successive profiles acquisition Δt is approximately equal to the exposure time of the detector, therefore 1 Δt is the frame rate of an array detector (or equivalently ‘A-scan’ rate). It is important to ensure that ΔΦ is less than 2π . Since the phase can be unambiguously determined in the range of 2π , and the phase difference is within the range of 4π a procedure of phase wrapping has to be performed to transform the phase differences to the range (−π , π ) [12, 13]. In the procedure adapted to recovery of bidirectional flows the following algorithm is used: if | ΔΦ |< π , the phase shift it is left as it is, and when π >1 the minimum phase difference can be determined as ΔΦ min = (SNR ) −1 2 [15]. Vakoc, et al., suggest that retinal blood flow velocities 0 ,
(22)
otherwise
Where I 0 (⋅) is a modified Bessel function of the first kind, zero order. As the length of the known phasor s increases, the shape of density function p A (a) changes from that of a Rayleigh density to approximately a Gaussian density with mean equal to s . Joint STdOCT uses the time dependent modulation of the signal, therefore it is successful, when the amplitude of signal is higher than the maximal amplitude of noise component. This occurs when the distributions of signal amplitude ( κ zω ≠ 0 , Eq. (21)) and noise ( κ zω = 0 ) are separated. The minimal value of κzω, which almost always meets this requirement is κ zω = 7 , Fig. 3.
Fig. 3. The probability density functions of amplitude for different value of parameter κzω. Black and red curve correspond to the distributions of amplitude of pure noise (κzω=0) and amplitude of signal for critical value of (κzω=7) that assure correct recovery of velocity in STdOCT.
When κ zω converges to zero, the probability of detecting the correct position of signal amplitude decrease. Every detection of noise causes indication of random velocity, therefore,
#92569 - $15.00 USD
(C) 2008 OSA
Received 7 Feb 2008; revised 7 Apr 2008; accepted 8 Apr 2008; published 14 Apr 2008
28 April 2008 / Vol. 16, No. 9 / OPTICS EXPRESS 6017
the distribution of recovered velocity broaden and its mean value converges to the center of the available velocity range (usually to zero). To determine critical values of SNR below which STdOCT and the phase-resolved method give false readings, we performed computer simulations based on provided theoretical model. In order to reconstruct the process of velocity estimation, multiple signals ( M = 30 ) were generated with respect to the shape of spectrum and the probability density function of amplitude (Eq. (22)) and phase (Eq. (21)). The magnitude of change in harmonic component between consecutive signals was set to correspond to 0.35 v max and 0.75 v max . Both methods operates on exactly the same amount of generated signals. The velocity was recovered for different SNR and the results are shown in Fig. 4.
Fig. 4. Computer simulations of velocity estimations obtained with STdOCT and phase-resolved SOCT for two different velocities (0.35vmax and 0.75 vmax – solid and dashed curve, respectively) under the conditions of decreasing SNR.
Both methods fail for certain SNR, however STdOCT is more robust under low SNR. The failure appears as an underestimate of retrieved velocity values. In joint STdOCT the critical SNR above which recovered velocity is reliable does not depend on the magnitude of the set velocity, whereas phase-resolved SOCT fails earlier for higher velocity. To explain this effect we analyzed the velocity recovery process in the phase-resolved method within the entire theoretical range. Four different velocities: 0.05, 0.35, 0.75 and 0.95 of vmax were chosen and the velocity estimation for each of them was performed. The dependence of velocity reading on proximity to the theoretical limits of velocity are shown in Fig. 5(a). One can see that there is no significant difference in critical values of SNR for velocities 30 dB (Fig. 5), whereas the corresponding value for STdOCT is >6 dB (Fig. 4). Additionally, since signal amplitude Azω used in joint STdOCT to retrieve velocity value depends on number of spectra registered in time ( M ), the critical SNR can be improved M 1 / 2 times with increasing number of measurements (Table 1.). In phase-resolved SOCT increasing number of spectra does not improve measurement sensitivity, however it facilitates detection of mean value of phase difference distribution. Performed simulations do not take into consideration the washout of interference fringes [17]. This phenomenon deteriorate SNR, hence in this way it affects velocity recovery. Because both methods suffer from the blurring of interference fringes in the same degree, the conducted comparison is still valid. 3. Experiment We use laboratory high resolution Spectral OCT system comprising a broadband light source (Broadlighter, Superlum, Δλ = 90nm , central wavelength 840 nm), a fiber Michelson interferometer with fixed reference mirror and custom designed spectrometer with a volume phase holographic grating DG (1200 grooves/mm) and an achromatic lens focusing spectrum on 12-bit CCD line-scan camera (Aviiva M2, Atmel), Fig. 6. The experiments were performed for three different objects: moving mirror, capillary flow and blood flow in human retina. In measurements of the velocity of moving mirror, a silver mirror was attached to a piezoactuator (Physik Instrumente) and it was driven by a triangular voltage signal. The exact velocity was calculated at the moment of a linear slope of the driving signal from trajectory registered by the position sensor mounted inside the actuator. Measurements were performed with A-scan rate of 40.4 μs. To investigate flows in scattering media, we used a water solution of Intralipid flowing through capillaries. Two 700 μm thick glass capillaries with flow in opposite directions were mounted at the angle of 88 deg to the direction of the probing beam ( z -axis) and stable, laminar flow was ensured by a medical drip system. The sets of 40 spectra were collected at the same transversal position of the light beam. The acquisition time was set to 52 μs including 10 μs dead time needed for stabilization of the position of galvo scanner driven by the stepwise signal. The optical power of the light illuminating the sample was 3.3 mW.
#92569 - $15.00 USD
(C) 2008 OSA
Received 7 Feb 2008; revised 7 Apr 2008; accepted 8 Apr 2008; published 14 Apr 2008
28 April 2008 / Vol. 16, No. 9 / OPTICS EXPRESS 6019
Fig. 6. Experimental Spectral OCT system : OI optical isolator, FC fiber coupler, PC polarization controller, DC dispersion compensator, NDF neutral density filter, X galvo-scanner, L1 lens, DG volume phase holographic grating, CCD line-scan camera. Three different objects are measured: moving mirror, capillary flow, blood flow in human retina.
For all retinal blood flow examination the optical power of light illuminating the cornea was set to 750 μW. In the case of regular exposure time (40 μs) the velocity recovery is based on 20 spectra, each recorded with 43 μs of repetition time. In measurements with short exposure time (5 μs and 1 μs) 40 spectra were collected for each lateral position of the scanning beam. Despite such a short exposure times, the repetition time was >41 μs due to the dead time of 40 μs between consecutive measurements. The velocity estimations in phase-resolved SOCT and joint STdOCT are always based on exactly the same registered data in all comparative experiments. This guarantees that the differences in velocity recovery are solely caused by the methods themselves, not by experimental environment or different amount of processed data. 4. Results and discussion 4.1 Moving mirror In order to validate provided theoretical analyses we performed an experiment with a moving mirror as an object. To investigate the relation between the velocity estimation and SNR, the light intensity in objective arm had been reduced by a neutral density filter from 20 dB to 6 dB (from =10 to =0.5) for two velocities 1.9 mm/s and 3.9 mm/s corresponding to 0.35 and 0.95 of vmax. The sets of 30 spectra were collected and then processed to obtain phaseresolved and STdOCT velocity estimations. Figures 7(a), 7(b) presents achieved velocity values, which are displayed together with theoretical results demonstrated in Fig. 4. In the next step we verified the capability of measuring velocities close to the upper limit. The objective mirror was driven with different velocities within the whole theoretical range. The intensity of the light in objective arm was constant and resulted in SNR=17.5 dB in structural tomogram. A single velocity value was calculated from 18 spectra. Retrieved values of velocities were marked in Fig. 7(c) together with theoretical prediction.
κ
#92569 - $15.00 USD
(C) 2008 OSA
κ
Received 7 Feb 2008; revised 7 Apr 2008; accepted 8 Apr 2008; published 14 Apr 2008
28 April 2008 / Vol. 16, No. 9 / OPTICS EXPRESS 6020
Fig. 7. Experimental and theoretical comparison of velocity estimations obtained with STdOCT and phase-resolved SOCT a, b. velocity readings for fixed velocities equal to 0,75 vmax and 0.35 vmax, respectively, under the conditions of decreasing SNR; c. velocity readings in whole theoretical velocity range for fixed SNR=17.5 dB.
The experimental results are in good agreement with the theoretical model. Although theoretical ranges for both methods are identical, the phase-resolved SOCT fails earlier than STdOCT and its correctness depends on the magnitude of measured velocity. STdOCT is able to detect a velocity for SNR ~30 dB lower than the phase-resolved method. This leads to conclusion that useful velocity range in phase-resolved SOCT is significantly narrower than in STdOCT. 4.2 Capillary flow Two experiments were designed, one to verify the method of STdOCT in case of bidirectional flow in scattering media and the other to compare with phase-resolved SOCT. Figure 8 presents STdOCT images achieved as individual steps during velocity recovery (section 2.2).
Fig. 8. Bidirectional flow of Intralipid: a. 2D structural tomogram; arrow indicates the direction of incident light b. 2D velocity map, c. Doppler shift distribution in depth calculated from the single set of 40 spectra corresponding to the green vertical line on panel (b). d. distribution of maximal intensity signals retrieved from panel (c); e. plot of the transverse distribution of velocity corresponding to the green horizontal line on panel (b).
The structural tomogram of the capillaries and the velocity map that indicates bidirectional flow are shown in Figs. 8(a), 8(b). Single lines in structural and velocity image are obtained from a set of 40 A-scans. As a first step in velocity recovery, the procedure of zero-padding to 128 points in time space was applied. The signal underwent 2D Fourier transformation and formed the Doppler shift distribution in depth as shown in Fig. 8(c). The positions of maximal intensities for each depth z were detected and points that most likely correspond to noise ( κ zω ≤ 2 ) were removed by thresholding procedure, Fig. 8(d). Images (c) and (d) correspond #92569 - $15.00 USD
(C) 2008 OSA
Received 7 Feb 2008; revised 7 Apr 2008; accepted 8 Apr 2008; published 14 Apr 2008
28 April 2008 / Vol. 16, No. 9 / OPTICS EXPRESS 6021
to a single line in 2D velocity map (Fig. 8(b)), on which the values of velocity are encoded using false colors. Figure 8(e) presents a single 1D velocity distribution along the transversal direction indicated by green horizontal line on the velocity map. All presented velocity distributions have parabolic shapes, what implies that measured flow is laminar. To compare the STdOCT and phase-resolved flow velocity estimation in scattering media, the flow measurements for different concentrations of Intralipid solution and different flow rates were performed. Concentration of scattering medium affects signal intensity and its change yields different characteristics of SNR decrease in depth. The acquisition parameters of OCT data were unchanged. To take full advantage from recording multiple spectra, we averaged all single A-scans and they were displayed as a single line in structural tomograms. The results of both methods of flow estimation are presented in Fig. 9. The experiment was performed under three different flow conditions. In the first case, Fig. 9(a), the concentration of the Intralipid solution was chosen in such a way that the SNR changes significantly between front and back side of the capillary. The flow velocity was set to approximately 0.75 of maximal velocity. Both methods return a parabolic distribution of flow velocity, however phase-resolved method exhibits a slight asymmetry, which increases with depth. Then the Intralipid concentration was changed to maximize SNR at the back side of capillary and the flow velocity was increased to exceed vmax. (Fig. 9(b)) We can observe that both methods give similar readings to approximately half of the velocity range. For higher velocities phaseresolved method dramatically underestimates the velocity values, and for vmax returns zero. In STdOCT velocities beyond the range are wrapped and found as negative values. The distortions of velocity distributions in phase-resolved method appear when SNR decreases or when velocities are too high (however still in the theoretically achievable range).
Fig. 9. Velocity measurements of Intralipid flow in capillaries for different flow rates and concentration of the scattering medium. Each panel consists of 4 images: i. structural tomogram of the inside of capillary; arrow indicates the direction of light and green line marks the position of 1D velocity distributions presented on the last image in the row, ii., iii. velocity maps reconstructed by phase-resolved SOCT and STdOCT, respectively, iv. comparison of 1D velocity distributions retrieved from both velocity maps: green – phase-resolved, red - STdOCT; arrow indicates the direction of light; a. high concentration of Intralipid, flow velocity set to be 0.75 vmax, b. lower concentration of Intralipid, flow velocity slightly higher than vmax, c. high concentration of Intralipid and flow velocity reaches the upper limit, vmax.
#92569 - $15.00 USD
(C) 2008 OSA
Received 7 Feb 2008; revised 7 Apr 2008; accepted 8 Apr 2008; published 14 Apr 2008
28 April 2008 / Vol. 16, No. 9 / OPTICS EXPRESS 6022
The result of velocity estimation in conditions when both effects occur is shown in Fig. 9(c). It is evident that for the illuminated side of the capillary, the SNR is sufficient to return accurate velocity values for both methods. With decreasing SNR and increasing velocity phase-resolved method starts failing, while STdOCT remains unaffected. These results are in a good agreement with theoretical predictions and with experiments performed with moving mirror as an object. 4.3 Retinal blood flow, in vivo As a final test of joint STdOCT and phase-resolved SOCT capabilities in velocity estimations, the measurement of blood flow in human retina in vivo was performed. Figure 10(a) demonstrates cross-sectional image of human retina scanned through the region of optic disc. Figures 10(b) and 10(c) show two-dimensional maps of the flow velocity distribution obtained with SOCT and STdOCT, respectively.
Fig. 10. Measurement of blood flow velocity in human retina in the region of optic disc; a. cross-sectional image of the retina performed in close proximity to optic disc, b., c. velocity maps based on phase-resolved SOCT and STdOCT methods respectively.
The velocity distributions inside the vessel indicated by green lines in Fig. 10 are presented in Fig. 11. Although blood flow in large vessels is evident in both methods, the quantitative velocity estimations differ significantly. The magnitude of the blood velocity in the center of the vessel is 1mm/s for the phase-resolved method and 4mm/s for STdOCT. The recovered shapes of velocity distribution also differ greatly. In the phase-resolved method we observe distortion in the center of the vessel, what may cause a misinterpretation of nature of flow.
Fig. 11. Comparison of two velocity estimations: green points indicate phaseresolved SOCT and the red one STdOCT. Diagrams are based on information extracted from data presented in Fig. 10 (green lines).
The underestimate in phase-resolved SOCT arises from getting out of the useful velocity range. Beyond the useful range, velocity estimators became progressively underestimated with decreasing signal and finally decayed to zero. The useful range in STdOCT is wider, but of course also limited. Another difference is that in STdOCT decreasing signal does not influence the accuracy of velocity estimation but only the probability of its detection. It gives confidence that if signal is distinguished from noise and measured velocity does not exceed #92569 - $15.00 USD
(C) 2008 OSA
Received 7 Feb 2008; revised 7 Apr 2008; accepted 8 Apr 2008; published 14 Apr 2008
28 April 2008 / Vol. 16, No. 9 / OPTICS EXPRESS 6023
the upper limit, measured velocity is correct. This feature is especially valuable in biomedical imaging, where diagnoses are based on measured functional parameters. 4.4 Retinal blood flow imaging with ultra-short CCD exposure time Joint STdOCT comparing to phase-resolved SOCT requires more data to be collected to estimate velocity value. This experiment is performed to give the proof of concept of time effective STdOCT. To compensate longer scanning protocols shorter CCD exposure time is proposed. We assume that higher sensitivity offered by this technique enables measuring the velocity without any time extension comparing to regular imaging. The capability of STdOCT to estimate the flow velocity in human retina with extremely short CCD exposure time of 5 and 1 μs is presented in Fig. 12.
Fig. 12. Velocity estimation of retinal blood flow with extremely short CCD exposure time: 5 μs and 1 μs for top and bottom line, respectively; ellipses mark blood vessels a., d. cross-sectional images of human retina in the area of optic disc; maps of the velocity distribution: b., e. Phase-resolved SOCT method, c., f. Joint STdOCT method.
The maps in the top line (5 μs) present spatial distribution of the blood flow velocities only in the larger vessels. The velocity underestimation in the phase-resolved method results in vanishing of the middle-size vessels in the flow image. STdOCT results obtained for 1 μs exposure time enable reconstructing one large and three smaller vessels. The map based on the phase-resolved method shows only faded velocity image of the large vessel. Unfortunately, the present state-of-the-art of CCD technique does not allow taking full advantage of extremely short exposure time because the relatively long dead time of the CCD camera limits the duty ratio to 0.025. In general in OCT studies, there is a pressing need to collect more data in examination time acceptable for patients. However, in many cases optical power delivered to the object has to be limited either by the safety regulations or by the power limitations of the light sources and/or optical components. Recent developments in CCD and CMOS technologies probably will be soon completed with ultra-fast line scan cameras. This experiment shows that, in contrast to the phase-resolved method STdOCT is able to benefit from these improvements. Phase-resolved SOCT requires higher SNR than it is possible to #92569 - $15.00 USD
(C) 2008 OSA
Received 7 Feb 2008; revised 7 Apr 2008; accepted 8 Apr 2008; published 14 Apr 2008
28 April 2008 / Vol. 16, No. 9 / OPTICS EXPRESS 6024
achieve with ultra-short exposure time in CCD technique, whereas STdOCT can still operate reliably in these conditions. 5. Conclusions We demonstrate the potential of Joint Spectral and Time domain OCT to estimate flow velocities accurately. In this approach, SOCT measurements are repeated in one position of scanning beam to register interferogram, that simultaneously depends on optical frequencies and time. Intensity modulation along axis of optical frequencies encodes information about structure and the modulation along time axis contains information on velocity. We analyze known phase-resolved SOCT method under low SNR conditions. It appears that retrieved velocities are dramatically underestimated and have tendency to decay to zero. This strong dependence on SNR is especially adverse in the case of measurements of highly scattering media, where the contribution of noise increases with depth. This may lead to considerable corruption of the velocity profile for points located deeper. Another cause of distortion may occur if the velocity is close to the maximal limit of velocity measurable by the phase-resolved method. Since velocity estimation in STdOCT is based on Doppler shifts, it is significantly less vulnerable to both effects and is more reliable for any qualitative and quantitative analysis as it is demonstrated using the same sets of OCT data. The possibility to unequivocal assessment of blood circulation in human retina renders STdOCT especially valuable. Additionally, STdOCT is more sensitive and it is able to detect a correct value of velocity for SNR lower at ~30 dB than the phase-resolved method. This unique feature can be used to compensate longer scanning procedure by shortening CCD exposure time. The proof of concept of time effective STdOCT is ascertained by presented measurements of blood flow in human retina in vivo for 5 μs and 1 μs exposure time. Acknowledgments This work was supported by EURYI grant/award funded by the European Heads of Research Councils (EuroHORCs) and the European Science Foundation (ESF). Maciej Wojtkowski acknowledges additional support of Foundation for Polish Science (Homing project and EURYI) and Rector of NCU for the scientific grant 504-F. Maciej Szkulmowski acknowledges support of Polish Ministry of Science, grants for years 2005/2008. Anna Szkulmowska acknowledges support of Polish Science Foundation FNP’2008 scholarship for young researchers.