JOURNAL OF CLINICAL MICROBIOLOGY, Oct. 2007, p. 3436–3438 0095-1137/07/$08.00⫹0 doi:10.1128/JCM.00973-07 Copyright © 2007, American Society for Microbiology. All Rights Reserved.

Vol. 45, No. 10

Human Immunodeficiency Virus Type 1 (HIV-1) Plasma Load Discrepancies between the Roche COBAS AMPLICOR HIV-1 MONITOR Version 1.5 and the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 Assays䌤 F. Damond,1* B. Roquebert,1 A. Be´nard,2 G. Collin,1 M. Miceli,1 P. Ye´ni,3 F. Brun-Vezinet,1 and D. Descamps1 Laboratoire de Virologie, Ho ˆpital Bichat Claude Bernard, Paris, France1; INSERM U593, ISPED Bordeaux F33076, France2; and Service Maladies Tropicales et Infectieuses, Hopital Bichat Claude Bernard, Paris, France3 Received 10 May 2007/Returned for modification 11 July 2007/Accepted 27 July 2007

We compared plasma viral load values obtained with COBAS AMPLICOR human immunodeficiency virus type 1 (HIV-1) MONITOR version 1.5 and with COBAS TaqMan HIV-1 assays. Mean values were 4.2 and 2.9 log10 copies/ml, respectively, showing the lack of agreement between the two assays. quantification and fewer time-consuming manipulations (2, 4, 6). Since June 2006, the 4,000 HIV-1-infected patients managed in the Department of Infectious Diseases of Bichat Claude Bernard Hospital (Paris, France), whose viral loads had previously been assayed with the COBAS AMPLICOR HIV-1 MONITOR version 1.5 method (Roche Diagnostics Systems, Inc., Branchburg,

Plasma RNA viral load quantification is a clinically validated tool for monitoring human immunodeficiency virus type 1 (HIV-1) infection (5). There are several commercially available PCR assays for HIV-1 RNA quantification. Most of them were recently automated for sample preparation and were coupled to a real-time PCR-based method, giving a broader range of HIV-1

FIG. 1. Degree of agreement in log10 copies/ml between the COBAS MONITOR 1.5 and the COBAS TaqMan assays for the 160 plasma specimens. The x axis of the Bland and Altman curves bears the mean values for each sample obtained by the two techniques. The y axis bears the differences between the values obtained by the two techniques. The solid lines represent the mean differences between the values, and the dotted lines represent the mean difference plus or minus 1.96 SD (95% limit of agreement). * Corresponding author. Mailing address: Laboratoire de Virologie, Ho ˆpital Bichat Claude Bernard, 75018 Paris, France. Phone: 33 1 40 25 61 51. Fax: 33 1 40 25 67 69. E-mail: [email protected]. 䌤 Published ahead of print on 22 August 2007. 3436

VOL. 45, 2007

NJ) and by automated extraction with the Magnapure largevolume kit (Roche Diagnostics, Meylan, France), have been monitored with the new RealTime COBAS TaqMan HIV-1 system assay, as recommended by the manufacturer. However, we were puzzled by significant discrepancies (more than 0.5 log10 copies/ ml) between the two assays of serial plasma samples from patients not taking antiretroviral therapy. We therefore compared HIV-1 RNA values obtained with the COBAS AMPLICOR HIV-1 MONITOR version 1.5 with those of the COBAS TaqMan HIV-1 assay in the routine clinical setting. Plasma EDTA samples stored at ⫺80°C were tested in parallel with the COBAS AMPLICOR MONITOR 1.5 assay (still available in our laboratory) and the COBAS TaqMan HIV system. Moreover, samples from all untreated patients with HIV-1 RNA values below 5,000 copies/ml with the COBAS TaqMan HIV system were also tested with the COBAS AMPLICOR HIV-1 MONITOR 1.5 assay. The HIV-1 subtypes were determined in all the patients by examining the relatedness of the pol sequences with reference sequences of HIV-1 genetic subtypes and circulating recombinant forms obtained from the Los Alamos database (http://hiv-web.lanl.gov). Mean values (log10 copies/ml) obtained with the two assays were compared with a t test for paired data. Discrepancies between the two assays were examined individually for each HIV-1 subtype by using the Kruskal-Wallis test. Statistical analyses were done with SAS version 8.2 software (SAS Institute Inc., Cary, NC), and P values of less than 0.05 were considered to indicate significant differences. Bland and Altman curves were used to represent the degree of agreement between the two techniques (Fig. 1). The HIV-1 plasma viral load was determined with both assays, using 160 plasma specimens obtained from 140 patients. The mean values (⫾ standard deviation [SD]) were 2.99 log copies/ml (⫾1.13) and 2.51 log copies/ml (⫾0.80), respectively, with the COBAS AMPLICOR HIV-1 MONITOR 1.5 assay and the COBAS TaqMan HIV system (Fig. 1). Among the 140 patients in whom viral load was determined with both assays, discrepancies of more than 0.5 log copies/ml were observed for 34 cases (24%). These patients were infected by subtypes A (n ⫽ 4), B (n ⫽ 9), D (n ⫽ 1), CRF_02 (n ⫽ 15), CRF_13cpx (n ⫽ 1), F (n ⫽ 1), and G (n ⫽ 3). The viral loads obtained with the 54 plasma samples from these 34 patients are summarized in Table 1. The mean values obtained with the COBAS TaqMan System and the COBAS AMPLICOR HIV-1 MONITOR 1.5 assay were 2.9 (⫾0.54) and 4.2 log10 (⫾0.82) copies/ml, respectively. The mean values obtained with the two assays differed by 1.3 log10 copies/ml (P ⬍ 0.0001). The viral load distributions obtained for B versus non-B subtypes did not differ significantly between the two assays (P ⫽ 0.32). Significant differences in viral load quantification, therefore, exist between these commercial assays from the same manufacturer. It has previously been shown that HIV genetic diversity can influence plasma HIV-1 RNA quantification in patients infected by non-B subtypes, and several authors have reported the failure of commercial assays for viral load monitoring in patients infected by non-B subtypes (1, 3, 4). This finding led the manufacturer to adapt the COBAS MONITOR version 1.0 assay to accept a broader range of viral diversity. Our results show that the COBAS TaqMan HIV-1 system

NOTES

3437

TABLE 1. Comparative plasma viral load values HIV-1 RNA viral load HIV-1 subtype

Patient

1a 1b 2 3a 3b 4 5a 5b 6a 6b 7a 7b 8 9a 9b 10a 10b 11a 11b 12a 12b 13a 13b 14 15 16 17 18a 18b 19 20 21 22a 22b 23a 23b 24a 24b 25 26a 26b 27 28a 28b 29a 29b 30a 30b 31a 31b 32 33 34a 34b

CRF02 CRF02 CRF02 A CRF02 B CRF02 CRF02 B CRF02 CRF02 CRF02 CRF13cpx G CRF02 CRF02 B A B B CRF02 CRF02 CRF02 G B F D A CRF02 A B B B G

MONITOR 1.5 (copies/ml)

MONITOR 1.5 (log10 copies/ml)

COBAS TaqMan (copies/ml)

COBAS TaqMan (log10 copies/ml)

13,700 66,400 254,000 38,300 142,000 122,000 5,960 4,250 22,500 150,000 2,840 54,200 416,000 139,000 51,200 58,300 29,000 1,340 1,000 17,300 15,800 1,964 3,080 3,090 46,200 1,210 5,800 6,540 3,900 13,500 2,750 3,720 2,880 2,930 7,220 5,910 9,600 13,600 6,810 22,200 10,400 61,500 3,440 2,680 4,890 5,320 5,700 6,920 430,000 53,900 381,000 11,000 35,900 14,400

4.14 4.82 5.40 4.58 5.15 5.09 3.78 3.63 4.35 5.18 3.45 4.73 5.62 5.14 4.71 4.77 4.46 3.13 3.00 4.24 4.20 3.29 3.49 3.49 4.66 3.08 3.76 3.82 3.59 4.13 3.44 3.57 3.46 3.47 3.86 3.77 3.98 4.13 3.83 4.35 4.02 4.79 3.54 3.43 3.69 3.73 3.73 3.84 5.63 4.73 5.58 4.04 4.54 4.16

207 787 4,980 1,393 1,396 1,824 692 555 1,291 2,192 ⬍40 159 34,300 2,894 1,889 3,271 1,914 140 272 1,887 835 100 671 ⬍40 1,840 ⬍40 1,535 418 989 2,650 301 476 705 388 1,780 1,590 512 2,188 802 3,175 3201 16,379 611 697 1,010 1,340 869 700 2,507 1,040 846 3,353 136 ⬍40

2.32 2.90 3.70 3.14 3.14 3.26 2.84 2.74 3.11 3.34 ⬍1.6 2.20 4.54 2.38 3.28 3.51 3.28 2.15 2.43 3.28 2.92 2.00 2.83 ⬍1.6 3.26 ⬍1.6 3.19 2.62 3.00 3.42 2.48 2.68 2.85 2.59 3.25 3.20 2.71 3.34 2.90 3.50 3.51 4.21 2.79 2.84 3.00 3.13 2.94 2.85 3.40 3.00 2.93 3.53 2.13 ⬍1.6

underquantifies not only divergent subtypes like CRF_02 but also HIV-1 subtype B. Recently, Gueudin et al. reported that viral load values obtained with the COBAS TaqMan system were 0.28 log10 lower on average than those obtained with the COBAS MONITOR version 1.5 assay: 23 (26%) of 88 patients had more than a 0.5 log10 difference between the values of the two methods, and nine had a difference exceeding 1 log10

3438

NOTES

copies/ml (4). It is noteworthy that the primers designed for the COBAS TaqMan system are located in the same highly conserved gag region of the HIV-1 genome as in the COBAS MONITOR 1.5 assay (6) (7). The most likely explanation for these differences in HIV-1 RNA quantification is a COBAS TaqMan system primer or probe mismatch. A problem during the automated nucleic acid extraction procedure on the COBAS AmpliPrep instrument is unlikely, as this would occur at random. Moreover, the discrepancies found here are not due to a problem with a particular lot, as several lots have been used in our laboratory since June 2006. However, we cannot rule out a problem with several lots, as these differences in HIV-1 quantification between the two assays were not observed during the premarketing evaluation of the COBAS TaqMan system (6). Indeed, during the initial evaluation, we tested 112 plasma samples from patients infected by HIV-1 subtype B (n ⫽ 54) and by various HIV-1 non-B subtypes (n ⫽ 58). We have also observed differences of more than 0.5 log10 copies/ml between the values of the two assays. These differences concerned seven plasma samples (7/112; 6%), as follows: COBAS MONITOR 1.5 ⬎ COBAS TaqMan (n ⫽ 5) and COBAS TaqMan ⬎ COBAS MONITOR 1.5 (n ⫽ 2). The frequency of discordant results from the initial evaluation was lower (6%) than that from the study presented here (25%). One possible explanation is that the kit used during the initial evaluation (C-Taq HIV-1 IVD⫹) was different from the one delivered by Roche after 2006 (CAP CTM HIV-1 IVD⫹).

J. CLIN. MICROBIOL.

These results question the reliability of the RealTime COBAS TaqMan system for viral load quantification in HIV1-infected patients. They also underline the importance of primer and probe design for real-time PCR quantification of viruses with high genetic diversity. Finally, they stress the need to harmonize viral load assays used in therapeutic trials. REFERENCES 1. Alaeus, A., K. Lidman, A. Sonnerborg, and J. Albert. 1997. Subtype-specific problems with quantification of plasma HIV-1 RNA. AIDS 11:859–865. 2. Berger, A., L. Scherzed, M. Sturmer, W. Preiser, H. W. Doerr, and H. F. Rabenau. 2002. Evaluation of the COBAS AmpliPrep/Cobas Amplicor HIV-1 Monitor Ultrasensitive Test: comparison with the Cobas Amplicor HIV-1 Monitor test (manual specimen preparation). J. Clin. Virol. 25(Suppl. 3): S103–S107. 3. Damond, F., C. Apetrei, D. Descamps, F. Brun-Vezinet, and F. Simon. 1999. HIV-1 subtypes and plasma RNA quantification. AIDS 13:286–288. 4. Gueudin, M., J. C. Plantier, V. Lemee, M. P. Schmitt, L. Chartier, T. Bourlet, A. Ruffault, F. Damond, M. Vray, and F. Simon. 2007. Evaluation of the Roche Cobas TaqMan and Abbott RealTime Extraction-Quantification systems for HIV-1 subtypes. J. Acquir. Immune Defic. Syndr. 44:500–505. 5. Mellors, J. W., A. Munoz, J. V. Giorgi, J. B. Margolick, C. J. Tassoni, P. Gupta, L. A. Kingsley, J. A. Todd, A. J. Saah, R. Detels, J. P. Phair, and C. R. Rinaldo, Jr. 1997. Plasma viral load and CD4⫹ lymphocytes as prognostic markers of HIV-1 infection. Ann. Intern. Med. 126:946–954. 6. Schumacher, W., E. Frick, M. Kauselmann, V. Maier-Hoyle, R. van der Vliet, and R. Babiel. 2007. Fully automated quantification of human immunodeficiency virus (HIV) type 1 RNA in human plasma by the COBAS AmpliPrep/COBAS TaqMan system. J. Clin. Virol. 38:304–312. 7. Sninsky, J. J., and S. Kwok. 1993. The application of quantitative polymerase chain reaction to therapeutic monitoring. AIDS 7(Suppl. 2):S29–S34.