Light-Addressable Potentiometric Sensor as a ... - Semantic Scholar

0 downloads 0 Views 2MB Size Report
Jul 1, 2016 - Abstract: A plug-based microfluidic system based on the principle of the light-addressable potentiometric sensor (LAPS) is proposed. The LAPS ...
micromachines Article

Light-Addressable Potentiometric Sensor as a Sensing Element in Plug-Based Microfluidic Devices Ko-Ichiro Miyamoto 1, *, Takuya Sato 1 , Minami Abe 1 , Torsten Wagner 2 , Michael J. Schöning 2,3 and Tatsuo Yoshinobu 1,4 1

2 3 4

*

Department of Electronic Engineering, Tohoku University, 6-6-05 Aza-Aoba, Aramaki, Aoba-ku, Sendai 980-8579, Japan; [email protected] (T.S.); [email protected] (M.A.); [email protected] (T.Y.) Institute of Nano- and Biotechnologies, Aachen University of Applied Sciences, Heinrich-Mußmann-Str. 1, Jülich 52428, Germany; [email protected] (T.W.); [email protected] (M.J.S.) Peter-Grünberg Institute (PGI-8), Research Centre Jülich, Jülich 52425, Germany Department of Biomedical Engineering, Tohoku University, 6-6-05 Aza-Aoba, Aramaki, Aoba-ku, Sendai 980-8579, Japan Correspondence: [email protected]; Tel.: +81-22-795-7075

Academic Editors: Manabu Tokeshi and Kiichi Sato Received: 18 May 2016; Accepted: 28 June 2016; Published: 1 July 2016

Abstract: A plug-based microfluidic system based on the principle of the light-addressable potentiometric sensor (LAPS) is proposed. The LAPS is a semiconductor-based chemical sensor, which has a free addressability of the measurement point on the sensing surface. By combining a microfluidic device and LAPS, ion sensing can be performed anywhere inside the microfluidic channel. In this study, the sample solution to be measured was introduced into the channel in a form of a plug with a volume in the range of microliters. Taking advantage of the light-addressability, the position of the plug could be monitored and pneumatically controlled. With the developed system, the pH value of a plug with a volume down to 400 nL could be measured. As an example of plug-based operation, two plugs were merged in the channel, and the pH change was detected by differential measurement. Keywords: chemical sensor; plug-based microfluidic device; light-addressable potentiometric sensor

1. Introduction Microfluidic devices are advantageous in various aspects for biological or clinical assays where it is necessary to handle small-volume samples. Common processes such as filtering, mixing, separating, heating, cooling, and sensing of the final products after a series of steps are realized by microfluidic systems constructed on a small substrate known as Micro-TAS or Lab-on-a-chip. They are expected to reduce the cost by minimizing the amount of required samples and reagents. In pursuit of the volume advantage of microfluidic devices, it is reasonable to divide the sample solution into small volumes and manipulate them in the form of plugs or droplets [1–4]. Consumption of reagents can be saved by reducing the volume, while a plug or a droplet can act as an independent reaction chamber by itself. To realize such a system, a sensing element is required that can probe a plug or a droplet inside the channel. In this study, we propose a plug-based microfluidic device based on a light-addressable potentiometric sensor (LAPS [5]), which is a chemical sensor based on the field effect in a semiconductor. In our previous papers, we applied a LAPS for measurement of continuously flowing samples in a microfluidic device [6,7], which featured (1) a complete flat sensor surface due to a simple structure consisting of silicon, insulator and electrolyte; (2) a flexible definition of measurement areas by

Micromachines 2016, 7, 111; doi:10.3390/mi7070111

www.mdpi.com/journal/micromachines

Micromachines 2016, 7, 111 Micromachines 2016, 7, 111

2 of 8

2 of 8

measurement areas by illumination; and (3) the label-free detection and visualization of chemical illumination; and (3) the label-free detection and visualization of chemical species. It was also possible species. It was also possible to visualize the spatial distribution of ions inside the flow [8]. In the case to visualize the spatial distribution of ions inside the flow [8]. In the case of continuously flowing of continuously flowing samples, however, the consumption of the solution was still large, which samples, however, the consumption of the solution was still large, which spoiled the main advantage spoiled the main advantage of microfluidic devices. To solve this problem, a plug-based version of a of microfluidic devices. To solve this problem, a plug-based version of a microfluidic device combined microfluidic device combined with LAPS is developed, which can (1) control the position of a plug with LAPS is developed, which can (1) control the position of a plug in the channel; (2) mix two plugs in the channel; (2) mix two plugs for reaction; and (3) detect the change by differential measurement. for reaction; and (3) detect the change by differential measurement. 2. Experiments 2. Experiments A microfluidic channel on the sensing surface of the LAPS is depicted in Figure 1a. It consists of A microfluidic channel on the sensing surface of thefilm LAPSand is depicted Figure 1a.an It consists a LAPS sensor plate, polydimethylsiloxane (PDMS) a glass in cover with Ag/AgClof a LAPS sensor plate, polydimethylsiloxane (PDMS) film and a glass cover with an Ag/AgCl electrode. electrode. ITO

Sensing area outlet at down-stream

Outlet Ag/AgCl ink

Glass substrate

Inlet

Sensing area

inlet

Thin-film PDMS

Sensing area at upper-stream

LAPS sensor (Si substrate)

(a) Air

Outlet Inlet

Mixing chamber

(b) Sample chamber

Sample chamber 1 (Urease) Sample chamber 2 (Urea) Outlet

Inlet

Sensing area Mixing chamber

Plug generation

(c)

Sensing area

(d)

Figure 1. (a) Test structure of the microfluidic device combined with LAPS; (b) Channel design with Figure 1. (a) Test structure of the microfluidic device combined with LAPS; (b) Channel design with a a chamber for merging and differential measurement; (c) Channel design to generate plugs on chip; chamber for merging and differential measurement; (c) Channel design to generate plugs on chip; (d) (d) Test structure with two sample chambers, one merging chamber, and one sensing area. Test structure with two sample chambers, one merging chamber, and one sensing area.

LAPS: The sensor substrate was n-type silicon with a size of 18 mm × 18 mm and a thickness of The sensor substrate was n-type size of layer 18 mm ˆ 18 mm a thickness 200LAPS: μm, insulated with a thermal oxide layersilicon and a with silicona nitride deposited byand LP-CVD [6–8]. of 200 µm,PDMS insulated with a thermal oxide layer and a silicon nitride layer deposited by LP-CVD [6–8]. film: A 500-μm-thick PDMS film to define the channel pattern was prepared by casting thePDMS PDMSfilm: (Silpot 184, Dow Corning Co., Ltd.,the Tokyo, Japan) in a polished aluminum mold,the A 500-µm-thick PDMSToray film to define channel pattern was prepared by casting which was pressed andCorning cured byToray heating. width andJapan) the height the U-shaped channel defined PDMS (Silpot 184, Dow Co.,The Ltd., Tokyo, in aof polished aluminum mold, which bypressed the PDMS were mm and The 500 μm, respectively. was andfilm cured by1 heating. width and the height of the U-shaped channel defined by the cover1 and To respectively. allow optical observation of a plug in the channel, a glass cover PDMS Glass film were mm assembly: and 500 µm, was usedcover as theand ceiling of the channel, whichobservation an Ag/AgClofelectrode was Glass assembly: To allowon optical a plug in theprepared channel,bya the glass following process. Firstly, the electrode pattern was defined by a masking tape on an indium tin cover was used as the ceiling of the channel, on which an Ag/AgCl electrode was prepared by (ITO)-coated plate, unnecessary removed etchingtape withonHCl after theoxide following process.glass Firstly, theand electrode patternITO waswas defined by a by masking an indium application of the mixture of zinc powder and glycerin (3 g/mL). Then, the patterned PDMS thin film tin oxide (ITO)-coated glass plate, and unnecessary ITO was removed by etching with HCl after was bonded onto the glass cover and Ag/AgCl ink for the reference electrode (No. 011464, ALS Co., application of the mixture of zinc powder and glycerin (3 g/mL). Then, the patterned PDMS thin film Ltd., Tokyo, Japan) was painted to cover the ITO pattern inside the channel. The width, the length was bonded onto the glass cover and Ag/AgCl ink for the reference electrode (No. 011464, ALS Co., and the thickness of the Ag/AgCl pattern were 1 mm, 2 mm, and about 70 μm, respectively. Finally, Ltd., Tokyo, Japan) was painted to cover the ITO pattern inside the channel. The width, the length and the thickness of the Ag/AgCl pattern were 1 mm, 2 mm, and about 70 µm, respectively. Finally,

Micromachines 2016, 7, 111

3 of 8

Micromachines 2016, 7, 111

3 of 8

the other side of the PDMS film was bonded onto the sensor surface after opening the inlet and outlet the other side of the PDMS film was bonded onto the sensor surface after opening the inlet and by an ultrasonic drill. Prior to each bonding process, the surfaces were treated by O2 plasma under a outlet by an ultrasonic drill. Prior to each bonding process, the surfaces were treated by O2 plasma pressure of 0.4 mbar with a power of 10 W for 1 min using a plasma cleaner (Zepto, Diener electronic under a pressure of 0.4 mbar with a power of 10 W for 1 min using a plasma cleaner (Zepto, Diener GmbH, Ebhausen, Germany). The bonding was finalized by heating at 95 ˝ C for 10 min. If an optical electronic GmbH, Ebhausen, Germany). The bonding was finalized by heating at 95 °C for 10 min. If observation the plug isof not whole area the ceiling mayceiling be covered Ag/AgCl an opticalofobservation thenecessary, plug is notthe necessary, the of whole area of the may bewith covered with so that the measurement be done atcan anybe point inside thepoint channel. In this case, the In measured points Ag/AgCl so that the can measurement done at any inside the channel. this case, the are defined bypoints light beams that induce photocurrent dependent on thedependent local valueon ofthe pH.local value measured are defined by lightabeams that induce a photocurrent ofFigure pH. 1b shows another channel pattern with a wider portion proposed by Itoh et al. [9], in which two successive plugs can be merged. In this structure, first proposed plug remains inside until Figure 1b shows another channel pattern with a wider the portion by Itoh et al. [9], the in second plug arrives, which is merged with the first plug and leaves the chamber together. Two which two successive plugs can be merged. In this structure, the first plug remains inside until the measurement areas are defined Ag/AgCl in theand upstream andchamber the downstream the second plug arrives, which isbymerged withelectrodes the first plug leaves the together. of Two measurement areasdifferential are definedmeasurement by Ag/AgCl electrodes in the upstream and the of the chamber, which allow of the pH change. In this study, thisdownstream configuration was chamber, which allow differential measurement of the pH In this study, an thisenzyme configuration applied to a measurement of enzymatic reactions, where the change. two plugs contained and its was applied to a measurement of enzymatic reactions, where the two plugs contained an enzyme substrate, respectively. and its substrate, Figure 1c shows respectively. a channel pattern with a sample chamber. The chamber acts as a dispenser, which Figure 1c shows a channel patternofwith a sample chamber. The chamber acts aschamber a dispenser, divides the sample solution into a series plugs. The sample solution in the sample was which divides the sample solution into a series of plugs. The sample solution in the sample pushed into the main channel by air from a syringe pump, and another continuous air flow inchamber the main was pushed into split the main channel sample by air from a syringe pump, another continuous flow in channel repeatedly the injected every time the mainand channel was occluded.air Figure 1d the main channel repeatedly split the injected sample every time the main channel was occluded. shows an integrated structure of two sample chambers, one merging chamber and one measurement Figure 1d shows an integrated structure of two sample chambers, one merging chamber and one point in the downstream, which was examined in this study. measurement point in the downstream, which was examined in this study. Measurement setup: The measurement setup used in this study is depicted in Figure 2. A droplet Measurement setup: The measurement setup used in this study is depicted in Figure 2. A of sample solution is supplied at the inlet using a micro-syringe (SGE analytical science, Victoria, droplet of sample solution is supplied at the inlet using a micro-syringe (SGE analytical science, Australia) and sucked into channel by aasperistaltic (AC-2120, ATTO Corp.,ATTO Tokyo, Victoria, Australia) andthe sucked into as thea plug channel a plug bypump a peristaltic pump (AC-2120, Japan) when the vent valve (PM-0815W, Takasago Electric Inc., Nagoya, in the Japan) downstream Corp., Tokyo, Japan) when the vent valve (PM-0815W, Takasago Electric Japan) Inc., Nagoya, in the is closed. The position of the plug is controlled by opening/closing the vent valve. The velocity downstream is closed. The position of the plug is controlled by opening/closing the ventline valve. The of the was of 0.78 by aspiration of the pump.ofThe areasensing of the area LAPS addressed lineplug velocity themm/s plug was 0.78 mm/s by aspiration the sensing pump. The of is the LAPS is by illumination a modulated from a light red-colored LED guided byguided an optical fiber (φ =fiber 1 mm). addressed byof illumination of alight modulated from a red-colored LED by an optical (ϕ The =photocurrent signal is amplified by a transimpedance amplifier (106 V/A) after 1 mm). The photocurrent signal is amplified by a transimpedance amplifier (106 and V/A)recorded and recorded 16-bit AD16-bit conversion at a sampling frequency of 100 kHz and a sampling number of 104 . of The after AD conversion at a sampling frequency of 100 kHz and a sampling number 104whole . The whole measurement is controlled by a homemade software developed with LabVIEW measurement process is process controlled by a homemade software developed with LabVIEW (National (National Austin, Instruments, Austin, TX, USA). Instruments, TX, USA).

Valve Open / close control PC / LabVIEW

Pump

electrode

plug

Bias voltage

A

Photocurrent

Fiber optics Modulated light

Figure 2. Schematic view Figure 2. Schematic viewofofthe themeasurement measurementsystem. system.

3. Results and Discussion 3.1. Control of the Plug Motion and LAPS Measurement After applying a droplet at the inlet and closing the vent valve, the plug moves through the 4 of 8 channel at a constant velocity. A bias voltage of −0.5 V is applied to the Ag/AgCl electrode with respect to the silicon substrate and the arrival of the plug at the electrode position is detected by monitoring the photocurrent signal as shown in Figure 3. The increase of the photocurrent was due 3. Results and Discussion to the arrival of the plug, which connected the Ag/AgCl electrode and the illuminated point of the sensor.ofWhen the Motion photocurrent exceeds a predefined threshold, the vent valve is automatically 3.1. Control the Plug and LAPS Measurement opened so that the plug stays at the electrode position during the LAPS measurement, in which the After applying droplet as at athe inlet and the vent valve,thethe through photocurrent (I) isa recorded function of theclosing bias voltage (V). When I-Vplug curvemoves is obtained, the the Micromachines 2016, 7,velocity. 111 4 of 8 vent is closed again andAthe plug is led toofthe outlet. channel atvalve a constant bias voltage ´0.5 V is applied to the Ag/AgCl electrode with 4a shows an example I-Varrival curves of obtained with in Figure 1a for of by respect toFigure the silicon substrate and of the the plug atthe thechannel electrode position is plugs detected andeach Discussion pH3.4Results tothe pHphotocurrent 10, with asignal lengthasofshown 2 mm and a volume of 1 increase μL. The bias voltage was sweptwas in the monitoring in Figure 3. The of the photocurrent due to range of −1.0 V to +1.0 V with a step width of 10 mV. In this series of experiments, different amounts the arrival the plug, which connected the Ag/AgCl electrode and the illuminated point of the sensor. 3.1.of Control of the Plug Motion and LAPS Measurement of NaCl were added to each pH buffer (Titrisol, Merck, KGaA, Darmstadt, Germany), so that the When the photocurrent exceeds a predefined threshold, thevent vent valve is automatically opened so that After applying a droplet inletwas and closing0.05 the moves through the total chloride concentration of at thethe plug always M, invalve, orderthe to plug avoid the influence of the the plug stays at the electrode position during the LAPS measurement, in which the photocurrent channel at a constant velocity. A bias voltage of −0.5 V is applied to the Ag/AgCl electrode with chloride sensitivity of the Ag/AgCl electrode. The inflection point of each I-V curve was calculated (I) is to the substrate and arrival the electrode position detected by recorded as a function of the bias voltage (V). When theplug I-V is obtained, theisvent valve closed andrespect plotted as asilicon function of pH asthe shown inofFigure 4b,atcurve inthe which a linear response with a ispH monitoring the photocurrent signal as shown in Figure 3. The increase of the photocurrent was due againsensitivity and the plug is ±led the outlet. of 45.9 1.0to mV/pH was observed. Micromachines 2016, 7, 111

to the arrival of the plug, which connected the Ag/AgCl electrode and the illuminated point of the sensor. When the photocurrent exceeds a predefined threshold, the vent valve is automatically opened so that the plug stays at the electrode position during the LAPS measurement, in which the photocurrent (I) is recorded as a function of the bias voltage (V). When the I-V curve is obtained, the vent valve is closed again and the plug is led to the outlet. Figure 4a shows an example of I-V curves obtained with the channel in Figure 1a for plugs of pH 4 to pH 10, each with a length of 2 mm and a volume of 1 μL. The bias voltage was swept in the range of −1.0 V to +1.0 V with a step width of 10 mV. In this series of experiments, different amounts of NaCl were added to each pH buffer (Titrisol, Merck, KGaA, Darmstadt, Germany), so that the total chloride concentration of the plug was always 0.05 M, in order to avoid the influence of the chloride sensitivity of the Ag/AgCl electrode. The inflection point of each I-V curve was calculated and plotted as a function of pH as shown in Figure 4b, in which a linear response with a pH sensitivity of 45.9 ± 1.0 mV/pH was observed.

Figure 3. Detectionofofthe theplug plug by signal. Figure 3. Detection by the thephotocurrent photocurrent signal.

Figure 4a shows an example of I-V curves obtained with the channel in Figure 1a for plugs of pH 4 to pH 10, each with a length of 2 mm and a volume of 1 µL. The bias voltage was swept in the range of ´1.0 V to +1.0 V with a step width of 10 mV. In this series of experiments, different amounts of NaCl were added to each pH buffer (Titrisol, Merck, KGaA, Darmstadt, Germany), so that the total chloride concentration of the plug was always 0.05 M, in order to avoid the influence of the chloride sensitivity of the Ag/AgCl electrode. The inflection point of each I-V curve was calculated and plotted as a function of pH as shown in Figure 4b, in which a linear response with a pH sensitivity of 45.9 ˘ 1.0 mV/pH was observed. Figure 3. Detection of the plug by the photocurrent signal.

(a)

(b)

Figure 4. (a) I-V curves measured for plugs with different pH values; (b) Inflection points of I-V curves plotted as a function of pH.

(a)

(b)

Figure 4. (a) I-V curves measured for plugs with different pH values; (b) Inflection points of I-V

Figure 4. (a) I-V curves measured for plugs with different pH values; (b) Inflection points of I-V curves curves plotted as a function of pH. plotted as a function of pH.

Micromachines 2016, 7, 111

5 of 8

Micromachines 7, 111 In order 2016, to further

5 of 8 reduce the volume of a plug, the height of the channel was reduced from 500 µm to 200 µm, while keeping the same width of 1 mm. A stable plug with a length of 2 mm and a In order to further reduce the volume of a plug, the height of the channel was reduced from volume of 400 nL could be reproducibly formed. The amplitude of the obtained photocurrent signal 500 μm to 200 μm, while keeping the same width of 1 mm. A stable plug with a length of 2 mm and a was almost the same as in the case of a 1 µL plug, as the contacting area of the plug with the sensing volume of 400 nL could be reproducibly formed. The amplitude of the obtained photocurrent signal surface and the intensity of illumination were the same. The shift of the I-V curve (data not shown) was almost the same as in the case of a 1 μL plug, as the contacting area of the plug with the sensing was againand linearly dependent on the pH value, with a pHThe sensitivity of 47.5 ˘ 1.9 (data mV/pH. surface the intensity of illumination were the same. shift of the I-V curve not shown)

was again linearly dependent on the pH value, with a pH sensitivity of 47.5 ± 1.9 mV/pH. 3.2. Merging Plugs and Differential Measurement 3.2.InMerging Plugs and Differential many assays, a test reagentMeasurement is added to the sample to cause some reactions to be detected. The addition of enzyme to its substrate, addition antibody to antigen, or vice to versa are commonly In many assays, a test reagent isthe added to theof sample to cause some reactions be detected. The used in such tests. To mimic such a situation, two plugs were merged inside the microfluidic channel addition of enzyme to its substrate, the addition of antibody to antigen, or vice versa are commonly using structure proposed in Reference and thewere pH change detected using the channel usedthe in such tests. To mimic such a situation,[9] two plugs merged was inside the microfluidic channel pattern 1b. proposed The measurement areas thethe upstream andwas the detected downstream thechannel chamber using in theFigure structure in Reference [9]in and pH change usingofthe were simultaneously by usingareas twoin optical fibers illuminating these areas with two light pattern in Figure 1b.monitored The measurement the upstream and the downstream of the chamber beams at different frequencies of 15optical kHz and 14.9 kHz, respectively. Bywith extracting each were modulated simultaneously monitored by using two fibers illuminating these areas two light beams modulated at different of 15 kHz and 14.9 kHz, respectively. By extracting frequency component from the frequencies obtained photocurrent signal, the pH values at both positionseach could component from the obtained signal, pH values at both positions befrequency independently determined [10]. Firstly,photocurrent a solution with the the total chloride concentration of could 0.154 M be prepared independently determined a solution withThen, the total chloride 0.154 M was by adding NaCl[10]. to 1Firstly, mM HCl solution. 1 µL of thisconcentration solution wasofintroduced was prepared NaCl to 1which mM HCl solution. 1 μL of this solutionThe wassecond introduced into into the channelby asadding the first plug, stayed insideThen, the merging chamber. plug, 1 µL the channel as the first plug, which stayed inside the merging chamber. The second plug, 1 μL of of 0.154 M NaCl solution, was introduced into the channel and delivered to the first measurement 0.154 M NaCl solution, was the introduced into channel andobtained. delivered to thehorizontal first measurement area in the upstream, where I-V curve inthe Figure 5a was The position area of this in the upstream, where the I-V curve in Figure 5a was obtained. The horizontal position thisposition, curve curve was shifted by ´76.8 mV with respect to that of a pH 7 buffer measured at the of same was shifted by −76.8 mV with respect to that of a pH 7 buffer measured at the same position, and and considering that the pH sensitivity at this position determined by a preliminary experiment was considering that the pH sensitivity at this position determined by a preliminary experiment was 51.5 ˘ 0.9 mV/pH, the pH value of the second plug was calculated to be 5.5. The second plug then 51.5 ± 0.9 mV/pH, the pH value of the second plug was calculated to be 5.5. The second plug then entered the merging chamber and the merged plug of 2 µL was delivered to the measurement area entered the merging chamber and the merged plug of 2 μL was delivered to the measurement area in the downstream. The I-V curve for the merged plug is shown in Figure 5b, which is shifted by in the downstream. The I-V curve for the merged plug is shown in Figure 5b, which is shifted by ´145.9 mV with respect to that of a pH 7 buffer measured at the same position. Using the pH sensitivity −145.9 mV with respect to that of a pH 7 buffer measured at the same position. Using the pH value of 43.7 ˘ 1.1 mV/pH thismV/pH position, value ofthe thepH merged wasmerged calculated be 3.7. sensitivity value of 43.7 ±at1.1 at the thispH position, valueplug of the plugtowas This value was slightly higher than 3.3, calculated for 1 mM HCl after dilution to double. The result calculated to be 3.7. This value was slightly higher than 3.3, calculated for 1 mM HCl after dilution to in this section confirms that proposed device used fordevice detecting a pH change induceda by double. The result in thisthe section confirms thatcan thebe proposed can be used for detecting pHthe addition a plug of to a plug sample, whichto is aexpected be applicable various types changeofinduced byreagent the addition of aofplug of reagent plug of to sample, which istoexpected to be of enzymatic and immuno-assays. applicable to various types of enzymatic and immuno-assays.

(a)

(b)

Figure 5. (a) I-V curves for the second plug measured in the upstream before merging; (b) I-V curves Figure 5. (a) I-V curves for the second plug measured in the upstream before merging; (b) I-V curves for the merged plug measured in the downstream. for the merged plug measured in the downstream.

Micromachines 2016, 7, 111 Micromachines 2016, 7, 111

6 of 8 6 of 8

3.3. Plug Generation in Channel the Channel 3.3. Plug Generation in the In channels the channels shown in Figure 1a,b, theplug plugwas wasmanually manually prepared prepared before In the shown in Figure 1a,b, the beforemeasurement; measurement; however, for practical it is desired can be generated onThus, a chip. Thus, plug however, for practical use, ituse, is desired that thethat plugthe canplug be generated on a chip. plug generation generation using the channel structure in Figure 1c was examined. Typical air flow in the main using the channel structure in Figure 1c was examined. Typical air flow in the main channel and the channel and of the sample 1.67 μL/s and 3.0 μL/s, channel injection speed ofthe theinjection sample speed were 1.67 µL/s andwere 3.0 µL/s, respectively. Therespectively. channel wasThe treated with was treated with a fluorine coating agent (FS-1060-2.0, Fluoro Technology Co., Ltd., Aichi, Japan) to a fluorine coating agent (FS-1060-2.0, Fluoro Technology Co., Ltd., Aichi, Japan) to make the surface make the surface hydrophobic. Figure 6 shows the effect of the treatment on the reproducibility of hydrophobic. Figure 6 shows the effect of the treatment on the reproducibility of plug generation. The plug generation. The variation of the plug volume was suppressed by the fluorine coating. variation of the plug volume was suppressed by the fluorine coating.

Figure 6. Effect of fluorine treatment ofof channel plug volume. volume. Figure 6. Effect of fluorine treatment channelon onthe thevariation variation of plug

3.4. Mixing of Two Plugs for Reaction 3.4. Mixing of Two Plugs for Reaction the channel shown in Figure 1d,sample the sample chambers a solution of(U0631, urea In theInchannel shown in Figure 1d, the chambers were were filled filled with awith solution of urea (U0631, Sigma-Aldrich, St. USA) Louis,and MO,that USA) and that of the urease Jack Sigma-Aldrich, bean (U1500, Sigma-Aldrich, St. Louis, MO, of urease from Jack from bean the (U1500, Sigma-Aldrich, 15,000–50,000 units/g), the latter of which catalyzes the hydrolysis of the former. 15,000–50,000 units/g), the latter of which catalyzes the hydrolysis of the former. Both of urea and Both of urea and urease were dissolved in 5 mM phosphate buffered saline (PBS), and the urease were dissolved in 5 mM phosphate buffered saline (PBS), and the concentration of urease concentration of urease was 2.8 g/L. The first plug from the chamber of the urea solution and the was 2.8 g/L. The first plug from the chamber of the urea solution and the second plug from that of second plug from that of the urease solution were generated, and they were mixed in the merging the urease solution werethe generated, and moved they were mixed in the chamber. After that,the the chamber. After that, merged plug to the sensing areamerging at the downstream, and then merged plug moved to the sensing area at the downstream, and then the change of the pH value due change of the pH value due to the enzymatic reaction in the mixed plug was observed as the change to theofenzymatic reaction The in thephotocurrent mixed plugwas was recorded observedatas athe change the photocurrent. The the photocurrent. fixed bias ofvoltage of −1.0 V. As photocurrent was recorded at a fixed bias voltage of ´1.0 V. As demonstrated in past studies [6–8], demonstrated in past studies [6–8], the response of the sensor was fast enough to monitor the the response of reaction. the sensor was fast enoughthe to time-course monitor the of enzymatic reaction. with Figure 7a shows the enzymatic Figure 7a shows the photocurrent different urea concentrations from 1 mM towith 8 mM. Although theconcentrations data contains noise to the of the time-course of the photocurrent different urea fromdue 1 mM to 8small mM.change Although the photocurrent, photocurrent with time after merging in all cases,increased which indicated data contains noisethe due to the smallincreased change of thethe photocurrent, the photocurrent with the of in theallpH value due to the production of ammonium In addition, theof time the afterincrease merging cases, which indicated the increase of the pH molecules value due[6]. to the production initial slopes of the photocurrent change in 0–25 s for 8 mM of urea and 0–50 s for other ammonium molecules [6]. In addition, the initial slopes of the photocurrent change in 0–25 s for 8 mM concentrations areother shown in Figure 7b. The slope varied depending the urea concentration. The of urea and 0–50 s for concentrations are shown in Figure 7b. Theon slope varied depending on the plug generation, mixing, and sensing of the pH change of the plug on the chip were successfully urea concentration. The plug generation, mixing, and sensing of the pH change of the plug on the chip demonstrated. were successfully demonstrated.

Micromachines 2016, 7, 111

7 of 8

Micromachines 2016, 7, 111

7 of 8

(a)

(b)

Figure 7. (a) Temporal change of photocurrent after merging a plug of urea solution and that of urease Figure 7. (a) Temporal change of photocurrent after merging a plug of urea solution and that of solution. The photocurrent response varied depending on concentrations of urea; (b) Initial slope of urease solution. The photocurrent response varied depending on concentrations of urea; (b) Initial photocurrent change as a function of the urea concentration. slope of photocurrent change as a function of the urea concentration.

4. Conclusions 4. Conclusions A plug-based microfluidic devicewas was developed developed on basis of a of LAPS. The motion of the plug A plug-based microfluidic device onthethe basis a LAPS. The motion of the was pneumatically controlled so that the plug stayed at the electrode position during the the LAPS plug was pneumatically controlled so that the plug stayed at the electrode position during LAPS measurement. A linear shift of the I-V curve depending on the pH value was observed for a plug measurement. A linear shift of the I-V curve depending on the pH value was observed for a plug with with a volume down to 400 nL. Two successive plugs were merged inside the channel and the pH a volume down to 400 nL. Two successive plugs were merged inside the channel and the pH values values before and after merging were successfully measured in the upstream and the downstream of beforethe andmerging after merging were successfully measured in the upstream and the downstream of the chamber. In addition, plug generation in the channel and detection of enzymatic merging chamber. addition, plug generation in thebychannel andofdetection of enzymatic reaction were In also examined. A pH change caused hydrolysis urea catalyzed by ureasereaction was were also examined. A pH change caused by hydrolysis of urea catalyzed by urease was within observed. observed. Based on the light-addressability of LAPS, many detection points can be defined Based the on the light-addressability LAPS,ismany detection points can be defined thewhere channel, channel, and, therefore, theofsystem expected to be applicable to various kindswithin of assays a small volume of samples are tested with reagents. to various kinds of assays where a small volume and, therefore, the system is expected to be applicable of samples are tested with reagents. Acknowledgments: This work was supported by JSPS Grant-in-Aid for Scientific Research (B) (contract No. 24310098). A part of work this research was carriedby out at the Machine Shop of Fundamental Acknowledgments: This was supported JSPS Grant-in-Aid forDivision Scientific Research (B)Technology (contract No. Center, Research Institute of Electrical Communication, University. 24310098). A part of this research was carried out at the Tohoku Machine Shop Division of Fundamental Technology

Center,Author Research Institute of Electrical Communication, Tohoku University. Contributions: T. Yoshinobu and M. J. Schöning conceived and designed the experiments; T. Sato and Abe performed T. theY.experiments; Miyamotoand anddesigned T. Wagnerthe analyzed the dataT. and theA.paper. AuthorM. Contributions: and M. J. S.K.conceived experiments; S. wrote and M. performed the experiments; K. M. and T. W. analyzed the data and wrote the paper. Conflicts of Interest: The authors declare no conflict of interest.

Conflicts of Interest: The authors declare no conflict of interest. References

References 1. Atencia, J.; Beebe, D.J. Controlled microfluidic interfaces. Nature 2005, 437, 648–655. 1. 2. 3. 4. 5. 6.

2. Garstecki, P.;D.J. Fuerstman, M.J.; Stone, H.A.;interfaces. Whitesides,Nature G.M. 2005, Formation of droplets and bubbles in a Atencia, J.; Beebe, Controlled microfluidic 437, 648–655. [CrossRef] [PubMed] microfluidic T-junction-scaling and mechanism of break-up. Lab Chip 2006, 6, 437–446. Garstecki, P.; Fuerstman, M.J.; Stone, H.A.; Whitesides, G.M. Formation of droplets and bubbles in a 3. Song, H.; Chen, D.L.; Ismagilov, R.F. Reactions in droplets in microfluidic channels. Angew. Chem. 2006, 45, microfluidic T-junction-scaling and mechanism of break-up. Lab Chip 2006, 6, 437–446. [CrossRef] [PubMed] 7336–7356. Song, H.; Chen, D.L.; Ismagilov, R.F. Reactions in droplets in microfluidic channels. Angew. Chem. 2006, 45, 4. Huebner, A.; Sharma, S.; Srisa-Art, M.; Hollfelder, F.; Edel, J.B.; deMello, A.J. Microdroplets: A sea of 7336–7356. [CrossRef] applications? Lab [PubMed] Chip 2008, 8, 1244–1254. Huebner, A.; Sharma, S.; Srisa-Art, M.; Hollfelder, F.; Edel, J.B.; deMello, A.J. sensor Microdroplets: A sea of 5. Hafeman, D.G.; Parce, J.W.; McConnell, H.M. Light-addressable potentiometric for biochemical applications? Chip1988, 2008, 8, 1182–1185. 1244–1254. [CrossRef] [PubMed] systems.Lab Science 240, Hafeman, D.G.; Parce, J.W.; McConnell, H.M. Light-addressable potentiometric sensor for biochemical systems. Science 1988, 240, 1182–1185. [CrossRef] [PubMed] Miyamoto, K.; Yoshida, M.; Sakai, T.; Matsuzaka, A.; Wagner, T.; Kanoh, S.; Yoshinobu, T.; Schöning, M.J. Differential setup of light-addressable potentiometric sensor with an enzyme reactor in a flow channel. Jpn. J. Appl. Phys. 2011, 50, 04dl08. [CrossRef]

Micromachines 2016, 7, 111

7. 8. 9. 10.

8 of 8

Miyamoto, K.; Hirayama, Y.; Wagner, T.; Schöning, M.J.; Yoshinobu, T. Visualization of enzymatic reaction in a microfluidic channel using chemical imaging sensor. Electrochim. Acta 2013, 113, 768–772. [CrossRef] Miyamoto, K.; Itabashi, A.; Wagner, T.; Schöning, M.J.; Yoshinobu, T. High-speed chemical imaging inside a microfluidic channel. Sens. Actuators B Chem. 2014, 194, 521–527. [CrossRef] Itoh, D.; Sassa, F.; Nishi, T.; Kani, Y.; Murata, M.; Suzuki, H. Droplet-based microfluidic sensing system for rapid fish freshness determination. Sens. Actuators B Chem. 2012, 171–172, 619–626. [CrossRef] Zhang, Q.; Wang, P.; Parak, W.J.; George, M.; Zhang, G. A novel design of multi-light LAPS based on digital compensation of frequency domain. Sens. Actuators B Chem. 2001, 73, 152–156. © 2016 by the authors; licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC-BY) license (http://creativecommons.org/licenses/by/4.0/).

Suggest Documents