Numerical Simulations of 3D Model of Knee-prosthesis ... - IFToMM 2015

3 downloads 4583 Views 921KB Size Report
effects of antero-posterior tibial slope on contact stresses ... contact area, the stress and displacements developed in the ... 1)danutcalafeteanu@yahoo.com.
The 14th IFToMM World Congress, Taipei, Taiwan, October 25-30, 2015 DOI Number: 10.6567/IFToMM.14TH.WC.OS1.008

Numerical Simulations of 3D Model of Knee-prosthesis Assembly with Antero-posterior Tibial Slope D. Calafeteanu1 Emergecy Hospital ` Craiova, Romania

D. Tarnita2 University of Craiova Craiova, Romania

Abstract: In this paper, starting from the virtual model of the human knee joint, and existent knee prosthesis, often used in total knee arthroplasty, we developed 3D models of all prosthetic components. The study investigates the effects of antero-posterior tibial slope on contact stresses in all components of total knee prosthesis using finite element analysis. Using AnsysWorkbench15.07 software, the stress and displacements maps are obtained for healthy knee, for osteoarthritic knee and for knee-prosthesis assemblies. Two loading force are considered: 800N and 2400N. For each prosthesis-knee assembly, two cases were considered: an antero-posterior slope of 00 and 50 respectively. The results, confirmed by clinical observations, suggest that the antero-posterior tibial slope of 5 0 is favourable. Keywords: 3D virtual knee prosthesis, total knee arthroplasty, finite element method, von Mises stress, antero-posterior tibial slope,

I. Introduction The knee joint is one of the most complex human joints taking into account the number of its components, their spatial geometry and their mechanical properties, the contacts between elements and the pressures acting on them. Virtual models were analyzed with finite element method in order to analyze much better the tibio–femoral contact area, the stress and displacements developed in the human knee joint under different solicitations. Virtual modeling of human knee joint has been addressed in several articles [1-19]. An important benefit of human joints numerical analysis lies in the flexibility of the modelling processes which can control of motion, loading and boundary conditions in the studies of joint response. The finite element method is becoming one of the most important tools in orthopaedic biomechanics. The finite element models of lower extremity are used to simulate a behavior of a healthy knee joint in-vivo [1-4]. In [5-10] studies regarding the development of osteoarthritis in knee joint which represent one of the most important disease which lead to total arthroplasty, are presented. In [11] the influence of the contact force model, contact geometry, and contact material properties on the dynamic response of a human knee joint model is studied. In the last decades major advancements in artificial knee replacement have improved the outcome of the surgery greatly. Finite element (FE) modeling allows implants to be tested and has been used in several biomechanical studies [12-31]. 1)

[email protected] [email protected] 3) [email protected] 2)

D.N. Tarnita3 University of Medicine &Pharmacy Craiova, Romania

FE method is used to evaluate biomechanical behavior of cancellous bone on patellofemoral arthroplasty [12], to evaluate and compare the proposed design of a custom femoral component with a conventional design [13-15], to evaluate the tibio-femoral contact stress [16, 17]. The influence of different designs of the joint area on tibial component fixation, kinematics and clinical outcome after a cemented total knee arthroplasty was studied in [18]. In [20-22, 26-27], the finite element method is used to simulate the loading conditions applied to the prosthesis device during the walking cycle in order to estimate the contact area, the contact pressure and the stress status of the polyethylene tibial components. Posterior tibial slope that is created during proximal tibial resection in total knee arthroplasty has emerged as an important factor in the mechanics of the knee joint and the surgical outcome [29-31]. In [31] the wear behavior of four different posterior tibial slopes to find out the appropriate posterior slope is presented. The objective of this study was to investigate the effects of posterior tibial slope on contact stresses in all components of total knee prosthesis using 3D finite element analysis. 2. Virtual modeling of the knee prosthesis components 2.1 Geometric modeling A knee prosthesis is made up of three parts: -The tibial component that replaces the top of the lower bone, the tibia, being attached directly to the bone; - Polyethylene tibial component that provides the bearing surface; -The femoral component that replaces the two femoral condyles and the groove where the patella runs on the distal femur. We used, as model, Scorpio Stryker type knee prosthesis, a common prosthesis used by orthopaedic surgeons in the Emergency Hospital of Craiova. The real models of the three prosthesis components are presented in Fig.1.

a)

c) b) Fig. 1. Physical prosthesis for human knee joint; a) femoral component; b) tibial component and polyethylene insert.

Fig. 3. Components of knee prosthesis: a) femoral component; b) tibial component; c) polyethylene insert.

To design the virtual model of the knee prosthesis, the DesignModeler application which is a pre-processor of AnsysWorkbench15.07 [32] is used. Using the effective measurement made on real components and after the identification of the simple shapes, the modeling operation was started. We have taken into account the 3D virtual model of healthy knee joint (Figure 2) elaborated by our team based on CT images [9,10]. a)

b)

Fig. 4. Virtual model of knee prosthesis: a) – Isometric view to the left. b) - Isometric view to the right.

After achieving the virtual model of the prosthesis, we aimed at an accurate positioning over bone ends, eliminating possible penetrations and gaps. Images for virtual models of knee joint with prosthesis are presented in Fig.5. To emphasize the polyethylene insert, we used a lighter colour than for the other 2 components of the prosthesis.

Fig.2. Virtual model of human knee joint [33, 34]

Solids were built over the geometrical model of healthy human knee joint, and they were cut with profiles made in various planes, in conformity with the surgical guides used in knee total replacement. Using „Slice, Body Operation, Extrude, Boolean Operation” commands and different features, the three prosthesis components were developed (Fig.3-4).We verified the positioning on bones to be without penetrations and gaps. Contact areas between the femoral component and the polyethylene insert were performed using operations such as “Substract, Check Body Entity”, which are particularly important for finite element analyses.

a)

b)

a)

b)

Fig. 5. Joint-prosthesis assembly at 176 : a) Anterior view; b) – Posterior isometric view o

2.2 Mesh generation For the realisation of numerical simulations of prosthesis joint, the following components that make up joint-prosthesis assembly have been considered: femur, tibia and the 3 components of the prosthesis: femoral, tibial and polyethylene insert. These components have been placed in a global XYZ system, which is required for proper placement of loads and degrees of freedom for the entire assembly. The structural analysis is statically nonlinear, nonlinearity being present by nonlinear contacts between the surfaces of polyethylene insert and femoral prosthesis component. Correct positioning of components and the analysis with finite element method has been achieved in the simulation environment of AnsysWorkbench15.07 application. This application is

intended for CAE field which allows advanced modeling and discretization using finite element method. The mesh geometry was realised using hexahedral elements of Solid186 type and tetrahedral elements of Solid 187 type, both of them being solid elements with middle node. Middle nodes are necessary for better approximation of results and their accuracy. For prosthesis components, “Sweep”, “Patch Conforming” and „Hex Dominant” methods were used for advanced discretization of complex areas in hexahedral elements, for control element formation. For an efficient discretization, elements with dimensions of 1 mm and 1.5 mm respectively, for the areas of maximum interest were used (femoral prosthesis-femur, femoral prosthesis - polyethylene insert, polyethylene insert - tibial prosthesis and tibial prosthesis - tibia contact areas). The transition from this dimension to a larger dimension of 4 mm for less complicated areas is possible by means of the options “Smoothing” on medium and “Transition” on fast. In Fig. 6 the mesh structure of the knee joint-prosthesis ensembly is presented. For each of studied cases: healthy knee (HK), osteoarthritic (OA) knee and knee-prosthesis assembly the virtual model is discretized individually, the contact areas are readjusted and the analysis is run. In Table 1 the numbers of nodes and elements for knee joint bones and for prosthesis component corresponding to the joint-prosthesis model at 176o are presented.

enter into contact. We must check these contacts to be closed using “Sticking” and, respectively, “Sliding” options (Fig. 7).

a)

b) Fig. 7. Visualisation of different control methods for the contacts between knee components: a) Visualisation of free spaces between components; b) Visualisation of penetrations; these may occur between the components participating in the contact.

The material properties are assigned to the different components of the assembly based on previously published data and they are found in Table 2. a)

b)

Fig. 6. a) Human knee joint with prosthesis –Local view of node and element network; b) Isometric view of prosthesis mesh network; Table 1. The network of nodes and elements

Component Femur Tibia Femoral component Tibia component Polyethylene component Total

Nodes 144024 85185 36180 30479 55456 351324

Elements 46321 26050 10605 8295 15857 107128

2.3. Boundary conditions “Bonded” and “No Separation” type contacts were used for the prosthesis components contacts. The calculation algorithm used for contacts is of “Augmented Lagrange” type. This algorithm adds an extra control to automatically reduce penetrations, being recommended because of its flexibility features. To impose a better interpretation of the areas that define the contact, "Bonded", "No Separation" and “On Gauss Points” commands were used. A contact radius of 0.1 mm was also assigned by means of the option “Pinball Region”, which supposes that the solver shall assess these areas more efficiently before the components

Table 2.Materials properties for knee-prosthesis components

Component

Femur Tibia Femoral component Tibial component Polyethylene component

Young’s Modulus MPa 17600 12500 210000 210000 1100

Poisson’s Ratio 0.3 0.3 0.3 0.475 0.42

The settings for the analysis and the boundary conditions are the following: -taking into account the large number of nodes and elements but, also, the presence of nonlinear contacts, for solving the analysis it is necessary to implement a “smaller steps” system; solver was set as Preconditioned Conjugate Gradient level 2 iterative types. -on the proximal head of the femur bone there is applied a 800N force and 2400 N force in the -Z axis direction; -on the same location it is applied the "Remote Displacement" that allows offset Z and RotY around the femur which allows movement of the hip; -on the tibia distal head is applied "Remote Displacement", which allows RotY (movement of the ankle around the tibia).

3.Results Using AnsysWorkbench 15.07 software, the numerical simulations and FEM analyses were processed for healthy knee joint, for OA knee joint with a deviation angle of 6 0 and for knee joint-prosthesis assemblies with antero-posterior slope of 0o and 5o respectively. The two variants of 0o and 5o tibial slope were chosen because Stryker Company offers both variants of complete surgical instruments for total knee arthroplasty for these two cases. Clinical observations show that for the case of 5o tibial slope is easier to mount the knee prosthesis components during total knee arthroplasty and, after surgical act, the flexion movement is improved (the range of motion increases). Two loading forces were taken into consideration: 800 N and 2400 N Applying a loading force equal to 800N for entire biomechanical system of healthy knee (HK) a maximum total displacement equal to 7.24 mm and a maximum von Mises stress equal to 6.98MPa, developed near the distal femoral head, are obtained (Fig.8). Von Mises stress was colour-coded where red represented the highest stress, while the blue represented the lowest.

c) Fig.9 Von Mises stress (Top view–left column; Bottom view–right column) in femoral cartilage (a), tibial cartilage (b) and menisci (c) for healthy knee Table 3. Von Mises stress for healthy knee (HK) and for osteoarthritic knee (OAK) with a varus inclination of 5o

HK

OAK

Femoral cartilage stress [MPa]

2.41

3.42

Tibial cartilage stress

[MPa]

2.17

3.01

Menisci stress

[MPa]

2.12

2.54

Applying a loading force equal to 800N, for entire biomechanical system of prosthetic knee (PK) with antero-posterior slope of 50, a maximum total displacement equal to 2.37 mm and a maximum von Mises stress equal to 18.14MPa are obtained (Fig.10). Table 4. Displacements and stress for entire biomechanical system

a

b

HK

PK 0O

PK 5O

Total displac. [mm]

7.24

2.37

2.624

Lateral displac. [mm]

7.17

2.35

2.47

von Mises stress [MPa]

6.98

18.14

14.08

c

) Fig. 8 – Total displacements (a), lateral displacements on X axis (b) and von Mises stress (c) for healthy knee joint assembly.

Von Mises stresses maps obtained for healthy knee corresponding to a loading force equal to 800 N in femoral cartilage, tibial cartilage and menisci are presented in Fig. 9. The images represent Top and Bottom views.

Fig. 10. Total displacements (a), and von Mises stress (b) for prosthetic knee at 176o with an antero-posterior slope of 5o.

a)

b)

Von Mises stresses maps (Top and Bottom views) obtained for a loading force equal to 800 N in polyethylene insert, tibial prosthesis and femoral prosthesis for an angle of 176o with 0o antero- posterior slope are presented in Fig. 11. For the case of 5o antero-posterior tibial slope, the maps are presented in Fig.12.

b)

c) Fig. 12. Von Mises stress (Top view–left; Bottom view–right) in polyethylene insert (a), tibial prosthesis (b) and femoral prosthesis (c) for knee-prosthesis assembly for an angle of 176o -5o antero-posterior slope.

a)

b)

c) Fig. 11 Von Mises stress maps (Top view – left; Bottom view– right) in polyethylene insert (a) tibial prosthesis (b) and femoral prosthesis (c) for knee-prosthesis assembly for an angle of 176o - 0o antero-posterior slope.

The maxmum stress values obtained for each analysed component of knee joint-prosthesis assembly for a load equal to 800N and 2400 N, in both cases: 0o and 5o antero-posterior slope, respectively, are presented in Table 5. The displacements values are presented in Table 6. Table 5. Von Mises stress maximum values Antero- Case Stress Stress Stress post. slope Insert Tibial Femoral [MPa] comp. comp [MPa] [MPa] 00 800N 18.14 16.82 17.28 2400N 53.78 50.12 51.84 50 800N 14.08 12.85 13.25 2400N 40.15 37.73 38.41 Table 6. Displacements maximum values Antero-post. Case Total slope Displ.[mm] 00 800N 2.37 2400N 7.13 50 800N 2.63 2400N 7.87

Stress Stress Femur Tibia [MPa] [MPa] 10.45 31.29 6.01 19.73

14.06 42.14 10.15 29.64

Lateral Displ.[mm] 2.35 7.51 2.47 7.46

Analyzing the results obtained by numerical simulations, presented in Table 5, we can see an increase of stress for OA knee joint versus the case of healthy knee joint for all components: femoral cartilage, tibial cartilage and menisci. The OA knee is exposed to greater stress at the medial compartment of the articular cartilage of the tibia and femur. The maximum values of von Mises stress increase with about 50% in OA case. The results show that misalignment (varus variation) could damage the articular cartilage because they increase the stress magnitude that progressively produces articular cartilage damage. In knee-prosthesis assemblies, we can observe that for a slope of 50 the stress decrease with about 20%. The decrease of stress values proves an improvement in knee stability and rehabilitation. The distribution of von Mises stress was more uniform with 50 than 00 posterior tibial slope and the peaks of von Mises stress were smaller with 50 slope, one of the explanation being that in this case the contact areas increase.

a)

4. Conclusion In this paper, starting from the virtual model of the human knee joint and existent knee prosthesis, often used in total knee arthroplasty, we developed 3D models of prosthetic components: femoral component, tibial component and polyethylene insert. Antero-posterior tibial slope that is created during proximal tibial resection in total knee arthroplasty represents an important parameter in the knee joint post surgical biomechanics for recovery of the knee joint function and preventions of complications. Using AnsysWorkbench15.07 software, the stress and displacements maps are obtained for healthy knee, for OA knee joint and for analyzed knee-prosthesis assemblies. The Finite Element Method is used to investigate the effects of antero-posterior tibial slope on contact stresses in the tibial polyethylene component of total knee prostheses and to obtain the diagrams and the maximum values of von Mises stress and displacements. The findings of this study confirms the clinical observations of surgeons, that the posterior tibial slope of 50 is favourable. The main advantage of the numerical simulations and FEM analysis consists in the fact they can be done in order to evaluate the biomechanical behaviour of human joints without an invasive intervention. A finite element analysis of the normal and prosthetic knee model will help surgeons and biomechanical researchers to develop improved devices for rehabilitation movements of patients suffering diseases.

[12]

[13]

[14]

[15]

[16]

[17]

[18]

[19]

[20]

[21]

Acknowledgements: The authors would like to thank Dr. Eng. Marius Catana for his technical support in the numerical simulations.

[22]

References [1]

Penrose J.M.T. Development of an accurate three dimensional finite element knee model. Comp. Meth. In Biomech. and Biomed. Eng. 5:291-300, 2002. [2] Donahue T.L H. et al. A finite element model of the Human knee joint for the study of tibio-femoral contact. J. Biomech. Eng. 124:279-280, 2002. [3] Beillas P. et al. A new method to investigate in vivo knee behavior using a finite element model of the lower limb. J Biomech. 37:019-1030, 2004. [4] Guess, T.M., 2012. Forward dynamics simulation using a natural knee with menisci in the multibody framework. Multibody System Dynamics, 28(1-2), 37-53. [5] Bae J.Y., Park K.S., Seon J.K., s.a., Biomechanical analysis of the effects of medial meniscectomy on degenerative osteoarthritis, Med Biol Eng Comput., 50:53–60, 2012. [6] Wilson W., C.C. et al., The Role of Computational Models in the Search for the Mechanical Behavior and Damage Mechanisms of Articular Cartilage, Med. Eng. Phys., 27(10), pp.810-826, 2005. [7] Yang N.H., The effect of the frontal plane tibiofemoral angle on the contact stress and strain at the knee joint, Mechanical Engineering Dissertations Department of Mech. and Ind. Engineering, Northeastern University, 2009. [8] Vidal A., Lesso R., Rodrighez R., Garcia S., Daza L., Analysis, simulation and prediction of contact stresses in articular cartilage of knee joint, Dept. of Mechanical Engineering, Inst.Tecnologico de Celaya, Mexico, International Conference Ansys, 2008. [9] Tarnita D., Catana, M., Tarnita, D.N., Contributions on the modeling and simulation of the human knee joint with applications to the robotic structures, In “New Trends on Medical and Service Robotics: Challenges and Solutions”, Mechanisms and Machine Science 20, pp. 283-297, Springer, 2014. [10] Tarnita, D., Catana, M., Tarnita, D.N., Modeling and Finite Element Analysis of the Human Knee Joint Affected by Osteoarthritis, Key Engineering Materials Vol. 601, pp 147-150, 2014. [11] Machado, M., Flores, P., Ambrósio, J., Completo, A., 2011. Influence of the contact model on the dynamic response of the

[23]

[24]

[25]

[26]

[27]

[28]

[29]

[30]

[31]

[32] [33]

human knee joint. Proceedings of the Institution of Mechanical Engineers, Part-K Journal of Multi-body Dynamics, 225(4), 344-358. Castro, A.P.G., Completo, A., Simões, J.A., Flores, P., Biomechanical behaviour of cancellous bone on patellofemoral arthroplasty with Journey prosthesis: a finite element study. Computer Methods in Biomechanics and Biomedical Engineering, 18(10),1090-1098, 2015. Zach L., Ruzicka P., Konvickova S. and L. Cheze, Stress analysis of knee joint endoprosthesis. Journal of Biomechanics 40(S2) XXI ISB Congress, Poster Sessions, July, 2007. Skallerud B., Wik T.S., Pettersen S.H., Subject specific finite element analysis of stress shielding around a cementless femoral stem. Clinical Biomechanics, 24:196–202, 2009. Harrysson O., Hosni Y., Nayfeh J., Custom-designed orthopedic implants evaluated using finite element analysis of patient-specific computed tomography data: femoral-component case study. BMC Musculoskeletal Disorders,8:91 doi:10.1186/1471-2474-8-91, 2007 Haut D.T., Hull M., Rashid M., Jacobs C., A finite element model of the human knee joint for the study of tibio-femoral contact. J. Biomech. Eng., 124:273-280, 2002. Soudry M, Walker PS, Reilly DT, Kurosawa H, Sledge CB Effects of total knee replacement design on femoral-tibial contact conditions, J Arthroplasty, 1(1):35-45, 1986. Villa T., Migliavacca F., Gastaldi D., Colombo M., Pietrabissa R., Contact stresses and fatigue life in a knee prosthesis: comparison between in vitro measurements and computational simulations. J. Biomech., 37:45-53, 2004. Soncini M., Redaelli A., Montevecchi F.M., FE dynamic analysis of a knee joint prosthesis during the walking cycle. Summer Bioengineering Conference, June 25-29, Key Biscayne, Florida, 2003. Ishikawa H., Fujiki H., Yasuda K., Contact analysis of ultrahigh molecular weight polyethylene articular plate in artificial knee joint during gait movement. Journal of Biomechanical Engineering 118:377-386, 1996. Sathasivam S., Walker P.S., Computer model to predict subsurface damage in tibial inserts of total knees. Journal of Orthopaedic Research 16, 564-571, 1998. Estupinan J.A., Bartel D.L., Wright T.M., Residual stresses in ultra-high molecular weight polyethylene loaded cyclically by a rigid moving indenter in nonconforming geometries. Journal of Orthopaedic Research 16:80-88, 1998. Reeves E.A., Barton D.C., FitzPatrick D.P., Fisher J., A two dimensional model of cyclic strain accumulation in ultra-high molecular weight polyethylene knee replacements. Proceedings of the Institute of Mechanical Engineers, 212(H):189-198, 1998. Taylor M., Barrett D.S., Explicit finite element simulation of eccentric loading in total knee replacement. Clin Orthop Relat Res 414:162–171, 2003. Godest A.C., de Cloke C.S., Taylor M., Gregson P.J., Keane A.J., Sathasivan S., Walker P.S., A computational model for the prediction of total knee replacement kinematics in the sagittal plane. J Biomech 33(4):435–442, 2000. Liau J.J., Cheng C.K., Huang C.H., Lo W.H., The effect of malalignment on stresses in polyethylene component of total knee prostheses—a finite element analysis. Clin Biomech (Bristol, Avon) 17(2):140–146, 2002. Yang R.S., Lin H.J., Contact stress on polyethylene components of a new rotating hinge with a spherical contact surface. Clin Biomech (Bristol, Avon) 16(6):540–546, 2001. Godest A.C., Beaugonin M., Haug E., Taylor M., Gregson P.J., Simulation of a knee joint replacement during a gait cycle using explicit finite element analysis. Journal of Biomechanics 35:267-275, 2002. Matsuda S., Miura H., Nagamine R., Urabe K., Ikenoue T., Okazaki K., Iwamoto Y., Posterior tibial slope in the normal and varus knee. Am J Knee Surg 12(3):165–168, 1999. Lee H.Y., Kim S.J., Kang K.T., Kim S.H., Park K.K., The effect of tibial posterior slope on contact force and ligaments stresses in posterior-stabilized total knee arthroplasty-explicit finite element analysis. Knee Surg Relat Res 24(2):91–98, 2012. Yi Shen., Xiaomiao L., Xiaodong F., Weili W., A 3D finite element model to investigate prosthetic interface stresses of different posterior tibial slope. Knee Surg Sports Traumatol Arthrosc DOI 10.1007/s00167-014-3144-9. www.ansys.com Morrison J.B, The mechanics of the knee joint in relation to normal walking, Journal of Biomechanical Engineering, 3, pp. 51-61, 1970.