Product Overview. 86 ... Off-the-shelf products or custom-designed solutions – ...
TEGO. R. IES. 8. WAVEFORM DIGITIZER. MODELS. □ PXD 721 x. 10↗. 16 Bit ...
PRODUCT CATALOG 2013/14 | English
Functional test equipment for laboratory and high speed production testing
nstruments
the test solutions provider
TABLE OF CONTENTS
2
3
Product Categories
7
Waveform Digitizer
9
PXD 721x High Resolution Waveform Digitizer Family
10
PXD 731x High Voltage Waveform Digitizer
12 NEW 2014 PXD 821x High Performance Waveform Digitizer 14 VX 734x High Resolution Waveform Digitizer Family
16
VX 2022 High Performance Digitizer
18
Timer/Counter option for Digitizers
20
Waveform Generator
23
PXA 722x Arbitrary Waveform Generator Family
24
PXA 732x Arbitrary Current Generator Family
26
VX 2026 Arbitrary Waveform Generator
28
VX 2701 Quad Sine Cosine Generator
30
High-Power SMU A 5710 AC Current Source
33
AXC 757x 250A High Current SMU Family
34 NEW 2014 36 NEW 2014 38
AXC 760x High Current SMU Family
NEW
AXC 755x 500A High Current SMU Family
AXC 76xx High Current SMU
42
AXV 7607 High Voltage SMU
44
Source & Measurement
Off-the-shelf products or custom-designed solutions – in the end it’s your satisfaction that matters
47
AXM 7720 Multichannel-Source Measurement Unit
48
AXS 772x System Power Supply
50
PX 773x PXI Source Measurement Unit
52
VX 2702 DC Source and Measurement Unit
56
VX 6620 PXI System Power Supply
58
VX 6625 Quad System Power Supply
60
Source only AXB 5721 Dual Precision Power Amplifier
63 NEW
66
VX 4616 Dual Precision Power Amplifier
68
VX 4620 Dual High Speed Power Supply
70
AXR 7510 Relay Switching Unit
73 NEW
74
PXT 1741 PXI Trigger Module
76
DTS 8710 Unclamped Inductive Switching Test System
78
VX 2210 High Precision Timer Counter
80
VX 3701 Synchro/Resolver Module
82
Test Systems & Custom Solutions VX 8800 Functional Test-System
http://www.vxinstruments.com/ catalog/Catalog2013.pdf
64
AXS 6710 AC/DC Source 550 V
Measurement & Others
Get the digital version here:
40
84 84
Product Overview
86
Glossary
88
Floating Measurement
88
Technical Terms
92
Contact
93
und Fertigung. In über 20 Ländern, von Europa über Amerika bis nach Asien (einschl. China) wissen Kunden unsere Systeme und deren Präzision zu schätzen. Wir werden auch in Zukunft daran arbeiten, diese Position weiter auszubauen.
4
5
VX Instrum
With great succe measurement in Our customers in appreciate our sy We are working
VX Instruments – the standard for quality With great success VX Instruments develops and produces highly innovative measurement instruments and test systems for automation and production.
Managing Director Johann Degenhart
Our customers in over 20 countries in Europe, America and Far East highly appreciate our systems with their outstanding precision and performance. The reliability and high quality of our products made in Germany is confirmed by certification according to ISO 9001:2008. We are working on strengthening our market position in the future.
6
7
PRODUCT CATEGORIES Waveform Digitizer
9
Waveform Generator
23
High-Power SMU
33
Source & Measurement
47
Source only
63
Measurement & Others
73
Test-Systems & Custom Solutions
84
8
WAVEFORMDIGITIZER DIGITIZER WAVEFORM
8
9
ProductCATEGORIES Categories PRODUCT
MODELS MODELS ■ PXD 721 x ■■ PXD 721 x 16 Bit; 100 MS/s; 100 MHz bandwidth; 15 MHz >15 MHz
Vertical resolution
12 Bit 12 Bit
Sampling times
1 ns, 2 ns, 4 ns, 8 ns, 16 ns, 32 ns, … 1 ns, 2 ns, 4 ns, 8 ns, 16 ns, 32 ns, …
Software selectable
Input impedance
1 MΩ || < 20 pF nom., 50 Ω 1 MΩ 20 pF 50 Ω
Software selectable
Input coupling
DC
Input ranges
50 Ω: 50 Ω: 1 MΩ: 1 MΩ:
DVM
100 mV, 300 mV, 1 V, 3 V, 10 V 100 mV, 300 mV, 1 V, 3 V, 10 V 100 mV, 300 mV, 1 V, 3 V, 10 V, 100 mV, 300 mV, 1 V, 3 V, 10 V, 30 V, 100 V 30 V, 100 V
0.25% of input + 0.25% of full scale
Filter
300 kHz, 1 MHz, 1 MHz, 20 MHz, 100 MHz 300 kHz, 20 MHz, 100 MHz
Input Internal function module Software PXI trigger Output Internal function module PXI trigger
0.25% of input + 0.25% of full scale
Measurement time
1 … 500 ms
One function module can trigger itself and the other one Via software command 0 … 7 Trigger 0 … 7 and star-trigger
From the PXI backplane
12 Bit 12 Bit
Level accuracy
0.6% + 0.3%
Software programmable
PXI Capabilities
Specification
Comment
PXI TTL/trigger usage
possible
PXI trigger 0 … 7; input and output
Software selectable
For example Marker Bit
±(of programmed value + of full range)
kS/s and active 300 kHz filter DC accuracy specified for an average value of 1000 samples with a sample rate of ≈50 ≈50 kS/s 300 kHz performed within 24 hours after an offset correction.
1
DC accuracy
Comment
DC accuracy specified with measurement time of 100 ms.
Comment
Level resolution
The dual-channel device PXD8214 High Voltage Waveform Digitizer needs two PXI slots (3 U).
2
Specification
Output to the other module Output each channel trigger 0 … 7 to PXI trigger 0 …7
Available with one and two channels The single channel-device PXD8213 High Voltage Waveform Digitizer uses only one PXI slot (3 U).
Specification 2
DC accuracy1
Trigger System
Comment
High throughput design for many applications The digitizers of the PXD821x-Family are designed for high throughput testing. This design guarantees highest quality measurements and is ideal for a wide range of application areas including automotive, communications, scientific applications, military/aerospace and consumer electronics.
www.VXInstruments.com
2
Download the complete Datasheet here: http://www.vxinstruments.com/ catalog/PXD821x.pdf
Ordering Option
15 Comment
DVM
DVM function
T/C
Timer/Counte
TCXO
TCXO Oscillat
DBW
Double Bandw
MEMSEG
Memory segm
feature results in additional test time improvement.
■■ ■■ ■■ ■■
VX 7340 VX 7342 VX 7344
Each channel is fully isolated. This results in a very high common mode rejection ratio (CMRR) compared to differential inputs. It allows floating low level signals to be measured with a very high accuracy and a maximum of resolution. This design of the VX7344 guarantees highest quality measurements.
Acquisition
Specification
Comment
Maximum sample rate
100 MS/s
Bandwidth
50 MHz, 100 MHz (with option DBW)
Vertical resolution
14 Bit, 16 Bit (with option RES16B)
Sampling times
10 ns, 25 ns, 50 ns, 100 ns, 250 ns, 500 ns, 1 µs, 2.5 µs, 5 µs, 10 µs, 25 µs, 50 µs, 100 µs, 250 µs, 500 µs, 1 ms, 2.5 ms, 5 ms, 10 ms, 25 ms, 50 ms, 100 ms, 250 ms, 500 ms, 1 s, 2.5 s, 5 s, 10 s
sSoftware selectable
Input impedance
1 MΩ || < 20 pF nom.; 50 Ω
Software selectable
Input coupling
DC
Maximum input voltage1
1 MΩ: 50 Ω:
60 VDC 250 VDC with option IN250 V 4 Vp
Input ranges
50 Ω: 1 MΩ:
250 mV, 500 mV, 1 V, 2 V, 4 V, 250 mV, 500 mV, 1 V, 2 V, 4 V, 8 V, 16 V, 32 V, 64 V, 128 V, 250 V
DC accuracy2
250 mV, 500 mV: 0.3% of input +2 mV others: 0.15% of input + 0.1% of full scale
Filter
30 kHz, 100 kHz, 300 kHz, 1 MHz
Waveform memory
2 MByte (1 MS)
2
Specification
Trigger from Internal External Software VXI-TTL trigger
PLL
0 … 100% of sampling depth
Level resolution
12 Bit
Level accuracy
0.6% + 0.3%
Trigger slope
positive or negative
Trigger hysteresis
0 … 100% of signal range
An. Trigger
Trigger Masterclock
Control logic
Amplifier
CLK
Input
50W
PLL
DC/DC
www.VXInstruments.com
2
Comment
Trigger is armed after presamples are capured Number of samples captured after trigger event Done by software
±(of programmed value + of full range)
Programmable via software
Comment
CMMR DC
>90 dB
Range 1 V, 5 kS/s, average of 100 samples, Common Mode DC Voltage 150 V, Input terminated by 50 Ω
Software selectable; filter 1st order
CMMR 50 Hz
80 dB typical
Range 0.25 V, 5 kS/s, average of 100 samples, Common Mode DC Voltage 150 V, Input terminated by 50 Ω
Specification
Comment
Family related VXI Digitizers
Accuracy
50 ppm, 1 ppm with option TCXO
In operating temperature range
VX7340
20MS / 16Bit
Aging per year
5 ppm, 1 ppm with option TCXO
VX7342
40MS / 16Bit
VX7344
100MS / 16Bit
Comment
HV
500Vpp Inputs
16BIT
16 Bit resolution
DBW
Double Bandwid
EXTTRG
Ext. analog trigg
17
OCXO
OCXO Oscillator
MEMSEG
Memory segmen
Family related VXI Digitizer
Memory
EEPROM
Specification
Time Base
Ordering Option
DC/DC
CH1 or CH2; each with its own trigger source input voltage range ±10 V Via software command Trigger 0 … 7 at the backplane
Modes Pre-trigger Post-trigger Continous
CH 1
16Bit-ADC
SerDes
Optocoupler
50W
Common Mode Rejection
The addition of common mode and differential input voltage may not exceed 60 V (250 V with option IN250V). DC accuracy specified for an average value of 100 samples with a sample rate of 5 kS/s performed 24 hours after offset correction. 1
Trigger System
Input
CH 2
■■
Amplifier
CLK
Masterclock
VXI Interface
Control logic
16Bit-ADC
■■
Trigger
SerDes
■■
The instrument calibration is done digitally and is fully automatic. The calibration data are stored in onboard EEPROM.
An. Trigger
Optocoupler
■■
Two independent channels Input voltages up to 500 Vpp 100 MS/s with up to 16 Bit resolution Up to 100 MHz bandwidth Fully isolated design/floating inputs High common mode rejection Designed for high throughput testing Additional external analog trigger option Many configuration possibilities according to trigger and sampling
Memory
EEPROM
SerDes
■■
The maximum voltage for each signal input is 60 V (120 Vpp). With the HV-option the range can be extended to 250 V (500 Vpp). This allows high voltage signals to be measured without signal conditioning.
SerDes
Waveform Digitizer
VX 734X HIGH RESOLUTION WAVEFORM DIGITIZER FAMILY
Control logic
16
Each channel is fully isolated. This results in a ve high common mode rejection ratio (CMRR) comp red to differential inputs. It allows floating low lev signals to be measured with a very high accuracy and a maximum of resolution. This design of the VX7340 guarantees highest quality measurement
The data acquisition may be triggered from the input itself or external inputs (optional) having programmable thresholds. Acquired data can be pre-trigger, post-trigger, or anywhere in between, with a programmable sample counter that controls the number of data points. An interrupt is issued when measurement is completed.
Download the complete Datasheet here: http://www.vxinstruments.com/ catalog/VX734x.pdf
VX7340
20MS / 16Bit
VX7342
40MS / 16Bit
VX7344
100MS / 16Bit
18
Waveform Digitizer
VX 2022 HIGH PERFORMANCE DIGITIZER ■■ ■■ ■■ ■■ ■■ ■■ ■■ ■■ ■■ ■■
VX 2022
Trigger Trigger Sources intern
High performance upgrade for HPE1429A Two channels with 20 MS/s and 12 Bit resolution Fast register based single slot VXI-C-size module Each channel is isolated to PE Channel 1 is isolated from channel 2 Each channel has its own internal analog trigger Each channel has an external trigger input Sample count for pre and post trigger programmable Pre and post trigger with the same time base High common mode rejection
General
Specification
Max. Sampling Rate
20 MS/s per channel
Min. Sampling Rate
0.05 S/s per channel
Sampling Times
50 ns, 100 ns, 200 ns, 500 ns, 1 µss, 2 µs, 5 µs, 10 µs, 20 µs, 50 µs, 100 µs, 200 µs, 500 µs, 1 ms, 2 ms, 5 ms, 10 ms, 20 ms, 50 ms, 100 ms, 200 ms, 500 ms, 1 s, 2 s, 5 s, 10 s, 20 s
extern VXI-TTL Software Trigger Modes
Specification
Comment
channel 1 or channel 2 ±11 Bit (12 Bit) ±10 V ±11 Bit (12 Bit)
Internal sources Voltage resolution Input voltage range Voltage resolution
Post-Trigger
Number of samples to be captured after a Trigger event. The Trigger is enabled after all pre-samples are captured. After the Trigger event the number of post samples is caputred and the Trigger is disabled again. This is done via software
Pre-Trigger
Continuous Trigger slope
pos., neg.
Trigger coupling
DC
Comment
Accuracy of Time Base
100 ppm
Voltage Resolution
12 Bit (±11 Bit)
Voltage Ranges1
0.25 V, 0.5 V, 1 V, 2 V, 4 V, 8 V, 16 V, 32 V, 64 V, 128 V, 250 V
Input impedance
1 MΩ
Common Mode Range1 (relative to PE)
0 … 250 V
Memory per Channel
4 MByte (2 MS)
Common Mode Rejection
Specification
Comment
CMRR DC
> 90 dB
Range 1 V, 5 kS/s, average value over 100 S (20 ms) Common mode DC voltage 150 V Input short circuited with 50 Ω resistor
CMRR 50 Hz
typ. 80 dB
Range 0.25 V, 5 kS/s, average value over 200 S (40 ms) Common mode AC voltage 250 Vs Input short circuited with 50 Ω resistor
Optional 8 MByte (4 MS)
The sum of common mode voltage and differential voltage between plus and minus input of the corresponding channels must not exceed 250 V. The maximum voltage difference between one pin of the front connector and PE must not exceed 250 V.
DC-Error2
19
Analog
Range
% of Meas Value
Offset
Bandwidth
±0.25 V
0.5
0.002 V
10 MHz
±0.5 V
0.4
0.004 V
10 MHz
±1 V
0.3
0.005 V
10 MHz
±2 V
0.3
0.010 V
10 MHz
±4 V
0.3
0.015 V
10 MHz
±8 V
0.3
0.025 V
10 MHz
±16 V
0.3
0.035 V
10 MHz
±32 V
0.3
0.070 V
4 MHz
±64 V
0.3
0.140 V
4 MHz
±128 V
0.3
0.250 V
2 MHz
±250 V
0.3
0.500 V
1 MHz
DC-error is specified for an average value of 100 measures with a sampling rate of 5 kS.
2
1
Download the complete Datasheet here: http://www.vxinstruments.com/ catalog/VX2022.pdf
20
Waveform Digitizer
TIMER/COUNTER OPTION FOR DIGITIZERS Timer/counter-functions for rise/fall time and frequency measurement ■■ 40 Bit counter width ■■ Time base accuracy up to 1 ppm ■■ Available as option for VX Instruments Waveform Digitizers ■■
Acquisition
Specification
Comment
Bandwidth
up to 100MHz
Depends on input bandwidth of digitizer
Time base accuracy
up to 1ppm
Depends on time base accuracy of digitizer
Counter width
40 Bit
Input impedance
1 MΩ || < 20 pF nom.; 50 Ω software selectable
Input coupling
DC
Maximum input voltage
1 MΩ: up to 250 VDC 50 Ω: up to 8 Vp
Depends on max. input voltage of digitizer
Input ranges
—
Depends on digitizer input ranges
Filter
30 kHz, 100 kHz, 300 kHz, 1 MHz, 20 MHz
Depends on digitizer input filters, software selectable
Available for VX Instruments waveform digitizers The timer/counter option (T/C) is available for almost every VX Instruments waveform digitizer. This functionality is implemented fully digitally in hardware to save components, space, time and costs. Many direct measurement possibilities collaterally with the digitizer sampling With the timer/counter option every digitizer has the ability to measure rise times, fall times, frequencies, period and pulse width very fast. This is also possible while the digitizers samples data after a trigger event. The timer/counter engine returns the results quickly after the measurement. There is no need to evaluate the sampled data points.
Depends on max. input impedance of digitizer
Measurement Modes
Specification
Frequency Counter width Range Minimum pulse width
40 Bit 0.1 Hz … 10 MHz 100 ns
Period Resolution Accuracy Range
10 ns ±10 ns 50ns … 10s
Time Interval and Pulse Width Resolution Accuracy Range
10 ns ±10 ns 50ns … 10s
Rise and Fall Time Resolution Accuracy Range
10 ns ±10 ns 50ns … 10s
Totalize Minimum pulse width Range
50 ns 0 … 240
Trigger
Specification
Level resolution
16 Bit
Level accuracy
0.6% + 0.3%
Trigger slope
positive or negative
Trigger hysteresis
0 … 100% of signal range
21
Comment
±(of programmed value + of full range)
Programmable via software
High time base accuracy and high input voltage range possible For digitizers with a time base accuracy of 1 ppm (may be optionable), the timer/counter time base accuracy will be also 1 ppm. Furthermore the input voltage range and input impedance depends on the input range of the digitizer, which has the timer/counter option implemented. For the PXD731x Waveform Digitizer family, the maximum input voltage can be 500 Vpp. The time base accuracy is 1 ppm.
Download the complete Datasheet here: http://www.vxinstruments.com/ catalog/TC_Option.pdf
WAVEFORMGENERATOR GENERATOR WAVEFORM
22 22
23 23
ProductCATEGORIES Categories PRODUCT
MODELS MODELS ■ PXA 722 x ■■ PXA 722 x 16 Bit; 100 MS/s; 40 MHz bandwidth; 2 ms >2 ms
AXC 7636
300 A
60 V
15 U
110 kg
High Power SMU 300 A / 60 V
AXC 7641
1600 A
15 V
19 U
140 kg
High Power SMU 1600 A / 15 V
AXC 7643
800 A
30 V
19 U
140 kg
High Power SMU 800 A / 30 V
AXC 7646
400 A
60 V
19 U
140 kg
High Power SMU 400 A / 60 V
AXC 7651
2000 A
15 V
23 U
170 kg
High Power SMU 2000 A / 15 V
AXC 7653
1000 A
30 V
23 U
170 kg
High Power SMU 1000 A / 30 V
AXC 7656
500 A
60 V
23 U
170 kg
High Power SMU 500 A / 60 V
Voltage Measurement Unit
Specification
Resolution
16 Bit 16 Bit
Filter frequencies
100 Hz, 1 kHz,10 kHz, 1 kHz,10 kHz, 100 kHz 100 kHz 100 Hz,
DC accuracy1 1V Range 1 V 10 V Range 10 V 100 V Range 100 V
±0.1% of full scale ±0.1% of full scale ±0.1% of full scale
Current Measurement Unit
Specification
Resolution
Comment
±0.1% of full scale with option HP
Comment
Comment
Option GPIB2
GPIB Interface
Option USB
USB2.0 Interface
2
Option LAN2
Ethernet Interface
16 Bit 16 Bit
Option FE 2
Front panel display
Filter frequencies
100 Hz, 100 Hz, 1 kHz,10 kHz, 1 kHz,10 kHz, 100 kHz 100 kHz
Option HP
High precision DC accuracy
DC accuracy1 Range *A
±0.3% of full scale max. output current
Hz-filter and 20 samples with an interval of 1 ms. With 100 100 Hz-filter 1 ms.
1
43
±0.1% of full scale with HP option (see ordering information table)
One of these interface options is mandatory.
2
Download the complete Datasheet here: http://www.vxinstruments.com/ catalog/AXC76xx.pdf
a programmable slew rate of up to 1000V/msec. With this fast slew rate the test throughput is much higher than with conventional high voltage solutions.
44
High-Power SMU
The AXV7607 provides five current ranges and is able to generate currents down to the nA range. Togetherhigh with the integrated voltage The AXV7607 voltage source andmeasureMeasurement (VMU) all for categories of high voltage ment Unit isunit developed fast semiconductor semiconductor test can be done. The integrated and cable testing. The AXV7607 guard amplifier can be used tocan drivebe theused cableas voltageshielding or as current source. With the bipolar and reduce the leakage current in the cabling. MOSFETDUT power amplifier the AXV7607 provides
AXV 7607 HIGH VOLTAGE SMU ■■ ■■ ■■ ■■ ■■ ■■ ■■ ■■ ■■
Extremely low noise with linear output stage Output voltage up to 2500 V Very fast rise and fall times Output current in pulse mode max ±30 mA Output current in continuous mode max ±8 mA Integrated voltage-Measurement Unit Integrated current-Measurement Unit Integrated voltage-monitor Integrated current-monitor
a programmable slew rate measure of up tounit 1000 V/ms The integrated current (CMU) is exand thus maximizes test throughput. tremely fast andthe is equipped with 7 measurement ranges. This allows current measurements
the AXV7607 is equipped with a fast voltage monitor, which provides a divided output voltage for save monitoring. Precise monitoring of very low currents down to pA with external equipment is usually very difficult and special equipment is needed. To simplify this task, the AXV7607 provideswith a fastthe current monitor. voltage In conjunction integrated
Measurement Unit relay (VMU) all requirements of high In the integrated matrix all high voltage signals (FORCE, SENSE, GUARD) are connected voltage semiconductor testing can be met. The via relays to the output. Additional and to drive integrated guard amplifier canrelays be used switchable resistors provide self test and conthe cable shielding and reduce the leakage curtact test capabilities. rent in the DUT cabling. All AXV7607 high voltage outputs (FORCE, SENSE, GUARD) and the monitor outputs are short circuit protected. Guard
UIG
HF
CMU
HS
Low Noise DC/DC VMU
DUT
Self Test LS LF
AXV 7607
Voltage Control Unit
Specification
Comment
Resolution
16 Bit
In all ranges
Compensation units
7
3 assembled, 4 arbitrary
Range
0 V … 2500 V
Programmable output voltage
DC accuracy Gain error Offset error
0.05% of final value 0.05% of full scale
Load
20 nF
Guard buffer
2 mA
Slew rate
10 V/ms … 1000 V/ms
Matrix High Speed CurrentMonitor
Current Control Unit
Specification
Comment
Resolution
16 Bit
In all ranges
Compensation units
7
3 assembled, 4 arbitrary
Range 1 Gain error Offset error
-10 μADC … +10 μADC 0.2% of value 0.2% of full scale
Programmable range
Range 2 Gain error Offset error
-100 μADC … +100 μADC 0.1% of final value 0.1% of full scale
Programmable range
Range 3 Gain error Offset error
-1 mADC … +1 mADC 0.1% of final value 0.1% of full scale
Programmable range
Range 4 Gain error Offset error
-10 mADC … +10 mADC 0.1% of final value 0.1% of full scale
Programmable range
Range 5 Gain error Offset error
-30 mADC … +30 mADC 0.1% of final value 0.1% of full scale
Programmable range (max. pulse length 10 ms1)
www.VXInstruments.com
2
Max. capacitive load
Cooling is designed for max. pulse/pause ratio of 1:5.
1
High Speed VoltageMonitor
Voltage Monitor
Specification
Current Measurement Unit
Specification
Output voltage
-5 V equivalent to +Full scale in each range
Resolution
16 Bit
Internal resistance
10 k
Overload protection
180 mA in all ranges
Accuracy
±5% of final value
Filter frequencies
100 Hz, 1 kHz, 10 kHz, 100 kHz
Current Monitor
Specification
Output voltage
+5 V equivalent to +Full scale in each range
Internal resistance
10 k
Accuracy
±5% of final value
Ordering Information
Comment
AXV 7607-1500
Range: -1500 V … +1500 V
AXV 7607-2500
Range: 0 … 2500 V
Option GPIB3
GPIB Interface
Option USB3
USB2.0 Interface
Option LAN3
Ethernet Interface
Option FE3
Front panel display
Option RMK
19” rack mounting kit
Option PBF
Completely RoHS compliant
Option CT
Contact Test
Specification takes effect with 100 Hz filter frequency and 20 consecutive measurement with an interval of 1 ms. 3 One of the interface options or the front-panel is mandatory. 2
DC accuracy2 Range 100 nA Range 1 μA Range 10 μA Range 100 μA Range 1 mA Range 10 mA Range 100 mA
±0.5% of full scale ±0.2% of full scale ±0.1% of full scale ±0.1% of full scale ±0.1% of full scale ±0.1% of full scale ±0.1% of full scale
Voltage Measurement Unit
Specification
Resolution
16 Bit
Maximum input voltage
2500 V
Filter frequencies
100 Hz, 1 kHz, 10 kHz, 100 kHz
DC accuracy2 Range 1 V Range 10 V Range 100 V Range 1 kV Range 10 kV
±0.3% of full scale ±0.1% of full scale ±0.1% of full scale ±0.1% of full scale ±0.1% of full scale
Download the complete Datasheet here: http://www.vxinstruments.com/ catalog/AXV7607.pdf
45
SOURCE&&MEASUREMENT MEASUREMENT SOURCE
46 46
ProductCATEGORIES Categories PRODUCT
MODELS MODELS ■ AXM 7720 48 ■■ AXM 7720 48 ±150 V; up to ±150 mA; source+sink; 16 Bit; 2 ch. ±150 V; up to ±150 mA; source+sink; 16 Bit; 2 ch. ■ AXS 772x ■■ AXS 772x max; ±15 max.; source+sink 2020……8080V Vmax; ±4±4 …… ±15 AA max.; source+sink
50 50
PX 773 x ■■■PX 773 x ±10……±60 ±60V;V; 200 mA 1A max.; source+sink ±10 200 mA …… 1A max.; source+sink
52 52
VX 2702 ■■■VX 2702 ±50V;V;upuptoto±100 ±100 mA; source+sink; ±50 mA; source+sink; 1616 BitBit
56 56
VX 6620 ■■■VX 6620
58 58
±10……±60 ±60V;V; 200 mA 1A max.; ±10 200 mA …… 1A max.; source+sink
■■■VX 6625 VX 6625
source+sink
60 60
250 mA; measurement; 2 ch. 0 0……1010V;V;0 0… … 250 mA; 1616 BitBit measurement; 2 ch.
SOURCE & & MEASUREMENT MEASUREMENT SOURCE
47 47
Product information:
48
Source & Measurement AXM 7720
AXM 7720 MULTICHANNELSOURCE MEASUREMENT UNIT ■■ ■■ ■■ ■■ ■■ ■■ ■■ ■■
Multichannel Source-Measurement-Unit The AXM7720 is a high-precision, high-speed multichannel Source-Measurement-Unit, which is designed for automated high throughput testing.
CMU and VMU with monitor outputs The integrated monitor outputs makes debugging very easy. With the integrated filter stages high precise measurements can be done even in high noisy environment.
Each function unit (VMU, CMU, Generator) is
Integrated high flexible relay matrix The integrated relayWide matrix is replaceable for range of components fast service and allows high sophisticated ■■ Bipolar transistors measurements without external connections
fully isolated to avoid ground loops and Areas of application common mode errors. ■■ High volume production testing Programmable rise- and falltimes ■ ■ Quality control The fast low noise linear bipolar power stage provides full 4-quadrant source and sink ■■ Failure analysis capability at very fast rise- and falltimes even at ■ ■ Engineering high capacitive loads.testing The rise- and falltimes are
Extremely low noise with linear output stage Fully isolated design, floating in- and outputs Fast measurement of current in nA-Range Specially designed for automatic test equipment and high throughput testing Short rise and fall times due to integrated sink capability Integrated matrix and digital I/Os Auto-sensing Digital calibration
programmable.
Two power ranges Test time With it's two power ranges, 50V/150mA and ■ ■ Bipolar 150V/50mA thetransistors AXM7720 covers a wide range of 25 ms different loads ■ ■ Diodes 15 ms
■ Diodes If the sense line is■not connected, automatically the output terminals are used as ■■ Zener diodes the sensing point . ■■
Voltage regulators Broken sense line. The output voltage will be reduced automatically aboutpin the voltage ■■ Low countdrop devices across the load line .
Block diagramm of one single source measurement unit
Areas of application:
Connector High volume production testing
CMU
Quality control
GEN
GEN HS
Bipolar transistors MOSFETs Diodes
GEN LS GEN LG GEN LF
Zener diodes VMU INT VMU INT
Resolution
16 Bit
Compensation units
4
DC accuracy Gain error Offset error
Voltage regulators
www.VXInstruments.com
2
0.05% of final value 0.05% of max. range (150 V)
Output ratings Power range 1 Power range 2
±50 V / ±150 mA ±150 V / ±50 mA
Programmable range Programmable range
Voltage ramp
1 … 1000 V/ms
Programmable range
Current Control Unit
Specification
Comment
Resolution
16 Bit
In all ranges
Compensation units
8 (2 x4)
4 per pos./neg. current controler
Range 1 Gain error Offset error
-10 μADC … +10 μADC 0.2% of value 0.2% of full scale
Programmable range
Range 2 Gain error Offset error
-100 μADC … +100 μADC 0.1% of value 0.1% of full scale
Programmable range
Range 3 Gain error Offset error
-1 mADC … +1 mADC 0.1% of value 0.1% of full scale
Programmable range
Range 4 Gain error Offset error
-10 mADC … +10 mADC 0.1% of value 0.1% of full scale
Programmable range
Range 5 Gain error Offset error
-150 mADC … +150 mADC 0.1% of value 0.1% of full scale
Programmable range
VMU EXT
Low pin count devices
VMU EXT
Matrix Test time
In all ranges
GEN HG GEN HS
20% -100%
-50%
GEN LS
50% -20%
GEN LG GEN LF
Iout
-60% -80%
float. shields VMU low
-100%
Bipolar transistors 25 ms
Integrated matrix and digital-I/Os
Diodes 15 ms
Output voltage (LF related)
-5 V equivalent to Resolution The current measurement unit has the +Full scale in each range capability to measure currents in nA-Range. Maximum input The leakage current of a capacitator canvoltage be measured very fast even in a noisy environment. 10 k Filter frequencies ±2% of final value
Voltage Measurement Unit Due to theSpecification integrated matrix the instruments
Fast measurement of current in nA-Range
Current Monitor
Specification
Output voltage (HF related)
+5 V equivalent to +Full scale in each range
Internal resistance
10 k
Accuracy
±2% of final value
can be connected flexible into the environment. Furthermore 7 digital outputs can be used to 16 Bit switch external relays on a specially designed DUT-board for example.
200 V
Calibration service
100 Hz, 1 kHz,
As every VX Instruments device the AXM7720 10 kHz, 100 kHz can be adjusted and calibrated in our laboratory. In this process a protocol containing preadjustment and post-adjustment data will be generated. This report be scale delivered with the ±0.4% ofwill full device after the adjustment-process. It will help ±0.2% of full scale your quality management to check and ensure the measurement ±0.1%quality. of full scale
DC accuracy1 Range 100 mV Range 1 V Range 10 V Range 100 V Range 1 kV
±0.1% of full scale ±0.1% of full scale
Current Measurement Unit Specification www.VXInstruments.com Resolution
16 Bit
Ordering Information
Comment
Overload protection
180 mA in all ranges
Option GPIB
GPIB Interface
Filter frequencies
100 Hz, 1 kHz, 10 kHz, 100 kHz
2
100%
-40%
VMU high
Specification
Accuracy
60% 40%
GEN HF
Voltage Monitor
Internal resistance
Uout
80%
CMU
GENto HG test Wide range of components
Selftest
100%
CMU low
GEN HF
VMU
Full 4-quadrant source and sink capability:
CMU guard
Engineering testing CMU
the test solutions provider
CMU high
Failure analysis
Selftest
Comment
nstruments
Shorted sense lines. The output voltage will be limited to about 5V above the prog. value.
(VCU/CCU)
Specification
49
■■ MOSFETs Auto-sensing protects units under test
Regulator
Voltage Control Unit
to test
Option USB2
USB2.0 Interface
Option LAN2
Ethernet Interface
Option FE 2
Front panel display
Option RMK
19” rack mounting kit
Option PBF
Completely RoHS compliant
Option CT
Contact Test
Specification takes effect with 100 Hz-filter-frequency and 20 consecutive measurement with an interval of 1 ms. 2 One of the interface options or the front-panel is mandatory.
DC accuracy1 Range 10 nA Range 100 nA Range 1 μA Range 10 μA Range 100 μA Range 1 mA Range 10 mA Range 150 mA
±0.5 nA ±2.0 nA ±10.0 nA ±50.0 nA ±0.5 μA ±5.0 μA ±50.0 μA ±800.0 μA
1
Download the complete Datasheet here: http://www.vxinstruments.com/ catalog/AXM7720.pdf
3
50
Source & Measurement
AXS 772X SYSTEM POWER SUPPLY ■■ ■■ ■■ ■■ ■■ ■■ ■■
Extremely low noise with linear output stage Fully isolated design, floating outputs Specially designed for automatic test equipment and high troughput testing Short rise and fall times due to integrated sink capability Built-in measurements Digital calibration New option: Digital Regulation Loop
Specifications
AXS 7720 AXS 7721 AXS 7722 AXS 7723
AXS 7720
AXS 7721
AXS 7723
Output ratings Output voltage1 Output ratings
0 to 20 V ±15 A
0 to 30 V ±10 A
0 to 40 V ±8 A
0 to 80 V ±4 A
Output accuracy Voltage offset error Voltage gain error Current offset error Current gain error
10 mV 0.05% of value 24 mA 0.15% of value
15 mV 0.05% of value 15 mA 0.15% of value
20 mV 0.05% of value 12 mA 0.15% of value
40 mV 0.05% of value 6 mA 0.15% of value
Programming accuracy Voltage Current
0.3 mV 0.24 mA
0.5 mV 0.15 mA
0.6 mV 0.12 mA
1.2 mA 0.06 mA
Ripple and Noise (20 Hz … 20 MHz) Voltage Current
0.3 mVeff 0.24 mAeff
0.5 mVeff 0.15 mAeff
0.6 mVeff 0.12 mAeff
1.2 mVeff 0.06 mAeff
Temperature drift Voltage Current
50 ppm/°C 150 ppm/°C
50 ppm/°C 150 ppm/°C
50 ppm/°C 150 ppm/°C
50 ppm/°C 150 ppm/°C
Readback accuracy Voltage 0.05% of value + Current 0.10% of value +
10 mV 12 mA
15 mV 10 mA
20 mV 8 mA
40 mV 4 mA
1 mV 200 µs to 0.1% of full s. 1 mA
1 mV 200 µs to 0.1% of full s. 1 mA
1 mV 200 µs to 0.1% of full s. 1 mA
1.5 mV 200 µs to 0.1% of full s. 1.5 mA
Line regulation
1 mV
1 mV
1 mV
1.5 mV
Sense voltage max.
3 V
3 V
3 V
3 V
Output capacitor
2200 µF
2200 µF
2200 µF
1000 µF
Load regulation Voltage (I = 0 … 90%) Voltage (0 … 90%) Current (U = 0 … 90%)
The addition of common mode and output voltage may not exceed 500 V.
1
AXS 7722
High-speed power supply The AXS772x is a single channel, high-speed power supply for high performance measurements. This reliable series, each a 3 U 19” module, is designed for high throughput testing by realizing short rise and fall times. The rise and fall times are programmable.
Various device variants available The programmable output voltage and current is available in various ranges (see specification). The AXS772x-Family supports current source and sink. This fact allows very fast fall times even with high capacity of the Units Under Test (UUT).
Fully isolated outputs to avoid ground loops All devices of the AXS772x series have one output which is fully isolated. The control loop for programmable output-voltage and -current is available as a standard analogue control and as a fully digital control. The digital control does not cause overshoot or ringing of the output and allows fast impedance-changing of loads.
Integrated measurement possibilities The AXS772x has an integrated current and voltage Measurement Unit to read back the actual current flow through the UUT and the actual voltage applied to the UUT.
Auto-sensing protects units under test An auto-sensing feature is integrated as a built-in security to protect Units under Test.
The instrument calibration is done digitally and is fully automatic. The calibration data are stored in on-board EEPROM.
DC floating voltage Output terminals can float up to ±500 VDC from chassis ground.
Industry standard SCPI commands and VXI Plug&Play drivers simplify system integration.
Supplemental characteristics for VX4620 D Full two-quadrant source and sink capability
Sink capability:
100%
Remote sensing Up to 3 V can be dropped in each load lead. The drop in the load leads subtracts from the voltage available for the load.
40% 20% -100%
-50%
50%
100%
-20%
Ordering Information Option DRL Option GPIB
Iout
-40%
Comment
-60%
Digital Regulation
2
Option USB2 Option LAN2
Ca So
Qu All bu str
60%
Low latency The average time required for the output voltage to begin to change following receipt of digital data is 20 ms for the power supplies connected directly to the GPIB.
GPIB interface capabilities SH1, AH1, T6, L4, SR0, RL1 (4), PP0, DC1, DT0 and C0; IEEE-488.2-compatible command set.
Uout
80%
Auto sensing Sense output terminals can be left open, sensing on force output terminals is guaranteed automatically.
Output programming response time The rise and fall time (10-90% and 90-10%) of the output voltage is less than 5 ms; The output voltage change settles within one LSB (0.025% x rated voltage) of final value in less than 20 ms.
51
-80%
GPIB Interface
-100%
USB2.0 Interface Ethernet Interface
Option FE Remote sensing: Front panel display 19” rack kit Up Option to 3VRMK can be dropped inmounting each load lead. The drop in the load leads subtracts from the voltaOption CT Contact Test ge available for the load One of the interface options is mandatory. 2
Auto sensing: Sense output terminals can be left open, sensing on force output terminals Download the complete is guaranteed auDatasheet here: tomatically http://www.vxinstruments.com/ Output programming response time: catalog/AXS772x.pdf The fastest rise and fall time (1-99% and 99-1%) of the output voltage is less than 1 ms; The output voltage change settles within 0.1% of the final value in less than 5 ms.
Interface capabilities: IEEE-488.2-compatible command set over GPIB, USB or RS232
Wa Tw
Sa Th IEC equ
Ins Th plu to tin
Pro V S D
■■ ■■ ■■ ■■ ■■ ■■
Digital-I/Os
■■
100%
PXI Interface
PX 7731 PX 7732 PX 7733 PX 7734 PX 7736
Specifications Output ratings Output voltage1 Output current Output ranges (DC) Measurement Unit Voltage range Current ranges (DC)
Specifications Output ratings Output voltage1 Output current Output ranges (DC) Measurement Unit Voltage range Current ranges (DC)
PX 7731
PX 7732
PX 7733
-10 VDC … 10 VDC -1.0 ADC … 1.0 ADC 1.0 A, 0.1 A, 10 mA, 1 mA, 100 μA, 10 μA
-20 VDC … 20 VDC -0.5 ADC … 0.5 ADC 0.5 A, 0.1 A, 10 mA, 1 mA, 100 μA, 10 μA
-30 VDC … 30 VDC -0.4 ADC … 0.4 ADC 0.4 A, 0.1 A, 10 mA, 1 mA, 100 μA, 10 μA
-20 VDC … 20 VDC 0.5 A, 0.1 A, 10 mA, 1 mA, 100 μA, 10 μA
PX 7734
PX 7736
-40 VDC … 40 VDC -0.25 ADC … 0.25 ADC 0.25 A, 0.1 A, 10 mA, 1 mA, 100 μA, 10 μA -40 VDC … 40 VDC 0.25 A, 0.1 A, 10 mA, 1 mA, 100 μA, 10 μA
The addition of common mode and output voltage may not exceed 60 V.
Configurable Digital Inputs/Outputs The PX773x has four included free configurable digital TTL I/O’s and four open drain outputs, for example to drive relays or leds.
Current- / VoltageMeasurement D
A
16bit ADC-M
(internal memory possible)
Umeas (I) Umeas (U)
Generator Specification
Specification
Comment
Number of outputs
1
Output Relays
yes
Resolution
16 Bit
Voltage accuracy
0.05% + 0.05%
±(of programmed value + of full range3)
Current accuracy Accuracy in highest range Accuracy all other ranges
0.1% + 0.1% 0.05% + 0.05%
±(of programmed value + of full range) ±(of programmed value + of full range)
Temperature drift Voltage Current
50 ppm/°C 150 ppm/°C
Ripple&Noise (20 Hz … 20 MHz) Voltage (highest I-range) Voltage (all other I-ranges)
< 12 mVRMS < 60 mVpp < 10 mVRMS < 40 mVpp
RMS Normal-Mode RMS Normal-Mode
Output Settling Time2 Rise Time Fall Time