recommend that the pET be used to monitor patients receiving oral enzyme replacement therapy, which is a major drawback for the clinical application of the pET. On the other hand, the antigenic substrate in the pET seems to be equally stable in relation to elastase 1. Only elastase concentrations determined with the pET in stools stored at 4 °C for 48 h were significantly higher than directly measured concentrations. Although both pET and mET assessments were performed on stool samples stored in continuously closed cups, partial evaporation with consecutive concentration of the substrate cannot be excluded. If we take into consideration the different aspects of fecal elastase testing highlighted in this study, the lack of specificity of the pET assay remains the crucial problem. We therefore support the proposal of renaming the polyclonal “elastase 1” ELISA because the test detects a molecule different from elastase 1 (7 ). Further studies should evaluate the effects of porcine enzymes on the results of the pET assay in patients with normal and mildly to moderately decreased exocrine pancreatic function. Additionally, the true antigen detected with the pET assay should be characterized to reliably define the role of this new assay in the diagnosis of exocrine pancreatic insufficiency.
References 1. Loeser C, Moellgaard A, Foelsch UR. Fecal elastase-1: a novel, highly sensitive and specific tubeless pancreatic function test. Gut 1996;39: 580 – 6. 2. Stein J, Spichez Z, Lembcke B, Caspary WF. Evaluation of fecal elastase as a new non-invasive test for exocrine pancreatic insufficiency. Z Gastroenterol 1997;35(Suppl 1):122–9. 3. Siegmund E, Loehr JM, Schuff-Werner P. The diagnostic validity of noninvasive pancreatic function tests—a meta-analysis. Z Gastroenterol 2004; 42:1117–28. 4. Lu¨th S, Teyssen S, Forssmann K, Kolbel C, Krummenauer F, Singer MV. Fecal elastase-1 determination: ‘gold standard’ of indirect pancreatic function tests? Scand J Gastroenterol 2001;36:1092–9. 5. Dominguez-Munoz JE, Hieronymus C, Sauerbruch T, Malfertheiner P. Fecal elastase test: evaluation of a new noninvasive pancreatic function test. Am J Gastroenterol 1995;90:1834 –7. 6. Keim V, Teich N, Moessner J. Clinical value of a new fecal elastase test for detection of chronic pancreatitis. Clin Lab 2003;49:209 –15. 7. Hardt PD, Hauenschild A, Nalop J, Marzeion AM, Porsch-Ozcurumez M, Luley C, et al. The commercially available ELISA for pancreatic elastase 1 based on polyclonal antibodies does measure an as yet unknown antigen different from purified elastase 1. Binding studies and clinical use in patients with exocrine pancreatic insufficiency. Z Gastroenterol 2003;41:903– 6. 8. Lankisch PG. Now that fecal elastase is available in the United States, should clinicians start using it? Curr Gastroenterol Rep 2004;6:126 –31. 9. Stein J, Jung M, Sziegoleit A, Zeuzem S, Caspary WF, Lembcke B. Immunoreactive elastase I: clinical evaluation of a new noninvasive test of pancreatic function. Clin Chem 1996;42:222– 6. 10. Waters DL, Dorney SF, Gaskin KJ, Gruca MA, O’Halloran M, Wilcken B. Pancreatic function in infants identified as having cystic fibrosis in a neonatal screening program. N Engl J Med 1990;322:303– 8. 11. Bronstein MN, Sokol RJ, Abman SH, Chatfield BA, Hammond KB, Hambidge KM, et al. Pancreatic insufficiency, growth and nutrition in infants identified by newborn screening as having cystic fibrosis. J Pediatr 1992;120:533– 40. Previously published online at DOI: 10.1373/clinchem.2004.046888
Soluble CD40 Ligand Measurement Inaccuracies Attributable to Specimen Type, Processing Time, and ELISA Method, Anna Margre´t Halldo´rsdo´ttir,1 Joshua Stoker,2 Rhonda Porche-Sorbet,1 and Charles S. Eby1* (Divisions of 1 Laboratory Medicine and 2 Cardiology, Washington University School of Medicine, St. Louis, MO; * address correspondence to this author at: Washington University School of Medicine, St. Louis, MO 63110; fax 314-362-1461, e-mail
[email protected]) CD40 ligand (CD40L) is a member of the tumor necrosis factor superfamily and is produced in a variety of cells, including platelets. The soluble form (sCD40L) is a mediator of both inflammatory and hemostasis processes and has been implicated in the pathogenesis of atherosclerosis. Clinical studies have revealed increased sCD40L in patients with unstable angina (1 ) and identified an association between increased sCD40L and future risk for death or nonfatal myocardial infarction (2, 3 ). While prospectively measuring sCD40L in a cohort of persons at risk for cardiovascular complications, we identified both preanalytical and analytical sources of error. This report documents the effects of specimen type (serum and plasma), processing (time and temperature), and commercial reagent selection on sCD40L ELISA results. These findings raise concerns about the accuracy of sCD40L results reported in recent clinical studies. After obtaining informed consent, we enrolled 147 patients older than 60 years referred for diagnostic cardiac catheterization in an Institutional Review Board-approved study to evaluate the value of clinical, echocardiographic, and biomarker variables for prediction of future cardiovascular complications. When combined with clinical predictors, B-type natriuretic peptide and C-reactive protein, but not sCD40L, were independent predictors of death or cardiovascular hospitalization at 6 months (data not shown). The unexpectedly poor correlation between undiluted plasma sCD40L results and clinical outcomes in this study motivated us to perform the following investigations. Venous blood from the 147 study participants [mean (SD) age, 70.8 (6.9) years] was collected into plastic tubes containing tripotassium EDTA (BD Vacutainer; Becton Dickinson) before catheterization and placed on ice for 1– 4 h before processing. Blood from 10 control individuals [mean (SD) age, 38.7 (8.4) years] was collected into both EDTA and plain glass tubes (Becton Dickinson) and maintained at room temperature for 30 min before processing. Study and control samples were centrifuged twice: first at 2790g for 5 min to separate cells from the plasma/serum and then at 16 000g for 3 min to remove any residual platelets. Supernatants were aliquoted and stored at ⫺70 °C. We assessed the effects of time and temperature on measured sCD40L concentrations by collecting whole blood from a single healthy individual into a syringe and immediately aliquoting it into EDTA-containing and plain glass tubes. For every time point analyzed, plasma and serum tubes were kept on cells at room temperature, and
Clinical Chemistry 51, No. 6, 2005
in 1 experiment they were also kept on ice. Samples were then centrifuged and stored as described above. Selected samples from the time–temperature experiment were either filtered through a 0.2 m syringe filter or were ultracentrifuged at 200 000g for 4 h at 4 °C to remove potential remaining platelet microparticles before repeat sCD40L testing. sCD40L concentrations were measured with an sCD40L assay (Quantikine®; R&D Systems). According to the package insert, the R&D ELISA is suitable for measuring sCD40L in serum and plasma and is linear within the analytical range of the assay (0.0625– 4 g/L). The stated lower limit of detection is 0.0042 g/L. Reference intervals for serum (0.675–38.373 g/L; mean, 8.273 g/L; n ⫽ 44) and platelet-poor EDTA plasma (0.106 –11.831 g/L; mean, 2.987 g/L; n ⫽ 16) were provided by the manufacturer. The manufacturer’s protocols were followed. We retested selected samples with 2 sCD40L assays (BMS 239 and BMS 293) from Bender MedSystems. The BMS 239 is suitable only for testing serum, whereas the BMS 293 is a high-sensitivity assay designed both for plasma and serum (package inserts). The lower limit of detection for the BMS 239 is 0.095 g/L, and the lower limit of detection for the BMS 293 is 0.005 g/L. The manufacturer’s protocols were followed. When establishing a central 95% interval for sCD40L with 10 control plasma samples diluted 1:5, per the R&D package insert, we found that all results were below the lowest point on the calibration curve (0.0625 g/L). Following discussions with the manufacturer’s technical consultants, we tested control and patient plasma samples undiluted. The distribution of sCD40L concentrations in undiluted plasma for the 147 patients is shown in Fig. 1A. When the 2 control and 8 patient plasma samples with sCD40L concentrations exceeding the upper limit of the calibration curve were diluted 1:2 and 1:5 in calibration diluent, the results were not linear (data not shown). To determine whether the nonlinear dilution response was a systematic analytical problem, we added recombinant sCD40L to serum and plasma samples with sCD40L concentrations ⬍0.2 g/L to produce a predicted concentration of 2 g/L. Serial 2-fold dilutions of these samples in the calibration diluent also produced nonlinear results (data not shown). Interestingly, when we compared the values for undiluted serum and plasma samples from the control group, the mean sCD40L concentration in undiluted serum (1.33 g/L) was 6-fold higher than in undiluted plasma (0.24 g/L). There was a weak correlation (r ⫽ 0.223) between serum and plasma concentrations for undiluted samples. To evaluate the correlation between sCD40L concentrations measured by ELISAs from different manufacturers, we compared results obtained for selected serum and plasma samples by the R&D ELISA and the Bender BMS MedSystems ELISAs. The R&D and Bender BMS 239 assays showed good correlation for serum samples (Fig. 1B), but for plasma, the correlation was poor between the
1055
new Bender high-sensitivity BMS 293 assay and the R&D ELISA (Fig. 1C). Finally, we investigated the impact of the processing variables time and temperature on sCD40L determinations. When serum from a healthy donor was stored on cells at room temperature, there was a 6- to 7-fold increase in sCD40L concentrations after 180 min (Fig. 1D). sCD40L concentrations in similarly treated plasma samples did not increase, and most values were below the analytical range. When either serum or plasma was stored on ice, no increase in sCD40L concentration was observed over 180 min (data not shown). To examine whether the time-dependent increase in serum sCD40L concentrations was attributable to release of platelet microparticles producing membrane sCD40L, we either filtered or ultracentrifuged specimens before repeat testing. No difference was observed (data not shown). Accurate measurement of an analyte is essential for its clinical diagnostic utility. Despite reports showing an association between increased sCD40L and cardiovascular complications (2, 3 ), there is poor agreement among studies regarding sCD40L ranges for controls or cases with similar cardiovascular risk factors (1–7 ). In addition, the methods sections in some reports fail to specify whether serum or plasma was tested or whether plasma was tested with an ELISA designed for testing serum, and they provide few sample-processing details (2, 3, 6, 8 ). After reviewing the literature and the manufacturers’ product specifications, we decided to measure plasma sCD40L with the R&D ELISA. We were disappointed to discover that the only sCD40L ELISA suitable for plasma testing at that time lacked the sensitivity to measure sCD40L in 100% of controls when plasma was diluted 1:5, according to the manufacturer’s recommendations. When we tested undiluted control and patient plasma samples, 10 of 157 (6%) gave exceedingly high results, and serial dilutions of these specimens produced nonlinear results. After we shared these findings with the manufacturer, the R&D sCD40L ELISA was briefly withdrawn from the market while changes were made in the assay diluent to address the presence of heterophilic antibodies in some samples. However, no changes were made to increase the sensitivity of the assay. Most clinical studies have measured sCD40L in serum with either the R&D or Bender 293 ELISAs, reducing the problem of analytical insensitivity. Thom et al. (9 ) reported that mean measured sCD40L concentrations were 9-fold higher in serum than in plasma when assayed with a Bender sCD40L ELISA, which is consistent with our results. In addition, we have shown that the agreement between the R&D and Bender 239 ELISA methods for measuring sCD40L in serum was excellent (Fig. 1B). However, we have also shown that the serum sCD40L concentration increases significantly with time in samples stored at room temperature (Fig. 1D). This is in agreement with previously published data (9, 10 ) and represents the combination of in vivo sCD40L, which is likely to be the
1056
Technical Briefs
physiologically relevant component, and ex vivo-released sCD40L. Platelets are activated during the process of clot retraction, and sCD40L shedding from the platelet surface probably accounts for the progressive increase in serum concentrations. Shed sCD40L could be bound to platelet
microparticles (7 ). However, in our experiments, filtration and ultracentrifugation did not lead to a decrease in serum concentrations of sCD40L, suggesting that ex vivoreleased sCD40L is not bound to intact membrane. It may therefore be impossible to distinguish between in vivo and ex vivo release of sCD40L.
Fig. 1. sCD40L results depend on specimen type/processing and ELISA method. (A), distribution of undiluted plasma sCD40L concentrations in 147 patients undergoing cardiac catheterization. Each data point represents 1 patient. Specimens were tested undiluted. Dotted lines represent the dynamic range of the assay. (B and C), comparison of 2 commercial sCD40L ELISA tests, using plasma and serum specimens from the time-temperature experiments. Dashed lines represent the lines of unity. Results have been corrected for dilution. (B), comparison of the R&D assay with the Bender BMS 239 for serum samples. The results of the regression analysis were as follows: y ⫽ 0.96x ⫹ 0.42 g/L. (C), comparison of the R&D ELISA with the Bender BMS 293 for plasma samples. The results of the regression analysis were as follows: y ⫽ 0.96x ⫹ 6.08 g/L. (D), effect of sample storage conditions on measured sCD40L concentrations. Each data point represents the mean of 2 experiments, the error bars represent SD. Serum (E) and plasma (F) samples drawn from a healthy donor were stored on cells at room temperature for different lengths of time before processing. The graph shows the change in measured sCD40L concentrations with time. Results have been corrected for dilution.
Clinical Chemistry 51, No. 6, 2005
The measurement of sCD40L concentrations in human blood with the R&D ELISA is therefore problematic for the following reasons: the assay lacks sensitivity for measuring sCD40L concentrations in diluted plasma samples; testing of serum is problematic because of ex vivo release of sCD40L; there is poor correlation between plasma and serum samples; and the linearity of measurements obtained with the reformulated assay reagents has not been evaluated. Recently, Bender MedSystems began selling a highsensitivity sCD40L ELISA (Bender 293) suitable for plasma and serum testing. A preliminary evaluation confirmed that it is more sensitive than the R&D sCD40L ELISA test for plasma, but no further studies have been performed. In summary, investigators should carefully consider the choice of specimen type, specimen-handling procedures, and properties of the commercial ELISA tests when measuring sCD40L concentrations in blood because each of these variables can critically affect measured sCD40L concentrations. The optimum strategy would be to measure sCD40L in platelet-free plasma by a sensitive analytical method.
We thank Bender MedSystems (Vienna, Austria) for supplying the high-sensitivity sCD40L ELISA.
References 1. Aukrust P, Mu¨ller F, Ueland T, Berget T, Aaser E, Brunsvig A, et al. Enhanced levels of soluble and membrane-bound CD40 ligand in patients with unstable angina. Circulation 1999;100:614 –20. 2. Heeschen C, Dimmeler S, Hamm C, van den Brand M, Boersma E, Zeiher A, et al. Soluble CD40 ligand in acute coronary syndromes. N Engl J Med 2003;348:1104 –11. 3. Varo N, de Lemos J, Libby P, Morrow D, Murphy S, Nuzzo R, et al. Soluble CD40L. Risk prediction after acute coronary syndromes. Circulation 2003; 108:1049 –52. 4. Scho¨nbeck U, Varo N, Libby P, Buring J, Ridker P. Soluble CD40L and cardiovascular risk in women. Circulation 2001;104:2266 – 8. 5. Viallard J, Solanilla A, Gauthier B, Contin C, De´chanet J, Grosset C, et al. Increased soluble and platelet-associated CD40 ligand in essential thrombocythemia and reactive thrombocytosis. Blood 2002;99:2612– 4. 6. Aggarwahl A, Schneider D, Terrien E, Sobel B, Dauerman H. Increased coronary arterial release of interleukin-1 receptor antagonist and soluble CD40 ligand indicative of inflammation associated with culprit coronary plaques. Am J Cardiol 2004;93:6 –9. 7. Jinchuan Y, Zonggui W, Jinming C, Li L, Xiantao K. Upregulation of CD40CD40 ligand system in patients with diabetes mellitus. Clin Chim Acta 2003;339:85–90. 8. Yan J, Zhu J, Gao L, Wu Z, Kong X, Zong R, et al. The effect of elevated serum soluble CD40 ligand on the prognostic value in patients with acute coronary syndromes. Clin Chim Acta 2004;343:155–9. 9. Thom J, Gilmore G, Yi Q, Hankey J, Eikelboom J. Measurement of soluble P-selectin and soluble CD40 ligand in serum and plasma. J Thromb Haemost 2004;2:2067–9. 10. Nannizzi-Alaimo L, Rubenstein M, Alves V, Leong G, Phillips D, Gold H. Cardiopulmonary bypass induces release of soluble CD40 ligand. Circulation 2002;105:2849 –54. DOI: 10.1373/clinchem.2005.048199
1057
Differences and Similarities between Two Frequently Used Assays for Amyloid  42 in Cerebrospinal Fluid, Niki S.M. Schoonenboom,1,2†* Cees Mulder,2† Hugo Vanderstichele,3 Yolande A.L. Pijnenburg,1 Gerard J. Van Kamp,2 Philip Scheltens,1 Pankaj D. Mehta,4 and Marinus A. Blankenstein2 (1 Alzheimer Center and Department of Neurology, and 2 Department of Clinical Chemistry, VU University Medical Center, Amsterdam, The Netherlands; 3 Innogenetics NV, Ghent, Belgium; 4 Institute for Basic Research in Developmental Disabilities, Department of Developmental Neurobiology, Division of Immunology, Staten Island, NY; † these authors equally contributed to the work; * address correspondence to this author at: Departments of Neurology and Clinical Chemistry, VU University Medical Center, PO Box 7057, 1081 HV Amsterdam, The Netherlands; fax 31-(0)204440715, e-mail
[email protected]) Amyloid  42 (A 42) concentrations in cerebrospinal fluid (CSF) are used to identify Alzheimer disease (AD) (1 ), but reported concentrations differ among studies, as does diagnostic accuracy (2 ). These differences may relate to the patient and control groups studied (3 ), processing and storage methods (4 ), intra- and interassay variation of the assays, or to the reagent antibodies used. A recent metaanalysis (2 ) stressed the importance of standardizing assays for A– 42 in CSF. In most studies, CSF A 42 was reported to be decreased, but in 2 studies, CSF A 42 was not significantly changed in AD (2 ), and in 1 study (5 ) even increased in the early stages of disease. These dissimilarities might reflect the specificities of the antibodies incorporated in the assays. The first aim of our study was to compare A 42 concentrations measured by 2 different assays in the same CSF samples. The first assay, widely used in Europe (6 ), uses 2 monoclonal antibodies (mAbs) and detects the full-length A 42 peptide, A (1– 42) (7 ). The second assay [A (N– 42)], used mainly in the United States (8 ), detects both full-length A 42 and A peptides truncated at the NH2 terminus (9 ). The second aim of our study was to compare diagnostic accuracies of the assays for patients with AD compared with controls without dementia and patients with frontotemporal lobar degeneration (FTLD). Finally, we investigated the relationship between CSF A (1– 42) and A (N– 42) concentrations and albumin ratio, age, disease duration, and disease severity. Between October 2000 and December 2002, we recruited 39 AD patients, 24 FTLD patients, and 30 nondementia controls at the Alzheimer Center of the VU University Medical Center (VUMC). All patients underwent a standardized investigative battery (3 ). A diagnosis of “probable” AD was made according to the NINCDSADRDA criteria (10 ); the clinical picture of FTLD (including frontotemporal dementia, semantic dementia, and progressive aphasia) was based on international clinical diagnostic criteria (11 ). Disease duration was defined as the time in years between the first symptoms by history and the lumbar puncture.