Pan Ki Kim, PhD1, Yoo Jin Hong, MD, PhD1, Dong Jin Im, MD1, Young Joo Suh ... Jin Hur, MD, PhD1, Young Jin Kim, MD, PhD1, Byoung Wook Choi, MD, PhD1.
We hypothesized that systolic T1 mapping with a heart-rate-dependent (HRD) pulse sequence .... scheme with different cardiac phases at second scan were.
A Modified Look-Locker Inversion Recovery (MOLLI) sequence was acquired before and 15 min after administration of 0.1 mmol/kg gadopentetate dimeglumine.
J Cardiovasc Magn Reson. 2013;15:13. doi:10.1186/1532-429X-15-13. 11. Dall'Armellina E, Piechnik SK, Ferreira VM, Si QL, Robson MD, Francis JM, et al.
Jan 16, 2014 - Methods. Canines (n = 29) underwent CMR at 7 days (acute) and. 4 months (chronic) following reperfused MIs at 3T (n = 19) and 1.5T (n = 10).
Reproducibility of multiple T1 Mapping techniques and ... - Springer Link
Jan 16, 2014 - The Look-Locker (LL) and Modified Look-Locker both precontrast (MOLLI_Pre) and postcontrast (MOLLI_Post). T1 time were measured by 3 ...
Yingchoncharoen et al. Journal of Cardiovascular Magnetic Resonance 2014, 16(Suppl 1):P18 http://www.jcmr-online.com/content/16/S1/P18
POSTER PRESENTATION
Open Access
Reproducibility of multiple T1 Mapping techniques and to ECV quantification Teerapat Yingchoncharoen*, Chirstine Jellis, Zoran Popovic, Scott Flamm, Deborah Kwon From 17th Annual SCMR Scientific Sessions New Orleans, LA, USA. 16-19 January 2014 Background T1 Mapping has emerged as a new marker to quantify myocardial extracellular volume (ECV), which is typically increased in the setting of diffuse fibrosis or infiltrative heart diseases. However the reproducibility of the technique is largely unknown. We sought to identify the interobserver, intraobserver variability of different T1 mapping techniques as well as validation of one technique against the standard technique for ECV quantification. Methods We selected 10 patients with the diagnosis of cardiomyopathy (5 with delayed contrast enhancement and 5 without). The Look-Locker (LL) and Modified Look-Locker both precontrast (MOLLI_Pre) and postcontrast (MOLLI_Post) T1 time were measured by 3 investigators blinded to clinical data at 2 different time points. Interobserver and Intraobserver reproducibility were assessed using 2-way ANOVA approach. The ECV was calculated using pre and post contrast T1 and hematocrit. Results There was a strong correlation between LL T1 and Postcontrast MOLLI(MOLLI_Post) (R2 0.82, p < 0.001) and a modest correlation between MOLLI_post and ECV (R2 = 0.42, p < 0.001). The interobserver and intraobserver variability of the measurements were expressed as standard errors of measurements (SEM) and 95% CI and were summarized in table 1. MOLLI_Post had the lowest intraobserver and intraobserver variability with minimal detectable change in T1 of 13.8 and 26.1 respectively. ECV showed minimal both interobserver and intraobserver variability.
Table 1 Interobserver and intraobserver variability of LL, MOLLI_Pre, MOLLI_Post and ECV Intraobserver
LL
MOLLI_Pre
MOLLI_Post
ECV
5.7
14.5
5.0
0.9
Min Δ detectable
15.7
40.1
13.8
2.6
Interobserver
29.1
15.7
9.1
1.9
Min Δ detectable
80.7
43.5
26.1
5.3
Conclusions Both LL and MOLLI are highly reproducible technique. LL was strongly correlated with post-contrast MOLLI which is moderately correlated to ECV. Funding None. Published: 16 January 2014
doi:10.1186/1532-429X-16-S1-P18 Cite this article as: Yingchoncharoen et al.: Reproducibility of multiple T1 Mapping techniques and to ECV quantification. Journal of Cardiovascular Magnetic Resonance 2014 16(Suppl 1):P18.
Submit your next manuscript to BioMed Central and take full advantage of: • Convenient online submission • Thorough peer review • No space constraints or color figure charges • Immediate publication on acceptance • Inclusion in PubMed, CAS, Scopus and Google Scholar • Research which is freely available for redistribution Submit your manuscript at www.biomedcentral.com/submit