SETA MULTIFLASH
34200-0
SETA MULTIFLASH ABEL CLOSED CUP TEST MODULE
TEST METHODS Abel (Closed Cup and Equilibrium) IP 170; ISO 13736; EN 13736; BS 2000 Part 170, 49/492; IP 304 Parts 1 and 2; IP 491; IP 492; ISO 1516; ISO 1523; BS 3900 Parts A8 and A9; BS 6664 Parts 1 and 2 • Precisely conforms to the test methods • Gas or electric ignition • Automatic heating control, ignition & flash detection • Automatic post test cool down • Ramp, Flash / No Flash and Equilibrium Modes • Optional sub-ambient tests • Optional liquid cooling 34000-0 + 34200-0 The test module comprises a DIPS pod (Dipping, Ignition, Pilot, Stirrer) and a Test Area. The DIPS pod contains an ignitor dipping mechanism, a coaxial gas / electric ignitor and a stirrer drive mechanism.
Stirrer Bath Prt Shutter
Automatic interface with advanced software
The Test Area comprises a liquid bath, bath temperature probes, sample temperature probe, flash detector and sample cup with lid and sliding shutter. The hotplate of the Universal Base Unit heats the bath.
The Universal Base Unit recognises the Test Module that has been installed and automatically loads the stirrer speed, temperature ramp, dipping interval parameters and calibration data all as specified in the test method.
Sub-ambient tests can be carried out using the Universal Cooling module (34008-0) and an external cryostat. Sample Prt
Lid
Flash Detector
Ignitor
DIPS Pod Stirrer drive Shutter actuator Flash detector Sample Prt
Draught Screen (Part of Universal Base Unit)
PLEASE NOTE IMAGE SHOWN WITHOUT JACKET
Bath Prt
Bath
TECHNICAL EXCELLENCE FROM STANHOPE-SETA
Simple operation All test parameters are controlled automatically via the Universal Base Unit and a test is initiated in just a few steps.
ABEL 1 Show Results 2 Run Test 3 Print Results 4 Alter Parameters
Safety features Bath over-temperature sensor. Sample Prt check to ensure that the probe is immersed in liquid. Ramp Mode; pre-heating dip to check that the sample is not above its flash point. Reliable operation & low service cost The test module is designed for reliable operation, the ignitor and other service components are low cost to replace.
Edit Sample Ident 123456789012 [SAMPLE-TANK2] Next
Automatic & precise flash detection Flash point detection is by a precision thermocouple.
34200-0/34000-0
Method:
Abel
Test reports Test results are automatically sent to an RS 232 port on the Universal Base Unit and displayed on the screen.
Fast cool-down after a test The required Universal Cooling Module (34008-0) allows water of a cryostat to be used for bath filling, rapid cooling and sub ambient tests. Unique ignition system The unique coaxial ignitor provides both gas and electric (hotwire) ignition sources.
PHYSICAL DATA
SPECIFICATIONS Seta Model:
ABEL:1.0°C/min,30rpm Sample (Wire) 32°C Expected Flash 70°C Status: Dipping Stop
Voltage:
Test Result Module: ABEL S/N : 0200439008 Test No. : 000012 Sample Id: OK RAMPWIRE 1.0°C/min,30rpm First Dip at: 61°C Expect Flash: 70°C Flash Point (Wire): Valid 73°C Corrected 73°C Barometric: 101kPa FLASH RESULT:
SAFETY FEATURES 115V or 230V ±15% (switchable), 50/60 Hz
Over-Temperature Cut-out:
Yes
Cup Type:
Closed Cup
Power:
1kW
Coolant Flow Switched Off (if fitted):
Test Modes:
Non-Equilibrium Equilibrium Flash/No Flash
Interface
RS232
Gas Shut Off:
Automatic
Memory
Last 5 tests
Integral Cooling Fan:
Automatic
Gas:
Laboratory gas, 3kPa (0.44psi) max pressure
Integral Safety Shield:
Yes
Software limits:
Yes
Temperature Range:
*5 to 93°C
*Extended lower range available
Barometric Correction:
Cooling Liquid: Automatic
Water (5°C and above), Water/Ethylene Glycol (below 5°), 125kPa (18psi) max pressure
Ignitor:
Hotwire or Flame
Temperature Probes
Class A Prt
Size (W x D x H): 47 x 42 x 34cm
Flash Detector:
Thermocouple
Weight:
Stirrer:
30 rpm
Heating Bath Type:
Liquid/Liquid
Integral Bath Cooling:
Fan
External Cooling (Assisted) Options:
External Cryostat or Water
Post Test Cool-down Time (90°C to 50°C):
45 min (Fan) 3 min (Assisted)
20kg
Remote End of Test Alert Use of the optional 00100-0 Setalert allows remote paging to the operator to signal the test has been completed or when operator assistance is required.