Simple Paper-Based Test for Measuring Blood ... - Clinical Chemistry

4 downloads 236 Views 1MB Size Report
coloring the paper faint red (pink). (D), The ..... 2013) and a laptop computer to digitize and analyze the ... and laptop could be used for many other purposes, not.
Clinical Chemistry 59:10 1506–1513 (2013)

Point-of-Care Testing

Simple Paper-Based Test for Measuring Blood Hemoglobin Concentration in Resource-Limited Settings Xiaoxi Yang,1 Nathaniel Z. Piety,1 Seth M. Vignes,1 Melody S. Benton,2 Julie Kanter,2,3 and Sergey S. Shevkoplyas1*

BACKGROUND: The measurement of hemoglobin concentration ([Hb]) is performed routinely as a part of a complete blood cell count to evaluate the oxygencarrying capacity of blood. Devices currently available to physicians and clinical laboratories for measuring [Hb] are accurate, operate on small samples, and provide results rapidly, but may be prohibitively expensive for resource-limited settings. The unavailability of accurate but inexpensive diagnostic tools often precludes proper diagnosis of anemia in low-income developing countries. Therefore, we developed a simple paperbased assay for measuring [Hb]. METHODS: A 20-␮L droplet of a mixture of blood and Drabkin reagent was deposited onto patterned chromatography paper. The resulting blood stain was digitized with a portable scanner and analyzed. The mean color intensity of the blood stain was used to quantify [Hb]. We compared the performance of the paperbased Hb assay with a hematology analyzer (comparison method) using blood samples from 54 subjects. RESULTS: The values of [Hb] measured by the paperbased assay and the comparison method were highly correlated (R2 ⫽ 0.9598); the standard deviation of the difference between the two measurements was 0.62 g/dL. The assay was accurate within 1 g/dL 90.7% of the time, overestimating [Hb] by ⱖ1 g/dL in 1.9% and underestimating [Hb] by ⱖ1 g/dL in 7.4% of the subjects. CONCLUSIONS: This study demonstrates the feasibility of the paper-based Hb assay. This simple, low-cost test should be useful for diagnosing anemia in resourcelimited settings, particularly in the context of care for malaria, HIV, and sickle cell disease patients in subSaharan Africa.

© 2013 American Association for Clinical Chemistry

1

Department of Biomedical Engineering, Tulane University, New Orleans, LA; 2 Sickle Cell Center of Southern Louisiana and 3 Department of Pediatrics, Section of Hematology/Oncology, Tulane University School of Medicine, New Orleans, LA. * Address correspondence to this author at: Tulane University, Department of Biomedical Engineering, 500 Lindy Boggs Building, New Orleans, LA 70118. E-mail [email protected].

1506

Hemoglobin (Hb) is the oxygen-transporting protein contained within red blood cells (RBCs). The mass concentration of Hb ([Hb]) is a measure of the oxygencarrying capacity of blood. The reference ranges of [Hb] in blood of adults are 12–16 g/dL in women and 14 –18 g/dL in men; values ⬍5 g/dL or ⬎20 g/dL are considered critical (1 ). A decreased [Hb], referred to as anemia, may be an indication of chronic blood loss, decreased RBC production, or abnormal destruction of RBCs (hemolysis). Hemolytic anemia may be associated with inherited blood disorders such as sickle cell disease, but can also be autoimmune in origin or result from exposure to certain drugs, toxins, infections, or transfusion of mismatched blood products (2 ). Other common causes of anemia include dietary deficiencies (3, 4 ), cirrhosis, and renal disease (2 ). Anemia reduces the oxygen-carrying capacity of blood, straining the cardiopulmonary system, which responds by increasing the heart rate and stroke volume to maintain oxygen delivery (5 ). Acute-onset anemia may be associated with fatigue, lethargy, and/or pallor (4 ). Patients with chronic anemia and having [Hb] ⬍5 g/dL may experience more serious complications such as angina, congestive heart failure, heart attack, and stroke (5 ). Depending on the clinical status of the patient, a [Hb] ⬍6 – 8 g/dL is currently recommended as the threshold trigger for initiating RBC transfusion (6 ). Increased [Hb] may indicate erythrocytosis due to smoking (7 ) or in response to hypoxia at high altitudes (8 ), polycythemia vera (9 ), congenital heart disease (10 ), severe chronic obstructive pulmonary disease (11 ), or severe dehydration (12 ). The increase in blood viscosity associated with an increased [Hb] may impair oxygen delivery by reducing the capillary perfusion, which may in turn increase the risk of hypoxia, as well as thrombotic and hemorrhagic complications (13 ). Because of its importance, [Hb] is measured routinely as a part of a complete blood cell count. Clinical

Received February 7, 2013; accepted June 11, 2013. Previously published online at DOI: 10.1373/clinchem.2013.204701 4 Nonstandard abbreviations: Hb, hemoglobin; RBC, red blood cell; POC, point-of-care; HbCS, Hb color scale; ␮PAD, microfluidic paper-based analytical device; RGB, red/green/blue; LOD, limit of detection; LOQ, limit of quantification.

Paper-Based Hb Test for Resource-Limited Settings

laboratories measure [Hb] with automated hematology analyzers as well as spectrophotometers, blood gas analyzers, and stand-alone CO-oximeters (14 ). The need for rapid and accurate measurement of [Hb] at the bedside has driven the development and implementation of [Hb] testing on the point-of-care (POC) platform (15 ). Currently available POC devices are designed to operate in close proximity to the patient, provide accurate results (within 1 g/dL of clinical laboratory analyzers), require small volumes of blood, and be used by persons without expert training, making the nearly instantaneous measurement of [Hb] in operating room or emergency care settings a reality (16 –18 ). Notwithstanding their obvious advantages, the relatively high per-test and analyzer costs make the use of conventional POC devices for measuring [Hb] prohibitively expensive in resource-limited settings of lowincome developing countries in sub-Saharan Africa, on the Indian subcontinent, and throughout Southeast Asia (19, 20 ). Currently available low-cost alternatives including the “hematocrit/3” method (21 ) and the WHO Hb color scale (HbCS) test (22 ) have several drawbacks that limit their use in the field. In addition to being somewhat inaccurate, hematocrit/3 relies on knowledge of hematocrit and, therefore, requires a centrifuge (21, 23 ). The HbCS test is highly sensitive to the ambient lighting conditions, operator bias, and deviations from the recommended readout time (22, 24 ). The lack of confidence in the quality of [Hb] measurements may result in clinicians relying exclusively on their clinical judgment to prescribe transfusions in resource-limited settings (25 ). Therefore, our goal was to develop a simple, low-cost, paper-based Hb assay that addressed these limitations. Methods BLOOD SAMPLES

The study protocol was approved by Tulane University Biomedical Institutional Review Board. After written informed consent was obtained, blood samples were collected into 4- or 9-mL Vacutainer tubes (K2EDTA, BD) during routine blood draws (January–April 2013) from patients of the Pediatric Hematology-Oncology Clinic (Tulane University Hospital) and the Sickle Cell Center of Southern Louisiana (New Orleans, LA). We measured standard hematological parameters including [Hb] using a Medonic M-Series hematology analyzer (Medonic M16C/M20C, Boule Medical). The Hb content of blood samples was adjusted artificially to prepare samples with [Hb] ranging from 2.5 to 25 g/dL by adding autologous plasma (for lower [Hb]) or pelleted RBCs (for higher [Hb]) to the original sample.

FABRICATION OF THE MICROFLUIDIC PAPER-BASED ANALYTICAL DEVICES

The microfluidic paper-based analytical devices (␮PADs) were fabricated by a previously published method (26, 27 ). Briefly, the pattern of the ␮PADs was designed with illustration software (Canvas 11, ACD Systems International). The devices were printed onto sheets of chromatography paper (No. 1, Whatman) with a solid-ink (wax) printer (Phaser 8560N, Xerox), and then heated on a hot plate at 150 °C for 3 min. QUANTIFICATION OF THE BLOOD STAIN COLOR

Blood samples were mixed at 1:10, 1:15, or 1:20 ratios (vol/vol) with Drabkin reagent (Ricca Chemical Co.) and incubated at room temperature (20 –25 °C) for 10 min. The components of Drabkin reagent lysed red blood cells and reacted with all forms of Hb except sulfhemoglobin present in the sample, converting them to cyanmethemoglobin, a stable brownishcolored compound. A 20-␮L drop of the mixture was placed onto the center of each ␮PAD. The resulting blood stain was allowed to dry for 25 min (28 ). Quantification of [Hb] was accomplished by scanning the sheets of chromatography paper containing arrays of ␮PADs on a portable flatbed scanner (CanoScan LiDE110, Canon USA) and then analyzing the images automatically with a custom-coded algorithm (MATLAB, The Math Works). The quantitative analysis of blood stains was based on the red/green/blue (RGB) color model values of the digitized images. The color data in the red, green, and blue channels were extracted from the RGB values of each pixel within a blood stain. We defined the color intensity of each channel as 255 – (R, G, or B); e.g., for red channel, color intensity ⫽ 255 – R. The mean color intensity of the blood stain for each color channel was calculated from the color values of all pixels within the blood stain, including the pixels corresponding to the darker ring on the periphery of the stain. The correlation between the mean color intensity and the actual [Hb] was evaluated for all color channels; the green channel showed the best linear fit and was selected to quantify [Hb] in the blood samples. STATISTICAL ANALYSIS

Pearson correlation coefficient and regression were calculated to directly compare data obtained from the paper-based Hb assay and the Medonic M-Series hematology analyzer (comparison method). Three research associates performed both types of measurements on the same day. A Bland–Altman plot (29 ) was also constructed, with consideration for limitations, to visualize the comparison. Clinical Chemistry 59:10 (2013) 1507

Fig. 1. Design and fabrication of the paper-based Hb assay. (A), The pattern of the ␮PAD was printed with a solid ink (wax) printer on chromatography paper. (B), The printed pattern was melted to create a 1-mm-wide hydrophobic barrier spanning the entire thickness of the paper substrate. (C), To perform the assay, a drop of blood mixed with Drabkin reagent was placed onto the center of the device. The lysate wicked laterally toward the periphery, coloring the paper faint red (pink). (D), The image analysis algorithm automatically detected the blood stain within each ␮PAD (dashed circle) and extracted the RGB color information for all pixels contained within the blood stain. (E), The color intensity of the pixels (between dashed lines) was used to calculate the mean color intensity (solid red line) for each blood stain. The values of the mean color intensity were used to calculate the [Hb] in the blood samples. au, Arbitrary unit.

Results DESIGN AND OPERATION OF THE PAPER-BASED Hb ASSAY

Figure 1 illustrates the design and operation of the paper-based Hb assay. Each individual ␮PAD comprised a 2.8-cm-diameter circle (Fig. 1a) designed to create a constant area for quantification with an image analysis algorithm and prevent potential crosscontamination of samples. Each sheet of chromatography paper contained a 5-by-4 array of ␮PADs and included an alignment mark to simplify automation of the subsequent image analysis. We fabricated the ␮PAD arrays by printing their pattern with a solid-ink (wax) printer (Fig. 1A). We then used a hot plate to heat the printed chromatography paper above the 1508 Clinical Chemistry 59:10 (2013)

melting point of wax to allow the wax to reflow and form hydrophobic barriers through the full thickness of the paper (Fig. 1B). When designing the pattern of the ␮PAD (Fig. 1A), we accounted for the natural widening of the printed lines (to about 1 mm thick) during the melting process (Fig. 1B). We performed this paper-based Hb assay by first mixing a sample of whole blood with Drabkin reagent and then depositing a 20-␮L droplet of the mixture onto the ␮PAD and letting the blood stain develop (Fig. 1C). The droplet spread radially from the center through the paper substrate toward the periphery of the ␮PAD due to wicking by capillary action (30, 31 ), forming the characteristic blood stain pattern (Fig. 1D) reminiscent of the stains produced by drying coffee (32 ). We selected the volume of the droplet (20 ␮L) and size of the ␮PAD (2.8 cm) so that the outermost margin of the stain could not reach the periphery of the ␮PAD (Fig. 1D). We used a portable flatbed scanner to digitize the sheets of chromatography paper carrying ␮PADs with developed blood stains. We analyzed the images to determine the mean color intensity for the blood stain contained within each ␮PAD using a simple image processing algorithm developed specifically for this purpose (Fig. 1, D and E). The algorithm used the alignment mark on the sheets of paper to determine the location of each ␮PAD (Fig. 1D). The algorithm applied a circular binary mask to the image to select the area within the hydrophobic border of the ␮PAD, and then isolated the pixels of the blood stain within the circular region by removing pixels below a threshold value on the basis of the difference in color intensity between the paper substrate and the blood stain (Fig. 1D, dashed circle). Finally, the algorithm used the color information from all pixels within the blood stain to calculate the mean color intensity of the blood stain (Fig. 1E, red solid line). We used the mean color intensity of the stain (Fig. 1E) and a calibration curve (Fig. 2B) to measure [Hb] in the blood sample. All operations of the paper-based Hb assay, including sample preparation, placement of the sample onto the ␮PAD, drying of the blood stain, image scanning, and the automated image analysis, could be completed within 35 min. EFFECT OF DILUTION ON THE MEAN COLOR INTENSITY OF THE BLOOD STAIN

Figure 2 illustrates the effect of dilution on the mean color intensity of the stains produced by blood on paper. To perform these experiments, we prepared a series of calibration samples with their [Hb] adjusted to 2.5, 5, 10, 15, 20, and 25 g/dL (as described in Methods) using blood from 3 consenting volunteers. We then mixed a small volume of each calibration sample with

Paper-Based Hb Test for Resource-Limited Settings

Fig. 2. Dependence of the mean color intensity of the blood stain on the dilution ratio. (A), Whole blood samples were diluted or concentrated with plasma or pelleted RBCs to achieve [Hb] ranging from 2.5 to 25 g/dL. A 20-␮L drop of blood mixed with Drabkin reagent at 1:10, 1:15, and 1:20 ratios (vol/vol) was placed at the center of each ␮PAD. (B–D), Mean color intensities for the red (B), green (C), and blue (D) channels of the blood stain in a ␮PAD produced by each sample at each dilution ratio. Data shown as mean (SD) (n ⫽ 3). au, arbitrary unit.

Drabkin reagent at 1:10, 1:15, and 1:20 ratios (vol/vol), waited 10 min, and placed a 20-␮L droplet of each mixture onto paper. Droplets with markedly different Hb content, due to either a difference in [Hb] or dilution, produced blood stains with different color intensities that were easily distinguishable visually. As expected, the blood stain color intensity increased with increasing [Hb] for each mixing ratio (dilution) and decreased with increasing dilution for each [Hb] value (Fig. 2A). The mean color intensity of each blood stain correlated very strongly with its [Hb] (Fig. 2B–D). The combination of 1:15 mixing ratio and green channel showed the best linear fit (mean color intensity ⫽ 2.0986 ⫻ [Hb] ⫹ 3.7226, R2 ⫽ 0.996). Therefore, we used the 1:15 mixing ratio to prepare samples in all further experiments, the mean color intensity in green channel, and the linear equation produced by this graph (Fig. 2C) as the calibration curve for estimating the [Hb] of the sample

on the basis of the mean color intensity of its blood stain. To determine the limit of detection (LOD) and the limit of quantification (LOQ) of the assay, we varied [Hb] of blood samples from 0.5 to 25 g/dL by diluting whole blood with autologous plasma or concentrating whole blood with packed RBCs (as described in Methods). We defined the LOD as the lowest [Hb] at which the blood stain could be identified and measured using our image processing algorithm, and the LOQ as the lowest [Hb] at which the disagreement with the comparison method (Medonic hematology analyzer) was within 1 g/dL. We found that the LOD for our paperbased Hb assay was 1 g/dL and the LOQ was 2.5 g/dL. RELATIVE ACCURACY AND PRECISION OF THE PAPER-BASED Hb ASSAY

We tested the relative accuracy of the paper-based Hb assay by comparing the values of [Hb] measured using Clinical Chemistry 59:10 (2013) 1509

Fig. 3. Comparison of the [Hb] measurements performed with the paper-based Hb assay and the Medonic hematology analyzer (n ⴝ 54 subjects). (A), Linear regression analysis of the correlation between the [Hb] values determined by a hematology analyzer (actual [Hb]) and by the paper-based Hb assay (estimated [Hb]). The diagonal (red dashed line) represents a 1:1 correspondence between the actual and the estimated [Hb] values. (B), Bland–Altman plots of the agreement between the paper-based Hb assay and the hematology analyzer. The solid red line is the difference between the 2 methods (mean bias); the dashed red lines are limits of agreement (within 2 SD). The mean [Hb] was 11.50 (2.86) g/dL for the paper-based Hb assay and 11.43 (3.05) g/dL for the Medonic hematology analyzer.

our assay and the Medonic M-Series hematology analyzer for blood samples from 54 subjects. Subjects included individuals with Hb AA, Hb AS, Hb SC, and Hb SS genotypes; the presence of Hb S in the blood samples did not affect the [Hb] measurements (data not shown). Figure 3 illustrates the results of this side-byside comparison. The data points representing these measurements were distributed in close proximity to 1510 Clinical Chemistry 59:10 (2013)

the diagonal line (Fig. 3A, red dashed line), indicating a good agreement between the [Hb] estimated from the mean color intensity of blood stains in our paper-based Hb assay and the actual [Hb] obtained from the standard analysis (Fig. 3A). The linear least-squares regression analysis (Fig. 3A, black solid line) showed a high degree of correlation between the 2 measurements (y ⫽ 1.05x – 0.58, R2 ⫽ 0.9598). We also constructed the Bland–Altman plot (29 ) to evaluate the agreement between the paper-based Hb assay and the Medonic hematology analyzer (Fig. 3b). The standard deviation (SD) of the difference between the comparison method (hematology analyzer) and the paper-based Hb assay was 0.62 g/dL. The limits of agreement between the 2 methods were ⫺1.30 and 1.18 g/dL. The paper-based Hb assay agreed with the hematology analyzer within 1 g/dL 90.7% of the time, overestimating [Hb] by ⱖ1 g/dL in 1.9% of subjects and underestimating [Hb] by ⱖ1 g/dL in 7.4% of subjects (Fig. 3B). In a separate experiment, we tested the precision of the paper-based Hb assay by measuring repeatedly (n ⫽ 5) the [Hb] for a series of blood samples obtained from the same subject with the actual [Hb] of each sample adjusted (as described in Methods) to 2.5, 5, 10, 15, 20, and 25 g/dL. The SD for the [Hb] measurements performed with different droplets from the same blood sample was consistently ⬍0.5 g/dL for all values of the true [Hb] tested: 0.21 g/dL (CV 8.3%) for the sample with true [Hb] of 2.5 g/dL; 0.19 g/dL (CV 3.9%) for 5 g/dL; 0.11 g/dL (CV 1.1%) for 10 g/dL; 0.21 g/dL (CV 1.4%) for 15 g/dL; 0.34 g/dL (CV 1.7%) for 20 g/dL; and 0.78 g/dL (CV 3.1%) for 25 g/dL. For comparison, the [Hb] measurement range for the Medonic hematology analyzer was 0 –35 g/dL, with CV ⬍1.0% (values provided by the manufacturer). EFFECT OF DROPLET VOLUME VARIATION AND LONG-TERM STABILITY OF THE PAPER-BASED Hb ASSAY

We measured the effect of variation in the volume of the sample droplet, which may occur during the use of the assay in the field, on the [Hb] measurement for samples with true [Hb] of 5, 10, 15, and 20 g/dL. A 20% variation in the volume of the sample droplet (16 –24 ␮L) resulted in a ⱕ0.21 g/dL SD (CV 4.3%) of the estimated [Hb] for samples with true [Hb] ⫽ 5 g/dL; ⱕ0.66 g/dL (CV 5.6%) for 10 g/dL; ⱕ0.67 g/dL (CV 4.2%) for 15 g/dL; and ⱕ0.93 g/dL (CV 4.4%) for 20 g/dL (see Supplemental Fig. 1A, which accompanies the online version of this article at http://www. clinchem.org/content/vol59/issue10). As expected, the size of the blood stain depended on the droplet volume. For droplets of the same volume, but with different true [Hb] values (5, 10, 15, and 20 g/dL) the normalized size of the blood stains produced by the droplets

Paper-Based Hb Test for Resource-Limited Settings

had an SD of ⱕ0.03 (CV 4.7%) for 16 ␮L; ⱕ0.03 (CV 3.9%) for 18 ␮L; ⱕ0.03 (CV 4.5%) for 20 ␮L; ⱕ0.04 (CV 4.1%) for 22 ␮L; and ⱕ0.03 (CV 3.7%) for 24 ␮L (see online Supplemental Fig. 1B). Finally, we tested the long-term stability of the paper-based Hb assay measurements by scanning the sheets of paper containing the blood stains 4 times during the first 20 min, while the stain was still drying, and 13 times over the next 24 h, for samples with true [Hb] of 5, 10, 15, and 20 g/dL. The color of the dried blood stain remained stable after the initial drying period. We therefore were able to scan and analyze the blood stains at any point over the next 24 h without any significant change in the [Hb] measurement (see online Supplemental Fig. 2). During this study, the room temperature in our laboratory varied from 18 °C to 30 °C, and humidity varied from 20% to 80%. We did not systematically test the effect of room temperature or humidity variations on the performance of the assay. Discussion The measurement of blood [Hb] is one of the most frequently performed laboratory tests within a wide range of medical settings, from acute trauma care to chronic disease management (1 ). While the primary reason for initial [Hb] screening is to diagnose (or rule out) anemia, frequent monitoring of [Hb] is often required in the context of many conditions to quantify blood loss, track disease progression, assess treatment efficacy, or monitor patients during procedures. When reliable diagnostic devices are available, they are often not used because they are considered to be prohibitively expensive for the resource-limited settings of developing countries (33 ). For example, in Kenyan district hospitals, nearly 15% of children with clinical histories indicative of malaria or anemia did not have their [Hb] tested (20, 34 ). In this context, a reliable, low-cost assay for measuring [Hb] would improve the diagnosis of anemia and could help minimize waste of limited resources on unneeded treatments and reduce the risk of complications associated with misdiagnosis (20, 25, 33 ). The urgent need for such an assay in sub-Saharan Africa (in areas endemic with sickle cell disease and malaria) was the primary motivation for this study. Our paper-based Hb assay measures [Hb] by simply quantifying the appearance of a stain produced by the blood sample on paper. Perhaps the most widely used alternative low-cost method for estimating [Hb] in resource-limited settings is the WHO HbCS. To perform this test, the user must visually compare the color of a paper strip stained with blood to a standard at a certain time after the blood was applied to the strip.

Unlike the HbCS, our paper-based Hb assay is not affected by changes in ambient light because the flatbed scanner itself provides consistent illumination, and the color can be read at any time within a 24-h period after collection of the blood stain. The result is not affected by operator bias because the measurement is performed completely automatically. Very little training is required to perform our paper-based Hb assay. All a user would need to do is to mix a finger-prick volume of blood with Drabkin reagent at the appropriate ratio and deposit a droplet of this mixture on paper. The complexity of these operations is about the same as that of the sample preparation steps required from consumers for performing athome rapid diagnostic tests for sperm count (35 ) or Hb A1c (36 ). All subsequent operations of the paper-based Hb assay are completely automated and userindependent, including the scanning and analysis of the stains. Notwithstanding its simplicity, our paperbased Hb assay demonstrated excellent comparison (SD ⫽ 0.62 g/dL, n ⫽ 54) with the Medonic hematology analyzer. The performance of our assay also compares favorably with conventional POC devices (14, 19, 37 ), particularly in the context of our target applications: (a) providing the measurement of [Hb] when no other methods are available, and (b) enabling public health screening in resource-limited settings where the lowest cost, rather than the highest accuracy, is most important. We used a USB-powered flatbed scanner (Canon CanoScan LiDE110, $44 on amazon.com as of April 21, 2013) and a laptop computer to digitize and analyze the blood stains. Any computer that has a functional USB port and complies with the system requirements of the scanner could in principle be used, including older salvaged or refurbished models. Importantly, the scanner and laptop could be used for many other purposes, not exclusively for performing our paper-based Hb assay. The primary reason for using these multipurpose, consumer-grade electronic devices as the capital equipment part of a low-cost diagnostic assay is the ability to tap into the already well-established market of consumer electronics. These devices are much more available in resource-limited settings than specialized medical equipment such as a spectrophotometer and can be used for performing the [Hb] measurements without any modification of their original function. This approach represents an important advantage of our work with respect to recent studies that follow the conventional paradigm of POC development and propose to measure [Hb] by use of a new, potentially low-cost electronic device (38, 39 ). Such a device, designed specifically for measuring [Hb], would have to be separately developed, manufactured, distributed, calibrated, and maintained. The use of these POC devices, Clinical Chemistry 59:10 (2013) 1511

particularly the maintenance and/or replacement of broken devices, requires a level of medical and engineering infrastructure that is largely unavailable in resource-limited settings (40 ). Another important advantage of our approach is that the use of existing scanners and laptops, which may already be present in even the most remote facilities because they were purchased by consumers for their personal use, could allow for an effectively “zerocost” analyzer, compared with analyzer costs for conventional POC devices ranging from $800 to $15 000. In addition to the prohibitively high cost of the analyzers, conventional POC devices often require proprietary disposable supplies with per-test costs ranging from $0.02 to $3.67 (19 ). In contrast, the components required to perform our paper-based [Hb] assay are available generically at an estimated combined cost of ⬍$0.007 (0.7 US cents) per test. In summary, we developed and validated a paperbased, point-of-care Hb assay for the low-cost assessment of [Hb]. This assay represents a major step toward effective intraoperative care as well as [Hb] testing at the bedside or in urgent care settings where traditional methods of the clinical laboratory are unavailable or prohibitively expensive. The paperbased Hb assay will be useful for diagnosing anemia in resource-limited settings of low-income develop-

ing countries (e.g., sub-Saharan Africa), particularly in the context of malaria, HIV, and sickle cell disease.

Author Contributions: All authors confirmed they have contributed to the intellectual content of this paper and have met the following 3 requirements: (a) significant contributions to the conception and design, acquisition of data, or analysis and interpretation of data; (b) drafting or revising the article for intellectual content; and (c) final approval of the published article. Authors’ Disclosures or Potential Conflicts of Interest: Upon manuscript submission, all authors completed the author disclosure form. Disclosures and/or potential conflicts of interest: Employment or Leadership: None declared. Consultant or Advisory Role: J. Kanter, ApoPharma, Adventrix Pharmaceuticals. Stock Ownership: None declared. Honoraria: None declared. Research Funding: This work was supported in part by a 2012 NIH Director’s Transformative Research Award (NHLBI R01HL117329, PI: S.S. Shevkoplyas). Expert Testimony: None declared. Patents: X. Yang, 61/692,994, 61/558,009; N.Z. Piety, 61/692,994; J. Kanter, 61/692,994; S.S. Shevkoplyas, 61/692,994, 61/558,009. Role of Sponsor: The funding organizations played no role in the design of study, choice of enrolled patients, review and interpretation of data, or preparation or approval of manuscript.

References 1. Pagana KD, Pagana TJ. Mosby’s manual of diagnostic and laboratory tests. Third edition. Philadelphia, PA: Mosby, 2010. 2. Shander A. Anemia in the critically ill. Crit Care Clin 2004;20:159 –78. 3. Aspuru K, Villa C, Bermejo F, Herrero P, Lopez SG. Optimal management of iron deficiency anemia due to poor dietary intake. Int J Gen Med 2011; 4:741–50. 4. McLean E, Cogswell M, Egli I, Wojdyla D, de Benoist B. Worldwide prevalence of anaemia, WHO vitamin and mineral nutrition information system, 1993–2005. Public Health Nutr 2009;12: 444 –54. 5. Weiskopf RB, Viele MK, Feiner J, Kelley S, Lieberman J, Noorani M, et al. Human cardiovascular and metabolic response to acute, severe isovolemic anemia. JAMA 1998;279:217–21. 6. Carson JL, Carless PA, Hebert PC. Outcomes using lower vs higher hemoglobin thresholds for red blood cell transfusion. JAMA 2013;309:83– 4. 7. Aitchison R, Russell N. Smoking: a major cause of polycythemia. J R Soc Med 1988;81:431. 8. Horstman D, Weiskopf R, Jackson RE. Work capacity during 3-wk sojourn at 4,300 m: effects of relative polycythemia. J Appl Physiol 1980;49: 311– 8. 9. Marchioli R, Finazzi G, Specchia G, Cacciola R, Cavazzina R, Cilloni D, et al. Cardiovascular events and intensity of treatment in polycythemia vera. N Engl J Med 2013;368:22–33. 10. MacNee W. Pathophysiology of cor pulmonale in

1512 Clinical Chemistry 59:10 (2013)

11.

12.

13.

14.

15.

16.

17.

18.

chronic obstructive pulmonary disease. Part one. Am J Respir Crit Care Med 1994;150:833–52. Vicari AM, Ponzoni M, Alberetto M, Martani C, Pontiroli AE, Folli F. Erythrocytosis in a patient with chronic obstructive pulmonary disease. Haematologica 1998;83:183– 6. Tefferi A. Polycythemia vera: a comprehensive review and clinical recommendations. Mayo Clin Proc 2003;78:174 –94. Lowe GD, Lee AJ, Rumley A, Price JF, Fowkes FG. Blood viscosity and risk of cardiovascular events: the Edinburgh Artery Study. Br J Haematol 1997; 96:168 –73. Myers GJ, Browne J. Point of care hematocrit and hemoglobin in cardiac surgery: a review. Perfusion 2007;22:179 – 83. Gehring H, Hornberger C, Dibbelt L, Rothsigkeit A, Gerlach K, Schumacher J, Schmucker P. Accuracy of point-of-care-testing (POCT) for determining hemoglobin concentrations. Acta Anaesthesiol Scand 2002;46:980 – 6. Price CP, Kricka LJ. Improving healthcare accessibility through point-of-care technologies. Clin Chem 2007;53:1665–75. Spielmann N, Mauch J, Madjdpour C, Schmugge M, Weiss M, Haas T. Accuracy and precision of hemoglobin point-of-care testing during major pediatric surgery. Int J Lab Hematol 2012;34:86 – 90. Tudos AJ, Besselink GJ, Schasfoort RB. Trends in miniaturized total analysis systems for point-ofcare testing in clinical chemistry. Lab Chip 2001;

1:83–95. 19. Patel KP, Hay GW, Cheteri MK, Holt DW. Hemoglobin test result variability and cost analysis of eight different analyzers during open heart surgery. J Extra Corpor Technol 2007;39:10 –7. 20. Petti CA, Polage CR, Quinn TC, Ronald AR, Sande MA. Laboratory medicine in Africa: a barrier to effective health care. Clin Infect Dis 2006;42: 377– 82. 21. Carneiro IA, Drakeley CJ, Owusu-Agyei S, Mmbando B, Chandramohan D. Haemoglobin and haematocrit: is the threefold conversion valid for assessing anaemia in malaria-endemic settings? Malar J 2007;6:67. 22. Stott GJ, Lewis SM. A simple and reliable method for estimating haemoglobin. Bull World Health Organ 1995;73:369 –73. 23. Quinto L, Aponte JJ, Menendez C, Sacarlal J, Aide P, Espasa M, et al. Relationship between haemoglobin and haematocrit in the definition of anaemia. Trop Med Int Health 2006;11:1295–302. 24. Paddle JJ. Evaluation of the haemoglobin colour scale and comparison with the Hemocue haemoglobin assay. Bull World Health Organ 2002;80: 813– 6. 25. Allain JP, Anokwa M, Casbard A, Owusu-Ofori S, Dennis-Antwi J. Sociology and behaviour of west African blood donors: the impact of religion on human immunodeficiency virus infection. Vox Sang 2004;87:233– 40. 26. Yang X, Forouzan O, Brown TP, Shevkoplyas SS. Integrated separation of blood plasma from

Paper-Based Hb Test for Resource-Limited Settings

27.

28.

29.

30.

31.

whole blood for microfluidic paper-based analytical devices. Lab Chip 2012;12:274 – 80. Carrilho E, Martinez AW, Whitesides GM. Understanding wax printing: a simple micropatterning process for paper-based microfluidics. Anal Chem 2009;81:7091–5. Yang X, Kanter J, Piety NZ, Benton MS, Vignes SM, Shevkoplyas SS. A simple, rapid, low-cost diagnostic test for sickle cell disease. Lab Chip 2013;13:1464 –7. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307–10. Khan MS, Thouas G, Shen W, Whyte G, Garnier G. Paper diagnostic for instantaneous blood typing. Anal Chem 2010;82:4158 – 64. Fu E, Ramsey SA, Kauffman P, Lutz B, Yager P. Transport in two-dimensional paper networks. Microfluid Nanofluid 2011;10:29 –35.

32. Deegan RD, Bakajin O, Dupont TF, Huber G, Nagel SR, Witten TA. Capillary flow as the cause of ring stains from dried liquid drops. Nature 1997;389:827–9. 33. Bates I, Maitland K. Are laboratory services coming of age in sub-Saharan Africa? Clin Infect Dis 2006;42:383– 4. 34. English M, Esamai F, Wasunna A, Were F, Ogutu B, Wamae A, et al. Assessment of inpatient paediatric care in first referral level hospitals in 13 districts in Kenya. Lancet 2004;363:1948 –53. 35. Coppola MA, Klotz KL, Kim KA, Cho HY, Kang J, Shetty J, et al. Spermcheck fertility, an immunodiagnostic home test that detects normozoospermia and severe oligozoospermia. Hum Reprod 2010;25:853– 61. 36. Lenters-Westra E, Slingerland RJ. Six of eight hemoglobin A1c point-of-care instruments do not meet the general accepted analytical perfor-

mance criteria. Clin Chem 2010;56:44 –52. 37. Gomez-Simon A, Navarro-Nunez L, PerezCeballos E, Lozano ML, Candela MJ, Cascales A, et al. Evaluation of four rapid methods for hemoglobin screening of whole blood donors in mobile collection settings. Transfus Apher Sci 2007;36: 235– 42. 38. Bond M, Elguea C, Yan JS, Pawlowski M, Williams J, Wahed A, et al. Chromatography paper as a low-cost medium for accurate spectrophotometric assessment of blood hemoglobin concentration. Lab Chip 2013;13:2381– 8. 39. Kim D-S, Choi J-H, Nam M-H, Yang J-W, Pak JJ, Seo S. LED and CMOS image sensor based hemoglobin concentration measurement technique. Sensors Actuators B Chem 2011;157:103–9. 40. Yager P, Domingo GJ, Gerdes J. Point-of-care diagnostics for global health. Annu Rev Biomed Eng 2008;10:107– 44.

Clinical Chemistry 59:10 (2013) 1513