Solvent Purification System Standard Operating Procedure (SOP)
Recommend Documents
**NOTE** If you are using the solvent system to obtain solvent in the glove box, skip to section 2 - Glove Box SOP Step 1. Step 1 - Check the nitrogen pressure.
â³PCR templates must be added in âDNA extraction laboratoryâ! Diagnostic .... The agarose gel is then allowed to cool down to room temperature. Then spacer ...
Jember. Buku ini berisikan SOP Sistem Peringatan Dini secara umum, untuk .....
Berkoordinasi dengan Polres, Satpol PP dan ... Melakukan patroli keliling. 5.
Jul 2, 2014 ... Circular saws are used for straight sawing. Depending ... Do not operate a
circular saw with the guard wired back, or with the guard removed.
Nov 10, 2015 - Respiratory secretions, pus, fluid and wound swab culture. ⢠Plate the neat sample (â¥10µl) or wound
Standard Operating Procedure (SOP) Miter Saw ... In simple terms a rip cut is one that is made parallel to the grain and is usually done with a table saw, ...
The JGI annotation process for fungal genomes uses an automated .... In addition, regions of genomic conservation determined by VISTA alignments [Ratnere &.
May 15, 2001 - STANDARD OPERATING PROTOCOL (SOP). Subject: Total Phenolic Contents in Green tea, grape products echiancea Products, et al., by UV ...
produced by PT PJC is sold to the market in three types of packing, one of then is the drum ... used by the company because it has several advantages, such as more flexible movement of material ..... Surabaya) (Written in Bahasa Indonesia).
EMR Design. Karanvir Singh. 25. 09. It For Healthcare Delivery and Management: The Concept of Health Account Scheme-Has,. a Web Based Application.
2859-1, i.e. HRN ISO 2859-1: «Sampling procedures for the inspection by ...
Sampling plans are indexed according to acceptable quality level (AQL) for the
lot- ...
Sampling – Sampling for Inspection and Control of Soil Treatment Depth on the.
Projects ... General principles and procedures for inspection and sampling were ...
the International Organization for Standardization (ISO), and those principles and
procedures accepted are published in ISO 2859-1. ISO procedures applied to ...
Comprehensive care, management and treatment. DBS. Dried blood spots. DNA
... taking blood from infants for the HIV DNA PCR test. This test is routine care for
babies ... Two types of blood samples can be used for an HIV. DNA PCR test: 1.
Nov 5, 2011 - 5. homogenize tissue (add a few ml of PBS) using 5ml syringe with big needle. (18-20G) (add some PBS). 6. 5 min 1500 rpm (=280g) at RT ...
Nov 5, 2011 - 3. prepare spinal cord (sc) /brain â transfer into tube filled with PBS ... this solution is considered as 100% and 30%, 37%, and 70% Percoll are.
NASA Information Technology Requirement. (NITR) Procedures ... This standard
operating procedure (SOP) explains the procedure for drafting, reviewing, and.
Mar 17, 2016 - Pitch New Works provides a forum for creators and arts ... Dance Group and Laurie Anderson (United States
Altamont Lodge Ltd. Nominated Occupier Booking Rules & Procedures. General.
The “nominated occupier” against a share in Altamont, or their partner is ...
The purpose of this standard operating procedure (SOP) is to provide guidance
to .... Includes documents such as the following examples: 1. Letter. 2. Memo. 3.
Nov 17, 2007 - Email: [email protected] http://www.crlcefas.org ..... Insert the knife in between the shells of the animal and separate the shells with a twisting ...
Jan 8, 2010 - Standard Operating Procedure (SOP). Surface Area & Pore Size Distribution Analyses. Name: BET - Autosorb-1C. Model #: AX1C-MP-LP.
1r'ralrtu penyusunan SOP bergantung pada kompleksitas bencana, kompleksitas
kegiatan mitigasi, ketersediaan data pendukung, ketersediaan SDM ...
Jan 8, 2010 - volume â 4L, and write Biswas on the log notebook). 2. Fill Cold Trap ... Choose appropriate cell size for sample (6, 9 or 12mm). * We only have ...
Solvent Purification System Standard Operating Procedure (SOP)
**NOTE** If no solvent is being dispensed then check the top valve (Fig. ... Step 4 - Once finished using the glove box and solvent system, turn the top valve back.
Solvent Purification System Standard Operating Procedure (SOP) Prior to using the solvent purification system (Fig. 1) you must obtain training from one of the instrumentation managers. Current Instrumentation managers: Dr. Wes Chalifoux and Michael Walley **NOTE** If you are using the solvent system to obtain solvent in the glove box, skip to section 2 - Glove Box SOP Step 1. Step 1 - Check the nitrogen pressure. If the tanks are low, inform the solvent system manager. Step 2 - Turn on vacuum pump. Step 3 - Place the braided line for the required solvent into the reaction vessel. Note: If using a 14/20 flask, use the provided 24/40 to 14/20 adapter. Step 4 - Turn the main valve counterclockwise to EVACUATE. Allow the pressure to go down until the pressure stays steady on the pressure gauge (located on the right side of bottom manifold). Turn the main valve clockwise to REFILL. Turn the main valve counterclockwise to EVACUATE. Repeat the evacuation/refill procedure a total of 3 times leaving the flask evacuated. Step 5 - Turn the main valve counterclockwise to DISPENSE. Step 6 - Slowly open the solvent metering valve to dispense the required volume of solvent then close the valve finger tight only. DO NOT OVERTIGHTEN THIS VALVE! **NOTE** If no solvent is being dispensed then check the top valve (Fig. 1); see section 2 - Glove Box SOP (step 3) to see procedure for switching this over. Step 7 - Turn the main valve clockwise to REFILL to allow the flask to be filled with nitrogen and to clear the line of solvent. Be careful that the line does not come out of the reaction flask due to the high nitrogen pressure. This can be done by either holding both the line and the flask together with your hand or by using a keck clip. Step 8 - Turn the main valve to CLOSED. Step 9 - Remove the line and place a septum on the reaction flask as soon as possible. Step 10 - Turn off the vacuum.
Figure 1. Solvent purification system. Valves at the top (red circle) to switch between dispensing from the solvent system braided lines or the glove box bulkhead.
Glove Box Solvent SOP If you have not received training for the Glove box do not set up the solvent system to dispense to the glove box. Step 1 - Check the nitrogen pressure prior to using the solvent purification system. If the tanks are low, inform the solvent system manager. Step 2 - Turn the valve on top of the front solvent purification column with the point going towards the line to the glove box (see Fig. 1). Note: You can easily check this by looking at the stainless steel line and seeing if it goes into the back of the glove box. Step 3 - See glove box SOP for dispensation of solvents in the glove box. Step 4 - Once finished using the glove box and solvent system, turn the top valve back to the correct position to allow for the normal use of the solvent system.