Stresses in the ankle joint and total ankle ... - BDH Medical

34 downloads 120 Views 551KB Size Report
Rahul Kakkar MRCSa,*, M.S. Siddique FRCS(Orth)b a ST6-Trauma .... A B S T R A C T. The ankle is a highly congruent joint with a surface area of 11– 13 cm2.
Foot and Ankle Surgery 17 (2011) 58–63

Contents lists available at ScienceDirect

Foot and Ankle Surgery journal homepage: www.elsevier.com/locate/fas

Review

Stresses in the ankle joint and total ankle replacement design Rahul Kakkar MRCSa,*, M.S. Siddique FRCS(Orth)b a b

ST6-Trauma and Orthopaedics, Newcastle upon Tyne, United Kingdom Consultant, Freeman Hospital, Newcastle upon Tyne, United Kingdom

A R T I C L E I N F O

A B S T R A C T

Article history: Received 19 March 2010 Received in revised form 9 February 2011 Accepted 19 February 2011

The ankle is a highly congruent joint with a surface area of 11–13 cm2. Total ankle replacements have been attempted since the early 1970s and design has continually evolved as the early designs were a failure. This was because the stresses involved and the mutiaxial motion of the ankle has not been understood until recently. It has been shown that the talus slides as well as rolls during the ankle arc of motion from plantarflexion to dorsiflexion. Furthermore, the articular surfaces and the calcaneofibular and tibiocalcaneal ligaments have been shown to form a four bar linkage dictating ankle motion. A new design ankle replacement has been suggested recently which allows multiaxial motion at the ankle while maintaining congruency throughout the arc of motion. The early results of this ankle replacement have been encouraging without any reported failures due to mechanical loosening. ß 2011 European Foot and Ankle Society. Published by Elsevier Ltd. All rights reserved.

Keywords: Total ankle replacement Biomechanics Stresses Failure Ligament balance Four bar linkage

Contents 1. 2.

3. 4.

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . The native ankle joint . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1. Anatomy and stability . . . . . . . . . . . . . . . . . . . . . . . . . 2.1.1. Bony configuration . . . . . . . . . . . . . . . . . . . . 2.1.2. Ligamentous configuration . . . . . . . . . . . . . . 2.2. Biomechanics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.2.1. Mechanical properties of bone and cartilage 2.2.2. Axis and range of motion . . . . . . . . . . . . . . . 2.2.3. Restraints of ankle motion . . . . . . . . . . . . . . 2.2.4. Forces and stresses in the ankle . . . . . . . . . . First generation ankle replacements . . . . . . . . . . . . . . . . . . . . Second generation total ankle replacements . . . . . . . . . . . . . References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

1. Introduction The ankle joint acts as a link between the leg and the foot and plays a role in transferring load from the leg to the foot. The most common pathologies which can disrupt ankle function are osteoarthritis, rheumatoid arthritis and posttraumatic arthritis. Saltzman et al. [40] showed that posttraumatic arthritis was the most common and tends to occur more commonly in younger patients. The classical treatment for ankle arthritis has been * Corresponding author. E-mail address: [email protected] (R. Kakkar).

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

. . . . . . . . . . . . .

58 59 59 59 59 59 59 59 60 60 61 61 63

arthrodesis but total ankle replacement has become popular over the last few decades. The success of the total hip and knee replacements in the 1970s prompted attempts at total ankle replacement (TAR) [12]. The short and intermediate term results of the TARs were quite satisfactory, but the long-term results were unsatisfactory [48]. Hence there was a need to improve the performance as well as the survival rate of the existing TAR by improving upon design, constraint and function. In this article, the biomechanics of the ankle joint is analysed and current designs of total ankle replacement explored to determine if the ideal requirements for ankle replacement have been met.

1268-7731/$ – see front matter ß 2011 European Foot and Ankle Society. Published by Elsevier Ltd. All rights reserved. doi:10.1016/j.fas.2011.02.002

[()TD$FIG]

R. Kakkar, M.S. Siddique / Foot and Ankle Surgery 17 (2011) 58–63

59

2. The native ankle joint 2.1. Anatomy and stability 2.1.1. Bony configuration The ankle consists of three bones (tibia, fibula and talus), collateral and syndesmotic ligaments and is a dynamic and highly congruent joint. The talar body articulates superiorly with the tibial plafond, medially with the medial malleolus and laterally with the lateral malleolus. Inman [24] showed that the talus is not a cylinder, but rather a section of a frustum of a cone, the apex of which is directed medially and thus has a smaller medial and larger lateral radius. The tibial plafond serves as the most superior aspect of the ankle joint and articulates with the dome of the talus. It is concave in the anteroposterior plane and elevated slightly on the medial side. This shape dictates an orientation of the distal tibial plafond that is slightly oblique from distal lateral to proximal medial [24]. Fig. 1. Medial ligaments of the ankle.

2.1.2. Ligamentous configuration The deltoid ligament stabilizes the ankle medially [10]. The contribution of the deltoid ligament to ankle joint contact has been reported by Earll et al. [13]. They report that deltoid ligament sectioning produces the greatest changes in both contact area size (decreased up to 43%) and peak pressure values (increased up to 30%) and this emphasises the fundamental role in ankle mobility played by the deltoid ligament (Fig. 1). The syndesmotic ligaments join the tibia to fibula and consist of an anterior and posterior tibiofibular ligament in addition to the transverse tibiofibular ligament and the interosseous ligament. The anterior talofibular ligament (ATFL), calcaneofibular ligament (CFL), and posterior talofibular ligament (PTFL) all contribute to the lateral stability of the ankle (Fig. 2). The position of the ankle determines the tautness or looseness of the lateral ligaments. The PTFL is maximally stressed and CFL taut in dorsiflexion while ATFL is taut and PTFL, CFL loose in plantarflexion [11,39]. 2.2. Biomechanics 2.2.1. Mechanical properties of bone and cartilage The average thickness of the ankle cartilage is approximately 1.6 mm whereas the thickness of the knee cartilage is 6–8 mm [43]. The yield strain of human trabecular tibial bone is 0.73  0.06% in compression and 0.65  0.05% in tension. The mean E (Young’s modulus) of the tibial cortex was found to be 34.11 GPa [21]. Morgan and Keaveny [36] showed that the Young’s modulus (E) of human trabecular tibial bone was 1091  634 MPa in compression and 1068  634 MPa in tension while Lowery [32] found that the subchondral bone of the distal tibia had an elastic modulus of 300– 450 MPa. After removal of the subchondral plate, compressive resistance was lowered by 30–50%, and with sectioning of the subchondral bone 1 cm proximal to the subchondral plate, by 70– 90%. Hvid et al. [22] found talar bone to be 40% stronger than tibial bone, while Calderale et al. [7] found that removing part of the cortical shell of the talus placed abnormal increased stress on the remaining talar cancellous bone. Therefore, ideally, the talar component of the prosthesis should be anatomically sized, fully cover the talar body, and have a wide support on talar neck. Thus, as much superficial subchondral talar bone should be saved as possible, particularly the anterior part of the talar body and talar neck. Hvid et al. [22] also found an eccentricity of the area of maximal bone strength of the distal tibia that is posteromedial and not central. This area of stiffer bone could act as a pivot point, with the risk of overloading the surrounding weaker anterolateral bone. To avoid ‘‘off centre’’ forces on a prosthesis and possible collapse of the weak lateral tibia, proper alignment of the prosthesis and adequate ligament balancing of the

[()TD$FIG]

Fig. 2. Lateral ligaments of the ankle [38].

ankle must be achieved. In particular, valgus malalignment should be corrected [20]. 2.2.2. Axis and range of motion The normal range of motion in the ankle ranges from 238 to 568 of plantarflexion, and from 138 to 338 of dorsiflexion [31,41]. Three distinct axes of the ankle joint have been reported during various motions based on the curvature of the talar trochlea [3] with the axis inclined upwards medially during Plantarflexion(PF) and upwards laterally during Dorsiflexion (DF) (Fig. 3). Although the ankle joint axis at the neutral position is often regarded as a single axis, dorsiflexion–plantarflexion hinge, the axis orientation may vary [26]. The axis has been described as a changing axis or changing instant centers of rotation due to the shape of the talar trochlea and the action of soft tissues [5]. In cadaveric and gait studies, the rotation has been shown to range between 108 and 128 [44]. It is the varying center of rotation that allows the talus to glide and slide within the ankle mortise during PF and DF [4,33]. Also, the curvature of the talus and the distal tibia show varying radii [4,50] that allow horizontal rotations to occur in the foot or leg with movements of the ankle and thus transverse plane motion is coupled with sagittal plane motion [33,41]. The axial rotation of the tibia with respect to the talus is reported to be between 68 and 128 [37,44]. Lundberg et al. [33] observed 8.98 of external rotation of the talus as the ankle moved from neutral position to 308 of dorsiflexion, whereas a small amount of internal

[()TD$FIG]

60

[()TD$FIG]

R. Kakkar, M.S. Siddique / Foot and Ankle Surgery 17 (2011) 58–63

Fig. 4. Four bar linkage model developed by Leardini et al. [27].

Fig. 3. Axes of ankle motion.

rotation occurred with plantarflexion from neutral to 108, followed by external rotation at terminal plantarflexion [33]. Michelson and Helgemo [35] reported that dorsiflexion resulted in an average of 7.28  3.88 of external rotation of the foot relative to the leg with ankle dorsiflexion, and 1.98  4.128 of internal rotation with plantarflexion. Leardini et al. [27] developed a mathematical model to explain the multiaxial motion of the ankle in the sagittal plane. A four-bar linkage model was described (Fig. 4) showing the talus/calcaneus and tibia/fibula rotating about one another on inextensible line segments that represent the calcaneofibular (line AB in Fig. 4) and tibiocalcaneal (deltoid) (line DC in Fig. 4) ligaments without resistance. Motion between the polycentric, polyradial trochlea consisted of a combination of ‘‘rolling’’ and ‘‘sliding’’ motions. In this model, rotation was dictated by the most anterior fibers of the anterior talofibular and calcaneofibular ligaments. Leardini [28] also showed that these specific fiber bundles were isometric through the range of sagittal motion of the ankle. The instant centre of rotation (shown by ‘‘star’’ in Fig. 4) translated from a posteroinferior to a superoanterior position, which is consistent with several studies that suggest that the ankle is congruent and rotates about a transient centre of rotation [42]. 2.2.3. Restraints of ankle motion The tibiotalar articulating surface contributes to 70% of the antero-posterior stability, 50% of inversion/eversion stability and 30% of internal/external rotation stability [47]. The rest of the stability is provided by the ligaments. Renstrom et al. [39] found that during various motions of the ankle joint, the anterior talofibular and calcaneofibular ligaments were synergistic, that is when one ligament is strained the other one is relaxed and vice versa and thus providing stability.

2.2.4. Forces and stresses in the ankle The ankle has a load bearing surface area of 11–13 cm2 [45] The tibiotalar area, however, accounts for only approximately 7 cm2 [45] while Calhoun et al. [8] found that during weight bearing 77–90% of the load is transmitted through the tibial plafond to the talar dome. With an interface area of 7 cm2, the average compressive load per unit area at the interface during gait would be approximately 3.5 MPa in a patient of 700 N body weight [20]. A vertical load on the ankle of 5.2 times body weight was found during gait [45]. The peak resultant force acting at the ankle joint during the stance phase during running was 9.0–13.3 BW [6], Landing from a jump generated 2–12 BW [34] while heel-toe running at 4.5 m/s generated a force of 2.8 BW [9,15]. In diseased ankles, the joint load decreased to approximately three times body weight; the same values were noted in replaced ankles [45] and anteroposterior and lateral shear forces during gait were estimated to reach levels of two and three times body weight, respectively. The vertical load that is transmitted to the trabecular bone at the prosthesis-bone interface may exceed the inherent trabecular bone strength in normal daily activities [20]. Having knowledge of the normal ankle joint, a discussion can now be carried about the various Total Ankle Replacement designs.

[()TD$FIG]

Fig. 5. First generation ankle replacement, which consisted of all polyethylene tibial component and metallic talar component (similar to hip replacement systems) fixed with cement.

[()TD$FIG]

R. Kakkar, M.S. Siddique / Foot and Ankle Surgery 17 (2011) 58–63

61

3. First generation ankle replacements Around twenty three total ankle replacements were developed during early 1970s [2]. The first generation TARs were of two types namely constrained and unconstrained. Constrained prostheses limited motion to the sagittal plane. They were typically spherical, spheroidal, conical, cylindrical or sliding cylindrical in shape [31]. They provided greater stability [50], minimized impingement of the malleoli against the talus [12], and decreased wear of the polyethylene insert due to larger contacting surfaces. The only disadvantage of the prostheses was increased stresses at the bone cement implant interfaces leading to early failure. Examples of this prostheses type were Imperial College London Hospital (ICLH), Conaxial, St. George/Buchholz, Thompson Parkridge Richards (TPR), TNK and Mayo designs. Unconstrained prostheses were typically trochlear, bispherical, concaveconvex, and, convexconvex in shape [31]. They provided improved range of motion but stability was greatly reduced. They also minimized the bone cement implant interface forces. Examples of unconstrained prostheses were Bath and Wessex, Irvine, Smith and Newton implants. Both the constrained and unconstrained prostheses were cement fixed (Fig. 5) during the 1970s [50]. The results of unconstrained prostheses proved to be a total failure due to instability and impingement [14].

4. Second generation total ankle replacements The second generation prostheses were developed to overcome the disadvantages and complications associated with the first generation prostheses [46]. These implants were intended to reproduce the anatomical characteristics of the ankle joint, joint kinematics, ligament stability and mechanical alignment. The sliding and rotational motions of the ankle joint were achieved by the two or three component designs (Fig. 6) [18]. Agility ankle was the prototype of the two component semiconstrained design. It replaced the medial, lateral and superior articulating surfaces of the ankle joint as fusion of the distal tibiofibular syndesmosis was carried out to stabilize the tibial component. The articular surface of the tibial component was larger than the surface of the talar component. The incongruency between the talar and tibial articulations theoretically allowed sliding as well as the rotational motions that supposedly mimic [()TD$FIG] ankle kinematics [40] (Fig. 7).

Fig. 6. Second generation ankle replacement consists of metallic tibial and talar components, stabilized with or without cement. A ‘meniscus-like’ polyethylene component is either fixed to the tibial component (A) or is mobile articulating with both components (B).

Fig. 7. The agility prosthesis, a two-component design, requires tibio-fibular fixation.

It was not successful as this semiconstrained design does not replicate normal ankle motion, as the ankle ‘slides’ from side to side during rotation and dorsi- and plantarflexion motion. This raised concerns regarding increased wear debris production and early loosening [18]. Furthermore, a non-union of the attempted tibio-fibular arthrodesis, risks loss of fixation on the proximal side, thus compromising stability [17]. Examples of three component total ankle replacements are: Scandinavian total ankle replacement (STAR) has a convex talar component with a longitudinal ridge, which is congruent with the distal surface of the mobile meniscus. Dorsi- and plantarflexion at the menisco–talar interface, but no talar tilt, are allowed while rotation is allowed at the (flat) menisco–tibial interface [20]. This presents a lower surface area for stress distribution in the distal tibia compared to some other ankle replacement designs but might also reduce stress shielding from a stem [17] (Fig. 8). The concerns regarding the small dimension of the tibial component and the lack of circumferential bone support makes the tibial component prone to subsidence (sinking) in the distal tibia cancellous bone. Furthermore, motion only in one axis, due to the non-anatomic cylindrical shape of the talus, may produce load transfer to the medial side and overstress the medial ligaments and capsule [20]. Valderabano et al. [49] reported tilting of the tibial component in 9 of 68 ankles (13%) within the first 3 months, without progression thereafter. This can be the result of shear forces and imperfect apposition of tibial prosthesis and bone. Ramses total ankle replacement: required wide talar bone resection therefore subsidence was a problem and frontal plane stability relies entirely on the medial and lateral ligamentous structures. This does not correspond to normal ankle biomechanics, where frontal plane stability relies mainly on joint congruency [20]. Hintegra ankle replacement: it consists of a flat tibial component, a polyethylene inlay and a convex conic talar component with a smaller medial radius. Both the talar and tibial components have ventral shields for screw fixation [20] (Fig. 9). The disadvantages are that it uses screws to obtain fixation in tibia which can loosen before bony ingrowth has occurred due to micromotion and inversion–eversion and rotational motion is not allowed, possibly causing higher stresses at the bone–implant interface [18]. Buechel Pappas low contact stress (LCS) TAR: its upper surface is flat, whereas its lower surface conforms to the trochlear surface, thereby providing unconstrained, sliding cylindrical motion with

[()TD$FIG]

62

[()TD$FIG]

R. Kakkar, M.S. Siddique / Foot and Ankle Surgery 17 (2011) 58–63

Fig. 8. The STAR prosthesis a three component design but no inversion/eversion is allowed.

Fig. 10. The tibial stem and the deep sulcus of the talar component accommodating a. matching polyethylene surface, allowing inversion/eversion motion, are characteristic features of the Buechel–Pappas ankle replacement.

LCS on the bearing surfaces, allowing inversion, eversion motion [5] (Fig. 10). The main concern regarding the Buechel Pappas implant and all implants having a tibial stem (SALTO, AES, Mobility) to facilitate fixation is weakening of the anterior tibial cortex, as an anterior cortical window has to be produced for insertion of the component. Load transmission in the weakened anterior cortex in the supramalleolar area is a concern as the removal of the subchondral plate lowers the compressive resistance by 30–50%, and with sectioning of the subchondral bone 1 cm proximal to the subchondral plate, by 70–90% [32] and therefore early failures are a risk of this prosthesis. Furthermore, stress shielding is possible due to the stem as compared to other designs without stems [17]. But, a broad comparison of their outcomes reveals that the results of most of the existing total ankle replacement models are unsatisfactory because of the inability to restore adequately the critical stabilizing role of the ligaments, poor reproduction of the normal mechanics of the ankle joint and the lack of involvement of the underlying subtalar joint in the coupled pattern of motion of the entire ankle complex [19,16]. In order to overcome the unsatisfactory outcomes of the existing TAR models, a novel TAR model was required which would ideally not have a tibial stem and

mimic the mutiaxial motion of ankle while maintaining congruency throughout the range of motion. Leardini et al. [27,28] had developed the four bar linkage system and using computer animation demonstrated that rolling as well as sliding occurred in ankle motion, that is the talus slided forward on the tibial mortise while rolling backwards during plantarflexion and vice versa in dorsiflexion. From the model it was deduced that the shape of the articulation surface compatible with the ligament rotation during various ankle motions was an arc of a circle which is polycentric and polyradial in nature. The demonstrated mutual action of the articular surfaces and the ligaments in the intact ankle joint was used to enhance the total ankle replacement design with the help of computer aided design [28]. A new design convex-tibia fully congruent three-component prosthesis was eventually preferred among all the possible pairs of articular surfaces tested by Leardini and Moschell [29]. This work also pointed out that the flat-tibia plus natural-talus designs [5,25] were unlikely to restore the original characteristic pattern of ligament tensioning and thus likely to fail because of abnormal stresses. As a result the BOX (Bologna Oxford (BOX)) total ankle replacement was developed as shown in Fig. 11. This Ankle replacement has a convex tibial component with bars to prevent

[()TD$FIG] [()TD$FIG]

Fig. 9. Hintegra uses screw fixation which can loosen before bony ingrowth occurs.

Fig. 11. The BOX ankle replacement (figure from [1]).

[()TD$FIG]

R. Kakkar, M.S. Siddique / Foot and Ankle Surgery 17 (2011) 58–63

Fig. 12. Model depicting congruency of the BOX prosthesis throughout the motion arc (figure modified from [30]).

stress shielding, which articulates with a biconcave meniscal bearing with a groove on the talar component to provide mediolateral stability. The shapes of the three components are compatible with physiologic ankle mobility and with the natural role of the ligaments. This prosthesis maintains full congruence at the articulating surfaces of the meniscal bearing over the entire motion arc (Fig. 12). There have been pilot studies done and initial results have shown no failures due to mechanical loosening [30]. Ingrosso et al. [23] have shown that the BOX prosthesis seems to contribute to an early functional recovery at 6 months which is maintained at one year. With reduction of pain and recovery of joint control and gait, variables of high clinical interest, such as stance balance and ability in propulsion, improve considerably. It remains to be seen whether the BOX meets the ideal requirements for a total ankle replacement and becomes successful. The search for the ideal ankle replacement continues. References [1] Affatato S, Leardini A, Leardini W, Giannini S, Viceconti M. Meniscal wear at a three component total ankle prosthesis by a Q4 knee joint simulator. J Biomech 2007;40:1871–6. [2] Alvine FG. Total ankle arthroplasty. In: Myerson MS, editor. Foot and ankle disorders vol. 2. Philadelphia: Saunders; 2000. p. 1087. [3] Barnett CH, Napier J. The axis of rotation at the ankle joint in man, its influence upon the talus and mobility of the fibula. J Anat 1952;86:98. [4] Bauer G, Eberhardt O, Rosenbaum D, Claes L. Total ankle replacement. Review and critical analysis of the current status. Foot Ankle Surg 1996;2(2):119–26. [5] Buechel FF, Pappas MJ, Iorio LJ. New Jersey low contact stress total ankle replacement: Biomechanical Rationale and Review of 23 cementless cases. Foot Ankle 1988;8(6):279–90. [6] Burdett RG. Forces predicted at the ankle during running. Med Sci Sports Exerc 1982;14(4):308–16. [7] Calderale PM, Garro A, Barbiero R, Fasiolo G, Pipino F. Biomechanical design of the total ankle prosthesis. Eng Med 1983;12:69–80. [8] Calhoun JH, Li F, Ledbetter BR, Viegas SF. A comprehensive study of pressure distribution in the ankle joint with inversion and eversion. Foot Ankle Int 1994;15:125–33. [9] Cavanagh PR, Lafortune MA. Ground reaction forces in distance running. J Biomech 1980;13(5):397–406. [10] Close JR. Some applications of the functional anatomy of the ankle joint. J Bone Joint Surg Am 1956;38:761–81. [11] Colville MR, Marder RA, Boyle JJ, Zarins B. Strain measurement in lateral ankle ligaments. Am J Sports Med 1990;18:196–200. [12] Conti SF, Wong YS. Complications of total ankle replacement. Clin Orthop 2001;391:105–14. [13] Earll M, Wayne J, Brodrick C, Vokshoor A, Adelaar R. Contribution of the deltoid ligament to ankle joint contact characteristics: a cadaver study. Foot Ankle 1996;17(6):317–24. [14] Evanski PM, Waugh TR. Management of arthrodesis of the ankle: an alternative to arthrodesis. Clin Orthop 1977;122:110–5.

63

[15] Frederick EC, Hagy JL. Factors affecting peak vertical ground reaction forces in running. Int J Sport Biomech 1986;2:41–9. [16] Giannini S, Leardini A, O’Connor JJ. Total ankle replacement: review of the designs and of the current status. Foot Ankle Surg 2000;6:77–88. [17] Gill LH. Challenges in total ankle arthroplasty. Foot Ankle Int 2004;25(4): p.195–207. [18] Gougoulias N, Khanna A, Maffuli N. History and evolution in total ankle arthroplasty. Br Med Bull 2009;89(1):111–51. [19] Guyer AJ, Richardson EG. Current concepts review: total ankle arthroplasty. Foot Ankle Int 2008;29:256–64. [20] Hintermann B. Total ankle arthroplasty: historical overview current concepts and future perspectives. Wien, New York: Springer; 2004. pp. 25–42. [21] Hunt KD, Dean O’Loughlin V, Fitting DW, Adler L. Ultrasonic determination of the elastic modulus of human cortical bone. Med Biol Eng Comput 1998;36(1): 51–6. [22] Hvid I, Rasmussen O, Jensen NC, Neilsen S. Trabecular bone strength profiles at the ankle joint. Clin Orthop 1985;199:306–12. [23] Ingrosso S, Benedetti MG, Leardini A, Casanelli S, Giannini S. Gait analysis of a novel design of ankle replacement. J Foot Ankle Res 2008;1(Suppl. 1):P1. [24] Inman VT. The joints of the ankle, 2nd ed., Baltimore: Williams & Wilkins; 1991. pp. 31–74. [25] Kofoed HA. A new total ankle joint prosthesis. In: Kossowky R, Kossowky N, editors. Material sciences and implant orthopaedic surgery. Dordrecht: Maxtinus Nijhoff; 1986. p. 75–84. [26] Kofoed H. Current status of ankle arthroplasty. Berlin: Springer; 1998. p. 157. [27] Leardini A, O’Connor JJ, Catani F, Giannini S. A geometric model of the human ankle joint. J Biomech 1999;32:585–91. [28] Leardini A, Catani F, Giannini S, O’Connor JJ. Computer-assisted design of the sagittal shapes for a novel total article replacement’. Med Biol Eng Comput 2001;39:168–75. [29] Leardini A, Moschell D. Dynamic simulation of the natural and replaced human ankle joint. Med Biol Eng Comput 2002;40:193–9. [30] Leardini A, Catani F, Romanogli M, Digennaro A, Sarti D, Giannini S. A novel total ankle replacement: from biomechanical research to clinical outcome. J Biomech 2007;40(S2):S93. [31] Lewis G. The ankle joint prosthetic replacement: clinical performance and research challenges. Foot Ankle Int 1994;15(9):471–6. [32] Lowery RB. Fractures of the talus and os calcis. Opin Orthop 1995;6:25–34. [33] Lundberg A, Svensson OK, Nemeth G, Selvik G. The axis of rotation of the ankle joint. J Bone Joint Surg Br 1989;71(1):94–9. [34] McNair PJ, Prapavessis H. Normative data of vertical ground reaction forces during landing from a jump. J Sci Med Sport 1999;2(1):86–8. [35] Michelson JD, Helgemo S. Kinematics of the axially loaded ankle. Foot Ankle Int 1995;16:577–82. [36] Morgan EF, Keaveny TM. Dependence of yield strain of human trabecular bone on anatomic site. J Biomech 2001;34:569–77. [37] Murray MP, Drought AB, Kory RC. Walking patterns of normal men. J Bone Joint Surg Am 1964;46:335–60. [38] Netter FH. Atlas of human anatomy. Philadelphia, PA: Saunders; 2006. p. 514. [39] Renstrom P, Wertz M, Incavo S, Pope M, Ostgaard HC, Arms S, et al. Strain in the lateral ligaments of the ankle. Foot Ankle 1988;9(2):59–63. [40] Saltzman CL, McIff TE, Buckwalter JA, Brown TD. Total ankle replacement revisited. Journal of Orthopaedic & Sports Physical Therapy 2000;30(2): 56–67. [41] Sammarco GJ, Burstein AH, Frankel VH. Biomechanics of the ankle: a kinematic study. Orthop Clin North Am 1973;4:75–96. [42] Siegler S, Chen J, Schneck CD. The three-dimensional kinematics and flexibility characteristics of the human ankle and subtalar joint. J Biomech Eng 1988;110:364–73. [43] Shepherd DE, Seedhom BB. Thickness of human articular cartilage in joints of the lower limb. Ann Rheum Dis 1999;58(1):27–34. [44] Stauffer RN. Total joint arthroplasty. The ankle. Mayo Clin Proc 1979;54(9): 570–5. [45] Stauffer RN, Chao EYS, Brewster RC. Force and motion analysis of the normal, diseased, and prosthetic ankle joint. Clin Orthop 1977;127:189–96. [46] Thomas RH, Timothy RD. Ankle arthritis. J Bone Joint Surg Am 2003;85: 923–36. [47] Tochigi Y, Rudert MJ, Saltzman CL, Amendola A, Brown TD. Contribution of articular surface geometry to ankle stabilization. J Bone Joint Surg Am 2006;88:2704–13. [48] Tooms RE. Arthroplasty of ankle and knee. In: Crenshaw AH, editor. Campbell’s operative orthopedics. St. Lois: C.V. Mosby Company; 1987 . p. 1145– 50. [49] Valderrabano V, Hintermann B, Dick W. Scandinavian total ankle replacement: A 3.7year average followup of 65 patients. Clin Orthop Relat Res 2004;424: 47–56. [50] Waugh TR, Evanski PM, McMaster WC. Irvine ankle arthroplasty. Prosthetic design and surgical technique. Clin Orthop 1976;114:180–4.