Garry E. Gold1, Daniel R.Thedens2, John M. Pauly2, Klaus R Fechner3, Gabrielle Bergman1, Christopher F.Beaulieu1, Albert Macovski2 ..... Burton-Wurster.
Technical MR Imaging ofArticular Methods Using UkrashortTEs
Cartilage
of the
Garry E. Gold1, Daniel R. Thedens2, John M. Pauly2, Klaus R Fechner3, Gabrielle
T
he T2 relaxation decreases
out
of cartilage and
calci-
fled zones near the bone-cartilage [1, 2]. T2 relaxation times in these
interface zones
time
in the radial
can
10 msec
be
all zones
or less [1, 2]. Through-
of
cartilage,
short T2 relaxation
times
components
with
may be important
in
early detection of cartilage abnormalities [3]. We have developed two methods for highresolution,
ultrashort
cartilage. scopic
data
The second space (3D)
TE imaging
The first technique
spectro-
with high-resolution images. technique uses a non-Cartesian K-
along
trajectory image
of articular
acquires
to acquire
in about
a three-dimensional
5 mm.
Both
ultrashort TEs (to obtain signal the short-T2 components of
methods intensity cartilage)
use from and
shift artispectro-
lipid suppression to avoid chemical facts. The pmjection-reconstruction
scopic imaging (PRSI) technique is best suited for further evaluation of a focal region of cartilage abnormality. quence is useful entire joint
knee
whereas the 3D cone seeither for a survey of the
or for high-resolution
imaging
of a
compartment.
Materials Rationale
and
Methods
and Theory
PRSI.-The projection-reconstniction method used here has been applied to study atherosclerotic
Bergman1,
plaques [4] and tendons and menisci [5]. We used cither a half-pulse excitation, in which two excitations are combined to form a slice [4], or a conventional excitation
pulse.
The excitation
is immediately
fol-
lowed by pmjection-reconstruction spectroscopic readout gradients. After excitation, an oscillating readout gradient repeatedly scans one diameter in Kspace. This diameter rotates through 2it radians over the course of the scan, covering an entire cylinder in kk,s space. Increasing time delays are used to fill in the spectroscopic data. Reconstruction of the resultant PRSI data set uses a gridding algorithm [61. The size of the data set is 33 MB per slice. Reconstruction took approximately 6 mm per slice on an U1traSPARC workstation (Sun Microsystems, Mountain View, CA). Three-dimensional cones-The 3D cone imaging sequence is based on a non-Cartesian cone Kspace
trajectory
[7].
The
design
is similar
to a 3D
more efficient in 3D K-space coverage. A series of nested cones is overlaid with interleaved spirals. resulting in spiral readouts in the ks-k., plane and sinusoidal readouts in the k5-k and k-k planes. The 3D cone sequence provides isotropic resolution in all three dimensions. Because readout began from the center of K-space, an ultrashort ‘FE was possible (Thedens DR et al., presented at the International Society for Magnetic Resonance in Medicine meeting. April 1996). The 3D cone method used either chemical saturation to reduce lipid signal intensity or a spectrally and spatially selective excitation [8, 9]. A gridding algorithm [6] is used for reconstruction of the 3D cone data set. Reconstruction of the projection-reconstruction
technique
Innovation
but
Knee:
Christopher
12-MB
data
New
F. Beaulieu1, Albert Macovski2
set took 6 rain on an U1traSPARC
workstation. Patient
Information
Ten healthy volunteers and 10 patients with knee cartilage damage were examined. Patients with cartilage damage were recruited from a group under
consideration
for
autologous
chondrocyte
All scans were obtained with version 5.5 software on 1.5-1 Signa whole-body imaging systems (General Electric Medical Systems, Milwaukee, WI) using either the extremity coil for the whole-knee images or the 3-inch (7.6-cm) surface coil for images of the patellofemoral joint. Four scans were obtained for each healthy volunteer: PRSI, half-pulse excitation, TB = 200 psec; PRSI, conventional excitation, TE = 1.6 msec; 3D cone, chemical saturation, TE = 0.6 msec; and 3D cone, spectral-spatial excitation, TB = 6.6 msec. Three scans were obtained for each patient: 3D fat-suppressed gradient-recalled echo (3D-FSGRE), TB = 12 msec; PRSI, half-pulse excitation, TB = 200 Jisec; and PRSI, conventional excitation, TE= 1.6msec. transplantation.
Imaging Techniques
3D-FS-GRE.-The 3D-FS-GRE scan was performed on the high-speed gradient system with a maximum gradient amplitude of 2.2 G/cm and a maximum slew rate of 12 0/cm . jp1 ‘fl 3j.. FS-GRE technique used a field ofview (FOV) of 16 cm, with a512 x l92matrix and a2-mm slice thickness. TRIFE was 50/12 msec, and imaging time was 6 mm for 28 sections. The flip angle was 20g.
Received March 13, 1997; accepted after revision October 20, 1997. Presented atthe annual meeting ofthe American Roentgen Ray Society. Boston, May 1997. Winner ofthe President’s Award. 1 Department of Radiology, Stanford University, Stanford, CA 94305-5488.Address correspondence to 6. E. Gold. 2Department of Electrical Engineering, Stanford University, Stanford. CA 94305.
3VA Health Care System of Palo Alto, Stanford, CA 94305. AJR1998;170:1223-1226
AJR:170, May 1998
0361-803X198/1705-1223
©American
Roentgen Ray Society
1223
Gold
PRSI.-The PRSI technique was implemented on the high-speed gradient system with a maximum gradient amplitude of 2.2 G/cm and maximum slew rate of I 2 G/cm . msec’ . The PRSI technique used a minimum FOV of 6 cm, with an in-plane spatial resolution of 184 pm. Four interleaves were used to obtain spectroscopic data. Spectral resolution was 63 Hz (1.0 ppm), with a spectral bandwidth of 667 Hz (12.2 ppm). With a TR of 60 msec, imaging time is 4 mm per slice for the conventional excitation (TE = 1.6 msec, 2-mm slice thickness) and 8 mm per slice for the halfpulse excitation (TE = 200 psec, 3-mm slice thickness). The flip angle was 15#{176}. Three-dimensional cone.-The 3D cone technique was implemented on the standard gradient system with a maximum gradient amplitude of 2.0 G/cm and maximum slew rate of I .6 Gkm . msec. The 3D cone technique used a 12.8-cm FOV and a 1.0-mm isotropic resolution for the entire knee, which took 5.1 mm with a TR of 70 msec. A highresolution compartment survey 3D cone scan with 05-mm isotropic resolution and a 6.4-cm FOV took 9.7 mm with a TR of 70 msec. The flip angle was 20#{176}. Chemical saturation was used to reduce lipid signal intensity (TE = 600 lisec)or a spectrally and spatially selective excitation (TE = 6.6 msec). The spectral-spatial excitation was also used to restrict the FOV in one direction.
et al.
Results
the spectra
Images of healthy volunteers showed that PRSI technique can obtain high-resolution
face are also broader maximum), indicating
images
times
of the
patellofemoral
with
the half-pulse
obtained 200
isec,
3-mm
slice
creased signal-to-noise with the conventional msec,
2-mm
frequencies
cal shift
artifacts
Acquisition
the
PRSI
Images
thickness)
thickness).
allowed
data
set with
display
of im-
frequency. images
Water(Fig. from the pa-
cartilage
were
of the same
data
so no additional
ning
was required. was basis.
zones
of healthy
structed
Spectra
at the
tamed. Spectra cartilage-bone
acquired
scan-
from articular
water
voxels
across
cartilage,
frequency,
from the cartilage interface show lower
using time
of 1 mm. was
ob-
near the peak ar-
eas than spectra from the area closer to the joint surface (Fig. IB). The line widths on
cone images of the envolunteers were acquired coil and an isotropic TE was 600
5. 1 mm
(Figs.
psec,
2A and
resscan
and 2B).
Con-
trast was seen between articular cartilage adjacent bone, but poorer contrast was between
and muscle.
cartilage
Some
and seen
blurring
of the margins of the cartilage was seen. Reformation into either the sagittal (Fig. 2A) or the coronal The
(Fig.
2B) planes
3D cone
technique
high-resolution
obtain lofemoral
the
inter-
(full width at half peak shorter T2 relaxation
regions.
the extremity
olution
and
reconwere
in those
to the cartilage-bone
Three-dimensional tire joint of healthy
as part
In addition, spectroscopic available on a voxel-by-
information
voxel
all
no chemi-
tellofemoral
set,
in-
Because
occurred. of a full spectral
ages at any spectral 1A) and lipid-frequency
=
compared (TE = I .6
are resolved,
technique
(TE show
ratio when excitation
slice
spectral
joint. excitation
closer
cm FOV Images
images
ofhealthy
joint
2D). These
data
were
and the 3-inch were
reformatted
such as axial (Fig.
was performed. was
also
used
of the
volunteers
(Figs.
acquired (7.6-cm)
using surface
into standard
2C) and sagittal
to
paid2C a 6coil. planes,
(Fig.
2D).
was 600 psec, and scan time was 9.7 mm. Isotropic resolution was 0.5 mm.
TE
eas
Patient images (Figs. of cartilage damage
3 and 4) showing arwere acquired with
Fig. 1.-Axial MR images of patellofemoral joint of 25-year-old healthy volunteer using projection-reconstruction spectroscopic imaging sequence (TE = 200 isec). A, Water-frequency image. B, Magnified image of articular cartilage from box in A, along with spectra of patellofemoral cartilage. Note decreasing line width and increasing peak area as voxels progress from cartilage-bone interface to articular surface.
Fig. 2.-MR images of articular cartilage of knee of 24-year-old healthy volunteer using three-dimensional cone technique. A, Sagittal image with isotropic 1-mm resolution (field of view [FOV] = 12.8 cm, TE = 600 psec, scan time = 5.1 mm). B, Coronal image from same data set as A. C, Axial high-resolution image of patellofemoral joint with isotropic 0.5-mm resolution (FOV = 6.4 cm, TE = 600 psec, scan time D, Sagittal image from same data set as C.
1224
=
9.7 mm).
AJR:170, May 1998
MR Fig. 3.-Sagittal
of Articular
Cartilage
of the
Knee
MR image of knee of 30-year-old pacompartment
tient with medial (arrows). A, Fat-suppressed called echo image
cartilage
three-dimensional (TE 12 msec).
B, Projection-reconstruction frequency
Imaging
image
damage gradient-re-
spectroscopic
of same slice (TE
=
water-
1.6 msec).
#{149}..:,;*rtr
,
.
.
.
.
,-
( 4”
Fig. 4.-Sagittal MR image of knee of 39-year-old patient who had osteochondral drilling procedure 18 months previously. Medial-compartment cartilage damage (arrows) is seen. A, Fat-suppressed three-dimensional gradient-recalled echo image (TE = 12 msec). B, Projection-reconstruction spectroscopic waterfrequency image of same slice (TE = 1.6 msec).
the
PRSI
technique
using
the
extremity
Image
coil.
These tients
areas were well shown on all the pastudied, and spectra from these areas
were
available
fluid
was
intense
for
seen
Adjacent
as isointense
with
compared
on these
analysis.
joint
or slightly
not
hyper-
the articular
ideal
PRSI
cartilage
images.
contrast
when
will
A recent by
sensitivity
study
of patellar
Brossmann
cartilage
et al.
[3]
of an ultrashort-TE
construction
imaging
suppressed
spoiled
speci-
compared
3D fatecho
with
sequence
gradient-recalled
better
niques. lieved
This improved to be due to
associated short
T2
with
did
collagen
relaxation
and 3D cone ages
than
time
techniques
of cartilage
AJR:170, May 1998
with
two
delineation detection
was beof water which
[3]. Both produce
similar
tech-
has
the PRSI in vivo
ultrashort
imTEs.
because intensity
RF
synovial
which
from
on
of RF contrast
will
subject
of future
spoiling
than
water
content
to
[
study.
(Gold
nance
in Medicine current
PRSI
existing
cartilage
sequence
defects.
evaluation
T2 relaxation
has
techniques The
of cartilage
times.
several
in the ultrashort components
The
advantages
evaluation TE with
use of spectral
of
allows
not
limited
face
coil
interface.
et al., for
meeting.
resolution
by or
an
a data 6
can he acquired
Reso-
or at
improved
antialiasing coil
at the
1996)
April
capacity.
extremity
per
at lower
presented
The filter.
is
FOV to prevent aliasing. A limited number of slices
strict
pro-
mm
Magnetic
with
and processing
re-
inhomoge-
and
approximately
GE
injury
readout
requires
Society
of
be important
cartilage
artifacts
slices
International
storage
The
over
of
Multiple
resolution
1 11
of early
technique
time
line
(51. Estimation
spectroscopic shift
a very
a broader
area could
sites The
PRSI
to have
and
cartilage
by peak
spe-
For example,
shown
at the cartilage-bone
The
slice.
PRS!
a
hyaline
characterize
detects. time
chemical
cessing
or mag-
subtraction
help
been
relaxation
edema.
neity
In addi-
of the T2 decay of cartilage during the spectroscopic
transfer
width
solves
relative
has
T2
and
long-Tl
[10]. PRSI
fibrocartilage short
in determining
echo
spoiling,
fluid
bright
netization be the
the 3D
and line widths could cific areas of cartilage
the
the other
fiber,
as
The
unlike
intensity
appears
present.
gradient-recalled
have
The addition
readout.
and magnetization transfer contrast subtraction MR imaging. The projection-reconstruction imaging technique delineated cartilage lesions
such
cartilage signal
not signal
fluid
tion,
projection-re-
and the 3D
is
sequences.
spoiled
do
reduce
species
the
the PRSI
effusion
and 3D cone
methods,
mens
an
fat-suppressed
Discussion
with
cone techniques is similar to that with the 3D-FS-GRE technique. However. the signal level of fluid relative to articular cartilage is
needed
data
FOV
is
so a surto
re-
the
short
quired.
so use
of this
areas
stricted
to a specific
sequence
region
can
must
he
ac-
be re-
of interest.
1225
Gold Three-Dimensional
Cone
The 3D cone
used,
technique
also has advantages
over existing techniques. Because the central K-space trajectory is similar to that of projecimaging
tion-reconstruction contrast
and lesion
3D cone technique than conventional technique about over
used
by Disler
The
scan about cm)
The
the same with
(1.0-mm3 voxel mm. Decreased
using
3D cone volume
1.0-mm
conven-
technique (12.8
can
x 12.8
isotropic
x
resolution
volume) in approximately scan times are possible
both the 3D fat-suppressed recalled echo and 3D cone
mm
spoiled techniques
5 for
gradientwith the
use of high-speed gradient systems. For exampie, using high-speed gradients and the 3D cone FOV
technique, one can image a 12.8-cm with 1.0-mm isotropic resolution in 2.8
mm. pulse
Use of the spectral-spatial in 3D cone may provide
pression
at TFs
tional techniques The 3D cone processing
time
similar
excitation better fat sup-
to those
of conven-
(TE = 6.6 msec) [9]. technique requires of 6 mm
a data
entire volume. The FOV is not limited by an antialiasing filter, so a surface coil or extremity coil
is needed
to prevent
aliasing.
Acqui-
sition of an anisotropic FOV may be possible with a modified 3D cone trajectory. Alternatively, if the spectral-spatial excitation is
1226
which
can
be rein the
images, can be corrected by the apof field-mapping techniques for off-
resonance correction subject of future study.
is also
seen
3D cone plication
be
the
In conclusion, in vivo high-resolution trashort TE imaging of articular cartilage
ulin
has 3D
been
images
isotropic
resolution.
be used
The water
peak
by line
We
plan
content
within
to study
with
can cartitimes
to distinguish and hyaline
between cartilage
ability
have
would
for the long-term
399-407
im-
success
of the transplant.
J, Macovski
spectroscopic
A, Herfkens
R. MR
of collagen: tendons and
imaging
knee menisci. Magn Reson Med 1995;34:647-654 6. Jackson JI, Meyer CH, Nishimura DG, Macovski A. Selection of a convolution function for Fourier inversion using gridding. IEEE Trans Med Imaging 1991;MI-10:165-174
P. Nishimura
7. Irarrazabal sional
autologous technique
with
site. This
implications
or failure
plane
technique
this
transplants
at the repair
in any
PRSI
echo time projection reconstruction MR imaging of cartilage with histopathologic correlation: comparison with fat-suppressed spoiled grass and magnetization contrast MR imaging. Radiology 1997;203:501-507 4. Gold G, Pauly J, Moreno J, Glover G, MacovskiA, Herfkens R. Characterization of atherosclerotic plaque at 1ST. J Magn Reson Imaging 1993;3: 5. Gold G, Pauly
on a conventional
area and relaxation
and hope to be able growth of fibrocartilage portant
will
with scan times of 5-10 technique provides mor-
lage by relative chondrocyte
and
ofcartilage
to assess width.
[12]
shown
scanner cone
phologic
magnetic
DG.
resonance
Fast three
imaging.
dimen-
Magn
Reson
1995;33:656-662
Med
8. Meyer
CH, Pauly
JM, Macovski
A, Nishimura
D.
Simultaneous spatial and spectral selective cxcitation. Magn Reson Med 1990;15:287-304 9. Block WH, Pauly JM, KerrA, Nishimura D. Consistent fat suppression with compensated spectral-spatial pulses. Magn Reson Med 1997;38: 198-206
10. Disler D, McCauley
T, Kelman C, et al. Fat-supthree-dimensional spoiled gradient-echo
pressed MR
References
imaging
of hyaline
knee: comparison
1. XiaY,
to reconstruct the
dimension
Blurring,
whole-body mm. The
required
of 0.93
in one
stricted.
the knee
echo
et al. [10]
a 16 x 16 x 9 cm volume, gradients.
times 3D fat-
imaging
volume
the The
gradient-recalled
10 mm for a voxel
tional 12.8
are similar.
has shorter 3D methods.
spoiled
suppressed
sequences,
depiction
the FOV
et al.
FarquharT,
Burton-Wurster
N, Ray E, Jelin-
L. Diffusion and relaxation mapping of cartilage-bone plugs and excised disks using microscopic magnetic resonance imaging. Magn Reson Med 1994;31:273-282 ski
2. Freeman DM, Bergman G, Glover G. Short TE MR microscopy: accurate measurement and
zonal differentiation Magn
Reson
3. Brossmann
Med
of normal hyaline cartilage. 1997;38:72-81
J, Frank
L, Pauly
J, et al. Ultrashort
arthroscopy. 1 1. Peterfy
AiR
with
cartilage standard
in the
defects MR imaging
and
P. vanDijke
CF.
1996;167:127-132
S. Lang
CG, Majumndar
Sack K, Genant HK. MR imaging of the arthritic knee: improved discrimination of cartilage, synovium, and effusion with pulsed saturation transfer and fat-suppressed Ti-weighted sequences.
1994;l9l:413-419
Radiology
12. Mann
LC,
Pauly
quency
interpolation
rection.
Magn
Reson
JM,
Macovski
for fast Med
A.
Multifre-
off-resonance
cor-
1997;37:785-792
AJR:170, May 1998