Effects of Applying Different Resonance Amplitude on the ... - MDPI

0 downloads 0 Views 2MB Size Report
Jul 13, 2018 - In this study, a structural health monitoring technique known as the ... range and use an averaging technique to distinguish between the.
sensors Article

Effects of Applying Different Resonance Amplitude on the Performance of the Impedance-Based Health Monitoring Technique Subjected to Damage Wongi S. Na *, Dong-Woo Seo, Byeong-Cheol Kim and Ki-Tae Park Sustainable Infrastructure Research Center, Korea Institute of Civil Engineering & Building Technology (KICT), Gyeonggi-Do 10223, Korea; [email protected] (D.-W.S.); [email protected] (B.-C.K.); [email protected] (K.-T.P.) * Correspondence: [email protected]; Tel.: +82-31-9100-155 Received: 15 June 2018; Accepted: 11 July 2018; Published: 13 July 2018

 

Abstract: Smart materials such as piezoelectric transducers can be used for monitoring the health of building structures. In this study, a structural health monitoring technique known as the electromechanical impedance (EMI) method is investigated. Although the EMI method has the advantage of using a single piezoelectric patch that acts both as the actuator and as the sensor, there are still many issues to be addressed. To further understand the problem, the performance of the EMI method on a structure subjected to progressive damage at different resonance frequency ranges and peak amplitudes was investigated using three different statistical metrics: root-mean-square deviation (RMSD), mean absolute percentage deviation (MAPD) and correlation coefficient deviation (CCD). Metal plates were used throughout the study. The results acquired could be used to further understand the damage identification performance of the EMI method. Keywords: electromechanical impedance; piezoelectric transducer; damage detection; structural health monitoring; nondestructive testing

1. Introduction As civil infrastructures age, factors in the surrounding environment eventually cause their components to deteriorate over time, weakening the integrity of the structure. Most of the buildings around us are made from concrete, metals and composite materials, and as such, maintenance has always been a vital issue. There are many non-destructive ways of checking the status of a structure, including acoustic emission, radiographic, ultrasonic, liquid penetrant, eddy-current testing, magnetic-particle, and more. As the number of structures requiring management is increasing and in order to minimize the need for scheduled maintenance, an effective structural health monitoring system is the preferred choice compared to conventional non-destructive methods, The electromechanical impedance (EMI) method is a structural health monitoring technique that uses a single piezoelectric (PZT) transducer to act as both sensor and actuator [1]. The EMI method uses high frequency structural excitation, usually higher than 20 kHz, through a bonded PZT transducer to detect changes in structural mechanical impedance. The application of high frequency ranges means this method can detect local damage up to a few meters depending on the properties of the host structure. When applying the EMI method, a suitable frequency range needs to be manually determined through trial and error. To accomplish this, various frequency ranges are swept in order to search for a range with multiple peaks, and this range is selected to perform the EMI method. This is a vital step, as the absence of any peaks can result in a failure to identify damage as virtually no variations in impedance signatures will be observed, such as in composite structures [2]. Since the existence of these peaks is usually difficult to predict, the effect of different amplitudes on the performance of the EMI method should be investigated to achieve better understanding of the method. Sensors 2018, 18, 2267; doi:10.3390/s18072267

www.mdpi.com/journal/sensors

Sensors 2018, 18, 2267

2 of 13

To date, many investigations have been performed regarding the EMI method. In [2], the author showed that it is possible to distinguish between crack damage and debonding damage using a metal disc, which is similar to the concept shown later in this study. The main idea of the study was to create a fixed resonance frequency range and use an averaging technique to distinguish between the two different types of damage. In [3–6], environmental factors such as temperature variation effect, corrosive solution exposure, and the durability of PZT transducers were investigated and showed that the EMI method has several issues that needed to be resolved. The idea of using steel wire to minimize the temperature effect was proposed in [3]. Here, a PZT transducer was attached at one end of the wire while the other end is attached to a pipe with temperatures increasing up to 300 ◦ C. The results showed that the PZT transducer did not exceed 40 ◦ C. Na et al. investigated the performance of the EMI method on adhesive joints of glass-epoxy composite plates subjected to corrosive solution [4]. The results showed that damage to the adhesive can be monitored with the EMI method. Wandowski et al. [5] conducted experiments to detect delamination of CFRP panels whereby the authors applied a technique to compensate the effects of signature variation due to change in temperature. Yang et al. [6] also investigated the EMI technique subjected to temperature variations. In addition, the durability of the PZT transducers was also tested by measuring the impedance signature for up to 15 months. The results showed that a silicone rubber layer over the PZT transducer reduced the impedance signature changes during this time period. In Baptista and Filho [7], Bhalla et al. [8], Panigrahi et al. [9] and Wandowski et al. [10], the studies focused on minimizing the cost of performing the EMI method, as an HP4294 impedance analyzer can cost up to US$40,000. This was achieved by using devices such as a function generator, oscilloscope, FFT analyzer and an AD5933 evaluation board. The experimental results have shown that such systems can also measure impedance signatures with promising repeatability and reliability. Wandowski et al. [10] compared the performance of the AD5933 evaluation board with the impedance analyzer, HIOKI IM3570 and the results indicated that the evaluation board showed promising results. Although there are many more studies that have been conducted related to the EMI method, no investigation has been carried out on the performance of the method when subjected to different amplitude sizes. Thus, it would be important to know how different statistical metrics perform when subjected to different peak amplitudes as the non-model-based EMI method relies heavily on the signature acquired after the attachment of the PZT transducer. In this study, amplitude is decreased by attaching additional PZT transducers in series where various statistical metrics are used to analyze the acquired impedance signatures subjected to damage. 2. The EMI Method The one-dimensional equation proposed by Liang et al. (1994) shows how the EMI method works. In Equation (1), the electrical admittance Y (ω ) is a combined function of Zs (ω ) and Za (ω ), which are the mechanical impedance of the host structure and the PZT transducer, respectively. Other variables E

in the equation, which are I, V, ω, a, ε T33 , δ, d3x , Y xx represent the PZT output current, PZT input voltage, input frequency, geometric constant, dielectric constant, loss tangent, piezoelectric constant and Young’s modulus, respectively.  Y (ω ) = iωa

ε T33 (1 − iδ) −

Zs (ω ) E d23x Y xx Zs (ω ) + Za (ω )

 (1)

To measure the electrical impedance of a PZT transducer, a commercialized AD5933 evaluation board is utilized throughout this study where all the experiments were conducted at a room temperature of 22 ± 0.2 ◦ C. The board is manufactured by Analog Devices Co., Norwood, MA, USA, and retails for less than US$100. The experimental setup is shown in Figure 1. The AD5933 evaluation board is connected to a laptop, which can operate the device using the software provided within the package. The advantage of this device is its weight, as it is very light and its small size makes it portable. It can measure impedance up to 100 kHz with 511 data points and is fully powered by a USB cable

Sensors2018, 2018,18, 18,x2267 Sensors FOR PEER REVIEW

3 3ofof1313

portable. It can measure impedance up to 100 kHz with 511 data points and is fully powered by a that cable is connected the laptop. used for PZT this study the model manufactured by USB that is to connected to The the PZT laptop. The used was for this study PSI-5A4E was the model PSI-5A4E Piezo Systems Inc., Cambridge, MA, USA. The size of the PZT sheet was 72.4 mm × 72.4 mm with manufactured by Piezo Systems Inc., Cambridge, MA, USA. The size of the PZT sheet was 72.4 mma 0.508 mm, and of it was cutmm, intoand the required study. Notice that wiring of the ×thickness 72.4 mm of with a thickness 0.508 it was cutsizes intofor thethe required sizes for thethe study. Notice test specimen is taped onto the edge of the high table to allow it to be left in the air. Since this study that the wiring of the test specimen is taped onto the edge of the high table to allow it to be left in the involves creating damage to acreating metal plate, one to needs to manually pick up the plate and pick replace on air. Since this study involves damage a metal plate, one needs to manually up itthe the table after creating damage. Leaving the specimen the air the canspecimen eliminate in any signature variations plate and replace it on the table after creating damage. in Leaving the air can eliminate caused by changes in boundary conditions, as the impedance signatures are extremely sensitive any signature variations caused by changes in boundary conditions, as the impedance signatures areto any changes. extremely sensitive to any changes.

Figure 1. Electromechanical impedance experiment configuration. Figure 1. Electromechanical impedance experiment configuration.

After measuring the electrical impedance of the PZT, the next step is to quantify the intensity of After measuring the electrical impedance of thethree PZT,different the next equations step is to quantify thethe intensity the damage using a statistical method. In this study, were used, rootof the damage using (RMSD), a statistical method. this study,deviation three different were used, mean-square deviation mean absoluteInpercentage (MAPD)equations and the correlation the root-mean-square deviation mean absolute percentage deviation In (MAPD) and the coefficient deviation (CCD), which(RMSD), are shown as Equations (2)–(4), respectively. the equations, correlation coefficient deviation (CCD), which are shown as Equations (2)–(4), respectively. Inthe the represents the reference impedance signature and isthe corresponding signature. N is equations, the reference impedance and ( Zmean corresponding signature. Zk )i represents ̅ signifying k ) j is the number of (impedance signatures with the symbolssignature values and signifying N is the number of impedance signatures thethe symbols Z signifying mean anddata σZ signifying standard deviation. For the study, the real with part of impedance signature wasvalues used for analysis, deviation. For the study, the partsensitive of the impedance signature was used data analysis, asstandard it has been experimentally proven toreal be less to temperature variations thanfor the imaginary as it has been experimentally proven to be less sensitive to temperature variations than the imaginary part of the signature [11]. In general, the single value obtained from any of the three equations will part of thewith signature [11]. Iningeneral, theassingle value obtained from any the three equations will be be higher an increase damage, the impedance signature will of change more severely. The higher with an EMI increase in damage, as the impedancevalue signature change more severely. The final final step in the method is to define a threshold whichwill is manually defined by experts in stepfield. in the EMIany method is that to define a threshold valuevalue which manuallyto defined by experts in this this Then, values exceeds this threshold is is considered be damaged. field. Then, any values that exceeds this threshold value is considered to be damaged. ⁄ (2) RMSD = − !1/2 i2 . N N h RMSD = ∑ Re( Zk ) j − Re( Zk )i (2) [ Re( Zk )i ]2 ∑ =11 k =1 ⁄ (3) MAPD k= − 1 MAPD = 1 N CCD = 1 −

N

∑ |[ Re(Zk ) j − Re(Zk )i ]/Re(Zk )i |

k =1



̅





̅

(3) (4)

Sensors 2018, 18, 2267

4 of 13

Sensors 2018, 18, x FOR PEER REVIEW

1

N

   4 of 13 Z j ]· Re( Zk )i − Re Z i (4) NσZj σZi k=1 3. Impedance Signature Peak Reduction Effect on EMI Performance 3. Impedance Signature Peak Reduction Effect on EMI Performance To evaluate the performance of the EMI method when the impedance signature is reduced, three 15 mm PZT were Three test cases were created and tested,iswhere the To square evaluate thetransducers performance of prepared. the EMI method when the impedance signature reduced, first “C_1” involved attaching were one of the three PZTtest transducers the center of thewhere 100 mm threecase 15 mm square PZT transducers prepared. Three cases wereto created and tested, the square metal plate with a thickness 0.3 mm, as shown in Figure 2. This configuration was then first case “C_1” involved attaching one of the three PZT transducers to the center of the 100 mm square connected the aAD5933 evaluation andinthen progressive damage waswas created a metal metal platetowith thickness 0.3 mm, board, as shown Figure 2. This configuration thenusing connected to cutter along evaluation the dotted board, line upand to 20 mm. During this process, signatures the AD5933 then progressive damage wasimpedance created using a metal were cuttermeasured along the by exciting thetoPZT transducer every 2 mm of damagesignatures in the frequency range ofby25exciting kHz to the 65 kHz dotted line up 20 mm. Duringat this process, impedance were measured PZT in the 80 Hzatinterval. frequency range was chosen ranges and transducer every 2 Here, mm ofthe damage in the frequency rangeby of sweeping 25 kHz to various 65 kHz in the 80 Hzselecting interval. the most range with multiple resonance peaks.ranges Thus,and including thethe reference signature Here, the appropriate frequency range was chosen by sweeping various selecting most appropriate measured anyresonance damage was introduced, 11 impedance signatures were acquired. second range withbefore multiple peaks. Thus, including the reference signature measuredThe before any test case was “C_2” was conducted by attaching an additional PZT transducer series the was C_1 damage introduced, 11 impedance signatures were acquired. The secondin test case to “C_2” configuration shown Figure 2).PZT Thistransducer additionalin PZT is left freely the air, as the(also attachment conducted by (also attaching an in additional series to the C_1inconfiguration shown will affect2).the amplitude of PZT the is impedance theattachment equivalentwill impedance the two in Figure This additional left freely signatures in the air, asasthe affect theof amplitude transducers is changed. In as this since impedance the capacitance the PZT transducer is directly of the impedance signatures thecase, equivalent of theof two transducers is changed. In this proportional its size, theofamplitude will decrease. For this second testto(C_2), progressive damage case, since thetocapacitance the PZT transducer is directly proportional its size, the amplitude will is introduced the same test manner. Again, impedance signatures were measured formanner. every 2 mm of decrease. For in this second (C_2), progressive damage is introduced in the same Again, damage up signatures to 20 mm were in themeasured same frequency range asof above, resulting in mm 11 impedance impedance for every 2 mm damage up to 20 in the samesignatures. frequency The test case “C_3” involved attaching another in series onto attaching the C_2 rangethird as above, resulting in 11 impedance signatures. ThePZT thirdtransducer test case “C_3” involved configuration, which further increases theC_2 overall area of the PZT further transducers. Again, another PZT transducer in series onto the configuration, which increases the progressive overall area damage wastransducers. introduced Again, along the dotted line and thewas impedance signatures were measured in the an of the PZT progressive damage introduced along the dotted line and identical manner to thewere previous two test cases. Through thesetothree experiments C_2 and C_3), impedance signatures measured in an identical manner the previous two(C_1, test cases. Through 33 impedance signatures(C_1, wereC_2 acquired in total. these three experiments and C_3), 33 impedance signatures were acquired in total. CCD = 1 −

∑ [ Re(Zk ) j − Re

Figure 2. The setup for three test cases for evaluating the EMI performance subjected to amplitude Figure 2. The setup for three test cases for evaluating the EMI performance subjected to reduction. amplitude reduction.

Figure 3 shows all the impedance signatures for the three cases where the difference in the height 3 shows impedance forinthe three cases where decrease the difference in the of theFigure amplitudes canall be the clearly seen. Thesignatures amplitudes, general, significantly from C_1 to height of the amplitudes can be clearly seen. The amplitudes, in general, significantly decrease from C_3. The frequency ranges around 30 kHz, 44 kHz and 65 kHz are a good example of this. First, C_1 to C_3.the The frequencysignatures ranges around 30 28 kHz, 44and kHz32and 65the kHz are aofgood example of this. examining impedance between kHz kHz, center the resonance peaks First, examining the impedance signatures between 28 kHz and 32 kHz, the center of the resonance seems to shift slightly rightward from C_1 to C_3 with the overall amplitudes decreasing, which peaks seems to shift rightward from C_1 to C_3 with the overall amplitudes possibly indicates thatslightly progressive damage is changing the resonance frequency of the hostdecreasing, structure. which possibly indicates that progressive damage is changing the resonance frequency of the7host For C_1, C_2 and C_3 the maximum peak-to-peak heights of the resonance are around 15 kΩ, kΩ structure. For C_1, C_2 and C_3 the maximum peak-to-peak heights of the resonance are and 5 kΩ, respectively. Next, examining the impedance signatures between 43 kHz and 46 kHzaround shows 15 kΩ, 7 kΩ and 5 kΩ,resonance respectively. Next, examining thedamage impedance signatures between kHz and a similar result where is shifting rightward as progresses. However, the43 maximum 46 kHz showsamplitudes a similar result where resonance shifting rightward However, peak-to-peak for the three test casesisare around 14 kΩ, 5 as kΩdamage and 1.5progresses. kΩ from C_1 to C_3, the maximum peak-to-peak amplitudes for the three test cases are around 14 kΩ, 5 kΩ and 1.5 kΩ from proving that an increase in PZT area does not have a linear relationship to the maximum height of the peak-to-peak amplitudes. Finally, at the frequency range between 63 kHz and 67 kHz, the resonance is again shifting in the rightward direction with damage. This can be clearly seen in the C_1 test. In addition, another observation made here is the decrease in the peak amplitudes with

Sensors 2018, 18, 2267

5 of 13

C_1 to C_3, proving that an increase in PZT area does not have a linear relationship to the maximum height of the peak-to-peak amplitudes. Finally, at the frequency range between 63 kHz and 67 kHz, the resonance is again shifting in the rightward direction with damage. This can be clearly seen in Sensors 2018, 18, x FOR PEER REVIEW 5 of 13 the C_1 test. In addition, another observation made here is the decrease in the peak amplitudes with progressive progressivedamage damagefor forC_3. C_3.The Thepeak peakthat thatexists existsatat66 66kHz kHzisisreduced reducedand andvirtually virtuallydisappears disappearsafter after 20 20mm mmofofprogressive progressivedamage. damage.

Figure 3. Impedance signatures for C_1, C_2 and C_3 experiment. Figure 3. Impedance signatures for C_1, C_2 and C_3 experiment.

Using the impedance signatures shown in Figure 3, three statistical metrics (RMSD, MAPD and Using impedance shown in Figure 3, first threecolumn statistical metrics the (RMSD, MAPD and CCD) werethe calculated andsignatures are displayed in Table 1. The represents damage intensity CCD) were calculated and are displayed in Table 1. The first column represents the damage intensity from 0 mm to 20 mm, the second to fourth columns are the RMSD values (RC_1, RC_2, RC_3), the from to 20 columns mm, the second fourthvalues columns are the RMSD valuesand (RC_1, RC_2, RC_3), the fifth fifth 0tomm seventh are theto MAPD (MC_1, MC_2, MC_3) the last three columns are tothe seventh columns are the MAPD values (MC_1, MC_2, MC_3) and the last three columns are the CCD CCD values (CC_1, CC_2, CC_3). For the RMSD values, RC_1 starts at 7.38% and ends at 13.53%, values CC_2, CC_3). For the RC_1 startsatat1.63% 7.38%and and ends ends at RC_2 starts RC_2 (CC_1, starts at 5.78% and ends at RMSD 7.27%, values, and RC_3 starts at 13.53%, 4.18%. In addition, ataveraging 5.78% and ends at 7.27%, and RC_3 starts at 1.63% and ends at 4.18%. In addition, averaging the the 10 RMSD values for each test cases results in 11.46%, 6.04% and 3.25% for RC_1, RC_2 10 RMSD for eachThese test cases results 11.46%, 3.25% forthe RC_1, RC_2 and will RC_3, and RC_3,values respectively. results proveinthat with 6.04% smallerand amplitudes, RMSD values be respectively. These results prove that with smaller amplitudes, the RMSD values will be lower lower when subjected to the same level of damage. Furthermore, the averaged RMSD values ofwhen 6.04% subjected the53% same level of damage. Furthermore, averaged RMSD values of 6.04% and 3.25% and 3.25%toare (6.04/11.46) and 28% (3.25/11.46), the respectively. are 53% (6.04/11.46) and 28% (3.25/11.46), respectively. Table 1. Three statistical metric values for C_1, C_2 and C_3 experiment. Table 1. Three statistical metric values for C_1, C_2 and C_3 experiment.

Damage RMSD (%) (mm) RC_1 RC_2 (%)RC_3 RMSD Damage 2 (mm) 7.38 1.63 RC_1 5.78 RC_2 RC_3 4 10.61 3.92 2.29 2 7.38 5.78 1.63 6 10.06 2.09 4 10.61 5.50 3.92 2.29 6 10.06 4.72 5.50 2.09 8 10.95 3.25 10.95 5.94 4.72 3.25 10 8 11.68 4.19 10 11.68 5.94 4.19 12 14.24 7.03 3.41 12 14.24 7.03 3.41 14 14 13.32 6.43 3.72 13.32 6.43 3.72 16 16 11.59 3.84 11.59 6.92 6.92 3.84 11.22 6.93 6.93 3.86 18 18 11.22 3.86 20 13.53 7.27 4.18 20 13.53 7.27 4.18 Average 11.46 6.04 3.25 Average 11.46 6.04 3.25

MAPD (%) CCD (%) MC_1MAPD MC_2 MC_3 CC_1 CC_3 (%) CCD (%)CC_2 11.51 MC_2 1.58 MC_30.71CC_1 0.90 MC_1 CC_2 4.66 CC_3 1.06 16.10 1.70 1.03 1.63 2.37 1.90 11.51 1.58 0.71 0.90 4.66 1.06 15.07 2.20 1.08 1.49 4.51 1.62 16.10 1.70 1.03 1.63 2.37 1.90 15.07 2.20 1.62 15.77 2.33 1.08 1.68 1.49 1.734.51 3.37 3.68 15.77 2.33 3.68 17.38 2.80 1.68 2.05 1.73 1.943.37 5.20 6.05 17.38 2.80 2.05 1.94 5.20 6.05 19.58 3.16 1.75 2.77 7.06 4.05 19.58 3.16 1.75 2.77 7.06 4.05 18.36 2.96 1.91 2.46 6.07 4.81 18.36 2.96 1.91 2.46 6.07 4.81 21.12 2.95 1.97 1.91 6.75 5.10 21.12 2.95 1.97 1.91 6.75 5.10 21.79 3.07 2.05 1.81 6.94 5.13 21.79 3.07 2.05 1.81 6.94 5.13 27.55 3.46 5.97 27.55 3.46 2.12 2.12 2.54 2.547.64 7.64 5.97 18.42 2.62 1.64 1.92 5.46 3.94 18.42 2.62 1.64 1.92 5.46 3.94

Numbers in different colors represent colors in picture.

Numbers in different colors represent colors in picture.

For values, MC_1 is generally larger than than the values of RC_1 average of 18.42%. Forthe theMAPD MAPD values, MC_1 is generally larger the values of with RC_1anwith an average of The first 2 mm results in a value of 11.51% and increases as the damage progresses, ending with a value 18.42%. The first 2 mm results in a value of 11.51% and increases as the damage progresses, ending

with a value of 27.55%. For the MAPD values of MC_2, the average value was significantly decreased to 2.62% (from 18.42%) where the values start from 1.58% and end at 3.46%. This shows that the amplitude reduction of the impedance signature has a vital impact on the MAPD values. Next, the values for MC_3 have an average of 1.64%, starting from 0.71% and ending at 2.12%. Compared to the RMSD and MAPD values, the CCD values, the results show a different pattern.

on the correlation coefficient. However, when comparing RC_3, MC_3 and CC_3 average values, CC_3 has the highest with 3.94%. This experimentally shows that when there are many peaks with only small amplitudes in the signature, CCD is the preferred choice of the three statistical metrics. 4. Creating Various Resonance Frequency Ranges Sensors 2018, 18, 2267

6 of 13

In the previous section, experiments were used to confirm that a decrease in the amplitude of the impedance signature will also decrease the RMSD and MAPD values (but not CCD values). Thus, of further 27.55%.the Forinvestigation, the MAPD values of MC_2,reduction the average value wasEMI significantly decreased 2.62% to the amplitude effect on the method subjected to to damage (from 18.42%) where the values start from 1.58% and end at 3.46%. This shows that the amplitude in various frequency ranges should be examined. To achieve this, the conventional method of reductionaof thetransducer impedancetosignature has vital impact on the MAPD values. the resonance values for attaching PZT the surface ofathe host structure is changed in orderNext, to create MC_3 have an average of 1.64%, starting from 0.71% and ending at 2.12%. frequency ranges in various regions. In Na et al. [4], the authors introduced a technique for creating Compared to theranges RMSDinand MAPD values, the CCD of values, the resultsofshow a different pattern. resonance frequency certain regions regardless the properties the host structure. The The highest average value is observed for CC_2 with a value of 5.46%, followed by 3.94% for CC_3 core of this technique involves attaching a PZT transducer on to a metal disc, then attaching the metal and onto 1.92% CC_1. As expected, CCD index is insensitive to be variations in amplitude as it is disc thefor host structure. Here, thethe resonance frequency range can either increased or decreased based on correlation RC_3, MC_3 average simply bythe changing the coefficient. thickness ofHowever, the metalwhen disc. comparing Thus, in this study, thisand ideaCC_3 is applied to values, create CC_3 has the highest with 3.94%. This experimentally shows that when there are many peaks with various frequency ranges where additional free PZT transducers were utilized to reduce peak only small amplitudes in the signature, CCD is the preferred choice of the three statistical metrics. amplitudes. This configuration can be seen in Figure 4a, where a 15 mm square PZT transducer is attached on top of the metal disc with a 25 mm diameter and 3 mm thickness, then, two more devices 4. Creating Various Resonance Frequency Ranges were created using two different metal discs with thicknesses of 4 mm and 5 mm. These three devices the to previous experiments used to confirm a decrease amplitude were In used create 9section, test cases referred towere as “T3_x”, “T4_x” andthat “T5_x”, where in thethe first number of the impedance signature will also decrease the RMSD and MAPD values (but not CCD values). represents the thickness of the metal in mm and variable x is the total number of PZT transducers Thus, to further the investigation, amplitude reduction effect on the EMI method subjected connected. For example, T3_1 is thethe 3 mm metal disc with only the PZT transducer attached, T3_2 to is damage various frequencyfree ranges be examined. achieveand this, the has conventional method the samein with an additional PZTshould transducer attached To in series, T3_3 another free PZT of attaching a PZT to transducer to the surface of the host structure is changed in order to create transducer attached T3_2 in series. resonance frequency in various regions.measurements In Na et al. [4], are the displayed authors introduced technique for In Figure 4b, theranges 9 impedance signature using theathree devices creating resonance frequency ranges in certain regions regardless of the properties of the host structure. that were created. First, examining the three impedance signatures of T3_1, T3_2 and T3_3, there is a The core of involves a PZT on to a metal disc,and then38attaching the decrease in this the technique amplitudes of theattaching resonance peaktransducer located between 33 kHz kHz when metal disc PZT ontotransducers the host structure. Here, Here, the resonance range can be is either or additional are attached. it can be frequency said that the resonance that increased of the metal decreased simply by changing the thickness of the metal disc. Thus, in this study, this idea is applied disc. The height of the amplitude for T3_1, which was the difference between the highest and lowest to create frequency ranges where additional PZT were utilized peak points of various the impedance signature for this study, isfree about 16transducers kΩ. For T3_2 and T3_3, to thereduce heights of amplitudes. This configuration be seen in Figure 4a, where a 15 mmthe square is the peak amplitudes are about 7 can kΩ and 3.5 kΩ, respectively. Regarding other PZT 6 testtransducer cases, T4_1, attached onT5_1, top ofT5_2, the metal disc with a 25 mm and 3 mm thickness, then, T4_2, T4_3, and T5_3, the heights of diameter the peak amplitudes are roughly 16 two kΩ, 8more kΩ, devices 4.5 kΩ, were twokΩ, different metal discs withthe thicknesses ofpeaks 4 mm are andconcentrated 5 mm. These in three devices 24 kΩ,created 16 kΩ,using and 10 respectively. Overall, resonance 3 different were used create kHz, 9 test41cases referred as 50 “T3_x”, “T4_x” wheretothe first number regions, 33 to kHz~38 kHz~46 kHz to and kHz~55 kHz, and thus“T5_x”, it is possible investigate the represents the of the metal in mm and variable x is the total number of Compared PZT transducers performance ofthickness the EMI method subjected to damage at various frequency ranges. to the connected. T3_1 the 3 mmlarger metalfor disc only PZT transducer attached, T3_2 is other cases, For the example, amplitudes areisrelatively thewith peaks at the 50 kHz~55 kHz frequency range. the same with an additional free PZT transducer attached in series, and T3_3 has another free PZT Thus, it can be assumed that the metal disc with a thickness of 5 mm might perform better when the transducer to T3_2 series. EMI methodattached is applied. The in results of this test are discussed further later in this paper.

Figure 4. Different PZT attachment (a) configuration; (b) device impedance signatures. Figure 4. Different PZT attachment (a) configuration; (b) device impedance signatures.

In Figure 4b, the 9 impedance signature measurements are displayed using the three devices that were created. First, examining the three impedance signatures of T3_1, T3_2 and T3_3, there is a decrease in the amplitudes of the resonance peak located between 33 kHz and 38 kHz when additional PZT transducers are attached. Here, it can be said that the resonance is that of the metal disc. The height

Sensors 2018, 18, 2267

7 of 13

of the amplitude for T3_1, which was the difference between the highest and lowest points of the impedance signature for this study, is about 16 kΩ. For T3_2 and T3_3, the heights of the peak amplitudes are about 7 kΩ and 3.5 kΩ, respectively. Regarding the other 6 test cases, T4_1, T4_2, T4_3, T5_1, T5_2, and T5_3, the heights of the peak amplitudes are roughly 16 kΩ, 8 kΩ, 4.5 kΩ, 24 kΩ, 16 kΩ, and 10 kΩ, respectively. Overall, the resonance peaks are concentrated in 3 different regions, 33 kHz~38 kHz, 41 kHz~46 kHz and 50 kHz~55 kHz, thus it is possible to investigate the performance of the EMI method subjected to damage at various frequency ranges. Compared to the other cases, the amplitudes are relatively larger for the peaks at the 50 kHz~55 kHz frequency range. Thus, it can be assumed that the metal disc with a thickness of 5 mm might perform better when the EMI method Sensors 2018, 18, x FOR PEER 7 of 13 is applied. The results ofREVIEW this test are discussed further later in this paper.

5. Conducting Conducting the the EMI EMI Method Method with with Various Various Signature Signature Amplitudes Amplitudes 5. cases in in order order to to evaluate evaluate the the damage damage detection detection Figure 5 shows the EMI method used to test 9 cases different resonance resonance frequency frequency ranges. ranges. Figure performance using 3 different Figure 5a 5a shows shows the the three test cases, which will be be referred referred to to as as T3_1, T3_1, T3_2 T3_2 and and T3_3 T3_3 hereafter. hereafter. Again, the first number number represents the which thickness of of the the metal metal disc, disc, with with the the second second number number representing representing the the number number of of PZT PZT transducers transducers thickness connected for the test. The PZT attached metal disc is adhered to the center of the metal plate using connected for the test. The PZT attached metal disc is adhered to the center of the metal plate using an an epoxy glue and was left at room temperature for 24 h to ensure full curing. For the test, damage epoxy glue and was left at room temperature for 24 h to ensure full curing. For the test, damage was was created the same method introduced in Section 2, where a metal wasto used the created usingusing the same method introduced in Section 2, where a metal cutter cutter was used cut to thecut metal metal plate up to 20 mm with impedance signatures being measured at every 2 mm of damage. These plate up to 20 mm with impedance signatures being measured at every 2 mm of damage. These three threecases test cases conducted onsame the same metal as shown in the figure. damage intensity test were were conducted on the metal plateplate as shown in the figure. TheThe damage intensity of of each of cases the cases assumed the as same, as the distance PZT transducers was each of the was was assumed to be to thebe same, the distance from thefrom PZT the transducers was identical. identical. Forcase, each11 test case, 11 impedance wereinacquired in total (including reference For each test impedance signaturessignatures were acquired total (including a referenceasignature) signature) in the frequency ranges25 between 65 kHz.5b, Figure 5b, cthe shows test specimen in the frequency ranges between kHz to25 65kHz kHz.to Figure c shows test the specimen for the for the remaining 6 tests (T4_1, T4_2, T4_3, T5_1, T5_2, and T5_3), where the tests were identical to remaining 6 tests (T4_1, T4_2, T4_3, T5_1, T5_2, and T5_3), where the tests were identical to Figure 5a Figure 5a with different metal disc thicknesses. with different metal disc thicknesses.

(b) T4_x; T4_x; (c) (c) T5_x. T5_x. Figure 5. Test setup for (a) T3_x; (b)

Figure 6 shows the 33 impedance signatures acquired from the T3_1, T3_2 and T3_3 experiments, Figure 6 shows the 33 impedance signatures acquired from the T3_1, T3_2 and T3_3 experiments, where resonance peaks are concentrated between 36 kHz and 44 kHz. Compared with the impedance where resonance peaks are concentrated between 36 kHz and 44 kHz. Compared with the impedance signature before the PZT device was attached onto the metal plate as shown in Figure 4b, one can signature before the PZT device was attached onto the metal plate as shown in Figure 4b, one can observe that the resonance frequency has shifted rightward, about 5 kHz. Also, additional resonance observe that the resonance frequency has shifted rightward, about 5 kHz. Also, additional resonance frequency ranges have appeared at around 31 kHz and 45 kHz. Since the largest resonance at 36 frequency ranges have appeared at around 31 kHz and 45 kHz. Since the largest resonance at kHz~44 kHz is the resonance of the metal disc, one can assume that the resonance at 31 kHz and 45 36 kHz~44 kHz is the resonance of the metal disc, one can assume that the resonance at 31 kHz and kHz is the resonance of the metal plate. However, regardless of the resonance being either the metal 45 kHz is the resonance of the metal plate. However, regardless of the resonance being either the metal plate or the disc, signature variations can be visually identified in the resonance frequency ranges. plate or the disc, signature variations can be visually identified in the resonance frequency ranges. Finally, the impedance signature beyond 46 kHz has no resonance where the T3_1 signature shifts in Finally, the impedance signature beyond 46 kHz has no resonance where the T3_1 signature shifts in the downward direction subjected to damage. For T3_2 and T3_3, the impedance signatures seemed the downward direction subjected to damage. For T3_2 and T3_3, the impedance signatures seemed to to be unaffected by damage, showing virtually no sign of any shift movement. be unaffected by damage, showing virtually no sign of any shift movement.

kHz~44 kHz is the resonance of the metal disc, one can assume that the resonance at 31 kHz and 45 kHz is the resonance of the metal plate. However, regardless of the resonance being either the metal plate or the disc, signature variations can be visually identified in the resonance frequency ranges. Finally, the impedance signature beyond 46 kHz has no resonance where the T3_1 signature shifts in the downward Sensors 2018, 18, 2267direction subjected to damage. For T3_2 and T3_3, the impedance signatures seemed 8 of 13 to be unaffected by damage, showing virtually no sign of any shift movement.

Figure 6. Impedance signatures for the T3_1, T3_2 and T3_3 experiment. Figure 6. Impedance signatures for the T3_1, T3_2 and T3_3 experiment.

Sensors 2018, 18, x FOR PEER REVIEW

8 of 13

Figure 7 7shows acquiredfrom fromthe theT4_1, T4_1,T4_2 T4_2 and T4_3 experiments Figure showsthe the33 33impedance impedance signatures signatures acquired and T4_3 experiments where resonance peaks are concentrated between 43 kHz and 49 kHz. Specifically, , there two where resonance peaks are concentrated between 43 kHz and 49 kHz. Specifically, , there areare two resonance kHzand and45 45kHz~49 kHz~49 kHz where first is the resonancefrequency frequencyregions regionswhich which are are at at 43 43 kHz~45 kHz~45 kHz kHz where thethe first is the resonance of the metal plate as shown in the previous figure. We know that the second resonance resonance of the metal plate as shown in the previous figure. We know that the second resonance is is thethe resonance proventhat thatthe theresonance resonance shifts rightward resonanceofofthe themetal metaldisc, disc,as asitit was was experimentally experimentally proven shifts rightward after the plate.InInaddition, addition, resonance is larger with more dynamic activities after theattachment attachmentof of aa metal metal plate. resonance is larger with more dynamic activities in figure, asas thethe resonance of of thethe metal disc hashas in the the 43 43 kHz~45 kHz~45kHz kHzrange rangecompared comparedtotothe theprevious previous figure, resonance metal disc influenced theoutcome. outcome.The Thesmall smallresonance resonance located located at when compared to to influenced the at 30 30kHz~33 kHz~33kHz kHzisissmaller smaller when compared previousfigure, figure,and andone onepossible possible cause cause of forfor thethe experiment, thethe previous of this this isisthe thethicker thickermetal metaldisc discused used experiment, T4_x mmthicker thickerthan thanT3_x, T3_x, thus thus making making the away from thethe host as as T4_x is is1 1mm thePZT PZTtransducer transducer1 1mm mmfurther further away from host structure. For the frequency range above 55 kHz where no resonance can be observed, the impedance structure. For the frequency range above 55 kHz where no resonance can be observed, the impedance signatures forT4_1 T4_1show showaadownward downward shift, shift, with signatures remaining signatures for with the theT4_2 T4_2and andT4_3 T4_3impedance impedance signatures remaining virtually unchanged. virtually unchanged.

Figure 7. Impedance signatures for the T4_1, T4_2 and T4_3 experiment. Figure 7. Impedance signatures for the T4_1, T4_2 and T4_3 experiment.

Figure 8 shows the 33 impedance signatures acquired from the T5_1, T5_2 and T5_3 experiments Figure 8 showspeaks the 33 signatures from T5_1, T5_2 and T5_3 experiments where resonance areimpedance concentrated betweenacquired 53 kHz and 59 the kHz. At first glance, the impedance where resonance are concentrated between 53 compared kHz and 59 At first two glance, the impedance signatures have peaks a relatively smaller number of peaks to kHz. the previous figures. As such, signatures have a relatively smaller number of peaks compared to the(RMSD, previous two figures. As such, one can assume that the values obtained from the statistical methods MAPD, CCD) would one can assume that the values obtained from the statistical methods (RMSD, MAPD, CCD) be smaller compared to the test cases, T3_x and T4_x. This will be further discussed in Section 6.would For small compared resonance around 32 kHz, theT3_x peaksand seem to decrease andbe almost disappear withinthe T5_3 6. bethe smaller to the test cases, T4_x. This will further discussed Section signatures. resonance here is than the resonance observedand in the previous figure, with as thethe For the smallThe resonance around 32smaller kHz, the peaks seem to decrease almost disappear PZTsignatures. transducer The is further away from thesmaller host structure due to the useobserved of a thicker disc. T5_3 resonance here is than the resonance in metal the previous figure, as the PZT transducer is further away from the host structure due to the use of a thicker metal disc.

where resonance peaks are concentrated between 53 kHz and 59 kHz. At first glance, the impedance signatures have a relatively smaller number of peaks compared to the previous two figures. As such, one can assume that the values obtained from the statistical methods (RMSD, MAPD, CCD) would be smaller compared to the test cases, T3_x and T4_x. This will be further discussed in Section 6. For the small resonance around 32 kHz, the peaks seem to decrease and almost disappear with the T5_3 signatures. Sensors 2018, 18, 2267 The resonance here is smaller than the resonance observed in the previous figure, as the PZT transducer is further away from the host structure due to the use of a thicker metal disc.

9 of 13

Figure 8. Impedance signatures for the T5_1, T5_2 and T5_3 experiment.

Figure 8. Impedance signatures for the T5_1, T5_2 and T5_3 experiment.

6. Results and Discussion 6.1. Analyzing the Data for T3_x, T4_x and T5_x Cases Table 2 shows the values calculated using the three statistical methods for Figure 6. When the values in Table 1 are compared with the C_1, C_2 and C_3 test cases, the averaged values for each column are generally lower. In addition, most of the values increase as the damage progresses. The RMSD values for RT3_1 range from 5.33% to 11.03% with an average value of 9.18%. The average values for RT3_2 and RT3_3 are 2.95% and 2.16%, respectively. This shows that it is best to have impedance signatures with a large amplitude for high RMSD values, as one of the final steps in identifying damage using the EMI method is to define a threshold value. Thus, large RMSD values subjected to damage can allow the user to define a threshold value which can maximize the chance of differentiating a damaged structure from an intact one. The averaged values for MT3_1, MT3_2 and MT3_3 are 13.56%, 1.30% and 0.93%, respectively, which is a similar result to the data in Table 1. Again, a decrease in the amplitudes of the impedance signatures had a significant impact on the MAPD values, as 1.3% is only approximately 1/10 of 13.56%. For the CCD values, the average values are 2.95%, 3.98% and 2.05% for CT3_1, CT3_2 and CT3_3, respectively. This result shows that the reduction in the amplitudes does not have as large an impact, as it did with the RMSD and MAPD values. Table 3 shows the values calculated using the three statistical methods for Figure 7. With the extra 1 mm thickness of the metal disc, one might expect that the values calculated using statistical methods would be lower compared to the T3_x cases. However, all the averaged values for each column are larger compared to those in Table 2. For the RMSD values, the averaged values for RT4_1, RT4_2 and RT4_3 are 9.52%, 5.61% and 3.28%, respectively. For the MAPD values, the averaged values are 19.11%, 2.33% and 1.31% for MT4_1, MT4_2 and MT4_3, respectively, experimentally proving once again that the height of the peaks of the impedance signatures are very important when using the MAPD method. For the CCD values, the averaged values are 6.15%, 10.00% and 3.58% for CT4_1, CT4_2 and CT4_3, respectively. One of the reasons why the calculated values are relatively larger in this case is the overlap of the resonance of the metal disc and the plate. As shown in Figure 7, this seems to amplify the height of the resonance peaks in the frequency range that corresponds to the resonance of the metal plate. Table 4 shows the values calculated using the three statistical methods for Figure 8. With the thicker metal disc, one can expect that most of the calculated values will be lower compared to the values in Tables 2 and 3. However, when we compare the values in Table 4 with those in Table 2, more than half of the averaged values are larger. These are the RMSD averaged values of 3.66%, 3.32%, all the MAPD averaged values of 35.95%, 3.09%, 1.29%, and the CCD value of 2.73%. This shows that while the 5 mm thick metal disc is roughly two times thicker than the 3 mm thick metal disc, this difference does not have too much of an effect on the statistical values. In addition, the averaged

Sensors 2018, 18, 2267

10 of 13

value of MT5_1 is 35.91%, approximately triple the averaged value of MT3_1 (13.56%). Next, comparing Table 4 with Table 3, only the three averaged values of RT5_3 (3.32%), MT5_1 (35.95%) and MT5_2 (3.09%) have higher values. For the RMSD and MAPD averaged values, the number decreases from 7.28% to 3.32% and from 35.95% to 1.29%, respectively. For the CCD values, the results are the opposite, as the value increases from 2.06% to 2.73%, regardless of the fact that the impedance signature amplitudes are decreasing. Table 2. Three statistical metric values for the T3_1, T3_2 and T3_3 experiment. RMSD (%)

MAPD (%)

Damage (mm)

RT3_1

RT3_2

RT3_3

2 4 6 8 10 12 14 16 18 20 Average

5.33 7.09 10.53 8.82 8.61 9.41 9.65 10.46 10.87 11.03 9.18

1.82 2.12 2.36 2.86 3.13 2.92 2.93 3.25 4.24 3.83 2.95

1.39 1.18 2.04 2.89 2.21 2.11 2.30 2.58 2.20 2.65 2.16

CCD (%)

MT3_1 MT3_2 MT3_3 CT3_1 6.21 10.87 19.77 13.50 13.24 14.91 12.76 16.29 15.66 12.34 13.56

0.69 0.88 0.99 1.28 1.37 1.33 1.33 1.46 1.90 1.79 1.30

0.55 0.55 0.88 0.99 0.86 0.94 1.01 1.15 1.08 1.27 0.93

1.11 1.81 3.73 2.69 2.53 3.00 3.12 3.61 3.88 4.02 2.95

CT3_2

CT3_3

1.56 2.03 2.49 3.51 4.25 3.75 3.74 4.57 7.56 6.31 3.98

0.92 0.73 1.78 3.37 2.04 1.89 2.22 2.72 2.03 2.76 2.05

Numbers in different colors represent colors in picture.

Table 3. Three statistical metric values for the T4_1, T4_2 and T4_3 experiment. RMSD (%)

MAPD (%)

Damage (mm)

RT4_1

RT4_2

RT4_3

2 4 6 8 10 12 14 16 18 20 Average

5.61 7.39 9.38 8.85 9.25 7.67 9.06 10.94 13.28 13.77 9.52

2.86 2.92 5.66 5.67 6.36 5.97 5.49 5.20 7.63 8.30 5.61

2.19 3.57 3.05 3.13 3.27 3.14 3.05 3.30 3.48 4.66 3.28

CCD (%)

MT4_1 MT4_2 MT4_3 CT4_1

CT4_2

CT4_3

8.85 14.82 13.76 15.61 15.05 17.54 18.54 24.93 29.70 32.27 19.11

2.46 2.55 9.48 9.14 11.47 10.20 8.68 7.93 17.55 20.56 10.00

1.66 4.14 2.98 3.14 3.43 3.18 3.01 3.52 3.90 6.88 3.58

1.13 1.21 2.15 2.34 2.57 2.27 2.23 2.34 3.44 3.60 2.33

0.81 1.27 1.25 1.18 1.37 1.30 1.25 1.37 1.37 1.89 1.31

2.12 3.56 5.69 5.06 5.41 3.81 5.21 7.62 11.12 11.90 6.15

Numbers in different colors represent colors in picture.

Table 4. Three statistical metric values for the T5_1, T5_2 and T5_3 experiment. RMSD (%)

MAPD (%)

Damage (mm)

RT5_1

RT5_2

RT5_3

2 4 6 8 10 12 14 16 18 20 Average

5.67 4.97 7.09 8.07 8.44 8.39 6.35 8.18 7.44 8.23 7.28

1.66 3.17 3.32 3.49 3.39 3.37 4.35 4.91 4.23 4.68 3.66

2.38 2.24 2.81 3.21 3.25 3.26 3.56 3.52 4.24 4.72 3.32

CCD (%)

MT5_1 MT5_2 MT5_3 CT5_1 22.36 25.02 35.83 43.96 42.95 29.65 26.86 37.23 50.54 45.11 35.95

0.89 1.39 1.88 2.15 2.48 2.29 3.43 5.75 4.90 5.73 3.09

0.80 0.85 1.01 1.17 1.26 1.38 1.37 1.45 1.69 1.88 1.29

1.29 1.04 1.92 2.43 2.64 2.62 1.58 2.50 2.09 2.53 2.06

Numbers in different colors represent colors in picture.

CT5_2

CT5_3

0.61 1.71 1.85 2.02 1.92 1.91 3.06 3.83 2.90 3.50 2.33

1.42 1.29 1.91 2.43 2.50 2.52 2.98 2.91 4.16 5.13 2.73

Sensors 2018, 18, 2267

11 of 13

6.2. Regression Analysis on Experimental Data In this section, a linear regression analysis is performed using the results from the previous subsection to evaluate the performance of the EMI method subjected to different frequency ranges and statistical Sensors 2018, 18, x metrics FOR PEER(RMSD, REVIEWMAPD and CCD). Figure 9a shows the scatter plot using T3_1 results 11 of 13 from Table 2 (the 2nd, 5th and 8th columns). Three lines of best fit are drawn where the coefficient of determination ) is 0.67 for RMSD, for MAPD and for 0.66CCD. for CCD. As damage increases, all determination (R2)(R is2 0.67 for RMSD, 0.13 0.13 for MAPD and 0.66 As damage increases, all three 2 2 three show lines show an increasing trend where slopes have virtually thesame samesteepness. steepness.The TheRR of of 0.13 0.13 lines an increasing trend where the the slopes have virtually the obtained from the MAPD values is considerably lower than the other two statistical metrics, and one of the reasons for this is the randomness of the MAPD values. This This randomness randomness can can be be caused caused by of the the metal metal plate, which can have have aa significant significant effect on the damage to either a node or an anti-node of impedance signatures. In Figure 9b, the remainder of the results from the previous subsection are impedance signatures. In Figure 9b, the remainder of the results from subsection 2 2 values, where where the the three bars (RMSD, (RMSD, MAPD MAPD and and CCD) CCD) at T3_1 is of used to plot a bar graph of R values, By observation, observation, while while the the T3_x, T3_x, T4_x T4_x and and T5_x T5_x cases cases have have different resonance frequency frequency Figure 9a. By ranges, it seems that this difference does not have a significant impact on the regression analysis ranges, it seems that among the T3_x cases, the T3_2 case seems to have results as it appears to be random. For example, among values. Among Among the the T4_x T4_x cases, cases, the the T4_1 T4_1 case case has has the highest R22 values, values, while while of of the the T5_x highest R22 values. cases, the T5_3 case has the highest R22 values. values. These These results results show show the the complexity complexity of the EMI method analysis. However, However, one common common feature feature that can be noted is that the MAPD values are always data analysis. exception of of the the T3_1 T3_1 case. case. In In addition, addition, higher compared to the RMSD and the CCD values, with the exception the RMSD values are always higher than the CCD values except for the T5_2 case. Overall, the three statistical metrics metrics have have similar similar values values for for most most of of the cases, with the T5_3 case having the highest statistical values out of all the values. Another observation observation that that can can be made here is that when the impedance values also also decrease, decrease, whereas whereas the T5_x cases signature amplitude decreases from from T4_1 to T4_3, the R22 values 2 values increasing increasing from from T5_1 T5_1 to T5_3). This This shows shows that that the the size of the result in the exact opposite (R2 values 2 2 amplitudes does not have that significant an effect on the R values.

Figure and T5_x T5_x R R22 values. Figure 9. 9. Results Results for for (a) (a) T3_1 T3_1 statistical statistical metric metric values; values; (b) (b) T3_x, T3_x, T4_x T4_x and values.

7. Conclusions 7. Conclusions In this study, a structural health monitoring method known as the electromechanical impedance In this study, a structural health monitoring method known as the electromechanical impedance method was investigated, by subjecting a plate to progressive damage under different resonance method was investigated, by subjecting a plate to progressive damage under different resonance frequency ranges and peak amplitude heights. The first part of the study consisted of using a square frequency ranges and peak amplitude heights. The first part of the study consisted of using a square metal plate with a 15 mm square piezoelectric (PZT) transducer attached at the center plate. The test metal plate with a 15 mm square piezoelectric (PZT) transducer attached at the center plate. The test was to create 20 mm of progressive damage with impedance signatures being measured at every 2 was to create 20 mm of progressive damage with impedance signatures being measured at every 2 mm mm step, with 11 impedance signatures being acquired in total. This data was then compared with step, with 11 impedance signatures being acquired in total. This data was then compared with the the second test in which the damage created was identical to the first test, but this time with a free second test in which the damage created was identical to the first test, but this time with a free PZT PZT transducer attached in series to reduce the amplitudes of the impedance signature peaks. The transducer attached in series to reduce the amplitudes of the impedance signature peaks. The third third test was conducted with damage created identically to the previous tests, with another free PZT test was conducted with damage created identically to the previous tests, with another free PZT transducer attached to the series circuit. From these three tests (C_1, C_2 and C_3), 33 impedance transducer attached to the series circuit. From these three tests (C_1, C_2 and C_3), 33 impedance signatures were measured and these were analyzed using three different statistical metrics, which were root-mean-square deviation (RMSD), mean absolute percentage deviation (MAPD) and correlation coefficient deviation (CCD). The results of these tests show that with the decrease in amplitude with the attachment of additional PZT transducers, the RMSD and MAPD values were also decreased in general. However, for the CCD values, this effect was not seen, which

Sensors 2018, 18, 2267

12 of 13

signatures were measured and these were analyzed using three different statistical metrics, which were root-mean-square deviation (RMSD), mean absolute percentage deviation (MAPD) and correlation coefficient deviation (CCD). The results of these tests show that with the decrease in amplitude with the attachment of additional PZT transducers, the RMSD and MAPD values were also decreased in general. However, for the CCD values, this effect was not seen, which experimentally showed that the height of the resonance amplitudes was not a significant factor when analyzing the impedance signature data with a CCD statistical metric. Here, the results show that MAPD is the most appropriate method to use with large amplitudes as MC_1 results in 11.51% increasing up to 27.55% with 20 mm of damage. However, with small amplitudes, CCD performed the best with CC_3 starting from 1.06% increasing up to 5.97% with 20 mm damage. The second part of the study involved investigating the performance of the EMI method at different frequency ranges. To achieve this, the conventional method of attaching the PZT transducer was changed by using a metal disc, which was first introduced in Na et al. [4]. Three metal discs (T3_1 to T5_3) with different thicknesses were used to create 9 test cases. For each test, progressive damage was introduced in the same manner as in the first part of the study, and three tests were conducted for each metal plate. Afterwards, three statistical metrics were calculated, and showed a similar pattern to the first part of the study. The RMSD and MAPD values generally decreased with a decrease in the resonance amplitudes while the CCD values had no specific pattern, experimentally proving once again that the resonance amplitudes had very little effect on the CCD values. Here, the MAPD method performed the best with largest amplitudes as MT3_1 resulted in 12.34%, MT4_1 with 32.27% and MT5_1 with 45.11% when the test specimen was damaged up to 20 mm. In addition, the CCD method performed the best at 20mm damage with small amplitudes as CT3_3 resulted in 2.76%, CT4_3 with 6.88% and CT5_3 with 5.13%. This demonstrates that it is better to acquire an impedance signature with large amplitudes. Finally, linear regression analysis on all the RMSD, MAPD and CCD values obtained from the above tests was conducted to evaluate the performance of the EMI method. By fitting a line of best fit on all cases, it was found that the coefficient of determination (R2 ) value was highest for the MAPD values (8 out of 9 cases) compared to both the RMSD and CCD values. In addition, no significant pattern was observed when conducting the EMI method at different frequency ranges, which demonstrates the complexity of analyzing impedance signature data. One of the reasons for this is the sensitivity of the EMI method as many factors can cause the impedance signatures to change. These include the bonding condition of the epoxy adhesive, small differences in the dimensions of the metal plates, small temperature differences, etc. For this reason, research to compensate unwanted signature changes is an important area which should be researched further. However, the knowledge gained from this study can be used to understand the EMI method even further, bringing us one step closer to using this method for practical applications. Author Contributions: W.S.N. is the principal author for the research involved in writing and planning the overall work, D.-W.S. and B.-C.K. conducted the experiments and K.-T.P. analyzed the results from the experiments. Funding: The research was supported by a grant from “Evaluation techniques for cable system/earth anchor by micro and macro measured data (20180009-001)” funded by the Korea Institute of Civil engineering and building Technology (KICT), South Korea. Conflicts of Interest: Authors declare no conflict of interest.

References 1.

2.

Liang, C.; Sun, F.P.; Rogers, C.A. Coupled electromechanical analysis of adaptive material system determination of the actuator power consumption and system energy transfer. J. Intell. Mater. Syst. Struct. 1994, 5, 12–20. [CrossRef] Na, W.S. Distinguishing crack damage from debonding damage of glass fiber reinforced polymer plate using a piezoelectric transducer based nondestructive testing method. Compos. Struct. 2017, 159, 517–527. [CrossRef]

Sensors 2018, 18, 2267

3. 4.

5. 6.

7. 8. 9. 10.

11.

13 of 13

Na, W.S.; Lee, H. Experimental investigation for an isolation technique on conducting the electromechanical impedance method in high-temperature pipeline facilities. J. Sound Vib. 2016, 383, 210–220. [CrossRef] Na, S.; Tawie, R.; Lee, H.K. Electromechanical impedance method of fiber-reinforced plastic adhesive joints in corrosive environment using a reusable piezoelectric device. J. Intell. Mater. Syst. Struct. 2012, 23, 737–747. [CrossRef] Wandowski, T.; Malinowski, P.H.; Ostachowicz, W.M. Delamination detection in CFRP panels using EMI method with temperature compensation. Compos. Struct. 2016, 151, 99–107. [CrossRef] Yang, Y.; Lim, Y.Y.; Soh, C.K. Practical issues related to the application of the electromechanical impedance technique in the structural health monitoring of civil structures: I. Experiment. Smart Mater. Struct. 2008, 17, 035008. [CrossRef] Baptista, F.G.; Vieira Filho, J. A new impedance measurement system for PZT-based structural health monitoring. IEEE Trans. Instrum. Meas. 2009, 58, 3602–3608. [CrossRef] Bhalla, S.; Gupta, A.; Bansal, S.; Garg, T. Ultra low-cost adaptations of electro-mechanical impedance technique for structural health monitoring. J. Intell. Mater. Syst. Struct. 2009, 20, 991–999. [CrossRef] Panigrahi, R.; Bhalla, S.; Gupta, A. A low-cost variant of electro-mechanical impedance (EMI) technique for structural health monitoring. Exp. Tech. 2010, 34, 25–29. [CrossRef] Wandowski, T.; Malinowski, P.; Ostachowicz, W. Calibration problem of AD5933 device for electromechanical impedance measurements. In Proceedings of the EWSHM-7th European Workshop on Structural Health Monitoring, Nantes, France, 8–11 July 2014. Sun, F.P.; Chaudhry, Z.; Liang, C.; Rogers, C.A. Truss structure integrity identification using PZT sensor-actuator. J. Intell. Mater. Syst. Struct. 1995, 6, 134–139. [CrossRef] © 2018 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).

Suggest Documents