J Solid State Electrochem DOI 10.1007/s10008-013-2205-1
ORIGINAL PAPER
Tolerance of glucose electrocatalytic oxidation on NiOx/MnOx/GC electrode to poisoning by halides S. M. El-Refaei & M. M. Saleh & M. I. Awad
Received: 6 June 2013 / Revised: 24 July 2013 / Accepted: 28 July 2013 # Springer-Verlag Berlin Heidelberg 2013
Abstract This paper is devoted to the investigation of the tolerance of glucose oxidation on NiOx/MnOx/GC binary catalyst to poisoning by halide ions. NiOx/MnOx/GC electrode was prepared by the consecutive electrodeposition of MnOx followed by the deposition of NiOx nanoparticles atop the previously deposited MnOx nanoparticles. The electrode was characterized electrochemically and morphologically. Interestingly, experimental results showed that the NiOx/ MnOx/GC electrode shows a superior electrocatalytic activity compared with the NiOx/GC electrode and a behavior comparable to the NiOx/GC electrode regarding tolerance to poisoning by halides, even in the presence of high concentrations of Cl− and Br− up to 0.3 M. On the other hand, it was found that iodide ions have a significant effect on glucose electrooxidation under the prevailing operating conditions and even at concentrations lower than those of the other two halide ions. Keywords Glucose . Manganese oxide . Nickel oxide . Halide . Nanoparticles
Introduction Glucose electrooxidation on non-precious metals, especially Ni and its composites, as an anodic reaction in direct alkaline fuel cells and also as basis for enzymeless detection has been greatly developed in recent years due to their low cost, high oxidation capacity, and high stability [1–5]. One of the most challenging problems of glucose fuel cells and non-enzymatic glucose sensors is poisoning by halides especially chloride ions. So, studying the effect of halide ions on fuel oxidation on S. M. El-Refaei : M. M. Saleh : M. I. Awad (*) Department of Chemistry, Faculty of Science, Cairo University, Cairo, Egypt e-mail:
[email protected] M. M. Saleh e-mail:
[email protected]
the anode is considered a very important topic in electrocatalysis [3, 6–8]. It has been reported that halide ions exhibit a strong poisoning effect on glucose and methanol electrooxidation on noble metal electrodes such as Pt, Au, and their alloys [9–14]. The high poisoning of noble metals by halide ions is attributed to the high specific adsorption of the halide ions on such metals. In this context, a number of articles regarding this issue have been reported [9–14]. On the other hand, metal oxides such as CuOx and NiOx are susceptible to poisoning albeit to a lower extent compared to their metal counterparts. This is attributed to the difference in the adsorbability of halide ions on metal and metal oxides [15, 16]. In a previous report, it has been shown that the binary catalyst NiOx/MnOx/GC showed extraordinary enhancement of glucose electrooxidation in alkaline medium compared with NiOx/GC, MnOx/GC, or MnOx/NiOx/GC [4, 5]. As a continuation of our previous work, studying the effect of halide ion poisoning on our fabricated catalyst is important not only for application systems but also from a theoretical point of view. Generally, using non-enzymatic sensors in the analysis of glucose often suffers from the drawbacks of low sensitivity and interference of other oxidizable components such as ascorbic acid and uric acid [15]. The adsorption of chlorides, which decrease significantly the operational stability, is another challenge. It has been reported that CuOx-modified electrodes are a candidate for non-enzymatic catalyst for glucose oxidation in alkaline medium, and thus, poisoning by chloride has been extensively studied [15]. However, to the best of our knowledge, the effect of halide ion poisoning on the electrocatalytic activity of NiOx and/or MnOx nanoparticle-modified electrodes toward glucose oxidation is considered to be scarce. The purpose of the present work is to study the tolerance of glucose electrooxidation reaction on the proposed catalyst to poisoning by halides. The fabricated catalysts NiOx/GC and NiOx/MnOx/GC will be subjected to different concentrations of halide ions to study their impact on the performance of such electrodes toward glucose oxidation in alkaline medium.
J Solid State Electrochem
Experimental All chemicals used in this work were of analytical grade and were purchased from Merck and Sigma-Aldrich, and they were used as received without further purification. All solutions were prepared using second distilled water. Measurements Electrochemical characterizations were performed using an EG&G potentiostat (model 273A) operated with E-Chem 270 software. An electrochemical cell with a three-electrode configuration was used in this study. A platinum spiral wire and an Ag/AgCl/KCl (sat.) were employed as counter and reference electrodes, respectively. All potentials are presented with respect to the reference electrode. The working electrode was glassy carbon (diameter=3.0 mm). It was cleaned by mechanical polishing with aqueous slurries of successively finer alumina powder (down to 0.06 μm) and then washed thoroughly with second distilled water. Scanning electron microscope (SEM) images were taken using a field emission scanning electron microscope (FEI, QUANTA FEG 250). Electrode modification NiOx/MnOx/GC was prepared by consecutive electrodeposition of MnOx followed by the deposition of NiOx nanoparticles atop the previously deposited MnOx nanoparticles onto a glassy carbon electrode (GCE). MnOx was electrodeposited by cycling the potential from 0 to 0.4 V at a scan rate of 20 mV/s in 0.1 M Na2SO4 containing 0.1 M Mn(CH3COO2)2·5H2O for 60 cycles, and then it was activated by cycling the potential for 5 cycles in 0.5 M NaOH solution in the potential range −0.2 to 0.6 V at a scan rate of 100 mV/s [17]. Next, modification with NiO x was achieved in two sequential steps, i.e., the potentiostatic deposition of metallic nickel on the previously deposited MnOx (i.e., onto MnOx/GCE) from an aqueous solution of 0.1 M acetate buffer solution (pH=4.0) containing 1 mM Ni(NO3)2·6H2O by applying constant-potential electrolysis at −1.0 V for different time durations. Then the passivation of the thus electrodeposited Ni was achieved in 0.1 M phosphate-buffered solution (pH=7) by cycling the potential between −0.5 and 1 V for 10 cycles at a scan rate of 200 mV/s. Next, activation for 5 cycles in 0.5 M NaOH solution in the potential range −0.2 to 0.6 V was conduted [18, 19].
Results and discussion Material characterization Figure 1 shows the energy-dispersive X-ray (EDX) spectrum of the NiOx/MnOx/GC electrode prepared as mentioned in the
“Experimental” section. This figure reveals the peaks of manganese and nickel at 5.9 and 7.5 keV in quantities of 5.4 and 0.5 by weight percent, respectively. This sustains that both NiOx and MnOx are exposed in the same sample and that NiOx does not fully cover the previously deposited MnOx, consistent with the SEM image shown as inset in which the wellknown porous texture of MnOx is obscured by the subsequent deposition of NiOx [20]. More discussion regarding the SEM images of the different catalysts can be found elsewhere [4, 5]. Glucose electrooxidation Glucose electrooxidation reaction on the NiOx and MnOx single and binary catalyst-modified glassy carbon electrode at different orders and loading levels of the two oxides has been previously reported by our group [4, 5]. It has been reported that the NiOx/MnOx/GC electrode (MnOx deposited first) shows an excellent electrocatalytic activity and stability toward glucose oxidation compared with NiOx/GC, MnOx/ GC, or MnOx/NiOx/GC electrodes (NiOx deposited first). In those reports, we also found that the optimum loading level is 60 potential cycles of MnOx followed by 10 min of nickel deposition. Thus, in the present study, those conditions were selected for the preparation of the NiOx/MnOx/GC electrode as cited in the “Experimental” section. The extraordinary activity obtained on this electrode was attributed to the compilation of the better adsorption of glucose molecules on the MnOx sites and the increase of conductivity of NiOx due to the increase of Ni3+ content, taking in consideration that both oxides are exposed to the solution as has been reported [4, 5]. In the present work, the poisoning effect of halide ions on glucose electrooxidation on the proposed electrocatalyst (NiOx/MnOx/GC) will be presented. It is noteworthy to mention that NiOx/MnOx/GC has been reported to show extraordinary electrocatalytic glucose oxidation compared with the single catalyst (NiOx/GC) [4, 5]. Thus, the study on halide tolerance is essential for this electrode being applied in fuel cells and in non-enzymatic detection of glucose. As the effect of chloride and bromide ions is expected to be significantly different from that of iodide, results regarding chloride and bromide ions will be presented as one set and then poisoning by iodide ions as another set. Poisoning effect of chloride and bromide ions Figure 2 shows cyclic voltammetry (CV) responses for glucose electrooxidation on the NiOx/MnOx/GC electrode in the presence of different concentrations of (a) chloride and (b) bromide ions, ranging from 0.1 to 0.3 M. Glucose oxidation is revealed at a potential around 0.53 V. It has been reported that NiOOH acts as a mediator of glucose oxidation. The redox transition of nickel species from Ni(II) to Ni(III) occurs, and then in a following step, glucose is oxidized on the modified
J Solid State Electrochem
Count
Fig. 1 EDX spectrum of the NiOx/MnOx/GC electrode and (inset) SEM image of the same electrode. NiOx/MnOx/GC was prepared as described in the “Experimental” section
surface, i.e., via EC mechanism. Gluconolactone as well as methanoates and oxalates, as a result of the further oxidation of gluconolactone, has been reported as the possible oxidation product of glucose electrooxidation [5, 21–23]. Interestingly, upon the addition of the corresponding halide ions, chloride or bromide ions, there is no observed significant effect on the forward and backward scans even at high concentrations of the two halide ions. It has been reported [16] that the NiOx/GC electrode has strong tolerance to chloride ions which is consistent with the present results. However, the present electrode has the advantage of higher electrocatalytic activity compared to the NiOx electrode. This indicates that the proposed binary catalyst exhibits good resistance to surface fouling and is a promising electrocatalyst for the development of enzymeless glucose sensors and alkaline fuel cells at low cost. The above recorded results are the first run after the direct addition of halide ions. In order to further demonstrate the poisoning effect, various numbers of potential scans (cycles) were recorded in the presence of both the glucose and the halide ions. Similar scans were recorded in halide-free 20 mM glucose in 0.5 M NaOH solution for the sake of comparison. Figure 3 shows consecutive linear sweep voltammetry (LSV) responses for the forward (I) and backward (II) scans obtained on the NiOx/ MnOx/GC electrode in 0.5 M NaOH containing 20 mM glucose (halide-free solution) at a scan of rate 100 mV/s. To quantitatively demonstrate the effect of halide ions during potential scan cycling, we can correlate the peak current, which corresponds to glucose oxidation, to the number of potential cycles as it appears in Fig. 4 (the data were taken from Fig. 3). In Fig. 3 the peak current firstly decreases to a minimum value at run 4, and then it increases again to values close to the
initial peak current. This may be attributed to the fact that the electrode surface contains more than one type of active sites with varied activity toward glucose electrooxidation. At the start of the first four runs, glucose oxidation begins to occur at all active sites, but some of the active sites are poisoned by the oxidation products much faster than the other sites. Hence, the peak current decreases reaching the minimum value, albeit it is still high due to glucose oxidation at the other active sites on the electrode surface. The current begins to increase again and continues in a varied manner as the scan number increases. This increase may be attributed to many factors such as the poisoning species on the electrode surface that might be oxidatively recovered after a number of runs due to the regeneration of poisoned active sites. Also, Ni+2/Ni3+ redox couple transformation increases in the alkaline medium by increasing the potential scan cycles. The last fact is illustrated in Fig. 5 in which further potential cycling of the previously activated NiO x /MnO x /GC electrode in 0.5 M NaOH solution is conducted. Run 1 is recorded after activation of the catalyst in 0.5 M NaOH for 25 cycles to obtain the well-defined redox waves of the Ni2+/Ni3+ transformation system. The peak current of the Ni+2/Ni3+ couple increases with increasing potential cycling in glucose-free 0.5 M NaOH. This provides more active sites to glucose oxidation, and thus, the current increases again. Also, Fig. 3 proves that the initial decrease in the peak current is about 10 μA (2 % decrease from the original peak current), and this is considered negligible. Thus, the results reveal that the electrode response toward glucose oxidation is mainly constant within 15 cycles. Figure 6a, b shows the consecutive LSV responses for the forward scan (I) and backward scan (II) obtained on the NiOx/
J Solid State Electrochem
600
500
I/ A
400
560
chloride free glucose 0.1 M Cl0.2 M Cl0.3 M Cl-
I/ A
a
300
200
540
520
100
0
500 0
-100 -400
-200
0
200
400
600
6
8
10
12
14
16
Number of runs
Fig. 4 Relation between peak current for the forward scan at each run and number of runs (data were taken from Fig. 3)
600
500
400
I/ A
4
800
E / mV vs. Ag/AgCl/KCl (sat.)
b
2
0.1 M Br0.2 M Br0.3 M Brbromide free glucose
ions, the oxidation peak currents for the forward scan in the halide-free glucose and after adding Cl− and Br− are plotted as a function of the number of potential cycles as shown in Fig. 7. Many features could be extracted from Fig. 7 in comparison with Fig. 6:
300
200
100
0
-100 -400
-200
0
200
400
600
800
E / mV vs. Ag/AgCl/KCl (sat.)
Fig. 2 CV responses for glucose electrooxidation on the NiOx/MnOx/GC electrode in 0.5 M NaOH containing 20 mM glucose in the presence of different concentrations of the halide ions, Cl− (a) and Br− (b), at a scan rate of 100 mV/s
MnOx/GC electrode in 0.5 M NaOH containing 20 mM glucose and 0.1 M Cl− (a) and 0.1 M Br− (b), respectively. To quantitatively compare the poisoning effect of Cl− and Br−
1. In the presence of Cl− or Br− ions, the oxidation peak current for the first cycle is similar to that in the absence of both species. With potential cycling, the peak current decreases in the presence of any of the two species, i.e., Cl− or Br−, albeit to different extent. 2. The decreasing rate of the oxidation peak current follows the sequence Br− >Cl− >halide-free (glucose only) solution. This expected behavior is due to the difference between Cl− and Br− ions in electronegativity, ionic radii, and solvation ability [24]. 3. The peak currents change with the potential cycles with a regime similar to that obtained in the case of the halidefree (glucose only) solution (see Fig. 4). However, 80
d c b a
60 40
200 A
20
I/ A
1-15 scan I
0 -20
1-15 scan -40
II
-60 -80 -100
-200
0
200
400
600
E / V vs. Ag/AgCl/KCl (sat.)
Fig. 3 Consecutive LSV responses for the forward scan (I) and backward scan (II) obtained on the NiOx/MnOx/GC electrode in 0.5 M NaOH containing 20 mM glucose (halide-free)
-200
0
200
400
600
E / V vs. Ag/AgCl/KCl (sat.)
Fig. 5 A sequence of 15 potential cycles obtained in 0.5 M NaOH on the NiOx/MnOx/GC electrode, previously activated by 25 potential cycles in 0.5 M NaOH. (a) First, (b) fifth, (c) tenth, and (d) fifteenth cycles
J Solid State Electrochem
a 200 µA
1-15 scan
I 1-15 scan II
The above results reveal the tolerance of glucose oxidation on the NiOx/MnOx/GC electrode to poisoning by chloride and bromide ions. This may be attributed to the low adsorbability and high solvation of Cl− and Br− ions. Also, it could be attributed to the high reversible oxidation potentials of Cl2/C1− and Br2/Br− reactions (1.36 and 1.06 V vs. normal hydrogen electrode (NHE), respectively) [25]. Effect of iodide ion
-200
0
200
400
600
E / mV vs. Ag/AgCl/KCl (sat.)
b 200 µ A
1-15 scan
I 1-15 scan II
-200
0
200
400
600
E / mV vs. Ag/AgCl/KCl (sat.)
Fig. 6 Consecutive LSV responses for the forward scan (I) and backward scan (II) obtained on the NiOx/MnOx/GC electrode in 0.5 M NaOH containing 20 mM glucose in the presence of 0.1 M Cl− (a) and 0.1 M Br− (b)
recovery in the presence of Cl− is faster than that in the presence of Br−. Note that in the presence of Cl−, the initial peak current is completely recovered, while in the case of Br−, it is not. 4. The largest decrease in the peak current is about 22 μA for the Cl− ions and 26 μA for the Br− ions, i.e., 4.0 and 4.7 % of the original current, respectively.
Figure 8 shows the consecutive LSV responses for the forward (I) and backward (II) scans obtained on the NiOx/MnOx/GC electrode in 0.5 M NaOH containing 20 mM glucose and different iodide ion concentrations ranging from 10−5 to 10−2 M. Note that the concentration range used for the effect of iodide is much lower than that used above for Cl− and Br−. This is because I− has significant adsorbability, and so at high concentrations, it blocks the electrode surface active sites. In the iodide-free glucose case (curve a), the forward scan current corresponds to the combination of two reactions, i.e., oxidation of Ni2+ to Ni3+ and glucose electrooxidation by interaction with Ni3+. The backward scan corresponds to the combination of Ni3+ reduction to Ni2+ and the glucose electrooxidation at the already present Ni3+ [1–5]. The literature reported that the iodide ion is oxidatively adsorbed by Pt and Au electrodes as zerovalent atomic iodine at potentials between −0.4 and +0.4 V (Ag/AgCl reference). The amount of adsorbed iodine increases upon scanning the potential to values more positive than −0.4 V until the surface is saturated with a monolayer of close-packed iodine atoms with coverage limited by van der Waals interactions. Additional iodine atoms forced into the already spacelimited interfacial layer lead to the formation of molecular iodine, which is evolved into the solution as aqueous I2 e d a, b, c
200 A
glucose
a, b c d I
Cl-
e II
Br-
-400
-200
0
200
400
600
800
E / mV vs. Ag/AgCl/KCl (sat.)
Fig. 7 The relation between peak current at each run and number of runs (data were taken from Fig. 6)
Fig. 8 Consecutive LSV responses for the forward scan (I) and backward scan (II) obtained on the NiOx/MnOx/GC electrode in 0.5 M NaOH containing 20 mM glucose at different iodide ion concentrations: (a) 0.0, (b) 10−5, (c) 10−4, (d) 10−3, and (e) 10−2 M. Scan rate, 100 mV/s
J Solid State Electrochem 140 120
e
100 80
I/ A
(2I−↔I2 +2e). On the surface of Pt and Au anodes, at considerably more positive potentials and before oxygen evolution, the adsorbed zerovalent iodine is oxidized to aqueous iodate (I2↔IO3−) along with the oxide formation of the electrode surface. Finally, it is quite probable that the iodates formed on the electrode diffuse back into the solution and react with the iodide, forming iodine again [26–28]. In Fig. 8 upon the addition of the iodide ions, a significant effect appears in the forward and backward scans at each concentration (curves b–e). For 10−5 M I− (curve b), the LSV responses for the forward and backward scans are similar to those for the iodide-free glucose solution (curve a), and there is no significant change in both the forward and backward scans. In general, there are two main reactions (stated above) that contribute to the forward current. The current in the forward and backward directions is considered as a combination of the previously mentioned processes. For 10−4 M I− (curve c), again the forward scan is similar to that for the iodide-free glucose solution while the backward current is lower than that for the iodide-free glucose solution (curve a). The forward scan current does not decrease as was expected by the poisoning effect of iodide. This can be understood if we consider that the decrease in the current of glucose oxidation due to poisoning by adsorbed iodide is overweighed by the oxidation of the adsorbed I2 to IO3−. Hence, the current in the forward scan is kept almost unchanged. On the other hand, the decrease in the backward current in the presence of iodide ions might be attributed to the adsorption of iodine or iodates formed in the forward scan that block some of the active sites. Another factor is the presence of another reduction reaction of iodide derivatives in parallel to Ni3+ reduction; the combination of these two processes can decrease the overall oxidation current of glucose in the backward scan, so glucose oxidation current in the backward scan decreases in comparison with the iodide-free glucose case. In the case of 10−3 M I− (curve d), the forward scan current is higher than that in the iodide-free glucose case and the backward scan current is lower than that in the iodide-free glucose case (curve a). The increase in the forward scan may be attributed to the oxidation of the adsorbed zerovalent iodine to aqueous iodine molecules and/or iodates as they are available in this potential range as revealed from Fig. 9 (curve d), in which the CV responses are obtained on NiOx/MnOx/GC in 0.5 M NaOH at different I− concentrations, where the current increases between 0.4 and 0.6 V on NiOx/MnOx/GC during the forward scan in the presence of 10−3 M I−. The decrease in the backward scan current (Fig. 8, curve d) is attributed to the same reasons as in the case of 10−4 M I−. However, the decrease in the present case is more significant as the adsorbed species in the forward scan is higher in the present case compared with the case of 10−4 M I−. In the case of 10−2 M I− (curve e), where the concentration of iodide ions becomes close to the glucose concentration, the
60 40
d
20
a, b, c
0 -20 -40
-200
0
200
400
600
E / mV vs. Ag/AgCl/KCl (sat.)
Fig. 9 CV responses obtained on NiOx/MnOx/GC in 0.5 M NaOH at different I− concentrations: (a) 0.0, (b) 10−5, (c) 10−4, (d) 10−3, and (e) 10−2 M. Scan rate, 100 mV/s
transformation from adsorbed iodine to aqueous iodine and/or iodates becomes more considerable. Thus, the forward scan current becomes higher than that in the iodide-free glucose and lower than that in the case of 10−3 M I−. The increase in curve e in Fig. 8 covers a wide potential range from the onset potential of the peak to 0.65 V, in accordance with that found in Fig. 9 where no glucose was added. The increase in current in the forward scan in the presence of 10−2 M I− is not much higher than that in the case of 10−3 M I−; this could be attributed to glucose oxidation being more inhibited in the forward scan than in the case of 10−3 M I−, so the current difference between the cases of 10−3 and 10−2 M I− in the presence of glucose is not significant. In the backward scan (Fig. 8, curve e), the peak current becomes much lower than that in the case of 10−3 M I−, as the poisoning in this case is much higher than that in the cases of 10−3 and 10−4 M I−. Figure 10a–c shows consecutive LSV responses for the forward scan (I) and backward scan (II) obtained on the NiO x /MnO x /GC electrode in 0.5 M NaOH containing 20 mM glucose and 10−4 (a), 10−3 (b), and 10−2 (c) M I−, respectively. We can see that with the increase in the number of potential cycles, both the forward and backward currents decrease, and this decrease depends on the iodide concentration. In addition, with the increase in the concentration of I−, a small peak appears in the backward scans (II) in cases (b) and (c) with increasing potential scan cycles especially in case (c). That peak may be attributed to the reduction of the remaining NiOOH, as the surface poisoning by iodates and iodine hinders glucose from interacting with Ni3+. Figure 11 shows the relation between the peak current for the forward scan (Ip) with the number of potential cycles (data were taken from Fig. 10). This figure demonstrates an interesting falling dependence of the curve of the oxidation peak current with the potential sweep number and with the I− concentration. It is noteworthy to mention that the peak
J Solid State Electrochem
Number of run increase 1
a
a b
15
c
1
I
d
15
II -200
0
200
400
Fig. 11 Relation between peak current for the forward scan at each run and number of runs (data were taken from Fig. 10)
600
E / mV vs. Ag/AgCl/KCl (sat.)
Number of run increase 1
b 200 µA
15 1 I 15 II -200
0
200
400
600
E / mV vs. Ag/AgCl/KCl (sat.)
Number of run increase 1
c 200 µA
15 I
1
II
15
-200
0
200
400
600
E / mV vs. Ag/AgCl/KCl (sat.)
Fig. 10 Consecutive LSV responses for the forward scan (I) and backward scan (II) obtained on the NiOx/MnOx/GC electrode in 0.5 M NaOH containing 20 mM glucose at 10−4 (a), 10−3 (b), and 10−2 (c) M I−. Scan rate, 100 mV/s
current for the first run in cases (c) and (d) is higher than the blank itself (curve a). This behavior was attributed to adsorbed iodine oxidation as it is available in this potential range. In the following potential cycles, the peak current decreases with the potential scan number to values lower than that of the blank. The fact that the peak current decreases again after the first run, although two reactions occur (glucose oxidation and adsorbed iodine oxidation), could be explained by considering the cross effect of both glucose and I− oxidation on each other. That is to say, the glucose oxidation reaction is retarded by the adsorbed products of iodine oxidation, and the iodide oxidation reaction may retard further oxidation of I− to I2. In Fig. 11 the three curves in the presence of I− ions exhibit two decay stages, an initial rapid stage which depends on the concentration of I− ions followed by a slow one. During the initial decay, the glucose oxidation peak current (Ip) rapidly falls to ca. 90.0, 84.4, and 65.4 % of its original value for cases (b), (c), and (d), respectively, suggesting a quick poisoning process especially with the high I− concentration. This rapid stage is followed by a slow one in which the decrease in the peak current is negligible. For instance, for case (d), the peak current decreases from 343 to 332 μA (about 11 μA) that represents about 2 % of the whole current value. It becomes clear that the poisoning effect of chloride and bromide ions differs greatly from the poisoning effect of the iodide ion, and this difference may be attributed to the difference in their adsorbability and reactivity. It is likely that the radii and the electronegativity of the halide ions have a profound influence on their adsorption process. Electronegativity increases from I− to Cl− (I− =2.5, Br− =2.8, Cl− =3.0), while atomic radius decreases from I − to Cl − (I − = 135 pm, Br− = 114 pm, Cl− = 90 pm). So, the iodide ion is more predisposed to adsorption than bromide and chloride, and this is the reason behind their difference in poisoning effect [24]. In parallel to these and from a thermodynamic point of view, potentials for reversible Cl2/C1−, Br2/Br−, and I2/I− reactions are 1.36, 1.06, and 0.54 V vs. NHE, respectively. This also
J Solid State Electrochem
explains why chloride and bromide ions do not affect the glucose oxidation reaction even at high concentrations. While for the iodide ion, the state is very different as iodide can be oxidized easily to iodine or iodates that are also possible in the present experimental conditions [25–28].
Conclusions Halide ions have a significant poisoning effect on the electrocatalytic activity of the NiOx/MnOx/GC electrode toward glucose oxidation in alkaline medium. The poisoning degree of halide ions largely depends on their chemical nature and concentration in solution. For chloride and bromide ions, our fabricated electrode has high tolerance toward them even at high concentrations up to 0.3 M. While for the iodide ion, it was found that it has a significant effect on glucose oxidation in our operating conditions at very low concentrations in comparison with the concentrations of the other two halide ions. The present catalyst could serve as a catalyst for glucose oxidation and non-enzymatic sensor for glucose even in the presence of a large concentration of Cl−.
References 1. Lee KK, Loh PY, Sow CH, Chin WS (2012) CoOOH nanosheets on cobalt substrate as a non-enzymatic glucose sensor. Electrochem Commun 20:128–132 2. Cherevko S, Chung CH (2009) Gold nanowire array electrode for non-enzymatic voltammetric and amperometric glucose detection. Sens Actuators, B 142:216–223 3. Jeong H, Kim J (2012) Electrochemical oxidation of glucose at nanoporous black gold surfaces in the presence of high concentration of chloride ions and application to amperometric detection. Electrochim Acta 80:383–389 4. El-Refaei SM, Saleh MM, Awad MI (2013) Enhanced glucose electrooxidation at a binary catalyst of manganese and nickel oxides modified glassy carbon electrode. J Power Sourc 223:125–128 5. El-Refaei SM, Awad MI, El-Anadouli BE, Saleh MM (2013) Electrocatalytic glucose oxidation at binary catalyst of nickel and manganese oxides nanoparticles modified glassy carbon electrode: optimization of the loading level and order of deposition. Electrochim Acta 92:460–467 6. Pasta M, Mantia FL, Cui Y (2010) A new approach to glucose sensing at gold electrodes. Electrochem Commun 10:1407–1410 7. Park S, Boo H, Chung TD (2006) Electrochemical non-enzymatic glucose sensors. Anal Chim Acta 556(1):46–57 8. Pasta M, Mantia FL, Cui Y (2010) Mechanism of glucose electrochemical oxidation on gold surface. Electrochim Acta 55(20):5561– 5568 9. Hsiao MW, Adzic RR, Yeager EB (1992) The effects of adsorbed anions on the oxidation of D-glucose on gold single crystal electrodes. Electrochim Acta 37(2):357–363
10. Hsiao MW, Adzic RR, Yeager EB (1996) Electrochemical oxidation of glucose on single crystal and polycrystalline gold surfaces in phosphate buffer. J Electrochem Soc 143(3):759–767 11. Li Y, Song Y, Yang C, Xia X (2007) Hydrogen bubble dynamic template synthesis of porous gold for non-enzymatic electrochemical detection of glucose. Electrochem Commun 9(5):981–988 12. Seo B, Kim J (2010) Electrooxidation of glucose at nanoporous gold surfaces: structure dependent electrocatalysis and its application to amperometric detection. Electroanalysis 22(9):939–945 13. Kamath VN, Lal H (1970) Halide adsorption and the anodic oxidation of chemisorbed methanol on platinum. J Electroanal Chem 24(1):125–135 14. Sobkowski J, Franaszczuk K, Dobrowolska K (1992) Effect of anions and pH on the adsorption and oxidation of methanol on a platinum electrode. J Electroanal Chem 330(1–2):529–540 15. Luo S, Su F, Liu C, Li J, Liu R, Xiao Y, Li Y, Liu X, Cai Q (2011) A new method for fabricating a CuO/TiO2 nanotube arrays electrode and its application as a sensitive nonenzymatic glucose sensor. Talanta 86:157–163 16. Liu Y, Teng H, Hou H, You T (2009) Nonenzymatic glucose sensor based on renewable electrospun Ni nanoparticle-loaded carbon nanofiber paste electrode. Biosens Bioelectron 24(11):3329–3334 17. El-Deab MS, Ohsaka T (2006) Manganese oxide nanoparticles electrodeposited on platinum are superior to platinum for oxygen reduction. Angew Chem Int Ed 45(36):5963–5966 18. Giovanelli D, Lawrence NS, Jiang L, Jones TGJ, Compton RG (2003) Electrochemical determination of sulphide at nickel electrodes in alkaline media: a new electrochemical sensor. Sensor Actuator B Chem 88(3):320–328 19. Giovanelli D, Lawrence NS, Wilkins SJ, Jiang L, Jones TGJ, Compton RG (2003) Anodic stripping voltammetry of sulphide at a nickel film: towards the development of a reagentless sensor. Talanta 61(2):211– 220 20. Meng M, Tu Y, Ding T, Sun Z, Zhang L (2011) Effect of synthesis pH and Au loading on the CO preferential oxidation performance of Au/ MnOx–CeO2 catalysts prepared with ultrasonic assistance. Int J Hydrogen Energ 36(15):9139–9150 21. Zhao C, Shao C, Li M, Jiao K (2007) Flow-injection analysis of glucose without enzyme based on electrocatalytic oxidation of glucose at a nickel electrode. Talanta 71:1769–1773 22. Becerik I, Kadirgan F (1992) The electrocatalytic properties of palladium electrodes for the oxidation of D-glucose in alkaline medium. Electrochim Acta 37:2651–2657 23. Mho SI, Johnson DC (2001) Electrocatalytic response of carbohydrates at copper-alloy electrodes. J Electroanal Chem 500:524–532 24. Khamis A, Saleh MM, Awad MI (2013) Synergistic inhibitor effect of cetylpyridinium chloride and other halides on the corrosion of mild steel in 0.5 M H2SO4. Corros Sci 66:343–349 25. Novak DM, Conway B (1981) Competitive adsorption and state of charge of halide ions monolayer oxide film growth processes at Pt anodes. J Chem Soc Faraday Trans 1 77(10):2341–2359 26. Anson FC, Lingane JJ (1957) Anodic chronopotentiometry with platinum and gold electrodes. The iodide–iodine–iodate system. J Am Chem Soc 79(5):1015–1020 27. Bravo BG, Michelhaugh SL, Soriaga MP, Villegas I, Suggs DW, Stickney JL (1991) Anodic underpotential deposition and cathodic stripping of iodine at polycrystalline and single-crystal gold: studies by LEED, AES, XPS, and electrochemistry. J Phys Chem 95(13):5245– 5249 28. Verhoef JC (1978) Electrochemical behaviour of iodide at a rotating platinum disk electrode in methanol. Electrochim Acta 23(5):433– 438