Clinical Chemistry / Reduction in Labeling Errors in Phlebotomy

Reduction in Specimen Labeling Errors After Implementation of a Positive Patient Identification System in Phlebotomy Aileen P. Morrison,1,2 Milenko J. Tanasijevic, MD, MBA,2 Ellen M. Goonan, MS, MT(ASCP)SH,2 Margaret M. Lobo,2 Michael M. Bates,1 Stuart R. Lipsitz, ScD,1 David W. Bates, MD, MSc,1 and Stacy E.F. Melanson, MD, PhD2

CME/SAM

Key Words: Phlebotomy; Positive patient identification; Medical errors; Specimen labeling DOI: 10.1309/AJCPC95YYMSLLRCX

Upon completion of this activity you will be able to: • describe specimen identification errors and their consequences. • discuss the attributes of a positive patient identification system and the effects of implementing such a system. • analyze a system’s ability to detect and prevent specimen identification errors.

Abstract Ensuring accurate patient identification is central to preventing medical errors, but it can be challenging. We implemented a bar code–based positive patient identification system for use in inpatient phlebotomy. A before-after design was used to evaluate the impact of the identification system on the frequency of mislabeled and unlabeled samples reported in our laboratory. Labeling errors fell from 5.45 in 10,000 before implementation to 3.2 in 10,000 afterward (P = .0013). An estimated 108 mislabeling events were prevented by the identification system in 1 year. Furthermore, a workflow step requiring manual preprinting of labels, which was accompanied by potential labeling errors in about one quarter of blood “draws,” was removed as a result of the new system. After implementation, a higher percentage of patients reported having their wristband checked before phlebotomy. Bar code technology significantly reduced the rate of specimen identification errors.

870 870

Am J Clin Pathol 2010;133:870-877 DOI: 10.1309/AJCPC95YYMSLLRCX

The ASCP is accredited by the Accreditation Council for Continuing Medical Education to provide continuing medical education for physicians. The ASCP designates this educational activity for a maximum of 1 AMA PRA Category 1 Credit ™ per article. This activity qualifies as an American Board of Pathology Maintenance of Certification Part II Self-Assessment Module. The authors of this article and the planning committee members and staff have no relevant financial relationships with commercial interests to disclose. Questions appear on p 978. Exam is located at www.ascp.org/ajcpcme.

The recognition that preventable medical errors cause tens of thousands of US deaths annually has spurred a patient safety movement that seeks to create systems that are inherently error-preventing.1,2 The application of technology is considered integral to this process.3 Within laboratory medicine, as analytic quality has improved, the preanalytic and postanalytic stages of testing have emerged as the largest sources of laboratory error and, thus, prime targets for improvement.4-7 Accurate patient and specimen identification within the preanalytic stage is crucial; misidentification can have numerous consequences, such as an invasive procedure performed on the wrong patient,8 a result reported for the wrong patient, and missed or delayed diagnoses. Worst-case scenarios include missed cancer, unnecessary mastectomy,9 and hemolytic transfusion reactions.10 Specimen identification errors, such as mislabeled or unlabeled specimens, are detected at a rate of 0.04% to 0.1%,11-13 and many more may go undetected.14 In 2003, the Joint Commission announced accurate patient identification as the first of 6 National Patient Safety Goals.15 The Joint Commission also suggests the use of automated systems to prevent patient misidentification.16 Positive patient identification (PPID) refers to the correct initial identification of a patient and the absolute linking of all specimens to that patient throughout the entire testing process, including collection, analysis, and reporting.17 Two major technologies, bar coding and radiofrequency identification, are currently used for electronic PPID (EPPID). A typical bar code–based EPPID system consists of bar coded patient wristbands, handheld computers onto which orders are downloaded, bar code scanners used to confirm patient identity before © American Society for Clinical Pathology

Clinical Chemistry / Original Article

blood specimen collection, and portable printers to generate labels at the bedside. Early results from hospitals that have implemented EPPID systems seem promising.18-22 In 2000, the Valley Hospital, Ridgewood, NJ, implemented an EPPID system and reported an 82% decrease in the rate of incorrect or incomplete labels from 0.017% to 0.003%.18 St Jude Children’s Research Hospital, Memphis, TN, reported an 84% decrease in the incidence of mislabeled specimens from 0.032% to 0.005%.22 However, no study that we are aware of has shown a statistically significant effect of using EPPID for general specimen collection in a nonpediatric, large, academic hospital. Therefore, we evaluated the impact of a bar code–based bedside labeling system on the rate of specimen labeling errors within the inpatient phlebotomy service of a large, academic medical center.

Materials and Methods Study Site The study was performed at Brigham and Women’s Hospital, Boston, MA, a 777-bed academic medical center. The inpatient phlebotomy service at Brigham and Women’s Hospital is a deployed team that covers all inpatient care areas except the neonatal intensive care units and obtains approximately 50% of inpatient blood samples. In 2008, nearly 215,000 blood samples were drawn by the inpatient phlebotomy team. The Partners Human Resource Committee found this study exempt from board oversight. Tracking Mislabeled and Unlabeled Specimens The primary outcome measure for this study was the number of inpatient phlebotomy samples reported as mislabeled or unlabeled by the laboratory. The laboratory typically recognized an incorrectly labeled sample at the time of specimen receipt when laboratory staff compared identifiers on the specimen label with those on the accompanying requisition form. The sample was recorded as mislabeled if patient identifiers on the specimen label did not match those on the accompanying requisition form or patient identifiers on the label and requisition matched but it was later determined that the sample was obtained from a different patient. If a specimen arrived with no label, it was recorded as unlabeled. The percentage of total inpatient phlebotomy blood samples incorrectly labeled was calculated for each study period: preimplementation, from October 1, 2007, to July 31, 2008; implementation, from August 1 to 31, 2008; and postimplementation, from September 1, 2008, to June 30, 2009. Only errors committed by the inpatient phlebotomy team were included in the data.

Estimation of Prevented Mislabeling Events We used 2 different methods to estimate the number of mislabeled specimens prevented by the EPPID system in 1 year. For the first estimate, we assumed that any difference in the number of errors per accession postimplementation vs preimplementation was attributable to errors prevented by EPPID. Accordingly, we calculated the number of errors that would have occurred in the 10-month postimplementation period if the preimplementation error rate had persisted unchanged. We then deducted the actual number of errors reported in the 10-month postimplementation period and extrapolated the result to 12 months. We used audit data collected by EPPID to produce the second estimate. If, when scanning a patient’s wristband before specimen collection, the system detected a mismatch with the patient initially selected in the personal digital assistant (PDA) whose sample was to be drawn, an alert sounded and the phlebotomist was required to resolve the mismatch before proceeding. We compiled a tally of all cases during the 10-month postimplementation period in which this alert prevented a wrongpatient “draw” and extrapolated this number to 12 months. Phlebotomy Workflow The specimen labeling policy before EPPID implementation relied on a nurse or unit coordinator to manually generate patient addressograph labels and attach each to its correct requisition in preparation for phlebotomy rounds. A phlebotomist making rounds would check that the preprinted labels matched the requisition, were legible, and were of sufficient number, and create new labels if necessary. After EPPID implementation, the practice of preprinting labels was discontinued and new steps specific to EPPID were introduced ❚Figure 1❚. Frequency of Mistakes Associated With Label Preprinting Before EPPID implementation, we assessed the frequency of mistakes made at any point during the label preprinting process. Very few of these mistakes actually resulted in a mislabeled or unlabeled specimen that reached the laboratory, but they were disruptive with respect to workflow and elevated the risk of labeling errors, which was the reason for measuring them. We observed 265 draws and determined the frequency of 4 mistakes: (1) The requisition provided by the unit had no preprinted labels or an insufficient number of preprinted labels. (2) One or more of the preprinted labels was illegible. (3) Too many preprinted labels were provided with the requisition. (4) The wrong patient’s labels were provided. Time Required for Specimen Collection After implementation, we determined the time required to perform specimen collections using the handheld system. Ten phlebotomists collecting a total of 116 inpatient blood

© American Society for Clinical Pathology

Am J Clin Pathol 2010;133:870-877 871

DOI: 10.1309/AJCPC95YYMSLLRCX

871 871

Morrison et al / Reduction in Labeling Errors in Phlebotomy

❚Figure 1❚ Blood collection process using identification system. (1) Phlebotomist picks up the requisition. (2) Patient from whom sample is to be drawn is manually selected in the personal digital assistant (PDA) based on identifiers on the requisition. (3) Patient verbal identification (ID) is manually compared with requisition and wristband. (4) Wristband is scanned with the PDA. If the ID does not match the patient selected in step 2, an error message is generated. (6) Employee ID badge is scanned. (7) Labels are generated from the portable printer. (8) Specimens are labeled at the bedside. (9) Specimens and the requisition are sent to the laboratory for processing. Before implementation of electronic positive patient identification (EPPID), a nurse or unit coordinator preprinted labels before step 1. Steps 2, 4, 6, and 7 contain new processes introduced with EPPID. DOB, date of birth.

samples were observed in May 2009. The time required for specimen collection was defined as the interval between obtaining the paper requisition and finally placing the properly labeled tubes in a biohazard bag. Times were recorded on a form during observation. Patient Survey We developed a 5-item survey to monitor the effect of EPPID on patient satisfaction during phlebotomy. The first round of surveys was conducted in September and October 2008; 49 patients were surveyed. At that time, EPPID had been implemented but was not being used for patients on contact precautions; thus, these patients were considered EPPID-naive and acceptable candidates for initial surveys. The second round of surveys was conducted in April 2009; 50 patients were selected from across all inpatient services. Each patient was surveyed in person and asked to report whether his or her wristband had been checked or scanned and whether the specimens had been labeled at the bedside. The patient also rated the following on a 5-point Likert scale23: (1) the professionalism of the phlebotomist, (2) the adequacy of the labeling technology, and (3) the overall experience during phlebotomy relative to expectations. From the survey responses, we calculated the percentage of optimal responses before and after EPPID. Optimal responses were defined as “Yes” for the 2 yes-no questions and “5” (“agree” or “excellent”) for the Likert items. Implementation and Use Several EPPID systems were evaluated by a multidisciplinary group that included members from pathology, nursing, 872 872

Am J Clin Pathol 2010;133:870-877 DOI: 10.1309/AJCPC95YYMSLLRCX

information technology, and administration. After considering performance characteristics, reliability, information technology considerations, and goodness of fit regarding our workflow, we selected the Lattice mobile handheld system (Lattice, Wheaton, IL). The equipment included a Symbol MC70 handheld computer with bar code scanner, a Zebra mobile thermal printer, and an integrated wireless radio ❚Image 1❚. Twenty handheld systems were purchased for a team of 39 inpatient phlebotomists covering 3 daily shifts.

❚Image 1❚ Lattice mobile handheld system (Lattice, Wheaton, IL). Personal digital assistant, docking cradle, and portable label printer.

© American Society for Clinical Pathology

Clinical Chemistry / Original Article

Training Training consisted of eight 1.5-hour group introductory sessions, followed by individual training in which each phlebotomist was accompanied by a “super-user.” Phlebotomists were then “shadowed” for up to 1 week, as needed. Following the initial training, some phlebotomists required additional education. Statistical Analysis The analyzed data consisted of the aggregate number of mislabeled and unlabeled samples and aggregate accessions for 21 one-month periods. For our statistical analyses, the main outcome was the rate of incorrectly labeled samples (estimated as the number of errors per 10,000 accessions) during each of the 21 one-month periods. Logistic regression was used to model the rates over time. In the logistic models, we tested for trends over time and changes before and after

implementation of the EPPID system via Wald statistics. We also calculated rates, rate ratios, and 95% confidence intervals (CIs) via logistic regression. All analyses were conducted using Proc Genmod in SAS, version 9.2 (SAS Institute, Cary, NC). All tests were 2-tailed, and a P value less than .05 was considered statistically significant.

Results Specimen Identification Error Rates Labeling errors became less frequent after implementation ❚Table 1❚ and ❚Figure 2❚. Of 181,758 phlebotomy accessions in the preimplementation period, 55 (0.030%) were reported as mislabeled and 44 (0.024%) as unlabeled (Table 1). This compares with 32 (0.017%) mislabeled and 27 (0.015%) unlabeled of 184,043 accessions after implementation; the percentage decreases were 43% and 38%, respectively. Combining mislabeled and unlabeled samples revealed an average labeling error rate of 5.45 per 10,000 phlebotomy

0.12 Percent of Accessions

We elected to implement EPPID in 2 phases, the second of which had not been implemented as of manuscript submission. The main difference between the 2 phases concerns the percentage of specimen collections in the hospital that are performed using EPPID and the extent to which the EPPID system communicates with other hospital information systems. In the first phase, completed in August 2008, all inpatient phlebotomy blood collections were performed using EPPID, except for neonates, whose data could not be interfaced. Patients on contact precautions were also excluded initially while the appropriate procedures were being developed in collaboration with the Infection Control Department. In this phase, the handheld system interfaced with the hospital information system to receive an admission/discharge/transfer feed with current inpatient demographic information. However, we were not able to interface with electronic physician order entry to download orders directly to the handheld system and continued to use paper requisitions. In the second phase, the identification system will interface with electronic physician order entry, and paper requisitions will no longer be used. We will also expand EPPID to all inpatient specimen collections, including those performed by nursing and other health care professionals.

Combined 0.10 Mislabeled 0.08

Unlabeled

0.06 0.04 0.02 0.00

Preimplementation

Oct 2007

Postimplementation

Aug 2008

Jun 2009

❚Figure 2❚ Rate of mislabeled and unlabeled specimens. The percentage of mislabeled (dashed line) and unlabeled (dotted line) samples is plotted each month before, during (vertical line), and after positive patient identification implementation. The aggregate percentage of mislabeled and unlabeled samples is plotted as a bold solid line.

❚Table 1❚ Number of Mislabels and Missing Labels Before and After Implementation of a Positive Patient Identification System in Phlebotomy*

Mislabeled samples Unlabeled samples Total No. of accessions *

Preimplementation Period (October 1, 2007–July 31, 2008)

Postimplementation Period (September 1, 2008–June 30, 2009)

55 (0.030) 44 (0.024) 181,758

32 (0.017) 27 (0.015) 184,043

Data are given as number (percentage).

© American Society for Clinical Pathology

Am J Clin Pathol 2010;133:870-877 873

DOI: 10.1309/AJCPC95YYMSLLRCX

873 873

Morrison et al / Reduction in Labeling Errors in Phlebotomy

accessions preimplementation (95% CI, 4.47-6.63) and 3.20 per 10,000 phlebotomy accessions postimplementation (95% CI, 2.48-4.14). The error rate decreased significantly postimplementation vs preimplementation (rate ratio, 0.59; 95% CI, 0.43-0.81; P = .0013). Before EPPID, the labeling error rate was on an increasing trend; errors were increasing by an average of 9% per month (rate ratio, 1.09; P = .0118). After implementation, the labeling error rate took on a downward trend; this rate fell by an average of 16% per month (rate ratio, 0.84; P = .0003). Estimated Prevented Mislabeling Events Based on the reduced error rate after implementation, we estimated that EPPID prevented 28 mislabeled samples during 1 year. From auditing records, we estimated that use of the EPPID prevented 108 wrong-patient draws during 1 year. Observed Label Preprinting Mistakes Before EPPID Before EPPID, an insufficient number of preprinted labels was provided in 10.9% of draws (29/265) ❚Table 2❚. Illegible preprinted labels were provided in 8.7% of draws (23/265), too many in 4.5% (12/265), and labels for the wrong patient in 0.4% of draws (1/265). In total, a mistake was made during the preprinting process in about one quarter of draws before EPPID. Median Time Required for Specimen Collection The median time required to collect specimens using the handheld device in the postimplementation period was 6 minutes (n = 116; interquartile range, 5-8 minutes).

% Optimal Responses

Patient Survey The percentage of patients who reported being treated in a courteous, professional manner at their last phlebotomy

100 90 80 70 60 50 40 30 20 10 0

❚Table 2❚ Frequency of Mistakes Made During Manual Label Preprinting Process* Specimen Label Preprinting Mistake

Observed Frequency

Insufficient No. provided Illegible labels provided Excessive No. provided Wrong patient’s labels provided Total

10.9 8.7 4.5 0.4 24.5

*

Observed frequencies are the percentages of 265 blood “draws.”

increased from 67% (33/49) to 84% (42/50) after implementation ❚Figure 3❚. The percentage who could confirm that their wristband had been checked or scanned by the phlebotomist increased from 69% (34/49) to 92% (46/50). After implementation, 68% (34/50) as opposed to 49% (24/49) of patients preimplementation agreed that the equipment and techniques used by the phlebotomist seemed sufficient to prevent their specimens from being mislabeled. The percentage of patients who could confirm that their specimens were labeled at the bedside was similar preimplementation and postimplementation. Responses about the overall experience did not change substantially.

Discussion We evaluated the effects of implementing EPPID and found it to be an effective technology for reducing specimen labeling errors within our hospital, as supported by the statistically significant decrease in mislabeled and unlabeled specimens reported by our laboratory and the record of 108 wrong-patient draws prevented by the EPPID system. Only

Pre-EPPID EPPID

Courtesy of Phlebotomist

Wristband Checked

Specimens Labeled at Bedside

Sufficient Equipment for Proper Labeling

Overall Experience

❚Figure 3❚ Responses to patient satisfaction surveys. The percentage of optimal (“yes” or “excellent”) responses for each survey item is shown before (white bars) and after (black bars) electronic positive patient identification (EPPID) implementation.

874 874

Am J Clin Pathol 2010;133:870-877 DOI: 10.1309/AJCPC95YYMSLLRCX

© American Society for Clinical Pathology

Clinical Chemistry / Original Article

a few studies on the effects of such a system for general specimen collection have been published.18,22 Our study, performed at a large, academic teaching hospital, should assist other large institutions considering implementation. Although most EPPID studies report on systems interfacing with electronic order entry, we report on a phase 1 implementation that used paper requisitions. Considering that 50% of institutions do not have electronic requisitions,13 phased implementation may be more realistic for some, especially those attempting to increase patient safety in a setting of limited resources.

is t Ph le b

ot

om

c ni tro m ec te El sys

Subtypes of Mislabeling Errors Within what we have called “mislabeled specimens” at least 3 types of errors exist,11,12 2 of which were primarily addressed by the first phase of EPPID and 1 that will be further addressed by phase 2. These 3 types of mislabeling events result from specific breakdowns in the labeling process and have different levels of detectability ❚Table 3❚. The labeling process can be reduced to 3 items—the requisition, patient specimen, and specimen label—linked in a way that allows or prohibits errors ❚Figure 4❚. A weak link between the requisition and patient specimen allows one to start with patient A’s requisition and draw a specimen from patient B, making “specimen-requisition mismatch” and “wrong-blood-in-tube” errors possible. In the same way, a weak link between the patient specimen and the specimen label allows one to label patient A’s specimen as patient B’s, making “wrong-label” and wrong-blood-in-tube errors possible. Before EPPID, these links depended on human performance and were, thus, vulnerable to human error. For example, a match between the requisition and specimen label depended on the phlebotomist’s ability to visually inspect each preprinted label. EPPID implementation removed preprinting, a step that was problematic in about one quarter of draws, and established a technology-based specimen-label

Patient specimen

Requisition

Phlebotomist

Label

❚Figure 4❚ Links in the specimen identification process. The 3 items in the specimen identification process are represented. A strong link (solid line) ensures that 2 items will belong to the same patient. A weak link (dashed line) allows 2 items to belong to different patients. The electronic identification system creates a technology-based link between the patient’s specimen and the applied label.

link, only allowing labels to be printed for the patient whose wristband is scanned. This strong link helps to prevent wronglabel and wrong-blood-in-tube errors. More patients reported having their wristband checked when specimen collection was performed with EPPID, confirming the ability of EPPID to enforce critical steps in the specimen identification process. Recurring Errors Following implementation, approximately 72% of labeling errors (23/32) were committed when using the handheld system, the majority being specimen-requisition mismatch errors. EPPID can prevent errors only insofar as key steps within the standard operating procedure are followed. Because the identification system did not yet interface with

❚Table 3❚ Subtypes of Mislabeling Errors* Error Type

A B A

A B B

A A B

Description

Ease of Detection

Effect of EPPID

Wrong label: requisition for patient A; specimen obtained from patient A; sample labeled for patient B

Sample obtained from correct patient, but the wrong label is applied to tube

High; requisition-label mismatch

Decreased frequency; EPPID does not permit printing of label mismatched with patient from whom sample was obtained

Specimen-requisition mismatch: requisition for patient A; specimen obtained from patient B; sample labeled for patient B

Specimen is correctly labeled but does not match the requisition

High; requisition-label mismatch

Smaller effect than with other error types; phase 2 will further reduce frequency

Wrong blood in tube: requisition for Requisition and specimen label Low; requisition and Decreased frequency; EPPID does not patient A; specimen obtained from match, but the tube contains label match, making permit printing of label mismatched patient B; sample labeled for patient A the wrong patient’s blood detection difficult with patient from whom sample specimen was obtained

EPPID, electronic positive patient identification. * Based on our laboratory workflow the second and third error subtypes present a higher risk of reporting results for the wrong patient. All 3 error subtypes increase the risk that a result will be reported for the wrong patient.

© American Society for Clinical Pathology

Am J Clin Pathol 2010;133:870-877 875

DOI: 10.1309/AJCPC95YYMSLLRCX

875 875

Morrison et al / Reduction in Labeling Errors in Phlebotomy

electronic order entry in phase 1, the phlebotomist still had the job of selecting the correct patient in the PDA based on identifiers on the paper requisition. It was possible to select the wrong patient in the PDA and subsequently obtain a specimen from this patient, especially if his or her name or room number resembled that of the correct patient. The patient’s tubes would be labeled correctly as belonging to him or her, but the specimen should not have been obtained from this patient. This existing weak link in the labeling process, which will be replaced by electronic order entry in phase 2, partially explains why we saw a 43% decrease in our mislabeling rate, whereas other hospitals have seen reductions of more than 80%.18,22 While still in the first phase of implementation, further education on proper use of the labeling system should help to reduce recurring errors.

Benefits of the Second Implementation Phase

Discrepancy Between Two Estimates of Prevented Errors The 2 methods we used to estimate mislabeling errors prevented by the EPPID system produced 2 considerably different figures. This is likely because the first method relied on the rate of reported errors, whereas the second method included errors that most often are not reported, such as wrongblood-in-tube errors, which are difficult to detect within our laboratory. Thus, although the number of errors reported in our laboratory may have decreased by 28 during 1 year, the number of actual prevented errors may be as high as 108. Most identification errors are thought to go undetected.14

EPPID represents an effective technology for preventing specimen identification errors in a clinical laboratory without slowing the collection process or negatively impacting the patient experience during phlebotomy. Even without interfacing with electronic order entry, the system reduced error rates significantly. This technology represents a key to ensuring patient safety in laboratory medicine.

Additional Benefits of EPPID Before EPPID, the phlebotomist ID and blood collection time were manually inscribed and often illegible or inadequate. EPPID has ensured that 100% of labels are legibly printed with this information and that accurate auditing data are automatically recorded. EPPID did not slow our collection process or negatively affect the patient experience during phlebotomy. The time required for specimen collection using the system was comparable to published times, which were 5.8 minutes18 and 6 minutes20 in other studies. Technical Issues Use of the handheld system was 85% in the eighth postimplementation month. Less than 100% use was due partly to technical issues; in particular, when the PDA battery was low, the scanner did not work. A phlebotomist at the end of his or her shift with a low battery might revert to the manual labeling method instead of traveling to the laboratory to pick up a new PDA. Phlebotomists now carry backup batteries on rounds, which should help to raise use of the system to 100%. Such issues occur routinely following implementation of electronic solutions, underscoring the importance of postmeasurement evaluation and trouble-shooting. 876 876

Am J Clin Pathol 2010;133:870-877 DOI: 10.1309/AJCPC95YYMSLLRCX

In the second phase of implementation, the system will automatically download electronic orders to PDAs. Paper requisitions will be eliminated, and phlebotomists will no longer be able to mistakenly obtain a specimen from a patient who has no orders, thus reducing the frequency of specimenrequisition mismatch errors. Furthermore, by implementing the bar code labeling system only within our phlebotomy service, we were targeting only 15% of all inpatient mislabeled specimens, the rest being committed primarily by health care professionals other than phlebotomists. Our aim is to expand the technology to all specimen collection in our hospital.

Conclusion

From the Departments of 1Medicine and 2Pathology, Clinical Laboratories Division, Brigham and Women’s Hospital, Harvard Medical School, Boston, MA. Address reprint requests to Dr Melanson: Brigham and Women’s Hospital, 75 Francis St, Amory 2, Boston, MA 02115.

References 1. Kohn LT, Corrigan JM, Donaldson MS, eds; Institute of Medicine, Committee on Quality of Health Care in America. To Err Is Human: Building a Safer Health System. Washington, DC: National Academy Press; 1999. 2. Leape LL, Berwick DM. Five years after To Err Is Human: what have we learned? JAMA. 2005;293:2384-2390. 3. Institute of Medicine, Committee on Quality of Health Care in America. Crossing the Quality Chasm: A New Health System for the 21st Century. Washington, DC: National Academies Press; 2001. 4. Blumenthal D. The errors of our ways [editorial]. Clin Chem. 1997;43(8 pt 1):1305. 5. Bonini P, Plebani M, Ceriotti F, et al. Errors in laboratory medicine. Clin Chem. 2002;48:691-698. 6. Stankovic AK. The laboratory is a key partner in assuring patient safety. Clin Lab Med. 2004;24:1023-1035. 7. Lippi G, Guidi GC, Mattiuzzi C, et al. Preanalytical variability: the dark side of the moon in laboratory testing. Clin Chem Lab Med. 2006;44:358-365. 8. Chassin MR, Becher EC. The wrong patient. Ann Intern Med. 2002;136:826-833. 9. Celizic M. Cancer-free woman underwent double mastectomy because of lab mix-up. MSNBC Interactive Web site. http:// www.msnbc.msn.com/id/21127917. Accessed August 31, 2009.

© American Society for Clinical Pathology

Clinical Chemistry / Original Article

10. Sazama K. Reports of 355 transfusion-associated deaths: 1976 through 1985. Transfusion. 1990;30:583-590. 11. Valenstein PN, Raab SS, Walsh MK. Identification errors involving clinical laboratories: a College of American Pathologists Q-Probes study of patient and specimen identification errors at 120 institutions. Arch Pathol Lab Med. 2006;130:1106-1113. 12. Wagar EA, Tamashiro L, Yasin B, et al. Patient safety in the clinical laboratory: a longitudinal analysis of specimen identification errors. Arch Pathol Lab Med. 2006;130:1662-1668. 13. Wagar EA, Stankovic AK, Raab S, et al. Specimen labeling errors: a Q-Probes analysis of 147 clinical laboratories. Arch Pathol Lab Med. 2008;132:1617-1622. 14. Valenstein PN, Sirota RL. Identification errors in pathology and laboratory medicine. Clin Lab Med. 2004;24:979-996. 15. Joint Commission. 2003 National Patient Safety Goals. http://www.jointcommission.org/PatientSafety/ NationalPatientSafetyGoals/03_npsgs.htm. Accessed June 25, 2009. 16. The Joint Commission, Joint Commission International, World Health Organization. Patient identification. Patient Safety Solutions. 2007. http://www.ccforpatientsafety.org/ common/pdfs/fpdf/presskit/PS-Solution2.pdf. Accessed June 25, 2009.

17. Lippi G, Blanckaert N, Bonini P, et al. Causes, consequences, detection, and prevention of identification errors in laboratory diagnostics. Clin Chem Lab Med. 2009;47:143-153. 18. Bologna LJ, Lind C, Riggs RC. Reducing major identification errors within a deployed phlebotomy process. Clin Leadersh Manag Rev. 2002;16:22-26. 19. Turner CL, Casbard AC, Murphy MF. Barcode technology: its role in increasing the safety of blood transfusion. Transfusion. 2003;43:1200-1209. 20. Chan JCW, Chu RW, Young BWY, et al. Use of an electronic barcode system for patient identification during blood transfusion: 3-year experience in a regional hospital. Hong Kong Med J. 2004;10:166-171. 21. Porcella A, Walker K. Patient safety with blood products administration using wireless and bar-code technology. AMIA Annu Symp Proc. 2005:614-618. 22. Hayden RT, Patterson DJ, Jay DW, et al. Computer-assisted bar-coding system significantly reduces clinical laboratory specimen identification errors in a pediatric oncology hospital. J Pediatr. 2008;152:219-224. 23. Likert R. A technique for the measurement of attitudes. Arch Psychol. 1932;140:1-55.

© American Society for Clinical Pathology

Am J Clin Pathol 2010;133:870-877 877

DOI: 10.1309/AJCPC95YYMSLLRCX